K262135 · Tianjin Huahong Technology Co., Ltd. · QRL · Aug 18, 2026 · General, Plastic Surgery
Device Facts
Record ID
K262135
Device Name
Lancing device
Applicant
Tianjin Huahong Technology Co., Ltd.
Product Code
QRL · General, Plastic Surgery
Decision Date
Aug 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4850
Device Class
Class 2
Indications for Use
Model: HH-XXXI-T Without AST cap The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. With AST cap The Lancing Device is used with lancets to draw capillary blood from the fingertip, palm (at the base of the thumb) or forearm, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. Model: HH-XXXII-T The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.
Device Story
Manually operated blood sampling device; uses sterile single-use lancets to obtain capillary blood; features cocking barrel and release button mechanism; includes depth adjustment settings; models HH-XXXI-T and HH-XXXII-T. Used by single patient in home or clinical setting; not for shared use. Device retracts lancet after use to prevent sharps injury. Healthcare provider or patient uses blood sample for glucose or other diagnostic testing. Benefits include safe, controlled capillary blood collection.
Clinical Evidence
No clinical data. Bench testing only: evaluated device functionality, penetration depth accuracy, and repeated-use durability. Simulated clinical use study (n=500) performed per ISO 23908 and FDA sharps injury prevention guidance; all samples met acceptance criteria with no failures.
Technological Characteristics
Manual lancing device; materials: ABS, POM, PC, PS resins; spring: carbon steel. Features cocking barrel, release button, and adjustable penetration depth (0.45–3.2mm depending on model/cap). Non-sterile, reusable for single-patient use. No software or electronics.
Indications for Use
Indicated for single-patient use to obtain capillary blood samples from the fingertip (all models) or palm/forearm (models with AST cap) for diagnostic testing requiring small blood volumes.
Regulatory Classification
Identification
The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.
Special Controls
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 18, 2026
Tianjin Huahong Technology Co., Ltd.
Ningning Wang
Registered Engineer
A01, Plant B, # 278, Hangkong Rd., Tianjin Pilot Free Trade Zone(Air Port Industrial Park)
Tianjin, 300308
China
Re: K262135
Trade/Device Name: Lancing device
Regulation Number: 21 CFR 878.4850
Regulation Name: Blood Lancets
Regulatory Class: Class II
Product Code: QRL
Dated: June 24, 2026
Received: June 25, 2026
Dear Ningning Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262135 - Ningning Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K262135 - Ningning Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K.
Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2026.08.18
11:16:36 -04'00'
Colin K. Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262135 | ? |
| Please provide the device trade name(s). | | ? |
| Lancing device | | |
| Please provide your Indications for Use below. | | ? |
| Model: HH-XXXI-T Without AST cap The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. With AST cap The Lancing Device is used with lancets to draw capillary blood from the fingertip, palm (at the base of the thumb) or forearm, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. Model: HH-XXXII-T The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? |
{5}
510(k) Summary
K262135
# 510(k) summary
## I Submitter
Tianjin Huahong Technology Co., Ltd.
A01, Plant B, No.278, Hangkong Road, Tianjin Pilot Free Trade Zone (Air Port Industrial Park), 300308 Tianjin, China
Establishment Registration Number: 3009498536
Contact person: Ms. Ningning Wang
Registered Engineer
Tel.: +86-13021381776
E-mail: ningning.wang@hh-technology.com
Preparation date: June 25, 2026
## II Proposed Device
Trade Name of Device: Lancing device
Common name: Multiple Use Blood Lancet For Single Patient Use Only
Regulation Number: 21 CFR 878.4850
Regulatory Class: Class II
Product code: QRL
Review Panel General & Plastic Surgery
## III Predicate Devices
510(k) Number: K243306
Trade name: Lancet, Lancing device
Classification: Class II
Product Code: QRL, QRK
Manufacturer: Tianjin Huahong Technology Co., Ltd.
## IV Device description
The Lancing Device is a manually operated blood sampling device intended for use with sterile single-use lancets to obtain capillary blood samples for glucose and other blood testing requiring small amounts of blood.
1 / 7
{6}
510(k) Summary
The device family includes models with and without an Alternate Site Testing (AST) cap. Models without an AST cap are intended for obtaining capillary blood samples from the fingertip only. Models supplied with an AST cap may be used to obtain capillary blood samples from the fingertip, palm (at the base of the thumb), or forearm.
The Lancing Device is intended for single-patient use and must not be shared between users.
The housing and internal plastic components are manufactured from ABS, POM, PC, and PS resins. The spring is manufactured from carbon steel.
The device is supplied non-sterile and is intended for repeated use by a single patient in conjunction with sterile single-use lancets.
The subject device includes two additional models, HH-XXXI-T and HH-XXXII-T. The HH-XXXI-T model is technologically equivalent to the previously cleared HH-XIII-T model (without AST cap) and HH-XXVII-T model (with AST cap). Differences include external housing appearance, overall dimensions, and the number of depth adjustment settings. The HH-XXXI-T model provides eleven depth adjustment settings, whereas the HH-XIII-T and HH-XXVII-T models provide five depth adjustment settings.
The HH-XXXII-T model is technologically equivalent to the previously cleared HH-XVII-T model. The primary difference is the internal core structure. All other technological characteristics remain unchanged.
The newly added models utilize the same intended use, operating principle, lancet firing mechanism, patient-contacting materials, lancet compatibility, and depth adjustment function as previously cleared models included in K243306. The identified differences do not affect the safety or effectiveness of the device and do not raise different questions of safety or effectiveness.
# V Indication for use
# Model: HH-XXXI-T
# Without AST cap
The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.
