The ND® Needle Guidance System is indicated for use in ultrasound-guided procedures where precise placement of common interventional devices is required, such as for vascular access.
Device Story
Sterile, single-use molded bracket and needle guide system; attaches to ultrasound probe to provide mechanical stability and directional alignment for needles/instruments. Features adjustable in-plane (10-35 mm) and out-of-plane (5-40 mm) configurations. Operates via geometric principles; translates ultrasound vessel depth measurements into specific access angles. Used by clinicians in clinical settings for percutaneous access; provides real-time visualization guidance. Once target reached, release mechanism allows guide removal without disturbing needle position. Benefits include improved procedural accuracy and stability during vascular access.
Clinical Evidence
No human clinical data. Evidence includes bench testing (depth accuracy, slide resistance, cycle testing, needle fit), biocompatibility (ISO 10993 series, ASTM F756), and phantom testing (N=8 procedures). An acute porcine study (N=1) demonstrated successful vascular access in 6/6 target sites (femoral arteries/veins) with no adverse outcomes or mechanical failures.
Technological Characteristics
Molded plastic bracket and needle guide; EtO sterilized (ISO 11135). Mechanical sensing/actuation via adjustable guide channels. Dimensions support 10-35 mm (in-plane) and 5-40 mm (out-of-plane) depths. Standalone mechanical device; no software or electronics. Biocompatibility per ISO 10993-5, -10, -11, -23; ASTM F756. Packaging/shelf-life per ISO 11607 and ASTM D4169/D4332/F1929/F2096/F88/F1886.
Indications for Use
Indicated for adult patients requiring ultrasound-guided procedures for precise placement of interventional devices, specifically for vascular access.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
Accurate Access, LLC
% Sam Murray
Head of Regulatory and Quality
PRIA Healthcare Management, LLC
30 Batterson Park Rd.
Suite 120
Farmington, Connecticut 06032
July 27, 2026
Re: K262474
Trade/Device Name: ND Needle Guidance System
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: ITX
Dated: July 17, 2026
Received: July 17, 2026
Dear Sam Murray:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262474 - Sam Murray
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262474 - Sam Murray
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.07.27 16:42:21 -04'00'
For
Yanna Kang, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K262474
Please provide the device trade name(s).
ND Needle Guidance System
Please provide your Indications for Use below.
The ND® Needle Guidance System is indicated for use in ultrasound-guided procedures where precise placement of common interventional devices is required, such as for vascular access.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Please select the age group(s) for which the device(s) is to be used.
☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☐ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater)
?
?
?
?
?
{4}
Accurate Access
ND Needle Guidance System
510(k) Summary
# 510(k) SUMMARY
# K262474
# CONTACT INFORMATION
Submitter and Address
Accurate Access, LLC
3951 E. Paradise View Dr.
Paradise Valley, AZ 85253
480-662-0722
Contact Person
Nabil Dib, MD
Interventional Cardiologist / Co-Manager
ndib@isctr.org
Date Summary was Prepared
January 26, 2026
# DEVICE NAME
Tradename
ND® Needle Guidance System
Common / Usual Name
Diagnostic Ultrasonic Transducer
Product Code
ITX
Regulation Number
892.1570
Regulatory Class
Class II
# PREDICATE DEVICES
K160806 CIVCO Medical Instruments
Verza Guidance System
K093713 CIVCO Medical Instruments
AccuSITE Needle Guides
# INDICATIONS for USE
The ND® Needle Guidance System is indicated for use in ultrasound-guided procedures where precise placement of common interventional devices is required, such as for vascular access.
# DEVICE DESCRIPTION
The ND Needle Guidance System is a sterile, single use device consisting of a molded bracket and two interchangeable needle guide configurations: in-plane and out-of-plane). The device mounts securely onto a commercially available ultrasound (US) probe to facilitate percutaneous access to targeted anatomical structures during diagnostic or therapeutic procedures.
The ultrasound probe is prepared with a sterile cover and allows real-time visualization while the ND Needle Guidance System provides mechanical stability and directional alignment for the needle/instrument to be inserted. The device provides an adjustable insertion depth range for both in-plane and out-of-plane configurations. Once needle or instrument placement is achieved, the guide's release mechanism enables removal of the guidance system without disturbing the position of the inserted device.
- 1 -
510(K) SUMMARY
{5}
Accurate Access
ND Needle Guidance System
510(k) Summary
Bracket

Needle Guide

The ND Needle Guidance System is a molded plastic device consisting of components designed to allow it to be securely attached to an ultrasound probe. The device includes two (2) adjustable needle guides, one in the in-plane and one in the out-of-plane of ultrasound field view.
The in-plane needle guide provides a depth range of 10-35 mm and out-of-plane guide allows for 5-40 mm.
The device uses simple geometric principles to allow the desired depth to be selected via a calculated angular translation based on the ultrasound vessel depth measurement.
Once the target access has been achieved, the device is removed from the needle/instrument leaving it fixed in the targeted location to continue with the intended interventional procedure.

