The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults.
Device Story
Automatic digital blood pressure monitor (LD-526) for non-invasive measurement of systolic/diastolic blood pressure and pulse rate in adults. Device consists of main unit (pump, valve, PCB, LCD) and arm cuff (22-42cm). Operates via oscillometric method; pump inflates cuff; sensor detects arterial pressure fluctuations; software algorithm analyzes pulse waveform data to calculate blood pressure. Used in home healthcare environments; operated by patient. Output displayed on LCD; assists users in monitoring cardiovascular health. Power via 4 AAA batteries or Type-C AC adapter.
Clinical Evidence
No new clinical studies performed. Substantial equivalence supported by clinical data from predicate device LD-537 (K170466) based on identical intended use, patient population, measurement principle, and accuracy specifications (±3mmHg static pressure, ±5% pulse rate).
Technological Characteristics
Oscillometric measurement; automatic inflation/deflation. Materials: cuff materials compliant with ISO 10993-5, 10993-10, 10993-23. Power: 4 AAA batteries or Type-C AC adapter. Standards: ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-30. Connectivity: None (Type-C port for power only).
Indications for Use
Indicated for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults (22 years and older).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 31, 2026
Honsun (Nantong) Co., Ltd.
Jian Shen
Regulatory Affairs Director
#8 Tongxing Rd., Economic & Technological, Development Area
Nantong, Jiangsu 226009
China
Re: K262658
Trade/Device Name: Automatic Digital Blood Pressure Monitor (LD-526)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: July 30, 2026
Received: July 30, 2026
Dear Jian Shen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262658 - Jian Shen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K262658 - Jian Shen
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN C. BROWNING -S
CDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | ? |
| Please provide the device trade name(s). | | ? |
| Automatic Digital Blood Pressure monitor (LD-526) | | |
| Please provide your Indications for Use below. | | ? |
| The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
K262658
Page 1 of 1
{4}
Sponsor: HONSUN (NANTONG) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
# Summary of special 510(K)
Date: 8-May-2026
## 1. Submitter
| Company: | HONSUN(NANTONG)Co., Ltd. |
| --- | --- |
| Establishment registration number: | 3007740533 |
| Address: | No.8, Tongxing Road, Nantong Economic & Technological development Area, Jiangsu, 226009, China |
| Phone: | +86-513-80580116 |
| Fax: | +86-513-80580080 |
| Contact Person: | Jian Shen |
| Email: | Jason@lordmed.com |
## 2. Modified Device information
| Name of Device: | Automatic Digital Blood Pressure Monitor |
| --- | --- |
| Model: | LD-526 |
| Common Name: | Non-invasive blood pressure measurement system |
| Classification Name: | System, Measurement, Blood-Pressure, Non-Invasive |
| Regulation number: | 21 CFR 870.1130 |
| Regulatory Class: | II |
| Product Code: | DXN |
| Review Panel: | Cardiovascular |
## 3. Cleared Device information
| Name of Device: | Automatic Digital Blood Pressure Monitor |
| --- | --- |
| Model: | LD-537 |
| 510k number: | K170466 |
| Common Name: | Non-invasive blood pressure measurement system |
| Regulation number: | 21 CFR 870.1130 |
| Regulatory Class: | II |
| Product Code: | DXN |
| Review Panel: | Cardiovascular |
## 4. Device description
The Automatic Digital Blood Pressure Monitor is an automatic, non-invasive, blood pressure measurement device that is intended to measure the systolic and diastolic arterial blood pressure and pulse rate. The systolic and diastolic pressure are determined using the oscillometric method, where the cuff is inflated with a pump and deflates via an automatic electronic valve. During the inflation measurements, a pump within the main unit slowly inflates the cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the sensor
1 / 6
K262658
Page 1 of 6
{5}
Sponsor: HONSUN (NANTONG) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
to determine systolic pressure and diastolic pressure.
The Automatic Digital Blood Pressure Monitor consists of two parts: main unit and cuff. The main unit is mainly composed of pump, valve, PCB, enclosure and LCD. The cuff, which is applicable to arm circumference approximately between 22 and 42cm, includes the inflatable bladder and the shell.
## 5. Intended use and indications for use
The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults.
