LECTEC NON-INVASIVE PACING ELECTRODE

K872012 · Lectec Corp. · DRO · Jun 29, 1987 · Cardiovascular

Device Facts

Record IDK872012
Device NameLECTEC NON-INVASIVE PACING ELECTRODE
ApplicantLectec Corp.
Product CodeDRO · Cardiovascular
Decision DateJun 29, 1987
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5550
Device ClassClass 2

Regulatory Classification

Identification

An external transcutaneous cardiac pacemaker (noninvasive) is a device used to supply a periodic electrical pulse intended to pace the heart. The pulse from the device is usually applied to the surface of the chest through electrodes such as defibrillator paddles.

Special Controls

*Classification.* Class II. The special controls for this device are:(1) “American National Standards Institute/American Association for Medical Instrumentation's DF-21 ‘Cardiac Defibrillator Devices’ ” 2d ed., 1996, and (2) “The maximum pulse amplitude should not exceed 200 milliamperes. The maximum pulse duration should not exceed 50 milliseconds.”