Gastric Alimetry

K262292 · Alimetry , Ltd. · MYE · Sep 4, 2026 · Gastroenterology, Urology

Device Facts

Record IDK262292
Device NameGastric Alimetry
ApplicantAlimetry , Ltd.
Product CodeMYE · Gastroenterology, Urology
Decision DateSep 4, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1735
Device ClassClass 2
AttributesPediatric

Indications for Use

The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders. The Gastric Alimetry System is indicated for patients 12 years and older.

Device Story

Gastric Alimetry is an electrogastrography (EGG) system for non-invasive measurement of gastric myoelectrical activity. Input: surface myoelectrical signals captured via a 64-channel disposable adhesive array. Operation: Alimetry Reader digitizes signals; Alimetry App (iPad mini) manages setup and patient-reported symptoms; Alimetry Cloud provides secure portal for clinician access. Output: Gastric Alimetry Report containing myoelectrical waveforms, spectral analysis, symptom scores, and automated phenotype classification (Auckland Classification v1.0). Used in healthcare facilities by clinicians. The system aids diagnosis by quantifying gastric rhythm, amplitude, and meal response (Meal Response Ratio). Benefits include objective assessment of gastric myoelectrical function to support clinical decision-making for gastric disorders.

Clinical Evidence

No new clinical trials were required for this submission. The device relies on previously established clinical data for the predicate system. Pediatric reference intervals (ages 12-17) were established via a new study, and the Auckland Classification v1.0 framework is derived from expert consensus (Varghese et al., 2026). Software verification and validation testing were performed to confirm the safety and performance of the updated reporting and app features.

Technological Characteristics

System includes Alimetry Reader, 64-channel disposable electrode array (ANSI/AAMI EC12:2000 compliant), and charging dock. Connectivity via iPad mini (MDM-managed) and secure cloud. Non-invasive, battery-powered. Software uses fixed, deterministic rules for signal processing and automated phenotyping. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for patients 12 years and older as an aid in the diagnosis of various gastric disorders.

Regulatory Classification

Identification

An electrogastrography system (EGG) is a device used to measure gastric myoelectrical activity as an aid in the diagnosis of gastric motility disorders. The device system includes the external recorder, amplifier, skin electrodes, strip chart, cables, analytical software, and other accessories.

Special Controls

*Classification.* Class II (Special Controls). The special controls are as follows:(1) The sale, distribution and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter. (2) The labeling must include specific instructions: (i) To describe proper patient set-up prior to the start of the test, including the proper placement of electrodes; (ii) To describe how background data should be gathered and used to eliminate artifact in the data signal; (iii) To describe the test protocol (including the measurement of baseline data) that may be followed to obtain the EGG signal; and (iv) To explain how data results may be interpreted. (3) The device design should ensure that the EGG signal is distinguishable from background noise that may interfere with the true gastric myoelectric signal. (4) Data should be collected to demonstrate that the device has adequate precision and the EGG signal is reproducible and is interpretable.

