Led and Laser Hair Growth Device (H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500)

K262834 · Shenzhen Jianchao Intelligent Technology Co., Ltd. · OAP · Oct 9, 2026 · Physical Medicine

Device Facts

Record IDK262834
Device NameLed and Laser Hair Growth Device (H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500)
ApplicantShenzhen Jianchao Intelligent Technology Co., Ltd.
Product CodeOAP · Physical Medicine
Decision DateOct 9, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Led and Laser Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of IIa-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Classification of Skin Phototypes I to IV.

Device Story

Dome-shaped, over-the-counter hair growth device; utilizes low-level laser diodes (660nm) and LEDs (660nm) to stimulate hair growth. Operated by patient at home; powered by internal rechargeable lithium battery. Device includes buzzer for user prompts; adjustable treatment times (15-30 mins) for most models. Safety verified via compliance with IEC 60601-1, IEC 60825-1, and IEC 60601-2-83. Output intensity (2.7mW/cm2) comparable to predicate. Intended to provide therapeutic light energy to scalp to address androgenetic alopecia.

Clinical Evidence

Bench testing only. No clinical data provided. Performance supported by biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), photobiological safety (IEC 62471), and laser safety (IEC 60825-1) testing.

Technological Characteristics

Dome-shaped helmet; 70 LEDs and 54 laser diodes (660nm ± 10nm); energy <5mW/laser, <10mW/LED; intensity 2.7mW/cm2. Powered by internal rechargeable lithium battery. Class 3R laser product. Biocompatible materials per ISO 10993. Software-controlled treatment duration.

