TeKBrace Solo Soft Tissue Reinforcement Device

K262729 · Theramicro · QUW · Aug 18, 2026 · General, Plastic Surgery

Device Facts

Record IDK262729
Device NameTeKBrace Solo Soft Tissue Reinforcement Device
ApplicantTheramicro
Product CodeQUW · General, Plastic Surgery
Decision DateAug 18, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TeKBrace Solo Soft Tissue Reinforcement Device is a single use device intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. The TeKBrace Solo Soft Tissue Reinforcement Device is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the repair.

Device Story

TeKBrace Solo is a non-resorbable, single-use surgical mesh device; central tubular weave tape design tapering to cords at both ends. Used by surgeons during tendon and ligament repair procedures to reinforce soft tissue; does not replace native structure or provide primary mechanical strength; primary strength provided by sutures/fixation devices. Subject device adds crimped needles to implant cords to assist surgical technique; needles are not implanted. Device manufactured from polyethylene terephthalate (PET).

Clinical Evidence

Bench testing only; no clinical data provided. Performance testing from predicate device (K261621) deemed representative of worst-case conditions for subject device.

Technological Characteristics

Material: Polyethylene terephthalate (PET). Design: Central tubular weave tape with tapered cords. Sterilization: Ethylene Oxide (EO). Connectivity: None. Software: None.

Indications for Use

Indicated for adults requiring soft tissue reinforcement during tendon and ligament repair surgery, including rotator cuff, patellar, Achilles, biceps, quadriceps, and various collateral/spring/deltoid ligaments.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION August 18, 2026 Theramicro Christine Scifert Head of RA/QA 51 Germantown Ct. Suite 200 Cordova, Tennessee 38018 Re: K262729 Trade/Device Name: TeKBrace Solo Soft Tissue Reinforcement Device Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: QUW, FTL, OWX Dated: July 31, 2026 Received: July 31, 2026 Dear Christine Scifert: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K262729 - Christine Scifert Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K262729 - Christine Scifert Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K262729 ? | | Please provide the device trade name(s). | | ? | | TeKBrace Solo Soft Tissue Reinforcement Device | | | | Please provide your Indications for Use below. | | ? | | The TeKBrace Solo Soft Tissue Reinforcement Device is a single use device intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. | | | | The TeKBrace Solo Soft Tissue Reinforcement Device is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the repair. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? | K262729 Page 1 of 1 {5} # 510(k) Summary # TeKBrace Solo Soft Tissue Reinforcement Device # 30 July 2026 Company: TheraMicro 51 Germantown Court, Suite 200 Cordova, TN 38018 Company Contact: Christine Scifert Head of RA/QA TheraMicro Phone: (901) 831-8053 Email: cscifert@theramicro.com Trade Name: TeKBrace Solo Soft Tissue Reinforcement Device Common Name: mesh, surgical, non-resorbable, orthopedics, reinforcement of ligament Classification: Class II Regulation: 21 CFR 878.3300 (Surgical mesh) Panel: General and Plastic Surgery Product Code: QUW, FTL, OWX Primary Predicate: TheraMicro TeKBrace Solo Soft Tissue Reinforcement Device – K261621 # Device Description: The TheraMicro TeKBrace Solo Soft Tissue Reinforcement Device is a single use device intended to be used for reinforcement of soft tissue. The product is offered in a central tubular weave tape that tapers down to cords at both ends. The implant is available in multiple size configurations. This submission introduces additional device configurations that vary in length and width. The product is offered in a central tubular weave tape that tapers down to cords at both ends. The product is substantially equivalent in material and manufacturing compared to the TeKBrace Solo Soft Tissue Reinforcement Device (K251063 & K261621). The TeKBrace solo configurations included in this submission are similar to the existing TheraMicro TeKBrace Solo device in intended use, material and design, including a central tubular section tapering down to cords at the end. The devices in this submission differ from the previously cleared devices in that the devices will come with needles crimped to the implant cords and will be sterilized via Ethylene Oxide. The needles are used during the surgical technique and are not implanted K262729 Page 1 of 3 {6} TeKBrace Solo is manufactured from non-absorbable polyethylene terephthalate (PET), commercially known as polyester. ## Indications for Use: The TeKBrace Solo Soft Tissue Reinforcement Device is a single use device intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. The TeKBrace Solo Soft Tissue Reinforcement Device is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the repair. ## Substantial Equivalence: The subject TeKBrace Solo Soft Tissue Reinforcement Device is substantially equivalent to the following predicate devices: ### Primary Predicate: TheraMicro TeKBrace Solo Soft Tissue Reinforcement Device – K261621 ### Additional Predicate Device: TheraMicro TeKBrace Solo Soft Tissue Reinforcement Device – K251063 The subject TeKBrace Solo Soft Tissue Reinforcement Device is substantially equivalent to the predicate TeKBrace Solo Soft Tissue Reinforcement Device (K261621 & K251063) in materials, manufacturing, and indications for use. The subject components are substantially equivalent to the primary predicate in indications, manufacturing, and materials. TeKBrace Solo is similar to predicate in design with the only difference being the addition of a needle on the ends, which is only used for the surgical technique and is not implanted. Additionally, the subject device is provided sterile via EO sterilization rather than gamma irradiation. These insignificant differences do not introduce a new worst case condition and therefore, the subject is substantially equivalent to the predicate device. ## Performance Testing: The changes to the predicate resulting in the subject device do not introduce a new worst case condition. Therefore, testing performed on the predicate device (K261621) remains representative of K262729 Page 2 of 3 {7} the worst case condition. The change in sterilization method is not expected to negatively impact device performance. Therefore, no additional performance testing is needed. # **Conclusion:** The subject device is determined to be substantially equivalent to the predicate devices. K262729 Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...