In addition to the general controls of the act, the 3CPM EGG Machine is subject to the following special controls: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include specific instructions (a) to describe proper patient set-up prior to the start of the test, including the proper placement of electrodes; (b) to describe how background data should be gathered and used to eliminate artifact in the data signal; (c) to describe the test protocol (including the measurement of baseline data) which may be followed to obtain the EGG signal; and (d) to explain how data results may be interpreted. (3) The device design should ensure that the EGG signal is distinguishable from background noise that may interfere with the true gastric myoelectric signal. (4) Data should be collected to demonstrate that the device has adequate precision and the EGG signal is reproducible and is interpretable.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 4, 2026 Alimetry, Ltd. % Lina Kontos Partner Hogan Lovells Cadwalader US LLP 555 Thirteenth St. NW Washington, District of Columbia 20004 Re: K262292 Trade/Device Name: Gastric Alimetry Regulation Number: 21 CFR 876.1735 Regulation Name: Electrogastrography System Regulatory Class: Class II Product Code: MYE Dated: July 6, 2026 Received: July 7, 2026 Dear Lina Kontos: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K262292 - Lina Kontos Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K262292 - Lina Kontos Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, SIVAKAMI VENKATACHALAM -S *for* Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262292 | ? | | Please provide the device trade name(s). | | ? | | Gastric Alimetry | | | | Please provide your Indications for Use below. | | ? | | The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders. The Gastric Alimetry System is indicated for patients 12 years and older. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? | {4} K262292 Page 1 of 6 ### 510(k) SUMMARY ### Alimetry's Gastric Alimetry ### Submitter Alimetry Ltd. Address: 86 Symonds St. Grafton Auckland 1010 New Zealand Phone: +64 27 622 2306 Contact Person: Yaara Yarmut, Chief Regulatory Officer Date Prepared: September 1, 2026 Name of Device: Gastric Alimetry Common or Usual Name: Gastric Alimetry System Classification Name: Electrogastrography system, 21 CFR 876.1735 Regulatory Class: Class II Product Code: MYE ### Predicate Devices Trade Name: Gastric Alimetry (K252504) (primary predicate) Manufacturer: Alimetry Ltd. ### Device Description The Gastric Alimetry System is an electrogastrography (EGG) device, used for non-invasively measuring the myoelectrical activity of the stomach at the surface of the abdomen. The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders. The device is used to acquire and digitize gastric myoelectrical data and movement artifacts using the Alimetry Reader connected to a single use Gastric Alimetry Array which includes electrodes on an adhesive patch used for recording the myoelectrical data from the skin surface. The Gastric Alimetry App runs on an iPad mini and is used to set up the device and capture patient-reported symptom data. The Gastric Alimetry Report is provided to the clinicians at the end of the test and includes myoelectrical signal data for manual analysis, together with computed data summaries and plots. A Supplementary Report is also routinely available to clinicians that includes signal data from all 64 channels on the array. The Alimetry Cloud acts as {5} K262292 Page 2 of 6 a secure website portal for physicians to access Gastric Alimetry Reports. The Gastric Alimetry Dock (Accessory) is used to guide alignment of the Array and Reader during the setup procedure and charge the Reader. The Gastric Alimetry System is non-invasive and used in healthcare facilities. This submission addresses the following minor modifications: - Updates to the Gastric Alimetry Report, including: - Addition of automated phenotyping as defined by the Auckland Classification v1.0, a published framework derived by expert consensus (Varghese et al., 2026) which formalizes the phenotype information provided with the predicate. - Addition of the Meal Response Ratio for quantifying the timing of the gastric response to a meal stimulus. - Addition of reference intervals for patients aged between 12-17 years. - Updates to the App to add an option for healthcare facilities to use iPads linked to their own mobile device management (MDM) software, instead of the Alimetry-provided MDM. Additional minor updates were made to the Reader, Cloud, Report, App and labeling. ### Intended Use / Indications for Use The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders. The Gastric Alimetry System is indicated for use in patients 12 years and older. There are no changes to the intended use in the modified Gastric Alimetry System, which has the same intended use as the predicate Gastric Alimetry System (K252504). ### Technological Characteristics / Substantial Equivalence The modified Gastric Alimetry System retains the same core technological characteristics as the predicate Gastric Alimetry system. Both systems utilize identical hardware components: the Alimetry Reader, Gastric Alimetry Array, Gastric Alimetry Dock, and a commercial off-the-shelf iPad mini managed by a mobile device management system (MDM). iPad management using healthcare facilities' MDM is substantially equivalent to the use of Alimetry's MDM as it still requires the iPad to be managed by an MDM with the same controls. Both systems also utilize an equivalent secure Alimetry Cloud for data storage and access. The fundamental function of recording, storing, viewing, and processing gastric myoelectrical activity remains unchanged, and the general content of the generated reports is similar, with minor modifications: Modifications to the Gastric Alimetry Report, including: Addition of automated phenotyping as defined by the Auckland Classification v1.0, a framework derived by expert consensus which formalizes the phenotype information provided with the predicate. The phenotypes according to the Auckland Classification include six main phenotypes (Dysrhythmic, High Frequency, Low Meal Response, Sensorimotor, Continuous, and Late Onset Symptoms) and two emerging phenotypes (Low Frequency and High Amplitude); Addition of the 'Meal Response Ratio' (MRR) to serve as the primary indicator of postprandial meal response; Addition of reference intervals for patients aged between 12-17 years. Adult reference intervals were previously established and following a {6} K262292 Page 3 of 6 new study conducted in participants ages 12–17 years, reference intervals have now been established for this population as well. Note that “automated” throughout this Summary refers to the application of fixed, deterministic rules, and does not denote the use of artificial intelligence or machine learning as part of the classification functions of the software. A table comparing the key features of the subject and predicate devices is provided below. | | Gastric Alimetry System | Gastric Alimetry System (predicate) | Discussion | | --- | --- | --- | --- | | Intended Use | To record, store, view and process gastric myoelectrical activity as an aid to the diagnosis of various gastric disorders. | To record, store, view and process gastric myoelectrical activity as an aid to the diagnosis of various gastric disorders. | Same as the predicate. | | Indications for Use | Patients 12 years and older | Patients 12 years and older | Same as the predicate. | | Technological Characteristics | | | | | Setup and Charging Dock | Gastric Alimetry Dock for charging and user convenience | Gastric Alimetry Dock for charging and user convenience | Same as the predicate. | | Mobile Device Management | iPad managed either by Alimetry's MDM, or by the clinic's MDM | iPad managed via Alimetry's MDM | Similar. iPad still has to be managed by an MDM with the same controls. | | Electrodes | Disposable; same components as ECG electrodes. Peel-and-stick patch. | Disposable; same components as ECG electrodes. Peel-and-stick patch. | Same as the predicate. | | Weight of device on patient | 225 grams - Reader and Array | 225 grams - Reader and Array | Same as the predicate. | | Sampling Frequency | raw sampling: 250 Hz logging: 4 Hz | raw sampling: 250 Hz logging: 4 Hz | Same as the predicate. | | Low Frequency Range | DC | DC | Same as the predicate. | | High frequency range | 2 Hz | 2 Hz | Same as the predicate. | | Number of channels | 64 (+2 reference) | 64 (+2 reference) | Same as the predicate. | | Electrode to recorder interface | Reader located on the Array and directly connected to it. | Reader located on the Array and directly connected to it. | Same as the predicate. | | Screen | Dedicated tablet for system operation | Dedicated tablet for system operation | Same as the predicate. | {7} K262292 Page 4 of 6 | Motion Sensor | Accelerometer | Accelerometer | Same as the predicate. | | --- | --- | --- | --- | | Power Source | Battery powered | Battery powered | Same as the predicate. | | Reporting Features | | | | | Visualization of myoelectrical waveforms | Yes, all channels. | Yes, all channels. | Same as the predicate. | | Spectral Analysis | Yes | Yes | Same as predicate. | | Frequency and Amplitude of Myoelectrical Activity | Yes | Yes. | Same as predicate. | | Other Myoelectrical Parameters | Measure of rhythmic stability Measure of gastric response to meal stimulus presented as 'Meal Response Ratio'. Other measures of myoelectrical activity can be assessed visually using the spectrogram, amplitude plot, and/or signal traces. | Measure of rhythmic stability. Measure of gastric response to meal stimulus presented as 'Fed:Fasted Amplitude Ratio'. Other measures of myoelectrical activity can be assessed visually using the spectrogram, amplitude plot, and/or signal