RIVA® Posterior Fixation System with NanoACTIV Surface Technology; RIVA® Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology
Applicant
Xenix Medical
Product Code
NKB · Orthopedic
Decision Date
Sep 14, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Xenix Medical RIVA Posterior Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion as a pedicle screw fixation system or anterolateral fixation, in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine: - Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) - Spondylolisthesis - Trauma (i.e. fracture or dislocation) - Spinal Stenosis - Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis) - Tumor - Pseudoarthrosis - Failed previous fusion Xenix Medical RIVA Posterior Fixation System fenestrated screws are intended to be used with saline or radiopaque dye. Xenix Medical RIVA Posterior Fixation System Navigation Instruments are intended to be used with the Xenix Medical RIVA Posterior Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic® StealthStation® S8 System, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy. The RIVA Anchor Sacroiliac Fixation System is intended for sacroiliac joint fusion for: - Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months. - To augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion. - Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint. When connected to the Xenix Medical RIVA Posterior Fixation System, the RIVA Anchor Sacroiliac Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion as a pedicle screw fixation system or anterolateral fixation, in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine: - Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) - Spondylolisthesis - Trauma (i.e. fracture or dislocation) - Spinal Stenosis - Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis) - Tumor - Pseudoarthrosis - Failed previous fusion Xenix Medical fenestrated screws are intended to be used with saline or radiopaque dye. Xenix Medical RIVA Posterior Fixation System Navigation Instruments are intended to be used with the Xenix Medical RIVA Anchor Sacroiliac Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic® StealthStation® S8 System, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
System comprises titanium/cobalt-chrome spinal implants (pedicle screws, rods, connectors) and navigation instruments; used in thoracic, lumbar, and sacral spine surgery. Implants feature NanoACTIV micro/nano-roughened surface (1-100 nm) to promote osteogenic differentiation of mesenchymal stem cells. Navigation instruments integrate with Medtronic StealthStation S8 for image-guided placement (CT/MRI/fluoroscopy). Surgeon operates system in OR to stabilize spine or sacroiliac joint; implants facilitate biological fusion. Fenestrated screws allow delivery of saline or radiopaque dye. System provides structural support during healing; NanoACTIV surface enhances bone-implant fixation.
Clinical Evidence
No clinical data. Evidence consists of in vitro evaluations and imaging studies demonstrating that the NanoACTIV surface (1-100 nm) supports mesenchymal stem cell differentiation and mineralized matrix production compared to non-treated surfaces. Mechanical testing per ASTM standards (F1717, F1798, F2193, F543) confirms structural equivalence.
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed fusion. Also indicated for sacroiliac joint fusion due to dysfunction/disruption, degenerative sacroiliitis, or fractures.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
September 14, 2026
Xenix Medical
Tyler Holschlag
Vice President Product Development
149 W. Michigan
Orlando, Florida 32806
Re: K262193
Trade/Device Name: RIVA Posterior Fixation System; RIVA Anchor Sacroiliac Fixation System
Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral pedicle screw system
Regulatory Class: Class II
Product Code: NKB, KWQ, OLO, OUR
Dated: June 29, 2026
Received: June 29, 2026
Dear Tyler Holschlag:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262193 - Tyler Holschlag
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K262193 - Tyler Holschlag
Page 3
Sincerely,
# COLIN
O'NEILL -S FDA
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K262193
Device Name
RIVA® Posterior Fixation System with NanoACTIV Surface Technology; RIVA® Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology
Indications for Use (Describe)
RIVA Posterior Fixation System with NanoActive Surface Technology
INDICATIONS:
The Xenix Medical RIVA Posterior Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion as a pedicle screw fixation system or anterolateral fixation, in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine:
- Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis
- Trauma (i.e. fracture or dislocation)
- Spinal Stenosis
- Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Failed previous fusion
Xenix Medical RIVA Posterior Fixation System fenestrated screws are intended to be used with saline or radiopaque dye. Xenix Medical RIVA Posterior Fixation System Navigation Instruments are intended to be used with the Xenix Medical RIVA Posterior Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic® StealthStation® S8 System, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
RIVA Anchor Sacroiliac Fixation System with NanoActive Surface Technology
INDICATIONS:
The RIVA Anchor Sacroiliac Fixation System is intended for sacroiliac joint fusion for:
- Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.
- To augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion.
- Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint.
When connected to the Xenix Medical RIVA Posterior Fixation System, the RIVA Anchor Sacroiliac Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion as a pedicle screw fixation system or anterolateral fixation, in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine:
- Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis
- Trauma (i.e. fracture or dislocation)
- Spinal Stenosis
- Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 445-6740
EF
{4}
Pseudoarthrosis
- Failed previous fusion
Xenix Medical fenestrated screws are intended to be used with saline or radiopaque dye.
