ELIMINATOR BALLOON DILATOR

K940965 · C.R. Bard, Inc. · FGE · May 20, 1994 · Gastroenterology, Urology

Device Facts

Record IDK940965
Device NameELIMINATOR BALLOON DILATOR
ApplicantC.R. Bard, Inc.
Product CodeFGE · Gastroenterology, Urology
Decision DateMay 20, 1994
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bard® Balloon Dilator is intended to be used for dilatation of strictures of the gastrointestinal tract, but not limited to strictures of the esophagus, pylorus, duodenum, sphincter of Oddi, biliary tree and colon.

Device Story

The Bard® Balloon Dilator is a 6Fr. polyurethane catheter with a polyethylene terephthalate (PET) balloon at the distal end. It is a line extension of the Bard® Balloon Dilatation System, featuring a 230cm shaft, 4cm balloon length, and modified cone angles. The device is used by physicians in clinical settings to dilate gastrointestinal strictures. The operator inserts the catheter and inflates the balloon to the required diameter (6mm to 18mm) to mechanically expand the stricture. The device is manually operated; it does not involve electronic processing or software. Clinical benefit is derived from the mechanical relief of gastrointestinal obstructions.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and performance comparisons to legally marketed predicate devices.

Technological Characteristics

6Fr. polyurethane catheter shaft; polyethylene terephthalate (PET) balloon; 230cm length; 6-18mm inflated balloon diameter; single lumen; manual inflation mechanism. No software or electronic components.

Indications for Use

Indicated for patients requiring dilatation of gastrointestinal tract strictures, including esophagus, pylorus, duodenum, sphincter of Oddi, biliary tree, and colon.