traces. | Similar. Presentation is comparable. The change from ff-AR to Meal Response Ratio as the primary measure of gastric response to meal stimulus reflects the International BSGM Working Group's recommendation (Varghese et al., 2026). This change does not raise different questions of safety or efficacy. The subject device includes reference intervals for patients aged 12 years and older, whereas the predicate limited the reference intervals to patients aged 18 and older. | | Symptom Outputs | Severity scores for each symptomTotal Symptom Burden ScoreIdentification of patterns of symptoms (Meal Induced, Meal | Severity scores for each symptomTotal Symptom Burden ScoreIdentification of patterns of symptoms (Meal Induced, Meal | Same as predicate. | {8} K262292 Page 5 of 6 | | Alleviated, Late Onset, Continuous, or none) • Summary of Gut-Brain Wellbeing symptom scores ('depressive symptoms', 'stress symptoms', and 'anxiety symptoms') • Gut-Brain Wellbeing Total Score | Alleviated, Late Onset, Continuous, or none) • Summary of Gut-Brain Wellbeing symptom scores ('depressive symptoms', 'stress symptoms', and 'anxiety symptoms') • Gut-Brain Wellbeing Total Score | | | --- | --- | --- | --- | | **Clinical Summary** | Report contains all key information on a summary page along with a prepopulated field with an assessment of the relevant patient classifications based on the Auckland Classification phenotypes. | Report contains all key information on a summary page. Spectral results are summarized with a summary sentence at the top of the page. A box at the bottom page is provided for the clinician to document their summary. Phenotype classification provided in supplemental labelling materials. | Similar. The added automated assessment based on test outputs existing in the predicate report, and all of these outputs are still displayed prominently in the subject report so as to enable independent interpretation as would be performed in using the predicate system. | | **Technical / recording quality outputs** | Yes. Impedance, artifacts, test protocol (test duration, meal type, and meal consumption), BMI. Movement, Array position and spatial distribution of amplitude. Test quality cautions are also included if quality parameters are outside the recommended range. | Yes. Impedance, artifacts, test protocol (test duration, meal type, and meal consumption), BMI. Movement, Array position and spatial distribution of amplitude Test quality cautions are also included if quality parameters are outside the recommended range. | Same as the predicate. | | **Safety Features** | | | | | **Biocompatibility** | Array and Reader in contact with the patient. | Array and Reader in contact with the patient. | Same as the predicate. No changes to any patient-contact materials. | | **Software** | Setup (including clinical data), device control, data acquisition software and | Setup (including clinical data), device control, data acquisition software and | Same as the predicate. | {9} K262292 Page 6 of 6 | | system checks via App running on a tablet. | system checks via App running on a tablet. | | | --- | --- | --- | --- | | System Checks | Impedance monitor and connectivity check displayed prior to recordings. | Impedance monitor and connectivity check displayed prior to recordings. | Same as the predicate. | | Sterilization | Electrodes are disposable, non-sterile. Reader and Dock are reprocessed, not supplied sterile. | Electrodes are disposable, non-sterile. Reader and Dock are reprocessed, not supplied sterile. | Same as the predicate. | | Standards with which the Device Complies | | | | | Electrodes | ANSI/AAMI EC12:2000 – compliance with relevant requirements. | ANSI/AAMI EC12:2000 – compliance with relevant requirements. | Same as the predicate. | | Medical Electrical Equipment | IEC 60601-1 | IEC 60601-1 | Same as the predicate. | | Electromagnetic Compatibility | IEC 60601-1-2 | IEC 60601-1-2 | Same as the predicate. | ### Performance Data The modifications to the device since the prior clearances, namely modifications to the App and the Report, including addition of the Auckland Classification V1.0, addition the Meal Response Ratio metric, addition of reference intervals for patients aged between 12-17 years and addition of an option for healthcare facilities to use iPads linked to their own mobile device management (MDM) software, did not significantly impact the safety or performance of the device and did not require submission of new bench testing data beyond software verification and validation testing. ### Conclusions The modified Gastric Alimetry System is as safe and effective as the Gastric Alimetry System. The Gastric Alimetry System has the same intended uses and indications, and similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the modified Gastric Alimetry System and its predicate device raise no new or different questions of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...