## 6. Operation principle
This device adopts the oscillometric technology with Fuzzy Algorithm to measure the arterial blood pressure and pulse rate. The cuff is wrapped around the arm and automatically inflated by the air pump. The sensor of the device catches weak fluctuation of the pressure in the cuff produced by extension and contraction of the artery of the arm in response to each heartbeat. The amplitude of the pressure waves is measured, converted in millimeters of the mercury column, and is displayed by digital value.
7. Comparisons of technological characteristic with the predicate device
| Items | Subject Device LD-526 | Predicate Device LD-537 | Remark |
| --- | --- | --- | --- |
| Device name | Automatic Digital Blood Pressure Monitor | Automatic Blood Pressure Monitor | Same |
| Classification | II | II | Same |
| Product code | DXN | DXN | Same |
| Regulation number | 870.1130 | 870.1130 | Same |
| Indications for use | The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults | The Scian Automatic Digital Blood Pressure Monitor LD-518&LD-537 are devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.Besides, they can detect the appearance of irregular heartbeat during measurement, and give a | Same |
2 / 6
K262658
Page 2 of 6
{6}
Sponsor: HONSUN (NANTONG) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
| | | warning signal once the irregular heartbeat is detected. | |
| --- | --- | --- | --- |
| Measuring method | Oscillometric Method | Oscillometric Method | Same |
| Cuff location | Upper arm | Upper arm | Same |
| Specification | | | |
| Measuring range | 40 to 180mmHg (DIA, diastolic pressure) 60 to 260mmHg (SYS, systolic pressure) 40 to 160 beats/minute (PUL, pulse rate) | 40 to 260 mmHg (blood pressure) 40 to 160 beats/minute (pulse rate) | Similar Note 1 |
| Measuring accuracy | ±3mmHg for static pressure ±5% of the reading for the pulse rate | ±3mmHg for static pressure ±5% of the reading for the pulse rate | Same |
| Inflation | Automatic by the pump | Automatic by the pump | Same |
| Rapid deflation | Automatic electronic valve | Automatic electronic valve | Same |
| Power source | Battery: 4 AAA*1.5V, AC Adapter (Type C) | Battery: 4 AA*1.5V, AC Adapter | Similar Note 2 |
| Cuff range | 22cm~32cm (Standard) 22cm~42cm(optional) | 22cm~42cm (Standard) 22cm~32cm(optional) | Same |
| Operation condition | 5°C to 40°C, 15% to 90%RH, 700hPa to 1060hPa | 10°C to 40°C, 15%~85%RH, 700hPa~1060hPa | Similar Note 3 |
| Transport and storage condition | -25°C to 70°C, 15% to 90%RH, 700hPa to 1060hPa | -20°C to 55°C, 85%RH and below, 500hPa to 1060hPa | Similar Note 4 |
| Main body dimension | 125(L)*96(W)*70(H)mm | 105(L)*120(W)*55(H)mm | Similar Note 5 |
| Weight | Approximately 220g without batteries | Approximately 310g without batteries | Similar Note 6 |
| Function | | | |
| Memory | 2*90 sets | 2*90 sets | Same |
| Date and time | Yes | Yes | Same |
| WHO | Yes | Yes | Same |
| Irregular heartbeat detector | Yes | Yes | Same |
| Voice broadcast | Yes | Yes | Same |
3 / 6
K262658
Page 3 of 6
{7}
Sponsor: HONSUN (NANTONG) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
| Applicable for standards | | | |
| --- | --- | --- | --- |
| Biocompatibility | All user directly contacting materials are compliance with ISO 10993-5 and ISO10993-10/ISO 10993-23 requirements | All user directly contacting materials are compliance with ISO 10993-5 and ISO10993-10/ISO 10993-23 requirements | Same |
| Electrical Safety | Comply with ANSI AAMI ES 60601-1 requirements | Comply with ANSI AAMI ES 60601-1 requirements | Same |
| EMC | Comply with IEC 60601-1-2 requirements | Comply with IEC 60601-1-2 requirements | Same |
| Others | Comply with IEC 60601-1-11 requirements | Comply with IEC 60601-1-11 requirements | Same |
# Difference Discussion:
# Note 1: Measuring range
The blood pressure measurement range of the predicate device is 40 to 260mmHg, while the measurement range of the proposed device under review has been further subdivided into 40 to 180mmHg (DIA diastolic pressure) and 60 to 260mmHg (SYS, systolic pressure). The actual measurement ranges for both are identical. Therefore, the difference is considered not raise any issues in safety and effectiveness.
