← Product Code [DRO](/productcode/DRO) · K844811

# PACE AID MODEL 53 (K844811)

_Cardiac Resucitator Corp. · DRO · Jan 24, 1985 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K844811

## Device Facts

- **Applicant:** Cardiac Resucitator Corp.
- **Product Code:** [DRO](/productcode/DRO.md)
- **Decision Date:** Jan 24, 1985
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5550
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Regulatory Identification

An external transcutaneous cardiac pacemaker (noninvasive) is a device used to supply a periodic electrical pulse intended to pace the heart. The pulse from the device is usually applied to the surface of the chest through electrodes such as defibrillator paddles.

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) “American National Standards Institute/American Association for Medical Instrumentation's DF-21 ‘Cardiac Defibrillator Devices’ ” 2d ed., 1996, and
(2) “The maximum pulse amplitude should not exceed 200 milliamperes. The maximum pulse duration should not exceed 50 milliseconds.”

---

**Source:** [https://fda.innolitics.com/device/K844811](https://fda.innolitics.com/device/K844811)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com