## COMPARISON OF DEVICE CHARACTERISTICS
A comparison of the Indications for Use and technological features of the predicate and subject devices is provided in the table below and on the next page.
| SPECIFICATION | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | SECOND PREDICATE DEVICE |
| --- | --- | --- | --- |
| 510(k) | K000000 TBD | K160806 | K093713 |
| SPONSOR | Accurate Access | CIVCO Medical Instruments | CIVCO Medical Instruments |
| DEVICE | ND Needle Guidance System | Verza Needle Guidance System | AccuSITE US Needle Guides |
| INDICATION for USE | The ND Needle Guidance System is indicated for use where precise placement of a needle/instrument is needed for vascular access, tissue biopsy, fluid aspiration, or other treatment. The needle guidance system is affixed to a diagnostic ultrasound probe which provides visualization and guidance during insertion. | The system provides guidance for precise instrument placement of common interventional devices by positioning the device relative to the ultrasound transducer and the resulting image during a diagnostic or therapeutic procedure. This guidance system is intended for use with pediatric and adult patients. | The guidance system is intended for directing instruments such as catheters, electrodes and needles into a targeted anatomical location of a patient relative to the imaging instrument for percutaneous procedures. Prior to use healthcare workers should be trained in ultrasonography. • Vascular Access • Regional Anesthesia |
| REGULATORY | ITX | 892.1570 | SAME | SAME |
- 2 -
510(K) SUMMARY
{6}
Accurate Access
ND Needle Guidance System
510(k) Summary
## Discussion of Similarities and Differences
The principles of operation associated with the ND Needle Guidance System are the same for the predicate devices as well as other similar devices on the market for ultrasound guided access. Although other systems may use various methods to provide and secure the specific targeted depth selection (whether in-plane of out-of-plane), all are based on the same principle of achieving a desired depth by translating the associated access angle under ultrasound visualization.
| Bracket + Needle Guide | The bracket is attached to a compatible OEM ultrasound transducer; the needle Guide is connected to the bracket. |
| --- | --- |
| Depth/Angle Selection | The subject and predicate devices all allow for depth/angle selection. |
| Needle Release | The subject and predicate devices all allow for the needle or instrument to be freely released from the guide channel. |
| In-Plane / Out-of-Plane | Only Subject device has both configurations in a single device. |
## PERFORMANCE DATA
### Summary of Non-Clinical Testing
The ND Needle Guidance System was tested to and complies with applicable voluntary standards. The needle guide was tested to assure that the device meets its design specifications. Testing and results are summarized below.
### Design Verification Testing
- Depth Scale Legibility
- Slide Resistance Testing
- Cycle Testing = Bracket Snapping on Probe + Lever Lock Testing
- Needle Fit Testing
- Probe Cover Durability
All acceptance criteria were met demonstrating that the ND Needle Guidance System meets the specified design input requirements.
### Performance Testing
To cover a wide range of target vessel depths and device settings, testing of the ND Needle Guidance System was performed on 4 use configurations:
- 10 mm / Long Axis
- 35 mm / Long Axis
- 5 mm / Short Axis
- 40 mm / Short Axis
For Needle Centerline Testing and Needle Depth Accuracy Testing, all samples were within the acceptance criteria.
### Biocompatibility
Biocompatibility testing was performed as summarized below.
| # | BIOCOMPATIBILITY TEST | STANDARD | RESULTS |
| --- | --- | --- | --- |
| 1 | Cytotoxicity | ISO 10993-5 | The ND Needle Guidance System passed all biocompatibility testing. |
| 2 | Sensitization | ISO 10993-10 | |
| 3 | Irritation - Intracutaneous Injection | ISO 10993-23 | |
| 4 | Acute Systemic Toxicity | ISO 10993-11 | |
| 5 | Material Mediated Rabbit Pyrogen | | |
| 6 | Hemolysis | ASTM F756 | |
- 3 -
510(K) SUMMARY
{7}
Accurate Access
ND Needle Guidance System
510(k) Summary
### Sterility Validation
Ethylene oxide (EtO) sterilization was validated achieving a sterility assurance level (SAL) of at least 1x10⁻⁶ in accordance with ANSI/AAMI/ISO 11135.
### Shelf Life
Shelf life was validated with accelerated aging testing of 1 year per ISO 11607-1/-2.
### Transportation Testing
The ND Needle Guidance System passed all transportation and package integrity testing per ASTM D4169/D4332/F1929/F2096/F88/F1886.
### Usability Testing
During design and prototype development, a Formative Usability Evaluation was performed which included comprehensive design evaluations to assess the device's safety and performance. Results of usability testing confirmed that there were no critical tasks associated with use of the device with the testing demonstrating acceptable performance and user interaction per IEC 62366. Risk assessment documentation identified no critical tasks within the device workflow and all residual risks were deemed acceptable.
### Phantom Testing
The ND Needle Guidance System was tested using a commercial phantom which included several "vessels" at anatomical depths, thus simulating conditions of normal clinical use. N=8 procedures were performed in the ultrasound phantom simulator including short and long axis testing. All acceptance criteria were met providing additional evidence of device performance.
### Animal Study
An acute animal study was performed on a single pig (N=1). The purpose of the study was to evaluate the ND Needle Guidance System in the acute porcine animal model by obtaining vascular access in femoral vessels to demonstrate that the device performs as intended under conditions of simulated human clinical use.
Device performance was assessed by determining whether using the device resulted in successful puncture of the target vessel. These included the right and left femoral arteries and veins. Success was achieved at all 6 prespecified target sites.
The animal study also evaluated mechanical performance of the needle guidance system which was reported as passing for all endpoints. Inspection of the probe cover and the ultrasound probe did not reveal any evidence of damage.
There were no occurrences of adverse outcomes, nor reports of user difficulty.
In conclusion, the safety and performance data from the animal study provides strong support for 510(k) clearance of the ND Needle Guidance System.
### CONCLUSION
Based on the results of the nonclinical and animal testing, the ND Needle Guidance System was demonstrated to be as safe as the predicates with performance data providing evidence supporting a determination of substantial equivalence.
- 4 -
510(K) SUMMARY
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.