K262301 · Ischemaview, Inc. · QAS · Oct 9, 2026 · Radiology
Device Facts
Record ID
K262301
Device Name
Rapid ICH (v4.0)
Applicant
Ischemaview, Inc.
Product Code
QAS · Radiology
Decision Date
Oct 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, PCCP, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K262301 · Oct 9, 2026
Rapid ICH (v4.0)
Ischemaview, Inc.
Retrospective clinical head CT images from 13 clinical institutions; Imaging core laboratories; Multicenter study datasets
Retrospective clinical data was used to validate the performance (sensitivity and specificity) of the AI algorithm for detecting intracranial hemorrhage in non-enhanced head CT images.
Patients ≥18 years old undergoing non-enhanced head CT; Sample Size: 311 cases; Number of Sites: 21 sources (13 clinical institutions and 8 data providers/labs)
Not applicable for this study
Sensitivity and Specificity for detection of ICH; processing time
Rapid ICH is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced head CT images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Intracranial Hemorrhage (ICH) pathologies in head CT images, including IPH, IVH, SAH, and SDH/EDH intracranial hemorrhages. Rapid ICH uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH findings. Notifications include compressed preview images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of Rapid ICH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified radiologists are responsible for viewing full images per the standard of care. Rapid ICH is for use in ≥18 yr olds.
Device Story
Rapid ICH (v4.0) is a radiological computer-aided triage and notification software; operates within the Rapid Platform; processes non-enhanced head CT (NCCT) images. AI/ML algorithm analyzes images in parallel to standard-of-care interpretation; detects suspected intracranial hemorrhage (ICH). Outputs include priority notifications to clinicians via PACS, workstation, email, or mobile; includes compressed preview images for informational purposes. Used in hospital networks; operated by radiologists. Does not alter original images; not for diagnostic use. Benefits include improved workflow triage and prioritization of urgent cases.
Clinical Evidence
Retrospective, blinded, multicenter, multinational study (n=311 analyzed cases). Primary endpoint: sensitivity and specificity for ICH detection. Results: Sensitivity 0.949 (95% CI: 0.899, 0.975); Specificity 0.977 (95% CI: 0.942, 0.991). Secondary endpoint: median processing time 127s (range 114s-773s). Subgroup analysis performed for age, gender, scanner manufacturer, and ICH subtype. Ground truth established by three expert neuro-radiologists.
Technological Characteristics
Software-based AI/ML/Neural Network module. Operates on server or standalone desktop. Compliant with EN ISO 14971:2019, IEC 62304:2006, IEC 62366:2015, NEMA PS 3.1-3.20 (DICOM), and UL 2900-1 (cybersecurity). No hardware components; no alteration of original image data.
Indications for Use
Indicated for use in adults (≥18 years) to assist hospital networks and radiologists in workflow triage by flagging suspected intracranial hemorrhage (IPH, IVH, SAH, SDH/EDH) in non-enhanced head CT images.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
October 9, 2026
Ischemaview, Inc.
James Rosa
Chief Regulatory and Compliance Officer
1120 Washington Ave.
Suite 200
Golden, CO 80401
Re: K262301
Trade/Device Name: Rapid ICH (v4.0)
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: September 7, 2026
Received: September 8, 2026
Dear James Rosa:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262301 - James Rosa
Page 2
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP), Rapid ICH Predetermined Change Control Plan, Version C. Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device
{3}
K262301 - James Rosa
Page 3
Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262301 | ? |
| Please provide the device trade name(s). | | ? |
| Rapid ICH (v4.0) | | |
| Please provide your Indications for Use below. | | ? |
| Rapid ICH is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced head CT images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Intracranial Hemorrhage (ICH) pathologies in head CT images, including IPH, IVH, SAH, and SDH/EDH intracranial hemorrhages. | | |
| Rapid ICH uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH findings. Notifications include compressed preview images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. | | |
| The results of Rapid ICH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified radiologists are responsible for viewing full images per the standard of care. | | |
| Rapid ICH is for use in ≥18 yr olds. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
iSchemaView - Traditional 510(k) Rapid ICH
510(k) Summary
K262301
### 510(k) Summary
### iSchemaView, Inc.'s Rapid ICH
This document contains the 510(k) summary for the iSchemaView Rapid ICH (v4) device. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
**Applicant Name and Address:**
| Name: | iSchemaView, Inc. |
| --- | --- |
| Address: | 1120 Washington St., Suite 200 Golden, CO 80401 |
| Official Contact: | Jim Rosa Phone: (303) 704-3374 Email: rosa@ischemaview.com |
| --- | --- |
**Summary Preparation Date:**
6 June, 2026
**Device Name and Classification:**
| Trade Name: | Rapid ICH (v4.0) |
| --- | --- |
| Common Name: | Radiological Computer Aided Triage and Notification Software |
| Classification: | II |
| Product Code: | Primary: QAS |
| Regulation No: | 21 C.F.R. §892.2080 |
| Classification Panel: | Radiology Devices |
**Predicate Device:**
The iSchemaView Rapid ICH (v4) Triage and Notification Module is claimed to be substantially equivalent to the legally marketed predicate, iSchemaView Rapid ICH (v3) (K221456).
