← Product Code [QLQ](/productcode/QLQ) · K263159

# prodisc® C; prodisc® C SK; prodisc® C Nova; prodisc® C Vivo Instruments (K263159)

_Centinel Spine, LLC · QLQ · Oct 7, 2026 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K263159

## Device Facts

- **Applicant:** Centinel Spine, LLC
- **Product Code:** [QLQ](/productcode/QLQ.md)
- **Decision Date:** Oct 7, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.4515
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Regulatory Identification

Orthopedic manual surgical instrumentation for use with total disc replacement devices are non-powered hand-held devices designed specifically for use with a total disc replacement device and interface with the associated implant for the purpose of insertion, removal, placement, or repositioning, or to cut, rasp, or create a defect specific to the features of the associated implant. This type of device includes instruments specific to the geometry of the implant.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Technical specifications regarding geometry of the instruments must be specified and validated to demonstrate that the instruments can safely position, place, or remove the implant.
(2) The patient contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate that reprocessing of reusable devices that are provided non-sterile, or sterilization of devices provided sterile, is validated.
(4) Labeling must include:
(i) Identification of implant(s) and instruments which have been validated for use together; and
(ii) Validated methods and instructions for reprocessing any reusable parts.

---

**Source:** [https://fda.innolitics.com/device/K263159](https://fda.innolitics.com/device/K263159)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