# With AST cap
2 / 7
{7}
510(k) Summary
The Lancing Device is used with lancets to draw capillary blood from the fingertip, palm (at the base of the thumb) or forearm, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.
# **Model: HH-XXXII-T**
The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.
# **VI Comparison of technological characteristics with the predicate devices**
The comparison and discussion between the proposed device and the predicate devices are listed in below table 1:
Table 1 Substantial Equivalence Comparison
| Item | Predicate device Model: HH-XIII-T, HH-XVII-T, HH-XXVII-T (K243306) | Proposed device Model: HH-XXXI-T, HH-XXXII-T | Comments |
| --- | --- | --- | --- |
| Product name | Lancing device | Lancing device | Same |
| Product Code | QRL | QRL | Same |
| Regulation No. | 21 CFR § 878.4850 | 21 CFR § 878.4850 | Same |
| Class | II | II | Same |
| Prescription/over-the-counter use | Over-The-Counter Use | Over-The-Counter Use | Same |
| Indication for use | *Model: HH-XIII-T, HH-XVII-T* The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. | *Model: HH-XXXI-T* **Without AST cap** The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single | Same |
3 / 7
{8}
510(k) Summary
| | *Model: HH-XXVII-T* The Lancing Device is used with lancets to draw capillary blood from the fingertip, palm (at the base of the thumb) or forearm, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. | patient and should not be shared. **With AST cap** The Lancing Device is used with lancets to draw capillary blood from the fingertip, palm (at the base of the thumb) or forearm, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. *Model: HH-XXXII-T* The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. | |
| --- | --- | --- | --- |
| Puncture device to obtain micro blood samples | Yes | Yes | Same |
| Lancet retracted after use to prevent sharp injure | Yes | Yes | Same |
| Materials | ABS, POM, PC and PS | ABS, POM, PC and PS | Same |
| Reuse durability | Reusable Single Patient Use Only | Reusable Single Patient Use Only | Same |
4 / 7
{9}
510(k) Summary
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| --- | --- | --- | --- |
| Mechanical loading and firing function | Cocking barrel with releasing button | Cocking barrel with releasing button | Same |
| Penetration Depth | *Model: HH-XIII-T* 0.60~2.00±0.30mm | *Model: HH-XXXI-T* **Without AST cap** 0.45~1.95±0.30mm | Different |
| | *Model: HH-XXVII-T, With AST Cap* 3.2±0.30mm | *Model: HH-XXXI-T* **With AST cap** 3.2±0.30mm | Same |
| | *Model: HH-XVII-T* 0.6~1.8±0.30mm | *Model: HH-XXXII-T* 0.6~1.8±0.30mm | Same |
| **Discussion:** The HH-XXXI-T model has a slightly different penetration depth range (0.45–1.95 ± 0.30 mm) compared to the predicate HH-XIII-T model (0.60–2.00 ± 0.30 mm). This difference does not affect the intended use, operating principle, or blood sampling function of the device. The penetration depth remains within the range suitable for capillary blood sampling and does not raise different questions of safety or effectiveness. | | | |
5 / 7
{10}
510(k) Summary
## VII. Non-Clinical Testing
Non-clinical verification and validation testing was conducted to support a determination of substantial equivalence for the proposed Lancing Device models HH-XXXI-T and HH-XXXII-T.
The non-clinical testing program was designed to evaluate the safety and performance of the proposed devices and to demonstrate that they perform as intended when compared to the legally marketed predicate device (K243306). Testing addressed device functionality, penetration depth performance, durability.
### Performance Testing
Performance testing was conducted to evaluate critical functional characteristics of the proposed lancing devices, including:
- Lancet loading and removal
- Device cocking and release mechanism operation
- Penetration depth adjustment performance
- Trigger mechanism functionality
- Penetration depth accuracy
- Repeated-use durability and life cycle performance
Performance testing confirmed that the modified depth adjustment configuration of HH-XXXI-T and the revised internal core structure of HH-XXXII-T do not adversely affect device functionality, penetration depth performance, or overall device reliability.
All tested samples met the predefined acceptance criteria and demonstrated performance equivalent to the predicate device.
### Simulated Clinical Use Testing
A simulated clinical use study was conducted using 500 samples of the proposed lancing device in accordance with FDA Guidance Medical Devices with Sharps Injury Prevention Features (August 9, 2005) and ISO 23908.
The study evaluated the effectiveness and reliability of the sharps injury prevention mechanism under simulated use conditions, including needle activation and automatic retraction. All tested devices met the pre-established acceptance criteria,
6 / 7
{11}
510(k) Summary
and no failures were observed.
### Summary of Non-Clinical Test Results
The results of non-clinical testing demonstrate that the proposed lancing device models HH-XXXI-T and HH-XXXII-T meet their design and performance specifications and are as safe and effective as the predicate device. The identified differences between the proposed and predicate devices do not raise different questions of safety or effectiveness.
### VIII Clinical Testing
No clinical testing was required for this submission.
### IX Conclusion
The proposed lancing device models HH-XXXI-T and HH-XXXII-T have the same intended use and technological characteristics as the predicate device (K243306).
Any differences between the proposed and predicate devices, do not affect the intended use, operating principle, or performance characteristics of the device.
Verification and validation testing demonstrate that the proposed devices meet all applicable design and performance requirements and perform as intended. The identified differences do not raise different questions of safety or effectiveness.
Based on the information provided in this submission, the proposed lancing device models HH-XXXI-T and HH-XXXII-T are substantially equivalent to the predicate device (K243306) and are as safe and effective for their intended use.
7 / 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.