OSSIX Plus

K262415 · Datum Dental, Ltd. · NPL · Oct 29, 2026 · Dental

Device Facts

Record IDK262415
Device NameOSSIX Plus
ApplicantDatum Dental, Ltd.
Product CodeNPL · Dental
Decision DateOct 29, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3930
Device ClassClass 2
AttributesTherapeutic

Indications for Use

OSSIX® Plus is a resorbable collagen membrane intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for: - Ridge augmentation for later implant insertions. - Simultaneous ridge augmentation and implant insertions. - Ridge augmentation around implants inserted in delayed extraction sites. - Ridge augmentation around implants inserted in immediate extraction sites. - Alveolar ridge preservation consequent to tooth (teeth) extraction(s). - Over the window in lateral window sinus elevation procedure. - In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved. - In intra bony defects around teeth. - For treatment of recession defects, together with coronally positioned flap. - In furcation defects in multi rooted teeth.

Device Story

OSSIX Plus is a resorbable, biocompatible collagen membrane derived from porcine tendons; cross-linked with ribose. Used by dental surgeons in clinical settings for guided bone and tissue regeneration. The membrane acts as a physical barrier to guide tissue growth; it is porous, fibrous, and absorbs fluids to adhere to surrounding tissues. It is slowly resorbed and replaced by new tissue. The device is provided sterile (ethylene oxide) in various sizes. It is a single-use device. The current submission supports a manufacturing facility transfer; the device design, materials, and clinical application remain unchanged from the predicate.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including physicochemical characterization (dimensional, water uptake, tensile strength, porosity, TGA, surface morphology), biocompatibility (ISO 10993 series), and sterilization/packaging validation. Viral safety is supported by porcine tissue sourcing and historical inactivation studies. Bench testing confirms the device meets specifications and is equivalent to the predicate.

Technological Characteristics

Resorbable membrane of cross-linked porcine type I collagen. Porous lattice network (approx. 86% porosity). Dimensions: 15x25 mm, 25x30 mm, 30x40 mm; thickness ~0.2 mm. Sterilized via ethylene oxide (SAL 10^-6). Biocompatible, non-pyrogenic, non-antigenic. Complies with ISO 10993, ISO 11135, ISO 22442, and ASTM standards for packaging (D4169, F2825, F1929, F2096, F88).

Indications for Use

Indicated for dental patients requiring guided bone or tissue regeneration, including ridge augmentation, alveolar ridge preservation, sinus elevation, and treatment of periodontal defects (intra-bony, recession, furcation). Contraindicated in patients with collagen or porcine hypersensitivity, autoimmune/connective tissue diseases, acute/chronic infection at the surgical site, or those receiving intravenous bisphosphonates.

