K262268 · Fisher and Paykel Healthcare Limited · BTT · Aug 27, 2026 · Anesthesiology
Device Facts
Record ID
K262268
Device Name
Optiflow Oxygen Kit (AA484J)
Applicant
Fisher and Paykel Healthcare Limited
Product Code
BTT · Anesthesiology
Decision Date
Aug 27, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This product delivers respiratory gases to adult patients. It is intended for use with an F&P 820 humidifier at flows from 5 to 70 L/min. This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup. This product is indicated for the delivery of nasal high flow (NHF) by appropriately qualified healthcare professionals under the direction of a physician anesthesiologist in a medical procedure or surgical room. Qualitative carbon dioxide sampling can be used at nasal interface flow rates from 5 to 50 L/min. This product can be used for preoxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms under the direction of a physician anesthesiologist. This product is not intended for apneic ventilation.
Device Story
AA484J Optiflow Oxygen Kit is a multi-patient use breathing set; delivers humidified respiratory gases to adult patients. Components include dry line, humidification chamber, inspiratory limb, tubing clips, date-change stickers. Used with F&P 820 humidifier; flow range 5-70 L/min. Enables nasal high flow (NHF) delivery; supports qualitative CO2 sampling (5-50 L/min) and preoxygenation/supplemental oxygenation during intubation. Operated by qualified healthcare professionals in hospital settings (OR/procedure rooms). Requires hydrophobic filter for multi-patient use; maximum 24-hour use after setup; reprocessed between patients (max 30 cycles). Output is humidified gas stream; assists clinicians in airway management and oxygenation; benefits patients by providing respiratory support during procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and biocompatibility evaluation per ISO 10993-1:2020 and ISO 18562-1:2024.
Technological Characteristics
Multi-patient use breathing set; non-sterile. Components: dry line, humidification chamber, inspiratory limb. Flow range: 5-70 L/min. Compatible with F&P 820 humidifier. Biocompatibility per ISO 10993-1 and ISO 18562-1. Reusable (max 30 cycles).
Indications for Use
Indicated for adult patients requiring delivery of nasal high flow (NHF) respiratory gases, preoxygenation, or short-term supplemental oxygenation during intubation. Used in medical procedure or surgical rooms by qualified healthcare professionals under physician anesthesiologist direction. Contraindicated for apneic ventilation.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 27, 2026
Fisher and Paykel Healthcare Limited
Amelia Rosario
Regulatory Affairs Associate
15 Maurice Paykel Pl., E. Tamaki
Auckland, 2013
New Zealand
Re: K262268
Trade/Device Name: Optiflow Oxygen Kit (AA484J)
Regulation Number: 21 CFR 868.5450
Regulation Name: Respiratory Gas Humidifier
Regulatory Class: Class II
Product Code: BTT
Dated: July 28, 2026
Received: July 29, 2026
Dear Amelia Rosario:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K262268 - Amelia Rosario
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262268 - Amelia Rosario
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assitant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K262268
?
Please provide the device trade name(s).
?
Optiflow Oxygen Kit (AA484J)
Please provide your Indications for Use below.
?
This product delivers respiratory gases to adult patients. It is intended for use with an F&P 820 humidifier at flows from 5 to 70 L/min.
This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup.
This product is indicated for the delivery of nasal high flow (NHF) by appropriately qualified healthcare professionals under the direction of a physician anesthesiologist in a medical procedure or surgical room. Qualitative carbon dioxide sampling can be used at nasal interface flow rates from 5 to 50 L/min.
This product can be used for preoxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms under the direction of a physician anesthesiologist.
This product is not intended for apneic ventilation.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Please select the age group(s) for which the device(s) is to be used.
☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☐ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater)
?
?
{4}
AA484J Optiflow Oxygen Kit
# 510(k) Summary
As Required by 21 CFR 807.92
## I. SUBMITTER
| Company Name and Address | Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place East Tamaki Auckland 2013, New Zealand Telephone: +64 9 574 0100 |
| --- | --- |
| Prepared and Submitted by | Amelia Rosario Regulatory Affairs Associate |
| --- | --- |
| Contact Person | Reena Daken Regulatory Affairs Manager – North America Telephone: +64 9 574 0100 Email: Reena.Daken@fphcare.co.nz |
| --- | --- |
| Date Prepared | 18 June 2026 |
| --- | --- |
## II. DEVICE
| Name of Device | AA484J Optiflow Oxygen Kit |
| --- | --- |
| Common/Usual Name | Respiratory Gas Humidifier |
| Classification Name | 21 CFR 868.5450 |
| Regulatory Class | II |
| Product Code | BTT |
## III. PREDICATE DEVICE
Predicate Device: F&P Optiflow Oxygen Kit (AA403), K211096.
Reference device 1: F&P Optiflow Oxygen Kit Valve Coated (AA404), K234053
Reference device 2: AA451J Optiflow Oxygen Kit, K233821
Reference device 3: MR325 F&P 820 Chamber Manual Fill 500ml, K223684
Page 1 of 5
{5}
AA484J Optiflow Oxygen Kit
#### IV. DEVICE DESCRIPTION
The AA484J Optiflow Oxygen Kit is a multiple-patient use breathing set.
