K262627 · Sanhe PERFECTLASER Technology Co., Ltd. · GEX · Oct 26, 2026 · General, Plastic Surgery
Device Facts
Record ID
K262627
Device Name
Fractional CO2 Laser Therapy System (Z121)
Applicant
Sanhe PERFECTLASER Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology, plastic surgery and general surgery. Fractional functionality is intended for skin resurfacing only.
Device Story
Fractional CO2 Laser Therapy System utilizes 10.6μm CO2 laser energy to perform soft tissue ablation, vaporization, excision, and coagulation. Device features fractional and normal operational modes (continuous, single pulse, repetitive pulse). User sets laser energy density via LCD touchscreen; system delivers energy through an articulated arm with counterweight. Aiming beam (red diode laser) assists targeting. Used in dermatology, plastic surgery, and general surgery settings by clinicians. Thermal effect reaches dermis to vaporize tissue, stimulating collagen proliferation and tissue repair. Output parameters include pulse energy (1-300mj/point) and adjustable spot density. Device benefits patients by facilitating controlled skin resurfacing and surgical tissue management.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), laser safety (IEC 60825-1), surgical laser performance (IEC 60601-2-22), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, -10, -23).
Technological Characteristics
CO2 laser (10.6μm), max 35W output. Air-cooled. Articulated arm delivery. LCD touchscreen/footswitch controls. Electrical: 100V-230V, 50/60Hz. Standards: IEC 60601-1 (safety), IEC 60825-1 (laser), IEC 60601-2-22 (surgical laser), IEC 60601-1-2 (EMC), ISO 10993 (biocompatibility).
Indications for Use
Indicated for body soft tissue ablation, vaporization, excision, and coagulation in dermatology, plastic surgery, and general surgery; fractional mode indicated for skin resurfacing.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Fractional CO2 Laser Therapy System (FG900-S); Fractional CO2 Laser Therapy System (FG900-S2) (K241670)
{1}
FDA U.S. FOOD & DRUG
ADMINISTRATION
October 26, 2026
Sanhe PERFECTLASER Technology Co., Ltd.
% Ray Wang
General Manager
Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Bldg. #15, Xiyuehui, #5, Yihe N. Rd.
Fangshan District
Beijing, 102401
China
Re: K262627
Trade/Device Name: Fractional CO2 Laser Therapy System (Z121)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: July 29, 2026
Received: July 29, 2026
Dear Ray Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262627 - Ray Wang
Page 2
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{3}
K262627 - Ray Wang
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L.
HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.10.26 22:47:19
-04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K262627
Device Name
Fractional CO2 Laser Therapy System (Z121)
Indications for Use (Describe)
Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology, plastic surgery and general surgery. Fractional functionality is intended for skin resurfacing only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
510(k) Summary
## 510(k) Summary K262627
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
1. Date of Preparation: 10/01/2026
2. Sponsor Identification
**Sanhe PERFECTLASER Technology Co., Ltd.**
Room 1003, Building A1 (A-1), Enterprise Headquarters, Yanjiao Modern Service Industry Park
Pilot Zone, South Side of Nancheng Street, East Side of Shuichang South Road, Yanjiao High-tech
Zone, Sanhe City, Langfang City, Hebei Province, China
Contact Person: Yongchao Lian
Position: Certification Engineer
Tel: +86-15201170689
Email: 876010234@qq.com
3. Designated Submission Correspondent
Mr. Ray Wang
**Beijing Believe-Med Technology Service Co., Ltd.**
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401,
China
Tel: +86-18910677558
Fax: +86-10-56335780
Email: information@believe-med.com
1
{6}
510(k) Summary
4. Identification of Proposed Device
Trade Name: Fractional CO2 Laser Therapy System
Common Name: Powered Laser Surgical Instrument (Z121)
Regulatory Information
Classification Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Classification: II
Product Code: GEX
Regulation Number: 878.4810
Review Panel: General & Plastic Surgery
5. Identification of Predicate Device(s)
510(k) Number: K241670
Product Name: Fractional CO2 Laser Therapy System (FG900-S); Fractional CO2 Laser Therapy System (FG900-S2)
Manufacturer: Beijing ADSS Development Co., Ltd.
Reference Device 1: K162169
Product Name: EdgeOne CO2 Laser
Manufacturer: Jeisys Medical Inc.
6. Device Description
Fractional CO2 laser therapy system uses the thermal effect of laser to directly reach the dermis, instantly vaporize the tissue in the affected area, and stimulate collagen proliferation, thereby promoting a series of skin reactions such as tissue repair and collagen reconstruction. The parameters such as laser energy density required for treatment are set by the user.
7. Indication For Use Statement:
Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology, plastic surgery and general surgery. Fractional functionality is intended for skin resurfacing only.
