K262153 · Modular Medical, Inc. · LZG · Sep 2, 2026 · General Hospital
Device Facts
Record ID
K262153
Device Name
Pivot Insulin Delivery System
Applicant
Modular Medical, Inc.
Product Code
LZG · General Hospital
Decision Date
Sep 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pivot Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.
Device Story
Software-controlled, programmable insulin infusion pump; delivers insulin subcutaneously via gear-motor-driven mechanism. System components: reusable pump, disposable insulin cartridge with integrated battery, disposable infusion set, reusable inserter, adhesive pad, and Pivot Diabetes App. User configures basal rates (0.5-4.0 U/hr) and bolus doses (up to 20 units) via app. Pump monitors delivery; provides alarms for occlusions, low volume, and interruptions. Used by patients in home/clinic settings. App facilitates setup, status review, and history tracking. Design modifications include updated priming timeout, flexible bolus increments, and improved alarm management. Benefits include precise, automated insulin delivery for diabetes management.
Clinical Evidence
No clinical data. Performance supported by bench testing, including basal and bolus delivery accuracy (0.5 U minimum bolus tested on 20 pumps), battery performance, reliability testing (simulated shipping, handling, and drops), and software verification. Human factors validation conducted with 16 participants (Type 1 or Type 2 diabetes, age 18+) to assess new features.
Technological Characteristics
Microprocessor-controlled, gear-motor-driven infusion pump. Programmable basal rates (0.5-4.0 U/hr) and bolus delivery. Connectivity via Pivot Diabetes App. Sterilization per ISO 11137. Packaging tested per ASTM D4332, D7386, F1886, F2096, F88. Biocompatibility and cybersecurity leveraged from predicate. Software developed per ANSI AAMI IEC 62304.
Indications for Use
Indicated for subcutaneous insulin delivery for diabetes mellitus management in patients 18 years or older. Contraindicated for patients with inadequate hearing/vision to recognize alarms, inability to manage therapy, unwillingness to perform 4+ blood glucose readings daily, inability to follow User Guide, or those requiring basal rates outside 0.5-4.0 U/hr. Incompatible with insulin brands other than Humalog U-100.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
### I Background Information:
#### A 510(k) Number
K262153
#### B Applicant
Modular Medical Inc.
#### C Proprietary and Established Names
Pivot Insulin Delivery System
#### D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LZG | II | 21 CFR 880.5725 - Infusion Pump | CH - Clinical Chemistry |
#### E Purpose for Submission:
Modifications to the Pivot Insulin Delivery System cleared under K253534.
### II Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
#### B Indication(s) for Use:
The Pivot Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.
#### C Special Conditions for Use Statement(s):
Rx – For prescription use only.
The Pivot System is contraindicated for:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
a. Use by patients who do NOT have adequate hearing and/or vision to allow recognition of all functions of the Pivot System including Status and Alarms
b. Use by patients who cannot manage their Diabetes therapy
c. Use by patients unwilling to take a minimum of four (4) blood glucose readings per day
d. Use by patients who are unable to use the Pivot System in accordance with this User Guide
e. Use by patients who are not capable of following the User Guide
f. Use by patient populations requiring basal rates greater than 4 U/hr or less than 0.5 U/hr
g. Use with brands or concentrations of insulin other than Humalog® U-100 insulin
DO NOT expose your Pump within 6 inches from common products with magnets, including cell phone cases or wireless charging cases. Exposure to magnets or products with magnets may interfere with the Pump motor. Damage to the motor can lead to moderate or severe hypoglycemia, moderate hyperglycemia, or Diabetic Ketoacidosis (DKA).
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Pivot System specified by the manufacturer. This could cause disruptions in device performance or a device malfunction, which may lead to moderate or severe hypoglycemia, moderate hyperglycemia, or Diabetic Ketoacidosis (DKA).
Do not expose the Pump to unique medical emitters such as electrocautery, MRI, electrosurgical units, and diathermy devices. Exposure to these devices can lead to moderate or severe hypoglycemia, moderate hyperglycemia, or Diabetic Ketoacidosis (DKA).
Do not expose your Pivot System to X-ray screening used for carry-on and checked luggage. Newer full-body scanners used in airport security screening are also a form of X-ray screening and your Pivot System should not be exposed to them. Notify the Transportation Security Administration (TSA) Agent that your Pivot System cannot be exposed to X-ray machines and request an alternate means of screening.