Xenix Medical RIVA Posterior Fixation System Navigation Instruments are intended to be used with the Xenix Medical RIVA Anchor Sacroiliac Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic®
StealthStation® S8 System, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
PSC Publishing Services (301) 443-6740
EF
{5}
## 510(k) Summary
| **Device Trade Name:** | RIVA® Posterior Fixation System with NanoACTIV Surface Technology; RIVA® Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology |
| --- | --- |
| **Manufacturer:** | Xenix Medical 149 W Michigan St. Orlando, FL 32806 |
| **Contact:** | Tyler Holschlag Vice President of Product Development Xenix Medical |
| **Date Prepared:** | September 11, 2026 |
| **Classifications:** | 21 CFR §888.3070; Thoracolumbosacral pedicle screw system 21 CFR §888.3060; Spinal intervertebral body fixation orthosis 21 CFR §888.3040; Smooth or threaded metallic bone fastener 21 CFR §882.4560; Stereotaxic Instruments |
| **Class:** | II |
Product Codes:
| ***RIVA Posterior Fixation System*** | NKB, KWQ, OLO |
| --- | --- |
| ***RIVA Anchor Sacroiliac Fixation System*** | OUR, NKB, OLO |
***RIVA Posterior Fixation System with NanoACTIV Surface Technology***
**INDICATIONS:**
The Xenix Medical RIVA Posterior Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion as a pedicle screw fixation system or anterolateral fixation, in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine:
- Degenerative disc disease (defined as discogenic back pain with
K262193 1 of 6
{6}
degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis
- Trauma (i.e. fracture or dislocation)
- Spinal Stenosis
- Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Failed previous fusion
Xenix Medical RIVA Posterior Fixation System fenestrated screws are intended to be used with saline or radiopaque dye.
Xenix Medical RIVA Posterior Fixation System Navigation Instruments are intended to be used with the Xenix Medical RIVA Posterior Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic® StealthStation® S8 System, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
### RIVA Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology
INDICATIONS:
The RIVA Anchor Sacroiliac Fixation System is intended for sacroiliac joint fusion for:
- Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.
- To augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion.
- Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint.
When connected to the Xenix Medical RIVA Posterior Fixation System, the RIVA Anchor Sacroiliac Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion as a pedicle screw fixation
K262193 2 of 6
{7}
system or anterolateral fixation, in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine:
- Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis
- Trauma (i.e. fracture or dislocation)
- Spinal Stenosis
- Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Failed previous fusion
Xenix Medical fenestrated screws are intended to be used with saline or radiopaque dye. Xenix Medical RIVA Posterior Fixation System Navigation Instruments are intended to be used with the Xenix Medical RIVA Anchor Sacroiliac Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic® StealthStation® S8 System, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
### RIVA Posterior Fixation System with NanoACTIV Surface Technology
### DEVICE DESCRIPTION:
The Xenix Medical RIVA Posterior Fixation System is a non-cervical spinal fixation system used to build constructs within the body to act as temporary or permanent non-cervical spinal fixation devices and is intended for use as a posterior pedicle screw fixation system, and/or an anterolateral fixation system to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusion to occur.
The Xenix Medical RIVA Posterior Fixation System includes a variety of single-use implants manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and/or cobalt chrome alloy (Co-28Cr-6Mo per ASTM F1537 or Co-35Ni-20Cr-10Mo per ASTM F562) and is comprised of polyaxial pedicle screws, as well as connecting spinal rods, connectors, crossbars, and a separate set screw locking element. The instruments included in the Xenix Medical RIVA Posterior Fixation System facilitate the placement, adjustment, final locking, and removal, if necessary, of the system implants, and accessories to the system include trays and
K262193 3 of 6
{8}
caddies for storage, protection, and organization prior to and during the steam sterilization process.
Devices incorporate Xenix Medical's proprietary NanoACTIV™ micro and nano-roughened surface designed to improve fixation to adjacent bone. The Xenix Medical implant surfaces have been engineered with surface features at a nanometer (10⁻⁹) level, which have demonstrated the ability to elicit an endogenous cellular and biochemical response as represented by mineralization in human mesenchymal stem cells in vitro. The implant surface demonstrates elements to be considered a nanotechnology as outlined in the FDA nanotechnology guidance document.
### RIVA Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology
DEVICE DESCRIPTION:
The Xenix Medical RIVA Anchor Sacroiliac Fixation System consists of Titanium Alloy Implants (Ti-6Al-4V ELI per ASTM F-136) and surgical instrumentation for implantation. Implants are provided sterile in various lengths and diameters. The reusable instrumentation is provided non-sterile in a steam sterilization instrument tray.
Devices incorporate Xenix Medical's proprietary NanoACTIV™ micro and nano-roughened surface designed to improve fixation to adjacent bone. The Xenix Medical implant surfaces have been engineered with surface features at a nanometer (10⁻⁹) level, which have demonstrated the ability to elicit an endogenous cellular and biochemical response as represented by mineralization in human mesenchymal stem cells in vitro. The implant surface demonstrates elements to be considered a nanotechnology as outlined in the FDA nanotechnology guidance document.
