# Fractional CO2 Laser Therapy System (Z121) (K262627)

_Sanhe PERFECTLASER Technology Co., Ltd. · GEX · Oct 26, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K262627

## Device Facts

- **Applicant:** Sanhe PERFECTLASER Technology Co., Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Oct 26, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology, plastic surgery and general surgery. Fractional functionality is intended for skin resurfacing only.

## Device Story

Fractional CO2 Laser Therapy System utilizes 10.6μm CO2 laser energy to perform soft tissue ablation, vaporization, excision, and coagulation. Device features fractional and normal operational modes (continuous, single pulse, repetitive pulse). User sets laser energy density via LCD touchscreen; system delivers energy through an articulated arm with counterweight. Aiming beam (red diode laser) assists targeting. Used in dermatology, plastic surgery, and general surgery settings by clinicians. Thermal effect reaches dermis to vaporize tissue, stimulating collagen proliferation and tissue repair. Output parameters include pulse energy (1-300mj/point) and adjustable spot density. Device benefits patients by facilitating controlled skin resurfacing and surgical tissue management.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), laser safety (IEC 60825-1), surgical laser performance (IEC 60601-2-22), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, -10, -23).

## Technological Characteristics

CO2 laser (10.6μm), max 35W output. Air-cooled. Articulated arm delivery. LCD touchscreen/footswitch controls. Electrical: 100V-230V, 50/60Hz. Standards: IEC 60601-1 (safety), IEC 60825-1 (laser), IEC 60601-2-22 (surgical laser), IEC 60601-1-2 (EMC), ISO 10993 (biocompatibility).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Fractional CO2 Laser Therapy System (FG900-S); Fractional CO2 Laser Therapy System (FG900-S2) ([K241670](/device/K241670.md))

## Reference Devices

- EdgeOne CO2 Laser ([K162169](/device/K162169.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

FDA U.S. FOOD & DRUG
ADMINISTRATION

October 26, 2026

Sanhe PERFECTLASER Technology Co., Ltd.

% Ray Wang

General Manager

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Bldg. #15, Xiyuehui, #5, Yihe N. Rd.

Fangshan District

Beijing, 102401

China

Re: K262627

Trade/Device Name: Fractional CO2 Laser Therapy System (Z121)

Regulation Number: 21 CFR 878.4810

Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology

Regulatory Class: Class II

Product Code: GEX

Dated: July 29, 2026

Received: July 29, 2026

Dear Ray Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993

www.fda.gov

{2}

K262627 - Ray Wang

Page 2

Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

{3}

K262627 - Ray Wang

Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L.
HITHE -S

Digitally signed by
TANISHA L. HITHE -S
Date: 2026.10.26 22:47:19
-04'00'

Tanisha Hithe

Assistant Director

DHT4A: Division of General Surgery Devices

OHT4: Office of Surgical and

Infection Control Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{4}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K262627

Device Name
Fractional CO2 Laser Therapy System (Z121)

Indications for Use (Describe)
Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology, plastic surgery and general surgery. Fractional functionality is intended for skin resurfacing only.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{5}

510(k) Summary

## 510(k) Summary K262627

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

1. Date of Preparation: 10/01/2026
2. Sponsor Identification

**Sanhe PERFECTLASER Technology Co., Ltd.**

Room 1003, Building A1 (A-1), Enterprise Headquarters, Yanjiao Modern Service Industry Park
Pilot Zone, South Side of Nancheng Street, East Side of Shuichang South Road, Yanjiao High-tech
Zone, Sanhe City, Langfang City, Hebei Province, China

Contact Person: Yongchao Lian

Position: Certification Engineer

Tel: +86-15201170689

Email: 876010234@qq.com

3. Designated Submission Correspondent

Mr. Ray Wang

**Beijing Believe-Med Technology Service Co., Ltd.**

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401,
China

Tel: +86-18910677558

Fax: +86-10-56335780

Email: information@believe-med.com

1

{6}

510(k) Summary

4. Identification of Proposed Device

Trade Name: Fractional CO2 Laser Therapy System

Common Name: Powered Laser Surgical Instrument (Z121)

Regulatory Information

Classification Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology

Classification: II

Product Code: GEX

Regulation Number: 878.4810

Review Panel: General & Plastic Surgery

5. Identification of Predicate Device(s)

510(k) Number: K241670

Product Name: Fractional CO2 Laser Therapy System (FG900-S); Fractional CO2 Laser Therapy System (FG900-S2)

Manufacturer: Beijing ADSS Development Co., Ltd.

Reference Device 1: K162169

Product Name: EdgeOne CO2 Laser

Manufacturer: Jeisys Medical Inc.

6. Device Description

Fractional CO2 laser therapy system uses the thermal effect of laser to directly reach the dermis, instantly vaporize the tissue in the affected area, and stimulate collagen proliferation, thereby promoting a series of skin reactions such as tissue repair and collagen reconstruction. The parameters such as laser energy density required for treatment are set by the user.

7. Indication For Use Statement:

Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology, plastic surgery and general surgery. Fractional functionality is intended for skin resurfacing only.