The Pivot System is incompatible with certain common products that are used on the skin. These include bug sprays with or without DEET, perfumes, deodorant sprays, body lotions and sunscreens. Use of these products on the Pivot System itself or where the System is adhered to the skin can cause cracks in the plastic used to manufacture the Pivot components. DO NOT allow these products to come in contact with the Pump, Cartridge or Inserter. ALWAYS remove your Pump before applying these products and ALWAYS wash your hands before handling your Pump, Cartridge or Inserter after using such products. ALWAYS change your Cartridge if it becomes exposed to such products and immediately clean your Pump. Failure to do so may result in damage to the Pump, Insulin Cartridge or Inserter and in some cases lead to moderate hyperglycemia, moderate hypoglycemia, or Diabetic Ketoacidosis (DKA).
Do not use Pivot System components if they have been dropped as there could be unseen damage to the device that could lead to moderate hypoglycemia, moderate hyperglycemia, Diabetic Ketoacidosis (DKA), or a delay in insulin therapy.
Do not use the System in oxygen-rich environments, environments rich in flammable anesthetics, or other volatile agents as this presents a risk of fire or explosion that could lead to injury.
K262153 - Page 2 of 7
{2}
### III Device Description
The Pivot Insulin Delivery System is a software-controlled, programmable insulin infusion pump intended for the subcutaneous delivery of insulin for the management of diabetes mellitus. The system consists of a reusable Pump, a disposable Insulin Cartridge containing an integrated battery, a disposable Integrated Infusion Set with a built-in cannula for subcutaneous insulin delivery, a reusable Pivot Inserter for cannula insertion, an adhesive pad for secure attachment to the patient's skin, and the Pivot Diabetes App.
The Pump delivers insulin using a microprocessor-controlled, gear-motor-driven mechanism that transfers insulin from the Insulin Cartridge through the Integrated Infusion Set for subcutaneous infusion. The Pump provides programmed basal insulin delivery and user-initiated bolus insulin delivery. Insulin delivery is continuously monitored by system software, which provides alarms and notifications for conditions such as delivery interruptions, occlusions, low reservoir volume, and other system conditions.
The Pivot Diabetes App is used for system setup, configuration of therapy settings, review of system status, alarm information, and insulin delivery history.
The system delivers insulin at programmable basal rates between 0.5 and 4.0 units per hour in 0.1-unit increments. User-selectable bolus increment settings allow bolus doses to be delivered in predefined increments. Individual bolus deliveries are limited to a maximum of 20 units. Basal insulin delivery may be temporarily suspended by the user for up to two hours and resumed at any time during the suspension period.
### IV Substantial Equivalence Information:
A Predicate Device Name(s):
Pivot Insulin Delivery System
B Predicate 510(k) Number(s):
K253534
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K262153 | K253534 |
| --- | --- | --- |
| Device Trade Name | Pivot Insulin Delivery System | Same |
| General Device Characteristic Similarities | | |
K262153 - Page 3 of 7
{3}
| Intended Use/Indications For Use | Indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater. | Same |
| --- | --- | --- |
| **General Device Characteristic Differences** | | |
| Priming initiation timeout, minutes | 45 | 30 |
| Quick Bolus | Max 20U or ≤10 button presses at configured bolus increment of 0.5, 1, 2, 3, or 4U | Min 2U and max 20U with fixed increment of 2U per button press |
| Basal delivery suspension | 2 hours, but can be resumed at any time using the Pump Control Button | Fixed 30 minutes |
| Resumable alarms | Yes | No |
| How to address alarms is on the app | Yes | No |
| Cartridge and infusion set package configuration | 5-pack | 10-pack |
### V Standards/Guidance Documents Referenced:
- ANSI AAMI ISO 11137-1:2006/(R)2015, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2019)]
- ISO 11137-2 Third edition 2013-06 [Including AMD1:2022], Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose [Including Amendment 1 (2022)]
- ISO 11137-3 Second edition 2017-06, Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine control
- ISO 11737-2 Third edition 2019-12, Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ANSI AAMI ST72:2019, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
- ASTM D4332-22, Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing
- ISO TS 21726 First edition 2019-02, Biological evaluation of medical devices - Application of the threshold of toxicological concern (TTC) for assessing biocompatibility of medical device constituents
- ASTM D7386-16, Standard Practice for Performance Testing of Packages for Single Delivery Systems
K262153 - Page 4 of 7
{4}
- ASTM F1886/F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM F2096-11 (Reapproved 2019), Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- ASTM F88/F88M-23, Standard Test Method for Seal Strength of Flexible Barrier Materials
- ISO 11607-1 Second edition 2019-02 [Including AMD1:2023], Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including Amendment 1 (2023)]
- USP-NF M98900_01_01, <151> Pyrogen Test (USP Rabbit Test)
- USP-NF M98830_02_01, <85> Bacterial Endotoxins Test
- USP-NF M98910_01_01, <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
- ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices
- AAMI TIR101:2021, Fluid delivery performance testing for infusion pumps
- IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, Medical devices - Part 1: Application of usability engineering to medical devices
- ANSI AAMI HE75:2009/(R)2018, Human factors engineering - Design of medical devices
- ANSI AAMI IEC 62304:2006/A1:2016, Medical device software - Software life cycle processes [Including Amendment 1 (2016)]
- IEC TR 80002-1 Edition 1.0 2009-09, Medical device software - Part 1: Guidance on the application of ISO 14971 to medical device software
- ANSI AAMI SW91:2018, Classification of defects in health software
## VI Performance Characteristics:
### A. Non-Clinical Performance
#### 1. Basal Accuracy
Basal accuracy is unchanged from the predicate device.
