CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing set
K262398 · CERENOVUS, Inc. · NRY · Aug 13, 2026 · Cardiovascular
Device Facts
Record ID
K262398
Device Name
CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing set
Applicant
CERENOVUS, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Aug 13, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Product Code NRY: The CEREGLIDE 57 Intermediate Catheter, with the Cerenovus Aspiration Tubing Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. The Cerenovus Aspiration Tubing Set is intended to connect the Cerenovus Large Bore Catheter or the CEREGLIDE™ 71, 57 or 42 Intermediate Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. Product Code QJP: The CEREGLIDE 57 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE 57 Intermediate Catheter is also indicated for use as a conduit for retrieval devices.
Device Story
Variable stiffness, single-lumen catheter; stainless-steel braid with PTFE liner; hydrophilic coating; radiopaque marker band. Used in neurovascular procedures; introduced over steerable guidewire/microcatheter. Facilitates insertion/guidance of interventional devices; acts as conduit for retrieval devices. For thrombus removal, connects to external vacuum source via Cerenovus Aspiration Tubing Set (integrated flow control switch). Operated by physicians in clinical settings. Provides continuous aspiration (-68 to -95 kPa) to revascularize occluded vessels. Output is physical removal of thrombus; visualized via fluoroscopy. Benefits include restoration of blood flow in acute ischemic stroke patients.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing and design verification based on predicate similarities.
Indicated for adults (22+) with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, vertebral arteries) within 8 hours of symptom onset; patients ineligible for or failed thrombolytic therapy. Also indicated for facilitating insertion/guidance of interventional devices in neurovasculature and as a conduit for retrieval devices.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 13, 2026
Cerenovus, Inc.
Mia Dudicova
Regulatory Affairs Contractor
6303 Waterford District Drive,
Suite 315
Miami, Florida 33126
Re: K262398
Trade/Device Name: CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: NRY, QJP
Dated: July 12, 2026
Received: July 15, 2026
Dear Mia Dudicova:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262398 – Mia Dudicova
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262398 – Mia Dudicova
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262398 | ? |
| Please provide the device trade name(s). | | ? |
| CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set | | |
| Please provide your Indications for Use below. | | ? |
| Product Code NRY: The CEREGLIDE 57 Intermediate Catheter, with the Cerenovus Aspiration Tubing Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. The Cerenovus Aspiration Tubing Set is intended to connect the Cerenovus Large Bore Catheter or the CEREGLIDE™ 71, 57 or 42 Intermediate Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. Product Code QJP: The CEREGLIDE 57 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE 57 Intermediate Catheter is also indicated for use as a conduit for retrieval devices. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
Cerenovus, Inc.
Special 510(k) Premarket Notification
CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
# 510(k) Summary – K262398
## I. Submitter
Cerenovus, Inc.
6303 Waterford District Drive,
Suite 315
Miami, FL 33126 USA
Contact Person: Mia Dudicova (Regulatory Affairs Contractor)
Email: MDudicov@its.jnj.com
Tel: 00353 87 466 0113
## II. Date
## Prepared
July 12 \( ^{th} \) , 2026
## III. Device
## Information
| Table 1. Device Information | |
| --- | --- |
| Device Proprietary Name | CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set |
| Common or Usual name | Catheter, Thrombus Retriever Catheter, Percutaneous |
| Classification Name | 21 CFR 870.1250 – Catheter, Thrombus Retriever 21 CFR 870.1250 – Catheter, Percutaneous |
| Regulatory Classification | II |
| FDA Product Codes | NRY, QJP |
## IV. Predicate
## Device
## Information
The predicates are listed below in Table 2.
| Table 2. Predicate Devices | | | |
| --- | --- | --- | --- |
| 510(k) Number | Date Cleared | Name | Manufacturer |
| K241221 | October 16, 2024 | CEREGLIDE 42 Intermediate Catheter; CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set | Cerenovus, Inc. |
| K241244 | October 16, 2024 | CEREGLIDE 42 Intermediate Catheter; CEREGLIDE 57 Intermediate Catheter | Cerenovus, Inc. |
Continued on the next page
Page 1 of 6
{5}
Cerenovus, Inc.
Special 510(k) Premarket Notification
CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
# 510(k) Summary, continued
# V. Device Description
The CEREGLIDE™ 57 Intermediate Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless-steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless-steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity to reduce friction during navigation in the vasculature. The proximal end of the catheter has a luer fitting located on the end of the catheter hub which can be used to attach accessories for flushing and aspiration. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a Tuohy Borst Rotating Hemostasis Valve (RHV) with a side port and two slit introducers as accessories.
