The TheraSphere 360™ Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Y-90 Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass. The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Y-90 Microspheres dose vial selection. Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Y-90 Microspheres. For post-TheraSphere Y-90 Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Y-90 Microspheres.
Device Story
Browser-based platform supporting radioembolization therapy; provides interactive tools for calculating Y-90 microsphere activity based on desired dose, lung shunt fraction, residual waste, and liver mass. Used by authorized clinicians for treatment planning, dose vial selection, and retrospective post-treatment dose analysis. Operates via MIRD (Medical Internal Radiation Dose) schema to transform clinical inputs into activity requirements. Facilitates clinical decision-making by ensuring accurate dose calculation and inventory management. Benefits patients by supporting precise delivery of radiation therapy.
Clinical Evidence
No clinical data. Performance evaluation consisted of bench testing, including software verification (unit and integration testing), cybersecurity evaluation, and design validation to ensure requirements and user needs were met.
Technological Characteristics
Browser-based software platform; utilizes MIRD (Medical Internal Radiation Dose) schema for dosimetry calculations. Functions include treatment planning, activity calculation, vial selection, and post-treatment analysis. Software version V2.1.0.
Indications for Use
Indicated for adult patients undergoing radioembolization with TheraSphere Y-90 Microspheres; provides tools for pre-treatment planning, activity calculation, dose vial selection, and retrospective post-treatment dose determination.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 24, 2026
Boston Scientific Corporation
Amy Mckinney
Fellow, Regulatory Affairs
300 Boston Scientific Way
Marlboro, Massachusetts 01752
Re: K262462
Trade/Device Name: TheraSphere 360™ Y-90 Management Platform
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: July 17, 2026
Received: July 17, 2026
Dear Amy Mckinney:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262462 - Amy Mckinney
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K262462 - Amy Mckinney
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262462 | ? |
| Please provide the device trade name(s). | | ? |
| TheraSphere 360™ Y-90 Management Platform | | |
| Please provide your Indications for Use below. | | ? |
| The TheraSphere 360™ Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Y-90 Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass. | | |
| The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Y-90 Microspheres dose vial selection. | | |
| Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Y-90 Microspheres. For post-TheraSphere Y-90 Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Y-90 Microspheres. | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
# **K262462
510(k) SUMMARY**
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| **Submitter:** | Boston Scientific Corp. |
| --- | --- |
| **Company Contact Person:** | Ms. Amy E McKinney, MS, RAC Fellow, Regulatory Affairs Boston Scientific Corp |
| Phone: | 651-287-5096 |
| Email: | Amy.mckinney@bsci.com |
| **Alternate Contact Person:** | Ms. Janet Bukovcan Director, Regulatory Affairs Boston Scientific Corp. |
| Phone: | 613-866-9312 |
| Email: | Janet.bukovcan@bsci.com |
| **Device Name:** | TheraSphere 360™ Y-90 Management Platform |
| **Device Classification Name:** | Medical image management and processing system (LLZ) 21 CFR 892.2050 |
| **Review Panel:** | Radiology |
| **Predicate Devices / Reference 510(k):** | TheraSphere 360™ Y-90 Management Platform (K260479) |
| **Date of Preparation:** | July 17, 2026 |
# **Device Description:**
The TheraSphere 360 Y-90 Management Platform is an end-to-end, browser-based platform that will host a wide range of resources (e.g. radioembolization activity calculations, ordering, tracking, and education) that support Authorized Users of TheraSphere Microspheres.
The TheraSphere 360 Platform includes treatment planning functionality, activity calculation functionality, vial selection and ordering, and post-treatment analysis functionality. The treatment planning functionality and the activity calculation functionality include an interactive tool intended for calculating the activity of TheraSphere Microspheres required at the treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass.
The Vial Selector function allows users to select and order TheraSphere Microspheres dose vials from inventory that match desired results.
The post-treatment analysis functionality is intended as an optional tool for post-treatment evaluation following TheraSphere Microspheres treatment.
Page 1 of 3
{6}
K262462 – Page 2 of 3
# Indications For Use:
No changes were made to the Indications for Use.
The TheraSphere 360™ Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass.
The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection.
Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres.
# Technological Comparison:
This Special 510(k) was submitted for TheraSphere 360 V2.1.0 software. Changes in this 510(k) included a modification to update the dose calculations for vials in the vial selector cart when a user has modified the activity calculator inputs after placing vials in the cart. Other minor software anomalies were corrected as well that did not impact the functionality of the device.
The table below provides a summary comparison of the submitted device compared to the predicate device:
| Characteristic | TheraSphere™ 360 Y-90 Management Platform V2.1.0 | TheraSphere™ 360 Y-90 Management Platform V2.0 | Comment |
| --- | --- | --- | --- |
| Indications for Use | The TheraSphere 360 Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass. The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection. Additionally, the TheraSphere | The TheraSphere 360 Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass. The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection. Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with | Identical |
{7}
K262462 – Page 3 of 3
| Characteristic | TheraSphere™ 360 Y-90 Management Platform V2.1.0 | TheraSphere™ 360 Y-90 Management Platform V2.0 | Comment |
| --- | --- | --- | --- |
| | 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres. | TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres. | |
| Principles of Operation | Medical Internal Radiation Dose (MIRD) schema for dosimetry calculations | Medical Internal Radiation Dose (MIRD) schema for dosimetry calculations | Identical |
| Device Functionalities | Contains medical device functionalities which utilize radioembolization activity calculations to determine the dose of TheraSphere Microspheres required at treatment time. | Contains medical device functionalities which utilize radioembolization activity calculations to determine the dose of TheraSphere Microspheres required at treatment time. | Identical |
## Summary of Performance Data
Performance testing confirmed that the TheraSphere 360 V2.1.0 software performs safely and effectively and does not introduce any new or different safety risks. The TheraSphere 360 V2.1.0 medical device functions underwent software verification testing, including unit and integration testing and cybersecurity evaluation, to ensure that the product meets the defined software requirements. A design validation evaluation was conducted to ensure that the platform meets defined user needs. Verification and validation testing, including a usability evaluation, successfully demonstrated that the medical device functions and algorithms meet requirements and user needs, confirming safety and effectiveness of TheraSphere 360 V2.1.0 software and substantial equivalence to the predicate device.
Clinical testing was not required to validate TheraSphere 360 V2.1.0 software.
## Conclusion:
In conclusion, the TheraSphere 360 Y-90 Management Platform V2.1.0 medical device functions were evaluated against the predicate device and found to be substantially equivalent on the basis that 1) the indications for use are unchanged, and 2) the technology of the TheraSphere 360 Platform medical device functions is the same as the predicate device and no new or different issues of safety and effectiveness were raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.