Regulatory Classification

Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{1} **FDA** U.S. FOOD & DRUG ADMINISTRATION Datum Dental, Ltd. % Kelliann Payne Partner Hogan Lovells US Llp 1735 Market St., Floor 23 Philadelphia, Pennsylvania 19103 October 29, 2026 Re: K262415 Trade/Device Name: OSSIX Plus Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: July 15, 2026 Received: July 31, 2026 Dear Kelliann Payne: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K262415 - Kelliann Payne Page 2 Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {3} K262415 - Kelliann Payne Page 3 Office of Industry and Consumer Education (OICE) to ask a question about a specific regulatory topic. See the OICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-office-industry-and-consumer-education-oice) for more information or contact OICE by email (AskOICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K262415 Device Name OSSIX® Plus Indications for Use (Describe) OSSIX® Plus is a resorbable collagen membrane intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for: - Ridge augmentation for later implant insertions. - Simultaneous ridge augmentation and implant insertions. - Ridge augmentation around implants inserted in delayed extraction sites. - Ridge augmentation around implants inserted in immediate extraction sites. - Alveolar ridge preservation consequent to tooth (teeth) extraction(s). - Over the window in lateral window sinus elevation procedure. - In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved. - In intra bony defects around teeth. - For treatment of recession defects, together with coronally positioned flap. - In furcation defects in multi rooted teeth. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} Dentsply | datum Sirona | dental # 510(k) SUMMARY OSSIX® Plus K262415 # Submitter: Datum Dental Ltd. 38 Hayarkon St., Yavne, Israel 8122741 # Submission Correspondent: Kelliann Payne Hogan Lovells Cadwalader US LLP kelliann.payne@hlc.com (267) 675-4687 Date Prepared: October 29, 2026 # Name of Device: OSSIX® Plus # Classification Name: 21 CFR 872.3930 (Bone grafting material) Class II Product Code NPL - Barrier, Animal Source, Intraoral # Predicate Device OSSIX® Plus (K160281) # Reference Devices None # Device Description OSSIX® Plus is a biodegradable and biocompatible collagen membrane intended for use during the process of guided tissue and bone regeneration. The collagen is derived from tendons of veterinary-certified pigs and is purified and cross-linked using ribose. OSSIX® Plus is packaged in a double blister and an outer cardboard box and is sterilized by ethylene oxide. Due to its porous and fibrous microstructure, the membrane readily absorbs fluids, adheres to the surrounding tissues and provides a barrier that guides bone and tissue regeneration. Available in sizes: 15x25 mm, 25x30 mm, and 30x40 mm. Intended for use by dental surgeons. Page 1 of 8 {6} [LOGO] Dentsply Sirona datum dental ### Intended Use / Indications for Use OSSIX® Plus is a resorbable collagen membrane intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for: - Ridge augmentation for later implant insertions. - Simultaneous ridge augmentation and implant insertions. - Ridge augmentation around implants inserted in delayed extraction sites. - Ridge augmentation around implants inserted in immediate extraction sites. - Alveolar ridge preservation consequent to tooth (teeth) extraction(s). - Over the window in lateral window sinus elevation procedure. - In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved. - In intra bony defects around teeth. - For treatment of recession defects, together with coronally positioned flap. - In furcation defects in multi rooted teeth. ### Comparison of Technological Characteristics with the Predicate Device OSSIX® Plus is a resorbable collagen membrane intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier. The subject device has the same intended use and identical indications for use as the predicate OSSIX® Plus device previously cleared under K160281. This submission supports the transfer of manufacturing facility, and does not reflect a change to the device's intended use, indications for use, design, raw materials, composition, primary barrier system, sterilization modality, or fundamental principles of operation. In addition, several minor manufacturing-related changes, including updates to in-process testing methods, suppliers, equipment, and secondary packaging configurations, were implemented under the company's quality system and validated as appropriate, with no adverse impact on device design, intended use, product quality, safety, or performance. | Characteristic | Subject Device | Predicate Device | Assessment | | --- | --- | --- | --- | | | OSSIX® Plus | OSSIX® Plus (K160281) | | | Intended use / Indications for use | OSSIX® Plus is a resorbable collagen membrane intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for: - Ridge augmentation for later implant insertions. | OSSIX® Plus is a resorbable collagen membrane intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier for: - Ridge augmentation