The kit consists of a dry line, humidification chamber, inspiratory limb, tubing clips and date-change stickers.
The AA484J delivers humidified respiratory gases at flows from 5 to 70 L/min. When used with an FDA-cleared hydrophobic filter, the AA484J kit can be used on multiple patients. The kit is reprocessed between each patient. The kit can be exposed to a maximum of 30 reprocessing cycles and used for a maximum of 24 hours after setup.
When used with the appropriate accessories, the AA484J enables the delivery of nasal high flow (NHF) to adult patients by appropriately qualified healthcare professionals who perform anaesthesia care and airway management.
The AA484J is prescription only and provided in a non-sterile state.
#### V. INDICATIONS FOR USE
This product delivers respiratory gases to adult patients. It is intended for use with an F&P 820 humidifier at flows from 5 to 70 L/min.
This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup.
This product is indicated for the delivery of nasal high flow (NHF) by appropriately qualified healthcare professionals under the direction of a physician anesthesiologist in a medical procedure or surgical room.
Qualitative carbon dioxide sampling can be used at nasal interface flow rates from 5 to 50 L/min.
This product can be used for preoxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms under the direction of a physician anesthesiologist.
This product is not intended for apneic ventilation.
Page 2 of 5
{6}
AA484J Optiflow Oxygen Kit
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Design/technological characteristics | Subject device AA484J Optiflow Oxygen Kit | Predicate device F&P Optiflow Oxygen Kit (AA403) (K211096) | Comments |
| --- | --- | --- | --- |
| **Indications for use and intended use** | | | |
| Indications for Use | This product delivers respiratory gases to adult patients. It is intended for use with an F&P 820 humidifier at flows from 5 to 70 L/min. This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup. This product is indicated for the delivery of nasal high flow (NHF) by appropriately qualified healthcare professionals under the direction of a physician anesthesiologist in a medical procedure or surgical room. Qualitative carbon dioxide sampling can be used at nasal interface flow rates from 5 to 50 L/min. This product can be used for preoxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms under the direction of a physician anesthesiologist. This product is not intended for apneic ventilation. | This product delivers respiratory gases to adult patients. It is intended for use with an MR810 humidifier at flows from 5 to 70 L/min. This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup. This product is indicated for the delivery of Nasal High Flow (NHF) by appropriately qualified healthcare professionals under the direction of a physician anesthesiologist in a medical procedure or surgical room. Qualitative carbon dioxide sampling can be used at nasal interface flow rates from 5 to 50 L/min. This product can be used for preoxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms under the direction of a physician anesthesiologist. This product is not intended for apneic ventilation. | Identical |
| Availability | Prescription use (Part 21 CFR 801 Subpart D) | Prescription use (Part 21 CFR 801 Subpart D) | Identical |
| Patient Population | Adult patients | Adult patients | Identical |
Page 3 of 5
{7}
AA484J Optiflow Oxygen Kit
| Design/technological characteristics | Subject device AA484J Optiflow Oxygen Kit | Predicate device F&P Optiflow Oxygen Kit (AA403) (K211096) | Comments |
| --- | --- | --- | --- |
| Intended Use Environment | Hospital Environment | Hospital Environment | Identical |
| **Operation and safety features** | | | |
| Ambient Operating Temperature | 18 to 26 °C (64.4 to 78.8°F) | 18 to 26 °C (64.4 to 78.8°F) | Identical |
| System Specifications | **Flow Range**: 5 to 70 L/min | **Flow Range**: 5 to 70 L/min | Identical |
| Shelf Life | 2 years | 2 years | Identical |
| Storage Temperature | 10 °C to +50 °C (14 to 122 °F) | -10 °C to +50 °C (14 to 122 °F) | Identical |
| Sterility | Device not provided sterile. | Device not provided sterile. | Identical |
| Reusability | Multi-patient use | Multi-patient use | Identical |
| Disposal | Dispose of product safely in accordance with standard hospital procedure. | Dispose of product safely in accordance with standard hospital procedure. | Identical |
| **Components and Materials** | | | |
| Inspiratory Tube | Included | Included | Identical |
| Water/Humidification Chamber | Included | Included | Similar |
| Clip | Included | Included | Identical |
| Dryline | Included | Included | Similar |
Page 4 of 5
{8}
Optiflow Switch+ Filtered Nasal Interface with CO₂ Sampling
## VII. PERFORMANCE DATA
### Summary of Biocompatibility Testing
Performance testing was completed using the same FDA recognized consensus standards as the predicate device and confirms the subject device does not raise new questions of safety and effectiveness. The testing demonstrates substantial equivalence of the subject device to the predicate device.
The subject device has been tested to the applicable requirements of the following standards:
- ISO 10993-1:2020 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 18562-1:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
## VIII. CONCLUSION
The subject device is substantially equivalent to the predicate based on patient population, intended uses, comparison of the technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the device is substantially equivalent to the legally marketed predicate devices.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.