Predicate Devices:
| Device Name(s) | Manufacturer | K-Number |
| --- | --- | --- |
| *Primary Predicate* | | |
| RIVA Posterior Fixation System | Xenix Medical | K251154 |
| *Reference Predicate Devices* | | |
| neoWAVE LS Lumbar Straight, neoWAVE C Cervical, and Ti3D Cervical | Xenix Medical | K222988 |
| SOLACE Sacroiliac Fixation System | Xenix Medical | K233241 |
| SI-Bone Bedrock Granite | SI-Bone | K220195 |
### Performance Testing Summary:
Engineering analysis and rationale was provided to demonstrate that the subject RIVA Fixation System NanoACTIV screws are substantially equivalent to the predicate RIVA Posterior Fixation System.
K262193 4 of 6
{9}
Engineering analysis and rationale was provided to demonstrate that the subject RIVA Anchor Sacroiliac Fixation System NanoACTIV Screws are substantially equivalent to the predicates RIVA Posterior Fixation System and SOLACE Sacroiliac Fixation System.
In vitro evaluations, imaging studies, and testing were performed to confirm that the deliberately modified NanoACTIV surface produces micro- and nano-scale surface roughness ranging in size between 1-100 nanometers, which exhibit specific osteogenic differentiation. The in vitro study results demonstrated that the modified NanoACTIV nano surface supports differentiation of mesenchymal stem cells through the osteogenic lineage and production of a mineralized matrix as compared to non-treated micro-rough AM titanium surfaces. These evaluations further demonstrated that the endogenous cellular and biochemical response elicited by the NanoACTIV surface on non-additively manufactured titanium was shown to be equivalent to that of the NanoACTIV surface on additively manufactured implants.
## Summary of Substantial Equivalence:
The Xenix Medical RIVA Posterior Fixation System with NanoACTIV Surface Technology is substantially equivalent to its primary predicate, the Xenix Medical RIVA Posterior Fixation System (K251154), with respect to intended use, indications for use, principle of operation, and design: both devices incorporate the same polyaxial pedicle screw, connecting rod, connector, crossbar, and set-screw construct, the same titanium alloy (Ti-6Al-4V ELI per ASTM F136) and cobalt chrome alloy (per ASTM F1537/F562) implant materials, the same subtractive manufacturing process, and the same implant size ranges (Ø4.5–12.5mm, 25–110mm). The principal technological differences are, first, an additional manufacturing step that imparts the proprietary NanoACTIV micro- and nano-roughened surface to the implant, a surface modification with the same nanoscale characteristics (1–100 nm) previously cleared under K233241 and K222988, and second, a change in implant sterility: the predicate device implants are supplied non-sterile for terminal moist-heat sterilization at the healthcare facility, while the subject device implants are supplied sterile via gamma irradiation, consistent with the implants cleared under K233241 and K222988. In vitro osteogenic differentiation and imaging studies characterizing the NanoACTIV surface, together with mechanical testing per ASTM F1717 and ASTM F1798, demonstrate that these differences do not raise new or different questions of safety or effectiveness.
The RIVA Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology is substantially equivalent to its primary predicate, the SI-BONE iFuse Bedrock Granite Implant System (K220195), and to the reference predicates SOLACE Sacroiliac Fixation
K262193 5 of 6
{10}
System (K233241) and RIVA Posterior Fixation System (K251154), with respect to intended use, indications for use, and principle of operation. Like the predicate and reference devices, the RIVA Anchor implant is a fully threaded, fenestrated, cannulated titanium alloy (Ti-6Al-4V ELI) screw with a self-tapping tip, intended for guidewire-assisted, image-guided placement across the sacroiliac joint, and it connects to the same RIVA modular head used in the RIVA Posterior construct. The device is supplied sterile via gamma irradiation, consistent with the predicate and reference devices, and incorporates the same NanoACTIV surface process as the SOLACE and neoWAVE/Ti3D devices. Differences in implant diameter and length ranges relative to the predicate and reference devices do not alter the fundamental design or principle of operation. Mechanical testing per ASTM F1717, F1798, F2193, and F543, together with engineering rationale, demonstrate that these differences do not raise new or different questions of safety or effectiveness. Accordingly, the RIVA Posterior Fixation System and RIVA Anchor Sacroiliac Fixation System, each with NanoACTIV Surface Technology, are substantially equivalent to the cited predicate devices.
## Conclusion:
The Xenix Medical RIVA Posterior Fixation System with NanoACTIV Surface Technology and RIVA Anchor Sacroiliac Fixation System with NanoACTIV Surface Technology are substantially equivalent to the cited predicate devices with respect to intended use, indications for use, design, function, materials, and performance. The differences in the technological characteristics between the subject device and the predicate devices do not raise new or different questions of safety and effectiveness.
K262193 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.