The EXG PulseLinQ Cables are a reusable electrode cable system intended to transmit signals from EXG electrodes to compatible ECG monitors/defibrillators for diagnostic and monitoring purposes. The system is intended for use in outdoor and indoor emergency care settings.
Device Story
EXG PulseLinQ Cables are reusable electrode cable systems; transmit physiological electrical signals from EXG electrodes to compatible ECG monitors/defibrillators. Used in indoor and outdoor emergency care settings by clinicians. Device acts as a passive conduit; facilitates secure electrical connection between patient-applied electrodes and monitoring equipment. Output is raw ECG signal displayed on monitor for diagnostic/monitoring purposes; assists clinicians in patient assessment and clinical decision-making. Benefits include reliable signal transmission for patient monitoring in emergency environments.
Clinical Evidence
No clinical data. Bench testing only, including compliance with ANSI/AAMI EC53:2013/(R) 2020, IEC 60601-1, and IEC 60601-1-2.
Indicated for patients of all ages (infants, children, adolescents, and adults) requiring ECG monitoring in indoor or outdoor emergency care settings.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 30, 2026
CB Innovations, LLC
Oxana Pantchenko
Director, R&D and Regulatory
5835 Avenida Encinas, Suite 118
Carlsbad, California 92008
Re: K262816
Trade/Device Name: EXG PulseLinQ Cables (EXGPL)
Regulation Number: 21 CFR 870.2900
Regulation Name: Patient Transducer And Electrode Cable (Including Connector)
Regulatory Class: Class II
Product Code: DSA
Dated: October 11, 2026
Received: October 13, 2026
Dear Oxana Pantchenko:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262816 - Oxana Pantchenko
Page 2
may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Office of Industry and Consumer Education (OICE) to ask a question about a specific regulatory topic. See the OICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance)
{3}
K262816 - Oxana Pantchenko
Page 3
assistance/contact-us-office-industry-and-consumer-education-oice) for more information or contact OICE by email (AskOICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K262816 ? |
| Please provide the device trade name(s). | | ? |
| EXG PulseLinQ Cables (EXGPL) | | |
| Please provide your Indications for Use below. | | ? |
| The EXG PulseLinQ Cables are a reusable electrode cable system intended to transmit signals from EXG electrodes to compatible ECG monitors/defibrillators for diagnostic and monitoring purposes. The system is intended for use in outdoor and indoor emergency care settings. | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
K262816
510(k) Summary
CBI
C-BOOTH INNOVATIONS
### 510(k) Summary
Date prepared: October 11, 2026
### Manufacturer and 510(k) Premarket Notification Owner
CB Innovations, LLC
5835 Avenida Encinas STE 118
Carlsbad, CA 92008
Official Contact:
Oxana Pantchenko, Ph.D.
Director of R&D and Regulatory
(650)269-4915
### Device Information
Name: EXG PulseLinQ Cable System
Regulation Number: 21 CFR 870.2900 – Patient transducer and electrode cable (including connector)
Class: II
Product Code: DSA - Cable, Transducer and Electrode, Patient, (including connector)
Medical Specialty: Cardiovascular
### Predicate/Device Identification
K240377 LIFEPAK® ECG Cables by Stryker Physio-Control
### Device Description
The following cables are available:
- EXGPLZ: Compatible with ZOLL X Series & Propaq MD/M
- EXGPL15: Compatible with STRYKER LIFEPAK 15
- EXGPL12L: Connects EXGPL15 or EXGPLZ to EXG electrodes
- EXG3L: Connects EXGPL15 or EXGPLZ with the EXGPL12L
### Indications for Use
The EXG PulseLinQ Cables are a reusable electrode cable system intended to transmit signals from EXG electrodes to compatible ECG monitors/defibrillators for diagnostic and monitoring purposes. The system is intended for use in outdoor and indoor emergency care settings.
### Comparison of Technological Characteristics
The proposed and predicate devices share the same fundamental technological characteristics. Both ECG cable systems transmit physiological signals from patient electrodes to an ECG monitor/defibrillator for patient monitoring. The EXG PulseLinQ Cables are substantially equivalent to the LIFEPAK ECG Cables with respect to intended use, safety, effectiveness, and performance.
