← Product Code [LXS](/productcode/LXS) · K895864

# MICRORUPTOR III SYST, Q-SWITCHED ND:YAG LASER COMP (K895864)

_Lasag AG · LXS · Dec 29, 1989 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K895864

## Device Facts

- **Applicant:** Lasag AG
- **Product Code:** [LXS](/productcode/LXS.md)
- **Decision Date:** Dec 29, 1989
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4392
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Regulatory Identification

The Nd:YAG laser for posterior capsulotomy and peripheral iridotomy consists of a mode-locked or Q-switched solid state Nd:YAG laser intended for disruption of the posterior capsule or the iris via optical breakdown. The Nd:YAG laser generates short pulse, low energy, high power, coherent optical radiation. When the laser output is combined with focusing optics, the high irradiance at the target causes tissue disruption via optical breakdown. A visible aiming system is utilized to target the invisible Nd:YAG laser radiation on or in close proximity to the target tissue.

## Special Controls

*Classification.* Class II (special controls). Design Parameters: Device must emit a laser beam with the following parameters: wavelength = 1064 nanometers; spot size = 50 to 100 micros; pulse width = 3 to 30 nanoseconds; output energy per pulse = 0.5 to 15 millijoules (mJ); repetition rate = 1 to 10 pulses; and total energy = 20 to 120 mJ.

---

**Source:** [https://fda.innolitics.com/device/K895864](https://fda.innolitics.com/device/K895864)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
