← Product Code [OCC](/productcode/OCC) · K262747

# ID NOW Influenza A & B 2; ID NOW RSV; ID NOW Strep A 2; ID NOW COVID-19 2.0 (K262747)

_Abbott Diagnostics Scarborough, Inc. · OCC · Aug 31, 2026 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K262747

## Device Facts

- **Applicant:** Abbott Diagnostics Scarborough, Inc.
- **Product Code:** [OCC](/productcode/OCC.md)
- **Decision Date:** Aug 31, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.3980
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Pediatric

## Indications for Use

ID NOW Influenza A & B 2: The ID NOW Influenza A & B 2 assay performed on the ID NOW Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in direct nasal or nasopharyngeal swabs and nasal or nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Performance characteristics for influenza A were established during the 2016-2017 influenza season when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. ID NOW RSV: The ID NOW RSV assay performed on the ID NOW Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection of respiratory syncytial virus (RSV) viral RNA in direct nasopharyngeal swabs and nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the diagnosis of RSV in children <18 years and adults ≥60 years in conjunction with clinical and epidemiological risk factors. ID NOW Strep A 2: ID NOW Strep A 2 is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections. ID NOW COVID-19 2.0: ID NOW COVID-19 2.0 performed on the ID NOW Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology (NAAT) intended for the qualitative detection of nucleic acid from SARS-CoV-2 in direct anterior nasal (nasal) or nasopharyngeal swabs from individuals with signs and symptoms of respiratory tract infection. ID NOW COVID-19 2.0 performed on the ID NOW Instrument is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. SARS-CoV-2 RNA is generally detectable in nasal and nasopharyngeal swab specimens during the acute phase of infection. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not preclude coinfection with bacteria or other viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. A negative test result is presumptive, and it is recommended these results be confirmed by another molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. This test is intended for prescription use only and can be used in Point-of-Care settings.

## Device Story

ID NOW system performs rapid, instrument-based isothermal nucleic acid amplification (NAAT) for respiratory pathogens; utilizes disposable Sample Receiver, Test Base, and Transfer Cartridge; instrument heats, mixes, and detects fluorescence; provides automated qualitative results; used in point-of-care or clinical labs by professionals; aids diagnosis alongside clinical/epidemiological factors; reduces contamination risk via closed-system Test Base.

## Clinical Evidence

No new performance data were reviewed; substantial equivalence based on design control activities and verification of software modifications.

## Technological Characteristics

Isothermal nucleic acid amplification; fluorescence-based detection; automated instrument-based processing; closed-system Test Base; qualitative output; professional use; point-of-care or laboratory environment.

## Regulatory Identification

A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.

## Special Controls

*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.

## Predicate Devices

- ID NOW Influenza A & B 2 ([K262747](/device/K262747.md))
- ID NOW RSV ([K262747](/device/K262747.md))
- ID NOW Strep A 2 ([K262747](/device/K262747.md))
- ID NOW COVID-19 2.0 ([K262747](/device/K262747.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

[LOGO]

FDA

U.S. FOOD & DRUG

ADMINISTRATION

# SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

# I Background Information:

A 510(k) Number

K262747

B Applicant

Abbott Diagnostics Scarborough, Inc.

C Proprietary and Established Names

ID NOW Influenza A & B 2; ID NOW RSV; ID NOW Strep A 2; ID NOW COVID-19 2.0

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  OCC | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology  |
|  OOI | Class II | 21 CFR 862.2570 - Instrumentation for clinical multiplex test systems | CH - Clinical Chemistry  |
|  OZE | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology  |
|  PGX | Class II | 21 CFR 866.2680 - Streptococcus spp. nucleic acid-based assay | MI - Microbiology  |
|  QWR | Class II | 21 CFR 866.3982 - Simple Point-Of-Care Device To Directly Detect SARS CoV-2 Viral Targets From Clinical Specimens In Near-Patient Settings | MI - Microbiology  |

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

{1}

## II Review Summary:

This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring 510(k). The following items are present and acceptable.

1. The name and 510(k) number of the **SUBMITTER'S** previously cleared device.
2. Submitter's statement that the **INDICATIONS FOR USE/INTENDED USE** of the modified device as described in its labeling **HAS NOT CHANGED** along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device **has not changed**. **This change was for updates to the instrument software that include an operating system update, cybersecurity updates, and connectivity improvements. No new performance data were reviewed.**
4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics.
5. A Design Control Activities Summary which includes:
   a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
   b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

K262747 - Page 2 of 2

---

**Source:** [https://fda.innolitics.com/device/K262747](https://fda.innolitics.com/device/K262747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com