Please see the device comparison table for additional details.
| | Subject Device | Predicate |
| --- | --- | --- |
| Device Name | EXG PulseLinQ Cables | LIFEPAK ECG Cables |
| 510(k) clearance | K262816 | K240377 |
| Regulation Number | 21 CFR 870.2900 | 21 CFR 870.2900 |
| Product Class | II | II |
| Product Code | DSA | DSA |
| Prescription Use | Rx | Rx |
| Indications for Use / Intended Use | The EXG PulseLinQ Cables are a reusable electrode cable system intended to transmit signals from EXG electrodes to compatible ECG monitors/defibrillators for diagnostic | The LIFEPAK ECG cables are reusable electrode cable systems intended to transmit signals from patient |
Page 1 of 3
{6}
K262816
510(k) Summary
C-BOOTH INNOVATIONS
| | and monitoring purposes. The system is intended for use in outdoor and indoor emergency care settings. | electrodes to the LIFEPAK monitor/defibrillator for monitoring purposes. |
| --- | --- | --- |
| Intended Environment | Indoor and outdoor emergency care settings. | Indoor and outdoor emergency care settings. |
| Patient Population | Any patient requiring ECG monitoring, according to the monitor's intended use. | Any patient requiring ECG monitoring, according to the monitor's intended use. |
| Anatomical Sites | Attaches to EXG electrodes placed at standard specified locations on chest or extremities. | Attached to electrodes placed at standard specified locations on chest or extremities. |
| Usage | Reusable | Reusable |
| Sterile | Not sterile | Not sterile |
| ECG Monitoring | 1, 3, 5, 6 and 12-lead | 3, 5, 6, 12 or 15-lead |
| Cable Materials | TPU, TPE, PBT | PU, PVC, TPU |
| Cable Length | 9 ft | 5 ft, 8 ft |
| Compatibility with monitors/defibrillators | STRYKER LIFEPAK 15, ZOLL X Series & Propaq MD/M | LIFEPAK 35 monitor/defibrillator |
The subject and predicate devices are based on the same fundamental technological principles and share comparable designs. Specifically:
1. Both devices passively conduct electrical signals from ECG electrodes to compatible ECG monitoring equipment.
2. Both devices incorporate interfaces and connectors that facilitate secure connections between ECG electrodes and ECG monitoring devices.
3. Both devices are constructed from similar core components, including connector plug-ins, contact pins, pre-molded and over-molded components, and insulated cable assemblies.
4. Both devices utilize similar cable insulation and jacket materials designed to provide flexibility, durability, and electrical isolation during use.
The following technological differences exist between the subject and predicate devices:
1. The subject device is designed for use exclusively with EXG disposable electrodes, whereas the predicate device is compatible with traditional ECG electrodes.
2. The predicate device supports multiple ECG lead configurations, including 3, 4, 6, 12, and 15-lead cable systems, while the subject device is limited to 1, 3, 5, 6 and 12-lead configuration.
3. The subject device incorporates a 9-foot cable assembly, whereas the predicate device is available with cable lengths of 5 feet and 8 feet.
4. The ECG monitor/defibrillator interface connectors differ between the subject and predicate devices, reflecting their compatibility with different ECG monitoring and defibrillation platforms.
While minor differences exist between the two devices, these differences do not alter the intended use of the subject device. Both devices share the same fundamental technological characteristics. Furthermore, the identified differences, together with the results of successfully completed testing, do not affect the safety and effectiveness of the subject device. Nor do they raise any new or novel questions of safety and effectiveness compared to those associated with the predicate device.
### Summary of Non-Clinical Testing
The following testing has been successfully completed to demonstrate that the proposed EXG PulseLinQ Cable System meets the safety and performance requirements. Comprehensive testing included the following:
1) ANSI/AAMI EC53:2013/(R) 2020
2) IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020
3) IEC60601-1-2:2014/A1:2020
### Summary of Clinical Testing
Clinical testing was determined not applicable for the EXG PulseLinQ Cable System.
Page 2 of 3
{7}
K262816
510(k) Summary
[LOGO]
C-BOOTH INNOVATIONS
## Conclusion
Based on the testing performed, it can be concluded that the EXG PulseLinQ Cable System does not raise new questions of safety or effectiveness when compared to the predicate device. The similar indications for use/intended use, technological characteristics, and performance characteristics for the proposed EXG PulseLinQ Cable System are assessed to be substantially equivalent to the predicate device.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.