**Device Description:**
Rapid ICH is a radiological computer-assisted triage and notification software device. The Rapid ICH module is a non-enhanced CT (NCCT) processing module which operates within the integrated Rapid Platform to provide triage and notification prioritization of suspected intracranial hemorrhage. The Rapid ICH module is an AI/ML module. The output of the module is a priority notification to clinicians indicating the suspicion of ICH based on positive findings. The Rapid ICH module uses the basic services supplied by the Rapid Platform including DICOM processing, job management, imaging module execution and imaging output including the notification and compressed image.
**Indications for Use:**
Rapid ICH is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced head CT images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive
Page 1 of 6
{6}
iSchemaView - Traditional 510(k) Rapid ICH 510(k) Summary
findings of Intracranial Hemorrhage (ICH) pathologies in head CT images, including IPH, IVH, SAH, and SDH/EDH intracranial hemorrhages.
Rapid ICH uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH findings. Notifications include compressed preview images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of Rapid ICH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified radiologists are responsible for viewing full images per the standard of care.
Rapid ICH is for use in ≥18 yr olds.
# Substantial Equivalence and Comparison of Technological Characteristics:
Rapid ICH (v4) does not raise new questions of safety or effectiveness compared to the previously cleared Rapid ICH (v3 under K221456). Both devices are radiological computer-aided triage and notification software applications for use with non-contrast CT input. The minor change for this filing is the update to include recent training data and validation. Thus, the Rapid ICH (v4) software is substantially equivalent.
The following table summarizes and compares data on the Rapid ICH (K221456) to the subject device of this Traditional 510(k) submission. A table comparing the key features of the subject and predicate devices is provided below:
| Substantial Equivalence Table | | |
| --- | --- | --- |
| Comparison Feature | Rapid ICH(v3) (K221456) | Rapid ICH (v4) – Subject Device |
| Indications for Use | Rapid ICH is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced head CT images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of pathologies in head CT images, for IPH, IVH, SAH, and SDH Intracranial Hemorrhages (ICH). Rapid ICH uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a server or standalone desktop application in parallel to the ongoing standard of | Rapid ICH is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced head CT images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of pathologies in head CT images, for IPH, IVH, SAH, and SDH/EDH Intracranial Hemorrhages (ICH). Rapid ICH uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a server or standalone desktop application in parallel to the ongoing standard of |
Page 2 of 6
{7}
# iSchemaView - Traditional 510(k) Rapid ICH
510(k) Summary
| Substantial Equivalence Table | | |
| --- | --- | --- |
| Comparison Feature | Rapid ICH(v3) (K221456) | Rapid ICH (v4) – Subject Device |
| | care image interpretation. The user is presented with notifications for cases with suspected ICH findings. Notifications include compressed preview images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of Rapid ICH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified radiologists are responsible for viewing full images per the standard of care | care image interpretation. The user is presented with notifications for cases with suspected ICH findings. Notifications include compressed preview images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of Rapid ICH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified radiologists are responsible for viewing full images per the standard of care |
| Anatomy | Head | Head |
| Removal of cases from worklist queue | No | No |
| Primary Imaging Modalities | NCCT | NCCT |
| Technical Implementation | AI/ML/Neural Network | AI/ML/Neural Network |
| Segmentation of ROI | No, the device does not highlight or direct a user’s attention to a specific location in the image file. | No, the device does not highlight or direct a user’s attention to a specific location in the image file. |
| Preview Images | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes. | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes. |
| Primary User(s) | Clinician | Radiologist |
| Alteration of original image data | No | No |
| Alters Standard of Care | In parallel to | In parallel to |
Page 3 of 6
{8}
iSchemaView - Traditional 510(k) Rapid ICH 510(k) Summary
| Substantial Equivalence Table | | |
| --- | --- | --- |
| Comparison Feature | Rapid ICH(v3) (K221456) | Rapid ICH (v4) – Subject Device |
| Workflow | | |
| Notification / Prioritization | Yes – PACS, Workstation, email, mobile | Yes – PACS, Workstation, email, mobile |
# **AI/ML Module Development:**
The training data (4399 cases) from a diverse group of sites in the US and OUS included selected confounders and phantoms to train against a range of cases for generalization to cover the subgroup demographics and equipment manufacturers. Phantoms were used to address specific technical characteristics such as motion detection.