for later implant insertions; | Identical | Page 2 of 8 {7} Dentsply Sirona datum dental | Characteristic | Subject Device | Predicate Device | Assessment | | --- | --- | --- | --- | | | OSSIX® Plus | OSSIX® Plus (K160281) | | | | - Simultaneous ridge augmentation and implant insertions. - Ridge augmentation around implants inserted in delayed extraction sites. - Ridge augmentation around implants inserted in immediate extraction sites. - Alveolar ridge preservation consequent to tooth (teeth) extraction(s). - Over the window in lateral window sinus elevation procedure. - In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved. - In intra bony defects around teeth. - For treatment of recession defects, together with coronally positioned flap. - In furcation defects in multi rooted teeth. | - Simultaneous ridge augmentation and implant insertions; - Ridge augmentation around implants inserted in delayed extraction sites; - Ridge augmentation around implants inserted in immediate extraction sites; - Alveolar ridge preservation consequent to tooth (teeth) extraction(s); - Over the window in lateral window sinus elevation procedure; - In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved; - In intra bony defects around teeth; - For treatment of recession defects, together with coronally positioned flap; - In furcation defects in multi rooted teeth. | | | **Contraindications** | OSSIX® Plus must not be used in: 1. Patients with known collagen hypersensitivity. 2. Patients with sensitivity to porcine-derived materials. | OSSIX® Plus must not be used in: 1. Patients with known collagen hypersensitivity. 2. Patients with sensitivity to porcine-derived materials. | Substantially equivalent; additional contraindications added to align IFU with EU labeling requirements. | Page 3 of 8 {8} [LOGO] Dentsply Sirona datum dental | Characteristic | Subject Device | Predicate Device | Assessment | | --- | --- | --- | --- | | | OSSIX® Plus | OSSIX® Plus (K160281) | | | | 3. Patients suffering from autoimmune diseases and connective tissue diseases, such as: systemic lupus erythematosus, dermatomyositis, etc. 4. Patients with acute or chronic infection (osteomyelitis) at the surgical site. 5. In patients that have received or are currently receiving treatment with intravenously administered bisphosphonates. | 3. Patients suffering from autoimmune diseases and connective tissue diseases, such as: systemic lupus erythematosus, dermatomyositis, etc. | | | **Composition of Materials** | Cross-linked porcine type I collagen | Cross-linked porcine type I collagen | Identical | | **Usage** | Single Use | Single Use | Identical | | **Classification name** | Barrier, animal source, intraoral | Barrier, animal source, intraoral | Identical | | **Product Code** | NPL | NPL | Identical | | **Class** | 21 CFR §872.3930, Class II | 21 CFR §872.3930, Class II | Identical | | **Product Design** | Resorbable dental membrane made of porous lattice network of collagen | Resorbable dental membrane made of porous lattice network of collagen | Identical | | **Properties** | - Biocompatible; - non-pyrogenic; - non-antigenic; - porous and fibered microstructure; - readily absorbs bodily fluid; - adheres well to the surrounding and underlying tissues; - is not disrupted by closure of the surgical site; | - Biocompatible; - non-pyrogenic; - non-antigenic; - porous and fibered microstructure; - readily absorbs bodily fluid; - adheres well to the surrounding and underlying tissues; - is not disrupted by closure of the surgical site; | Substantially equivalent, as established through bench and biocompatibility testing. | Page 4 of 8 {9} [LOGO] Dentsply Sirona datum dental | Characteristic | Subject Device | Predicate Device | Assessment | | --- | --- | --- | --- | | | OSSIX® Plus | OSSIX® Plus (K160281) | | | | • slowly resorbed and replaced by new tissue. | • slowly resorbed and replaced by new tissue. | | | **Dimensions** | 15x25 mm; 25x30 mm; 30x40 mm | 15x25 mm; 25x30 mm; 30x40 mm | Identical | | **Thickness** | About 0.2 mm | About 0.2 mm | Identical | | **Porosity** | About 86% | About 80% | Substantially equivalent; porosity percentage remained comparable and average pore diameter remained within the same sub-micrometer range for cell occlusion and barrier function. | | **Wettability** | Wettable and completely soaked in aqueous solution in less than one minute | Wettable and completely soaked in aqueous solution in less than one minute | Identical | | **Cross-linking** | Chemically cross-linked with ribose | Chemically cross-linked with ribose | Identical | | **Trimmable** | Trimmable with surgical instruments | Trimmable with surgical instruments | Identical | | **Source** | Porcine tendons | Porcine tendons | Identical | | **Fat (Lipids)** | <0.1% | <0.1% | Identical | | **Purified Collagen Intermediate** | PUC | PUC | Identical | | **pH** | 6-8 | 6-8 | Identical | | **Packaging** | Double blisters system | Double blisters system | Substantially equivalent. Primary barrier system unchanged. Minor modifications to secondary shipping configurations, including transition from the historical 