2
{7}
510(k) Summary
### 8. Substantially Equivalent (SE) Comparison
Tab 1 General Comparison
| Item | Proposed Device | Predicated Device K241670 | Reference Device 1 K162169 | Remark |
| --- | --- | --- | --- | --- |
| Manufacturer | Sanhe PERFECTLASER Technology Co., Ltd. | Beijing ADSS Development Co.,Ltd. | Jeisys Medical Inc. | / |
| Device Name | Fractional CO2 laser therapy system | Fractional CO2 Laser Therapy System | EdgeOne CO2 Laser | / |
| Intended use | Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology,plastic surgery and general surgery.Fractional functionality is intended for skin resurfacing only. | Fractional CO2 Laser Therapy System is used for body soft tissue ablation, vaporization, excision and coagulation in dermatology,plastic surgery and general surgery. | It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology(ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing. | Same |
| Laser Wavelength | 10.6μm | 10.6μm | 10.6μm | Same |
| Max. Output Power Watts | Max 35W | Max 35W | 30W | Same as K241670 |
| Operational Mode | Fractional mode, and normal mode (Continuous output , Single pluse, Repetitive pulse ) | Fractional mode, and normal mode (Continuous output , Single pluse, Repetitive pulse ) | Fractional mode, normal mode (CW, Pulse, Single Pulse, Repeat, Group pulse, Ultra) | Same |
| Aiming beam | Red diode laser,Max | Red diode laser,Max | Diode laser(Red) 655 | Analysis1 |
3
{8}
510(k) Summary
| | 3mW | | 5mW | | +/-10nm, Max 1mW | |
| --- | --- | --- | --- | --- | --- | --- |
| Cooling | Air cooling | | Air cooling | | Water | Same as K241670 |
| Beam Delivery Handpiece | Articulated arm with counterweight | | Articulated arm with counterweight | | Articulated arm | Same |
| Laser firing Controls | LCD color Touchscreen Footswitch | | LCD color Touchscreen Footswitch | | LCD touch screen | Same |
| Electrical Requirements | 100V-230V,50/60Hz | | 230V,50/60HZ | | 230V~, 50/60Hz, 500VA(max. laser output),100VA(standby) | Analysis2 |
| Pulse Energy | 1-300mj/point | | 1-300mj/dot | | 1-300mJ | Same |
| Spot size | 0.12 mm | | 0.12mm | | 120um, 350um, 800um | Same as K241670 |
| Spot Density (DPA)/ cm2 | 25, 36, 49, 64, 81, 100, 121, 144, 169, 196, 225, 256, 289, 324, 361, 400, 441, 484, 529, 784, 1024, 1521, 2025, 2500, 3025 dots. | | 25, 36, 49, 64, 81, 100, 121, 144, 169, 196, 225, 256, 289, 324, 361, 400, 441, 484, 529, 784, 1024, 1521, 2025, 2500, 3025 dots. | | Spot density 0-42.4% | Same as K241670 |
| Scan Area Size | 2x2 mm-20x20 mm | | 2x2mm-20x20mm | | 15 x 15mm2 | Same as K241670 |
| Pulse Duration (normal mode) | Continuous output | N/A | Continuous output | N/A | 1-1000ms | Analysis3 |
| | Single pulse | 1-10 ms | Single pulse | 1~375 ms | | |
| | Repetitive pulse | Pulse width: 1-1000ms Delay time: 1-1000ms | Repetitive pulse | Pulse width: 0.1~375 ms Delay time: 1-500ms | | |
# NOTE:
For fractional skin-resurfacing functionality, the recommended parameter combination is:
Pulse energy : < 70 mJ/dot;
Uniform spot density : < 500 dots/cm2 leading to skin coverage of ~5% or less;
Spot size : 0.12 mm.
4
{9}
510(k) Summary
## Analysis
Analysis 1: The aiming beam power of proposed device does not cause heat or energy hazard to the patients during the treatment procedure. Therefore, this difference will not affect safety and effectiveness of the proposed device.
Analysis 2: The electrical requirements for the proposed device is different from the predicate device. However, electrical safety and EMC test has been conducted on the proposed device and the test result show that the device can work normally under this electrical requirements. Therefore, this difference will not affect safety and effectiveness of the proposed device.
Analysis 3: In Single Pulse mode, the pulse width of the Proposed Device (1–10 ms) falls within the pulse width range of the Predicate Device (1–375 ms), ensuring no higher risk of thermal injury. For the pulse width difference in Repetitive Pulse mode, we have introduced Reference Device K162169. Under the same intended use, the pulse width range of the Reference Device (1–1000 ms) fully covers that of the Proposed Device, guaranteeing equivalent safety and effectiveness. Therefore, this difference will not affect the safety and effectiveness of the Proposed Device.
### 9. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1 Edition 3.2 2020-08, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification, and requirements
IEC 60601-2-22: 2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
ISO 10993-10 Fourth edition 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation
ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
IEC 60601-1-2 Edition 4.1 2020-09, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Test
### 10. Clinical Test Conclusion
No clinical study is included in this submission.
### 11. Conclusion
Based on the non-clinical tests performed, the subject device is as safe, as effective, and
5
{10}
510(k) Summary
performance as well as the legally marketed predicate device, Fractional CO2 Laser Therapy System cleared under K241670.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.