Medical Monitor (32MD08WC)

K262985 · Lg Electronics.Inc · PGY · Oct 2, 2026 · Radiology

Device Facts

Record IDK262985
Device NameMedical Monitor (32MD08WC)
ApplicantLg Electronics.Inc
Product CodePGY · Radiology
Decision DateOct 2, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesPediatric

Indications for Use

This Medical Monitor is indicated for use in displaying radiological images and enhancing the image quality for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Device Story

Medical Monitor (32MD08WC) is a 31.5-inch TFT LCD display module; utilizes a-Si thin-film transistors; operates in normally black transmissive mode. Inputs radiological image data via eDP (HBR2) interface; processes signals to drive 3,840 x 2,160 pixel array; supports 10-bit grayscale/brightness per sub-pixel for >1.07 billion colors. Used in clinical settings by trained medical practitioners for review, analysis, and diagnosis of radiological images. Output is a high-resolution visual display of medical images; assists clinicians in diagnostic decision-making. Benefits include enhanced image quality for accurate radiological assessment.

Clinical Evidence

No clinical studies were conducted. Evidence consists of bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC TR 60601-4-2), and software validation per IEC 62304.

Technological Characteristics

31.5-inch TFT LCD panel; a-Si thin-film transistors; normally black transmissive mode; 3,840 x 2,160 UHD resolution; 10-bit signal processing; eDP (HBR2, 5.4 Gbps) interface. Basic software documentation level; verified per IEC 62304.

Indications for Use

Indicated for use by trained medical practitioners for displaying and enhancing radiological images for review, analysis, and diagnosis. Suitable for all age groups including neonates, infants, children, adolescents, and adults. Not intended for mammography.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1} **FDA** U.S. FOOD & DRUG ADMINISTRATION October 2, 2026 Lg Electronics.Inc Wonhi Lee RA Specialist 168, Suchul-Daero Gumi-Si, Gyeongsangbuk-do 39368 Republic Of Korea Re: K262985 Trade/Device Name: Medical Monitor (32MD08WC) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: August 27, 2026 Received: August 27, 2026 Dear Wonhi Lee: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2} K262985 - Wonhi Lee Page 2 Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {3} K262985 - Wonhi Lee Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262985 | ? | | Please provide the device trade name(s). | | ? | | Medical Monitor (32MD08WC) | | | | Please provide your Indications for Use below. | | ? | | This Medical Monitor is indicated for use in displaying radiological images and enhancing the image quality for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. | | | | Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? | {5} 510(k) #: K262985 # 510(k) Summary Prepared on: 2026-08-27 Contact Details 21 CFR 807.92(a)(1) | Applicant Name | LG Electronics Inc. | | --- | --- | | Applicant Address | 168, Suchul-daero Gumi-si Gyeongsangbuk-do 39368 Korea, Republic of | | Applicant Contact Telephone | +82-1091523670 | | Applicant Contact | Mr. Wonhi Lee | | Applicant Contact Email | wonhi.lee@lge.com | Device Name 21 CFR 807.92(a)(2) | Device Trade Name | Medical Monitor (32MD08WC) | | --- | --- | | Common Name | Medical image management and processing system | | Classification Name | Display, Diagnostic Radiology | | Regulation Number | 892.2050 | | Product Code(s) | PGY | Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K232127 | 32HL512D | PGY | Device Description Summary 21 CFR 807.92(a)(4) The 32MD08WC is a display module featuring a TFT LCD panel without an integrated LED driver. It utilizes a-Si thin-film transistors (TFTs) as active elements and operates in "normally black" mode as a transmissive display. It offers an active display area with a diagonal size of 31.5 inches and supports UHD resolution (3,840 x 2,160 pixel array). Each pixel consists of red, green, and blue sub-pixels (or dots) arranged in a vertical stripe pattern. The gray scale or brightness of each sub-pixel is determined by an 10-bit signal, enabling the display of over 1.07 billion colors. Additionally, it is designed to utilize an eDP (HBR2, 5.4 Gbps) interface. Intended Use/Indications for Use 21 CFR 807.92(a)(5) This Medical Monitor is indicated for use in displaying radiological images and enhancing the image quality for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. Indications for Use Comparison 21 CFR 807.92(a)(5) <Indication Comparison> When comparing the indications for use statements, the terminology demonstrates that the proposed device shares the same indications for use as the predicate device. Therefore, it supports the determination of substantial equivalence between the proposed device and predicate device. Technological Comparison 21 CFR 807.92(a)(6) The comparison table shows the proposed device (32MD08WC) has shares the same indications for use as the predicate device. Although there are some differences, they do not compromise the safety and performance of the proposed device. The major change includes its module, which has been verified and validated through physical laboratory testing conducted according to the FDA {6} Guidance, "Display Devices for Diagnostic Radiology". Based on the physical laboratory test results and software validation reports, all the differences between the proposed and predicate device do not raise any concerns regarding safety and effectiveness of the device. Therefore, the proposed device is substantially equivalent to the predicate device in terms of its indications for use and technological characteristics. # Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The following data were provided in support of the substantial equivalence determination: 1) Electrical Safety and Electromagnetic Compatibility Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: - IEC60601-1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems 2) Software Validation 32MD08WC contain software that belongs to Basic Documentation Level. The software was designed and developed according to a software development process and was verified and validated according to IEC 62304. Software information is provided in accordance with FDA guidance: - The content of premarket submissions for Device Software Functions (June 14, 2023) 3) Performance Testing The major change includes its module, which has been verified and validated through physical laboratory testing conducted according to the FDA Guidance, "Display Devices for Diagnostic Radiology". No clinical studies were considered necessary, and therefore, none were conducted. Based on the information provided in this premarket notification and in accordance with the Federal Food, Drug, and Cosmetic Act, 21 CFR Part 807, LG Electronics Inc. has determined that there are no significant differences between the proposed device and the predicate device that would adversely affect its effectiveness and safety. The proposed device has been proved to be as safe and effective as the predicate devices, which was previously cleared in K232127. They shares the same indications for use and similar technological characteristics, with any differences supported by software validation reports that demonstrate the proposed device's safety and effectiveness. Therefore, the technological differences between the proposed device and its predicate device do not raise any new concerns regarding safety or effectiveness, and the proposed device can be considered substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...