Indications for Use

Indicated for males with androgenetic alopecia (Hamilton-Norwood IIa-V) and females with androgenetic alopecia (Ludwig-Savin I-II) with Fitzpatrick Skin Phototypes I-IV.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION October 9, 2026 Shenzhen Jianchao Intelligent Technology Co., Ltd. % Riley Chen RA specialist Feiying Drug & Medical Consulting Technical Service Group Rm.2401 Zhenye International Business Center, # 3101-90, Qianhai Rd. Shenzhen, Guangdong 518052 China Re: K262834 Trade/Device Name: Led and Laser Hair Growth Device (H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500) Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: August 11, 2026 Received: August 11, 2026 Dear Riley Chen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K262834 - Riley Chen Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {3} K262834 - Riley Chen Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.10.09 14:40:16 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K262834 Device Name Name of Device: Led and Laser Hair Growth Device Model(s): H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500 Indications for Use (Describe) The Led and Laser Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of IIa-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Classification of Skin Phototypes I to IV. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} Shenzhen Jianchao Intelligent Technology Co., Ltd. 510(k)s –510(k) Summary # 510(k) Summary - K262834 "510(k) Summary" as required by 21 CFR Part 807.92. # I. Submitter Shenzhen Jianchao Intelligent Technology Co., Ltd. Rm101, 201, 301, Bldg.25, No.68 Hexiu West Road, Fuhai St., Baoan, Shenzhen, Guangdong, 518103, CHINA Fred Li Title: Director Tel: +86-18675507170 Email:fred@jianchaosmart.com Preparation date: 2026-10-08 # II. Device Information Name of Device: Led and Laser Hair Growth Device Model(s): H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500 Common or Usual Name: Laser, Comb, Hair Regulation Name: Infrared lamp Regulatory Class: II Product Code: OAP Regulation Number: 21 CFR 890.5500 # III. Predicate Device | Manufacturer | Trade Name | 510(k) Number | Cleared Date | | --- | --- | --- | --- | | Shenzhen Kaiyan Medical Equipment Co., Ltd | NOOANCE LED AND LASER HELMET (Model: M-120, M-282 PRO) | K231321 | July 31, 2023 | # IV. Device Description Led and Laser Hair Growth Device is an over-the-counter dome-shaped device designed to promote hair growth in person with androgenetic alopecia in women and men. The device is a combination of laser diodes (wavelength 660nm±10nm) and LEDs (wavelength 660nm±10nm). The device has a total of 17 models, including model H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500. The main differences among these models lie in the treatment time and indicators function. Page 1 of 5 {6} Shenzhen Jianchao Intelligent Technology Co., Ltd. 510(k)s –510(k) Summary--- Led and Laser Hair Growth Device is powered by internal rechargeable lithium battery that can be charged by external power adapter. The device equips a buzzer to prompt the user during operation. ## V. Indications for Use The Led and Laser Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of IIa-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Classification of Skin Phototypes I to IV. ## VI. Comparison of Technological Characteristics With the Predicate Device The Led and Laser Hair Growth Device has the same intended use as the predicate device. The technological characteristics such as energy type, wavelength, intensity, are similar to the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, Led and Laser Hair Growth Device can be found substantially equivalent to its predicate device. | Comparison Elements | Subject Device | Predicate Device | Remark | | --- | --- | --- | --- | | 510(k) Number | K262834 | K231321 | / | | Trade name | Led and Laser Hair Growth Device | Nooance Led And Laser Helmet | / | | Model | H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100, ID-FLH-500 | M-120, M-282 PRO | / | | Regulation number | 21 CFR 890.5500 | 21 CFR 890.5500 | Same | | Product code | OAP | OAP | Same | | Device classification | II | II | Same | | Prescription or OTC | OTC | OTC | Same | | Indication for use/ Intended use | The Led and Laser Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of IIa-V and | The NOOANCE LED AND LASER HELMET is indicated to promote hair growth in males with androgenetic alopecia who have HamiltonNorwood Classifications of IIa-V and | Same | Page 2 of 5 {7} Shenzhen Jianchao Intelligent Technology Co., Ltd. 510(k)s -510(k) Summary | Comparison Elements | Subject Device | Predicate Device | Remark | | --- | --- | --- | --- | | | females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Classification of Skin Phototypes I to IV. | Fitzpatrick Classification of Skin Phototypes I to IV, and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV. | | | Type of laser | Low-level laser diodes and light emitting diodes | Low-level laser diodes and light emitting diodes | Same | | Wavelength | Red LED: 660nm ± 10nmLaser Diodes: 660nm ± 10nm | Laser: 650±5nmRed light LED: 650±5 nm | Similar | | Amounts of laser | LED: 70 pcsLaser: 54 pcs | M-120:LDs: 51pcsLEDs: 69pcsM-282 PRO:LDs: 82pcsLEDs: 200pcs | DifferentNote 1 | | Energy of per diode | <5mW per laser diode<10mW per LED diode | < 5mW from each laser or LED | DifferentNote 1 | | Power intensity | 2.7mW/cm2 | M-120: approximately 1.3mw/cm2M-282 PRO: approximately 2.8mw/cm2 | SimilarNote 2 | | Classification according to IEC60825-1 | Class 3R | Class 1C | DifferentNote 3 | | Power supply | Lithium battery | Lithium-ion battery | Same | | Treatment time | H5520, H5521, H5522, H5523, H5524, H5525, H5530, H5605, H5610, H5615, H5620, H5625, H5630, H6510, H6520, INIA-HM100: 15/ 20/ 25/ 30min adjustableID-FLH-500: 25 min fixed | 25 minutes every other day for 16 weeks | SimilarNote 4 | | Applicable people | Norwood-Hamilton IIa-V (males)Ludwig-Savin I~II (females) | Norwood-Hamilton IIa-V (males)Ludwig-Savin I~II- (females) | Same | | Applicable skin | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Same | Page 3 of 5 {8} Shenzhen Jianchao Intelligent Technology Co., Ltd. 510(k)s -510(k) Summary | **Comparison Elements** | **Subject Device** | **Predicate Device** | **Remark** | | --- | --- | --- | --- | | Appearance design | Helmet | Helmet | Same | # **Comparison in details:** # **Note 1:** The light diodes quantity and energy of per diode of the subject device is different from the predicate device, however, the power intensity of the subject device is very close to that of the predicate device (model M-282 PRO), and the subject device complies with IEC 60601-1, IEC 60825-1 and IEC 60601-2-83 requirements, therefore, these differences will not raise any safety or effectiveness issue. # **Note 2:** The power intensity of the subject device is very close to the predicate device (model M-282 PRO), and the subject device complies with IEC 60601-1, IEC 60825-1 and IEC 60601-2-83 requirements, so this difference will not raise any safety or effectiveness issue. # **Note 3:** Though the laser classification of the subject device is different from the predicate device, the subject device complies with the IEC 60825-1 requirements, and relevant warning labels have been added on the product, so this difference will not raise any safety or effectiveness issue. # **Note 4:** The treatment frequency of model ID-FLH-500 of the subject device is the same as the predicate device, though the others of the subject device is a slightly different from that of the predicate device, theirs treatment time is adjustable, users can adjust the treatment duration according to their own needs. The power intensity of the subject device and the predicate device is very close, and the subject device complies with IEC 60601-1 and IEC 60825-1 requirements, so this difference in treatment frequency will not raise any safety or effectiveness issue. # **VII. Non-Clinical Testing** The following performance data were provided in support of the substantial equivalence determination. # **1) Biocompatibility Testing** The biocompatibility evaluation for the body-contacting components of the device was conducted in accordance with the “Use of International Standard ISO 10993-1, ‘Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process’, Document Issued on September 8, 2023”, as recommended by FDA. The following testing was performed to, and passed, including: - ➤ ISO 10993-5: 2009, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity Page 4 of 5 {9} Shenzhen Jianchao Intelligent Technology Co., Ltd. 510(k)s –510(k) Summary ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation # 2) Electrical Safety and EMC Electrical safety and EMC testing was performed to, and passed, as per the following standards: IEC 60601-1:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC 60601-1-2:2020, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility IEC 60601-1-11:2020, Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment IEC 60601-2-83: 2022, Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment IEC 62471: 2006, Photobiological safety of lamps and lamp systems IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification, and requirements # 3) Software Verification and Validation Software documentation consistent with Basic Documentation was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. # VIII. Conclusions Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device Led and Laser Hair Growth Device is as safe, as effective, and performs as well as the legally marketed predicate devices. Page 5 of 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...