#### 2. Bolus Delivery Accuracy
Bolus accuracy at intermediate and maximum bolus sizes (i.e., 10 and 20 U) is unchanged from the predicate device.
To assess bolus delivery accuracy of the new programable minimum bolus size of 0.5 U, 20 pumps (10 new and 10 aged) were tested by delivering the minimum bolus amount (0.5 U). The actual bolus volume delivered was compared to the expected bolus volume. The tables below show the number (and %) of boluses within the specified range of each target bolus volume for the new and the aged pumps separately.
Table 1: Amount of fluid delivered after a 0.5 U bolus request, 298 boluses, 10 new pumps
| Units delivered after a 0.5 U bolus request (% of commanded units) | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <25% | 25-75% | 75-90% | 90-95% | 95-105% | 105-110% | 110-125% | 125-175% | 175-250% |
K262153 - Page 5 of 7
{5}
| Units delivered after a 0.5 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Number and percent of boluses | 0/298 - | 0/298 - | 29/298 (9.7%) | 53/298 (17.8%) | 200/298 (67.1%) | 16/298 (5.4%) | 0/298 - | 0/298 - | 0/298 - | 0/298 - |
Table 2: Amount of fluid delivered after a 0.5 U bolus request, 297 boluses, 10 aged pumps
| Units delivered after a 0.5 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <25% | 25-75% | 75-90% | 90-95% | 95-105% | 105-110% | 110-125% | 125-175% | 175-250% | >250% |
| Number and percent of boluses | 0/297 - | 0/297 - | 31/297 (10.4%) | 72/297 (24.2%) | 191/297 (64.3%) | 3/297 (1.0%) | 0/297 - | 0/297 - | 0/297 - | 0/297 - |
### 3. Occlusion Detection
Occlusion detection is unchanged from the predicate device.
### B. Clinical Studies:
NA
### C. Other Supportive Device Performance Characteristics Data
### 4. Additional Bench Testing
In addition to the performance testing described above, other device verification testing was conducted, including:
- a. Battery performance testing was conducted to verify that changes made to battery connection, piezo and new motor do not negatively impact battery performance. 29 finished goods with build and sterilization records were able to operate beyond the acceptance criteria 80 hours + 3 minutes.
- b. Reliability verification was conducted using 29 devices to verify that the design changes to battery connection, piezo and new motor do not have an impact on the performance, durability, and robustness of the Pivot Insulin Delivery System after simulated shipping and handling, repeated insulin cartridge connection and disconnection, and drops.
- c. Software testing was conducted to verify that the software meets the performance requirements in monitoring the essential performance of the pump, that the controller software will trigger and log the designed alarms, and that the software meets controller states, state transitions, and logging requirements.
### 5. Sterility
The provided information about sterility is acceptable.
K262153 - Page 6 of 7
{6}
# 6. Human Factors
Human Factors validation testing was conducted in the United States and assessed use of the modified device with 16 people diagnosed with either type 1 or type 2 diabetes aged 18 years or older. Use tasks associated with the new features of the modified device were evaluated. Critical tasks associated with these design changes were identified based on the clinical harm severities that could occur due to use errors. K253534 HF data are leveraged for unchanged tasks.
# 7. Software
The software documentation provided was determined to be adequate to support substantial equivalence.
# 8. Leveraged Information from the predicate device cleared under K253534
a. Biocompatibility and drug compatibility are not impacted by the proposed changes in this submission.
b. Cybersecurity is unaffected by the changes proposed in this submission.
c. Interoperability is unchanged.
d. Wireless coexistence is unchanged from the predicate device.
e. IEC 60601-1 and its collateral standards including EMC were leveraged from the predicate device.
# VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
# VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K262153 - Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.