The RHV with side port is used for flushing, insertion of catheters, and connection to an external aspiration system. The slit introducers are designed to introduce the catheter into the base catheter and protect the distal tip of the catheter during insertion into the RHV of the base catheter.
The CEREGLIDE™ 57 Intermediate Catheter is designed for use in facilitating the insertion and guidance of appropriately sized interventional devices and as conduit for retrieval devices, and in the removal of thrombus from the neurovasculature using continuous aspiration from an external vacuum source when used with the Cerenovus Aspiration Tubing Set. The CEREGLIDE™ 57 Intermediate Catheter and Cerenovus Aspiration Tubing Set are intended to be used with a compatible aspiration pump as a vacuum source that provides a minimum vacuum pressure of -68 kPa [-20 inHg] and a maximum vacuum pressure of -95 kPa [-28 inHg], and minimum 17 L/min flow rate, with a collection canister of at least 1000 mL volume.
The CEREGLIDE™ 57 Intermediate Catheter is intended to be connected to the vacuum source using the Cerenovus Aspiration Tubing Set which is an externally manufactured tubing set with an internal diameter (ID) of 0.110" and overall length of 112", containing an integrated flow control switch.
# VI. Indications for Use
# Product code NRY:
The CEREGLIDE™ 57 Intermediate Catheter, with the Cerenovus Aspiration Tubing Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
The Cerenovus Aspiration Tubing Set is intended to connect the Cerenovus Large Bore Catheter or the CEREGLIDE™ 71, 57 or 42 Intermediate Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow.
# Product code QJP:
The CEREGLIDE™ 57 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE™ 57 Intermediate Catheter is also indicated for use as a conduit for retrieval devices.
Page 2 of 6
{6}
Cerenovus, Inc.
Special 510(k) Premarket Notification
CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
### VII. Predicate Comparison
A comparison of product features between CEREGLIDE™ 57 Intermediate Catheter and Cerenovus Aspiration Tubing Set and the predicate devices is presented in Table 3.
Continued on next page
Page 3 of 6
{7}
Cerenovus, Inc.
Special 510(k) Premarket Notification
CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
## 510(k) Summary, continued
| Table 3. Predicate and Subject Device Comparison | | | | |
| --- | --- | --- | --- | --- |
| Description | Predicate Device: CEREGLIDE 42 Intermediate Catheter; CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set (K241221, NRY) | Predicate Device: CEREGLIDE 42 Intermediate Catheter; CEREGLIDE 57 Intermediate Catheter (K241244, QJP) | Subject Device: CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set (NRY) | Subject Device: CEREGLIDE 57 Intermediate Catheter (QJP) |
| Indications For Use | The CEREGLIDE 42 Intermediate Catheter and the CEREGLIDE 57 Intermediate Catheter, with the Cerenovus Aspiration Tubing Set and a compatible aspiration pump, are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. The Cerenovus Aspiration Tubing Set is intended to connect the Cerenovus Large Bore Catheter or the CEREGLIDE™ 71,57 or 42 Intermediate Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. | The CEREGLIDE 42 Intermediate Catheter and the CEREGLIDE 57 Intermediate Catheter are indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE 57 Intermediate Catheter is also indicated for use as a conduit for retrieval devices. | The CEREGLIDE 57 Intermediate Catheter, with the Cerenovus Aspiration Tubing Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. The Cerenovus Aspiration Tubing Set is intended to connect the Cerenovus Large Bore Catheter or the CEREGLIDE™ 71,57 or 42 Intermediate Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. | The CEREGLIDE 57 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE 57 Intermediate Catheter is also indicated for use as a conduit for retrieval devices. |
| Product Code | NRY | QJP | NRY | QJP |
| Regulation Name | Catheter, Thrombus Retriever | Catheter, Percutaneous | Catheter, Thrombus Retriever | Catheter, Percutaneous |
| Classification | Class II - 21 CFR 870.1250 | | Same as predicates | |
| Basic Design | Variable stiffness single lumen catheter | | Same as predicates | |
Page 4 of 6
{8}
Cerenovus, Inc.