2

{7}

510(k) Summary

### 8. Substantially Equivalent (SE) Comparison

Tab 1 General Comparison

|  Item | Proposed Device | Predicated Device K241670 | Reference Device 1 K162169 | Remark  |
| --- | --- | --- | --- | --- |
|  Manufacturer | Sanhe PERFECTLASER Technology Co., Ltd. | Beijing ADSS Development Co.,Ltd. | Jeisys Medical Inc. | /  |
|  Device Name | Fractional CO2 laser therapy system | Fractional CO2 Laser Therapy System | EdgeOne CO2 Laser | /  |
|  Intended use | Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology,plastic surgery and general surgery.Fractional functionality is intended for skin resurfacing only. | Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology,plastic surgery and general surgery. | It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology(ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing. | Same  |
|  Laser Wavelength | 10.6μm | 10.6μm | 10.6μm | Same  |
|  Max. Output Power Watts | Max 35W | Max 35W | 30W | Same as K241670  |
|  Operational Mode | Fractional mode, and normal mode (Continuous output , Single pluse, Repetitive pulse ) | Fractional mode, and normal mode (Continuous output , Single pluse, Repetitive pulse ) | Fractional mode, normal mode (CW, Pulse, Single Pulse, Repeat, Group pulse, Ultra) | Same  |
|  Aiming beam | Red diode laser,Max | Red diode laser,Max | Diode laser(Red) 655 | Analysis1  |

3

{8}

510(k) Summary

|   | 3mW |   | 5mW |   | +/-10nm, Max 1mW |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Cooling | Air cooling |   | Air cooling |   | Water | Same as K241670  |
|  Beam Delivery Handpiece | Articulated arm with counterweight |   | Articulated arm with counterweight |   | Articulated arm | Same  |
|  Laser firing Controls | LCD color Touchscreen Footswitch |   | LCD color Touchscreen Footswitch |   | LCD touch screen | Same  |
|  Electrical Requirements | 100V-230V,50/60Hz |   | 230V,50/60HZ |   | 230V~, 50/60Hz, 500VA(max. laser output),100VA(standby) | Analysis2  |
|  Pulse Energy | 1-300mj/point |   | 1-300mj/dot |   | 1-300mJ | Same  |
|  Spot size | 0.12 mm |   | 0.12mm |   | 120um, 350um, 800um | Same as K241670  |
|  Spot Density (DPA)/ cm2 | 25, 36, 49, 64, 81, 100, 121, 144, 169, 196, 225, 256, 289, 324, 361, 400, 441, 484, 529, 784, 1024, 1521, 2025, 2500, 3025 dots. |   | 25, 36, 49, 64, 81, 100, 121, 144, 169, 196, 225, 256, 289, 324, 361, 400, 441, 484, 529, 784, 1024, 1521, 2025, 2500, 3025 dots. |   | Spot density 0-42.4% | Same as K241670  |
|  Scan Area Size | 2x2 mm-20x20 mm |   | 2x2mm-20x20mm |   | 15 x 15mm2 | Same as K241670  |
|  Pulse Duration (normal mode) | Continuous output | N/A | Continuous output | N/A | 1-1000ms | Analysis3  |
|   |  Single pulse | 1-10 ms | Single pulse | 1~375 ms  |   |   |
|   |  Repetitive pulse | Pulse width: 1-1000ms Delay time: 1-1000ms | Repetitive pulse | Pulse width: 0.1~375 ms Delay time: 1-500ms  |   |   |

# NOTE:

For fractional skin-resurfacing functionality, the recommended parameter combination is:

Pulse energy : < 70 mJ/dot;

Uniform spot density : < 500 dots/cm2 leading to skin coverage of ~5% or less;

Spot size : 0.12 mm.

4

{9}

510(k) Summary

## Analysis

Analysis 1: The aiming beam power of proposed device does not cause heat or energy hazard to the patients during the treatment procedure. Therefore, this difference will not affect safety and effectiveness of the proposed device.

Analysis 2: The electrical requirements for the proposed device is different from the predicate device. However, electrical safety and EMC test has been conducted on the proposed device and the test result show that the device can work normally under this electrical requirements. Therefore, this difference will not affect safety and effectiveness of the proposed device.

Analysis 3: In Single Pulse mode, the pulse width of the Proposed Device (1–10 ms) falls within the pulse width range of the Predicate Device (1–375 ms), ensuring no higher risk of thermal injury. For the pulse width difference in Repetitive Pulse mode, we have introduced Reference Device K162169. Under the same intended use, the pulse width range of the Reference Device (1–1000 ms) fully covers that of the Proposed Device, guaranteeing equivalent safety and effectiveness. Therefore, this difference will not affect the safety and effectiveness of the Proposed Device.

### 9. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1 Edition 3.2 2020-08, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification, and requirements
IEC 60601-2-22: 2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
ISO 10993-10 Fourth edition 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation
ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
IEC 60601-1-2 Edition 4.1 2020-09, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Test

### 10. Clinical Test Conclusion

No clinical study is included in this submission.

### 11. Conclusion

Based on the non-clinical tests performed, the subject device is as safe, as effective, and

5

{10}

510(k) Summary

performance as well as the legally marketed predicate device, Fractional CO2 Laser Therapy System cleared under K241670.

6

---

**Source:** [https://fda.innolitics.com/device/K262627](https://fda.innolitics.com/device/K262627)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).