Regulatory Classification

Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{1} MAY 20 1994 K940965 6.0 ### 510(k) SUMMARY FOR THE BARD® BALLOON DILATOR As required under Section 12, part (a)(i)(3A) of the Safe Medical Device Act of 1990, an adequate summary of any information respecting safety and effectiveness follows. ### 6.1 General Information Name and address of submitter: Interventional Products Division, C.R. Bard, Inc. 200 Ames Pond Drive Tewksbury, MA 01876 Contact: John A. DeLucia Director, Quality Assurance and Regulatory Affairs Date of Summary: 1 March 1994 Name of Device: Bard® Balloon Dilator Predicate Devices: Bard® Balloon Dilatation System (K863437) Microvasive Achiever™ (Boston Scientific, Inc.) Description and Intended Use of Device: The Bard® Balloon Dilator is a 6Fr. polyurethane 230cm shaft catheter containing a polyethylene terephthalate (PET) balloon. Inflated balloon diameters range in size from 6mm to 18mm. The Bard® Balloon Dilator is intended to be used for dilatation of strictures of the gastrointestinal tract, but not limited to strictures of the esophagus, pylorus, duodenum, sphincter of Oddi, biliary tree and colon. 000037 {2} 6.2 ## Summary of Similarities and Differences Please refer to Exhibit 6.2 for a Table of Similarities and Differences. The Bard® Balloon Dilator is a line extension to the currently marketed Bard® Balloon Dilatation System (K863437). Slight design modifications have been made to the predicate Bard® Dilator to expand the product line in order to present a greater selection of dilatation catheters to physicians. These modifications are: 1) extension of the catheter shaft from 180cm to a length of 230cm 2) reduction of the balloon length from 8cm to 4cm 3) modification of balloon geometry and cone angles 4) shorter distal tip 5) alternate supplier of polyurethane for the catheter shaft which contains a different colorant In all other respects, the proposed device is the similar to the currently marketed Bard® Balloon Dilatation System. The device subject to premarket notification is also similar to the balloon dilatation catheter, Achiever, Disposable TTS® Balloon Dilator, marketed by Microvasive (Boston Scientific, Inc.). Their catheter length, size, balloon length, inflated diameter, and balloon material are equivalent. The catheter shaft material is different in that the Bard® proposed device consists of polyurethane while the Microvasive Achiever™ is made of polyvinyl chloride. {3} Exhibit 6.2 Table of Substantial Equivalence* | Characteristics | Current Marketed Device Bard® Balloon Dilatation Catheter - K863437 | Microvasive Achiever™** | Proposed Bard® Balloon Dilators | | --- | --- | --- | --- | | Intended Use | For use in the dilatation of strictures of the gastrointestinal tract | same | same | | Catheter Length (cm) | 180 | 180 and 240 | 230 | | Catheter Size (Fr.) | 6 | 5 - 14 | 6 | | Balloon Length (cm) | 8 | 5.5 and 8 | 4 | | Inflated Balloon Diameter (mm) | 6 - 18 | 6 - 25 | 6 - 18 | | Shaft Material | Polyurethane/Reed Blue Colorant | Polyvinyl Chloride | Polyurethane/Phthalocyanine Blue | | Balloon Material | Polyethylene terephthalate | Polyethylene | Polyethylene terephthalate | | Lumen | Single | Single | Single | Recommended Maximum Balloon Inflation Pressure (psi) | Balloon Diameter (mm) | Current Marketed Device Bard® Balloon Dilatation Catheter - K863437 | Microvasive Achiever™ | Proposed Bard® Balloon Dilators | | --- | --- | --- | --- | | 6 | 120 | 75 | 100 | | 8 | 120 | 60 | 120 | | 10 | 90 | 50 | 90 | | 12 | 90 | 50 | 90 | | 15 | 60 | 45 | 60 | | 18 | 50 | 35 | 50 | * All dimensions are nominal. ** We could not ascertain the 510(k) number for this device; however, we consider it to be a legally marketed device since it is in commercial distribution. 000039 {4} ### 6.3 Substantial Equivalence Decision Tree The 510(k) "Substantial Equivalence" Decision-Making Process (Detailed) in ODE Guidance #K86-3, Guidance on the CDRH Premarket Notification Review Program, was used to determine substantial equivalence. Please refer to Exhibit 4.2 for a diagram of the 510(k) Decision Tree. The answers to the questions lead to a determination of substantial equivalence 1. Does the new device have same indication statement? Yes. There is no change in the intended use for the Bard® Balloon Dilatation System as a result of this product line extension. The previously reviewed and currently marketed Bard® Balloon Dilatation Catheter System and its proposed product line extension are intended for use in the dilatation of strictures of the gastrointestinal tract including, but not limited to, strictures of the esophagus, pylorus, duodenum, sphincter of Oddi, biliary tree and colon. 2. Does the new device have same technological characteristics, e.g. design, materials, etc.? Yes. The proposed Bard® Balloon Dilators like the currently marketed Bard® Balloon Dilatation Catheter System are substantially equivalent with respect to design, in that they consist of a catheter to which an inflatable balloon is bonded to the distal end of the catheter shaft. A slight design modification for the proposed catheters entails longer catheter shafts (180 to 230cm), shorter balloons (4cm), having slightly different cone angles, a shorter distal catheter tip, and a different blue colorant in the polyurethane shaft material. 000040 {5} The proposed design for the product line extension to the Bard® Balloon Dilatation System is also substantially equivalent to the Achiever™ balloon dilator catheters marketed by Microvasive. Microvasive's longest catheter has a usable length of 240cm, the catheter shaft sizes are between 5 and 14Fr, the lengths of the balloon range from 5.5 to 8cm, and the inflated balloon diameters range between 6 and 25mm. ### 3. Are the descriptive characteristics precise enough to ensure equivalence? Yes. A comparison of the key properties of intended use, design and materials are presented in the "Table of Substantial Equivalence", Exhibit 4.1. This table presents the descriptive and performance data of the currently marketed device, the Bard® Balloon Dilatation System (K863437), the proposed product line extension and Microvasive Achiever™ balloon dilator catheters. The table has been developed to support the performance of the proposed Bard® Balloon Dilatation System product line extension and demonstrates that the proposed Bard dilatation catheters are substantially equivalent to those currently marketed. 000041 {6} [LOGO] DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center – WO66-G609 Silver Spring, MD 20993-0002 March 25, 2015 C.R. Bard, Inc. Bard Interventional Products Division John A. DeLucia Director, R.A./Q.A. 200 Ames Pond Drive Tewksbury, MA 01876 Re: K940965 Trade/Device Name: Bard® Balloon Dilators Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE, KNQ Dated (Date on orig SE ltr): March 1, 1994 Received (Date on orig SE ltr): March 2, 1994 Dear John A. DeLucia, This letter corrects our substantially equivalent letter of May 20, 1994. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {7} Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Joyce M. Whang -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...