# **Clinical Characteristics:**
The primary users of Rapid ICH are radiologists to support workflow prioritization.
# **Performance Standards:**
Rapid ICH has been developed in conformance with the following standards, as applicable:
EN ISO 14971:2019 (R2021)
Application of Risk Management to Medical Devices
IEC 62304:2006 (R2015)
Medical device software – Software lifecycle processes
IEC 62366:2015 (R2020)
Application of Usability Engineering to Medical Devices
NEMA PS 3.1 - 3.20
Digital Imaging and Communications in Medicine (DICOM)
UL 2900-1 (2017)
Standard for Safety for Cybersecurity Network-Connected Products
# **Performance Data:**
Performance validation was conducted through a retrospective, blinded, multicenter, multinational study with Rapid ICH with the primary endpoint to evaluate the software’s performance in CT scans containing intracranial hemorrhage (ICH). The performance data is derived from 320 samples with 311 used for the primary endpoint, 9 Cases were removed due to exclusion criteria. The primary endpoint was exceeded with Se: 0.949 (0.899,0.975) and Sp: 0.977 (0.942, 0.991). Secondary endpoint was met, with a median processing time of 127s and a range of 114s to 773s; 90% of cases processed in under 3 minutes. Reported processing times are observed values for this version and dataset and are not acceptance criteria.
All validation data was Truthed using three (3) expert neuro-radiologists. The data was obtained from 21 sources: 13 named clinical institutions, of which 4 are outside the United States, and 8 data providers, imaging core laboratories or multicentre study datasets. Sources of the latter kind aggregate more than one contributing institution, so the number of contributing sites exceeds the number of sources listed. Of the 311 analysed cases, 250 (80.4%) are of United States origin and 61 (19.6%) are not. Confounders included:
Page 4 of 6
{9}
iSchemaView - Traditional 510(k) Rapid ICH
510(k) Summary
Calcification, Foreign Object(s), Hyperdense Vasculature, Infarct, Meningeal Hyperdensity and Tumor. As a sub-analysis, Truthers identified 47 cases specific to trauma related injury which showed Se: 1.00 (0.896,1.0) and Sp: 0.929 (0.685,0.987).
# **Subgroup Performance:**
Additional subgroup analysis was performed against age, gender, manufacturer, scanner model, reconstruction kernel, US/OUS, slice thickness, ICH subtype, hemorrhage volume and confounder presence, showing acceptable performance for generalization. Subtype and volume results rest on small samples in some categories and are reported descriptively; the weakest is subarachnoid hemorrhage at Se 0.800 (0.490, 0.943) on 10 cases.
# **Predetermined Change Control Plan (PCCP):**
Rapid ICH has an authorized Predetermined Change Control Plan (PCCP). All modification types are restricted to non-enhanced head CT; no modification will introduce a new imaging modality or a new acquisition type.
**Approved Planned Modifications:** (1) Device Input Compatibility — DICOM pre-processing and compatibility adjustments, no algorithm weight change; (2) CT Scanner Expansion — algorithm retraining to add new CT scanner models or reconstruction methods within CT modality only; (3) Performance Improvement via Data Expansion — algorithm retraining on expanded CT datasets triggered by >10% Se or Sp drift; (4) Processing Time Optimization — latency reduction (DICOM pipeline refactoring, inference quantization, parallelization, model pruning with confirmed output equivalence) with no change to algorithm weights or clinical outputs.
**Testing Methods and Acceptance Criteria:** Retraining modifications (Types 2 and 3): three-part sub-analysis (Original / New Cases / Integrated); primary endpoint aggregate Se and Sp ≥80% lower bound 97.5% CI; non-inferiority per subgroup (scanner manufacturer; age cohort 18–21, 22–50, 51–69 and ≥70 years; sex; ICH subtype IPH, IVH, SAH, SDH, EDH) with approved margins. Processing Time (Type 4): mean Time to Notification non-inferior to the prior authorized version.
**User Notification:** Software updates are deployed manually; updated labeling accompanies each update with a description of the modification, current Se/Sp and Time to Notification metrics, and a version history of all PCCP-authorized changes. Hospital installation personnel are notified at time of deployment.
# **Prescriptive Statement:**
Caution: Federal law restricts this device to sale by or on the order of a physician.
# **Safety & Effectiveness:**
Rapid ICH is designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 (risk management). The Rapid ICH performance has been validated with case data.
Page 5 of 6
{10}
iSchemaView - Traditional 510(k) Rapid ICH
510(k) Summary
**Conclusion:**
In conclusion, the iSchemaView Rapid ICH (v4) software device is substantially equivalent in intended use, technological characteristics, safety and performance characteristics to the legally marketed predicate device, Rapid ICH (K221456).
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.