10-box configuration to validated 100-unit and 200-unit master shipper | Page 5 of 8 {10} [LOGO] Dentsply Sirona datum dental | Characteristic | Subject Device | Predicate Device | Assessment | | --- | --- | --- | --- | | | OSSIX® Plus | OSSIX® Plus (K160281) | | | | | | configurations, do not affect product performance or package integrity. | | **Sterilization** | Ethylene oxide | Ethylene oxide | Identical | | **Sterility** | SAL 10^{-6} | SAL 10^{-6} | Identical | Therefore, OSSIX® Plus manufactured at the Yavne facility has the same intended use and indications for use, and similar technological characteristics as the predicate OSSIX® Plus device previously cleared under K160281. ### Summary of Data to Support Substantial Equivalence Datum Dental conducted a comparability program assessing whether OSSIX® Plus manufactured at the new facility continues to meet established specifications and maintains the same safety and performance profile as product manufactured at the old facility. The evidence includes specification testing, physicochemical characterization, biocompatibility testing, and viral safety support. ### Bench Testing Datum Dental conducted bench testing to support the site transfer and demonstrate that the OSSIX® Plus device manufactured in the new facility is equivalent to the previously cleared OSSIX® Plus. This bench testing included the following categories: Specification testing: - Appearance - Weight - Heavy metal content (USP <233>) - Collagenase resistance and trypsin resistance - Endotoxins (ANSI AAMI ST72:2019; USP <85>) Physicochemical characterization: - Dimensional checks - Water uptake - Denaturation temperature (USP <891>) - Ethanol residuals - Tensile strength - Porosity - Thermogravimetric Analysis (TGA) - Surface morphology Page 6 of 8 {11} [LOGO] Dentsply Sirona datum dental ### Biocompatibility Testing OSSIX® Plus was subjected to biocompatibility studies according to ISO 10993-1, including: - Cytotoxicity (ISO 10993-5); - Rabbit pyrogenicity (ISO 10993-11; USP <151>); - Sensitization (ISO 10993-10); - Intracutaneous reactivity (ISO 10993-23); - Acute systemic toxicity (ISO 10993-11); - Genotoxicity (ISO 10993-3); and - Ethylene oxide (EO) and ethylene chlorohydrin (ECH) residuals (ISO 10993-7). These results, together with historical biocompatibility data previously provided in K053260 and K160281, supported the conclusion that the transferred process does not introduce new or increased biological risk and that the subject device can be considered biocompatible for use as intended. ### Viral Safety Viral safety is supported through controlled porcine tissue sourcing and viral inactivation and clearance studies for the predicate device, previously provided in K053260 and K160281. OSSIX® Plus continues to meet the predefined viral safety acceptance criteria. Risk management, animal sourcing and collection, and process validation were conducted in accordance with ISO 22442 Parts 1, 2, and 3, respectively. ### Packaging Validation Packaging validation testing demonstrated that the updated secondary packaging configurations maintain package integrity, sterile barrier performance, and product protection under simulated shipping conditions. Shipping simulation was performed per ASTM D4169 (Distribution Cycle 13) and ASTM F2825, followed by sterile barrier integrity and seal strength testing per ASTM F1929 (dye penetration), ASTM F2096 (bubble leak), and ASTM F88 (seal strength). ### Shelf Life The 36-month shelf life was previously established on real-time stability data (K160281). Product stability comparability testing and packaging stability testing in accordance with ISO 11607-1 and ISO 11607-2, confirm that the labeled 36-month shelf life remains applicable to the subject device. ### Sterilization Validation The post-sterilization aeration conditions were modified while maintaining the existing EO sterilization method previously validated in K160281. Validation testing per ISO 11135 demonstrated that the revised process maintains a sterility assurance level (SAL) of 10⁻⁶ and EO and ECH residuals met the limits of ISO 10993-7, supporting the continued safety and performance of the device. ### Animal Testing Page 7 of 8 {12} Dentsply | datum Sirona | dental No new animal testing was conducted; the in vivo performance previously established for OSSIX® Plus (K053260, K160281) remains applicable, as the subject changes are limited to the manufacturing site and associated processes and were addressed through comparative nonclinical testing. ## Conclusion The information provided in this submission demonstrates that the subject OSSIX® Plus has the same indications for use and similar technological characteristics as the predicate OSSIX® Plus device. The manufacturing facility transfer and other minor manufacturing changes do not raise new questions of safety or effectiveness, and performance testing demonstrates that the subject device is as safe and effective as the predicate device. Therefore, the subject device is substantially equivalent to OSSIX® Plus (K160281). Page 8 of 8
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...