Special 510(k) Premarket Notification
CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
| Table 3 Predicate and Subject Device Comparison | | | | |
| --- | --- | --- | --- | --- |
| Description | Predicate Device: CEREGLIDE 42 Intermediate Catheter; CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set (K241221, NRY) | Predicate Device: CEREGLIDE 42 Intermediate Catheter; CEREGLIDE 57 Intermediate Catheter (K241244,QJP) | Subject Device: CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set (NRY) | Subject Device: CEREGLIDE 57 Intermediate Catheter (QJP) |
| Length | CEREGLIDE 42: 115 cm, 125 cm, 132 cm, 144 cm, 152 cm, 160 cm | | CEREGLIDE 57: 115 cm, 125 cm, 132 cm, 137 cm, 152 cm | |
| | CEREGLIDE 57: 115 cm, 125 cm, 132 cm, 137 cm | | | |
| Inner Diameter (ID) | CEREGLIDE 42: 0.042" CEREGLIDE 57: 0.057" | | 0.057" | |
| Distal Outer Diameter (OD) | CEREGLIDE 42: 0.053" CEREGLIDE 57: 0.068" | | 0.068" | |
| Proximal OD | CEREGLIDE 42: 0.0535" CEREGLIDE 57: 0.0685" | | 0.0685" | |
| Catheter Coating | Hydrophilic | | Same as predicates | |
| Coating Length | 55 cm | | Same as predicates | |
| Tip Configuration | Non-shapeable tip | | Same as predicates | |
| Marker Band | Platinum/ Iridium | | Same as predicates | |
| Braid | Stainless Steel | | Same as predicates | |
| Liner | PTFE Liner | | Same as predicates | |
| Hub | Polycarbonate | | Same as predicates | |
| Strain Relief | Polyolefin | | Same as predicates | |
| Outer Jacket | Pebax, Urethane | | Same as predicates | |
| Hemostasis Valve | Included | | Same as predicates | |
| Introducer Sheath | Slit Introducer (2) | | Same as predicates | |
| Sterilization Method | Ethylene Oxide | | Same as predicates | |
| Sterility Assurance Level (SAL) | \( 10^{-6} \) | | Same as predicates | |
| Packaging | Polyethylene Hoop and Mounting Card, Tyvek® Pouch, Carton | | Same as predicates | |
| Shelf Life | 3 years | | Same as predicates | |
| | Predicate device (NRY) | | Subject Device (NRY) | |
| Required Additional Accessories | Cerenovus Aspiration Tubing Set Compatible Aspiration Pump | | Same as predicate | |
| Minimum Aspiration Pressure | -20 inHg (-68 kPa) | | Same as predicate | |
| Maximum Aspiration Pressure | -28 inHg (-95 kPa) | | Same as predicate | |
| Flowrate (Air) | 0 to 60 LPM | | Same as predicate | |
| Tubing ID | 0.110 in minimum | | Same as predicate | |
| Tubing Length | 112 in | | Same as predicate | |
| Flow Control Mechanism | Flow Control Switch | | Same as predicate | |
Continued on next page
Page 5 of 6
{9}
Cerenovus, Inc.
Special 510(k) Premarket Notification
CEREGLIDE™ 57 Intermediate Catheter; Cerenovus Aspiration Tubing Set
## 510(k) Summary, continued
VIII. Non-Clinical Performance Data
**Performance Testing – Bench**
No additional bench testing was required for the addition of 152 cm working length CEREGLIDE 57 Intermediate Catheter, as previously generated performance testing for predicate devices was deemed sufficient and appropriate for substantial equivalence determination.
**Performance Testing – Animal**
No animal study was deemed necessary to evaluate the changes to the subject device and demonstrate substantial equivalence to the predicate devices.
**Performance Testing – Clinical**
Clinical studies were not required as appropriate verification and validation of the design were achieved based on the similarities of the proposed device and the predicate devices, and previous non-clinical testing.
**Sterilization**
No additional sterilization cycle validation was deemed necessary to adopt the subject 152 cm working length model into the existing EO sterilization cycle.
**Shelf-Life**
Changes did not impact the established shelf-life of the product.
**Biocompatibility Testing**
Evaluation of the device change compared to the predicates showed the biological risk of the subject device is unchanged compared to the predicates.
IX. Conclusion
Based upon the intended use, design, materials, function, and previous non-clinical testing, it is concluded that the subject CEREGLIDE™ 57 Intermediate Catheter and Cerenovus Aspiration Tubing Set are substantially equivalent to the predicate devices CEREGLIDE™ 42 Intermediate Catheter, CEREGLIDE™ 57 Intermediate Catheter, and Cerenovus Aspiration Tubing Set (K241221) and CEREGLIDE™ 42 Intermediate Catheter and CEREGLIDE™ 57 Intermediate Catheter (K241244). The difference in length does not raise any new questions regarding the safety and effectiveness of the subject device. The subject device, as designed, manufactured, packaged, and sterilized, is substantially equivalent to the predicate devices currently marketed under the Federal Food, Drug and Cosmetic Act.
End of document
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.