# Note 2: Power source
The charging method of the subject device is same as predicate device. While the specification of batteries and adaptors are different. The charging interface of the proposed device has been changed to Type-C interface (Note: Type-C port is limited to power delivery only and has no actual communication functionality). And the battery of the target device has been changed to 4* AAA batteries. The test results in IEC 60601-1 demonstrated that the proposed device is safety and the performance satisfied the specification. The difference will not raise new issue of safety and effectiveness.
# Note 3: Operation condition
The operation condition of the subject device is different with predicate device LD-537 (K170466), the difference introduces risks tested and evaluated by verifying test in accordance with IEC 60601-1-11 and ANSI AAMI ES 60601-1 provided in this submission, therefore the difference will not raise new issues of safety and effectiveness.
# Note 4: Transport and storage condition
The transport and storage condition of the subject device are different with predicate device LD-537 (K170466), the difference introduces risks tested and evaluated by verifying test in accordance with IEC 60601-1-11 and ANSI AAMI ES 60601-1 provided in this submission, therefore the difference will not raise new issues of safety and effectiveness.
4 / 6
K262658
Page 4 of 6
{8}
Sponsor: HONSUN (NANTONG) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
#### Note 5: Main body dimension
As for LD-526, its dimension is different with predicate device LD-537 (K170466) will not affect the safety and effectiveness.
#### Note 6: Weight
The weight of subject device is different with predicate device LD-537 (K170466) will not affect the safety and effectiveness.
## Conclusion
Based on above comparative analysis, compared with the predicate device LD-537 (K170466), the subject device will not raise new issues of safety and effectiveness. Thus, the subject device is as safe & effective and substantially equivalent to the predicate device.
## 8. Assessment of Non-clinical testing
### 8.1 Electrical safety and Electromagnetic Compatibility Testing
The electrical safety and EMC testing were conducted on the subject device. The related standards are shown as follows:
| Standard | Descriptions |
| --- | --- |
| ANSI AAMI ES 60601-1 | Medical electrical equipment-Part 1: General requirements for basic safety and essential performance |
| IEC 60601-1-2 | Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances -Requirements and tests |
| IEC 60601-1-11 | Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment |
| ISO 80601-2-30 | Medical electrical equipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers |
### 8.2 Biocompatibility Evaluation
The biocompatibility evaluations of the subject device were conducted in accordance with the International Standard ISO 10993-1 “Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing within a Risk Management Process” and FDA biocompatibility guidance, the Cytotoxicity test, Sensitization test, Irritation test are needed. The subject devices use the same patient-contact materials(cuff) with the predicate device, so the biocompatibility is not affected and it will not affect the effectiveness or raise a new safety issue.
### 8.3 Software Verification and Validation Evaluation
Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained In Medical
5 / 6
K262658
Page 5 of 6
{9}
Sponsor: HONSUN (NANTONG) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
Devices-Guidance for Industry and FDA Staff, May 2005.
## 9. Assessment of clinical testing
The subject device leverages the clinical testing data from its predicate device, model LD-537 (K170466). As demonstrated in the comparison table (Section 7), the subject device shares the same intended patient population (adults), identical blood pressure and pulse rate measurement ranges, and identical accuracy specifications ( \( \pm \) 3 mmHg for static pressure, \( \pm \) 5% of reading for pulse rate) as the predicate device. Both devices utilize the same fundamental oscillometric measurement principle and are intended for use in the same environments.
Given these core equivalencies in design, intended use, and performance specifications, the clinical data supporting the safety and effectiveness of the predicate device LD-537 is directly applicable and sufficient to support the substantial equivalence of the subject device model LD-526. The minor modifications identified in Discussions 1 through 6 are not related to clinical performance and do not affect the validity of the leveraged clinical data. Therefore, no new clinical studies are required for the subject device.
## 10. Conclusion
The subject devices have same intended use and similar characteristics as per the cleared predicate device. Moreover, bench testing contained in this submission supplied demonstrate that the differences between subject device and predicate device will not raise any safety and effectiveness issue.
The nonclinical tests support the safety of the device and the hardware and software verification and validation, demonstrate that the subject device performs as intended in the specified use conditions are same with predicate device. The performance tests demonstrate that the subject device performs comparably to the predicate device that is currently marketed for the same intended use. Thus, the subject device is substantially equivalent to the predicate device.
6 / 6
K262658
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.