K262091 · Ortler Global Tibbi Gerecler Anonim Sirketi · NKB · Sep 17, 2026 · Orthopedic
Device Facts
Record ID
K262091
Device Name
FRAXION Spinal Fixation System
Applicant
Ortler Global Tibbi Gerecler Anonim Sirketi
Product Code
NKB · Orthopedic
Decision Date
Sep 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FRAXION SPINAL FIXATION SYSTEM is intended for use in the non-cervical spine. When used as an anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the FRAXION SPINAL FIXATION SYSTEM is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: * Degenerative disc disease (DDD), defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; * Spondylolisthesis; * Trauma, including fracture or dislocation; * Spinal stenosis; * Curvatures, including scoliosis, kyphosis, and/or lordosis; * Tumor; * Pseudoarthrosis; and * Failed previous fusion.
Device Story
Top-loading, multi-component posterior spinal fixation system; provides immobilization/stabilization of non-cervical spinal segments as adjunct to fusion. Components include polyaxial/monoaxial/cannulated/fenestrated pedicle screws, rods, setscrews, connectors, and hooks. Surgeon places screws/hooks into vertebral anatomy; seats rods into heads; tightens setscrews to lock construct. Creates rigid construct to limit motion, maintain alignment, and support load sharing during osseous fusion. Used in OR by surgeons. Benefits patient by providing mechanical stability to unstable/deformed spinal segments.
Clinical Evidence
Bench testing only. Substantial equivalence relies on mechanical evaluation data previously established for the predicate device (K212220), including testing per ASTM F1717 (static/dynamic compression bending, static torsion) and ASTM F1798 (axial/torsional grip, static flexion-extension).
Technological Characteristics
Non-powered, mechanical spinal fixation system. Materials: Ti6Al4V-ELI titanium alloy (ASTM F136). Components: polyaxial/monoaxial/cannulated/fenestrated screws, rods (5.5/6.0 mm), connectors, hooks, setscrews. Top-loading design. Supplied non-sterile, single-use. No software or energy source.
Indications for Use
Indicated for skeletally mature patients requiring non-cervical spinal fixation (anterior/anterolateral/posterior) as an adjunct to fusion (autograft/allograft) for DDD, spondylolisthesis, trauma, spinal stenosis, spinal curvatures (scoliosis/kyphosis/lordosis), tumor, pseudoarthrosis, or failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
September 17, 2026
Ortler Global Tibbi Gerecler Anonim Sirketi
Victor Weaver
Regulatory Consultant
Ivedikkoy Mah., 1492 Cadde, 17 B, 43, Yenimahalle
Ankara, 06378
Turkey
Re: K262091
Trade/Device Name: FRAXION Spinal Fixation System
Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral Pedicle Screw System
Regulatory Class: Class II
Product Code: NKB, KWP, KWQ
Dated: June 22, 2026
Received: June 22, 2026
Dear Victor Weaver:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262091 - Victor Weaver
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K262091 - Victor Weaver
Page 3
Sincerely,

For Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262091 | ? |
| Please provide the device trade name(s). | | ? |
| FRAXION Spinal Fixation System | | |
| Please provide your Indications for Use below. | | ? |
| FRAXION SPINAL FIXATION SYSTEM is intended for use in the non-cervical spine. When used as an anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the FRAXION SPINAL FIXATION SYSTEM is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: * Degenerative disc disease (DDD), defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; * Spondylolisthesis; * Trauma, including fracture or dislocation; * Spinal stenosis; * Curvatures, including scoliosis, kyphosis, and/or lordosis; * Tumor; * Pseudoarthrosis; and * Failed previous fusion. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
K262091, Page 1 of 3
| 510(k) #: | 510(k) Summary | Prepared on: 2026-06-22 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Ortler Global Tibbi Gerecler Anonim Sirketi |
| --- | --- |
| Applicant Address | IVEDIKKOY MAH., 1492 CADDE, 17 B, 43, YENIMAHALLE Ankara 06378 Turkey |
| Applicant Contact Telephone | +905356949313 |
| Applicant Contact | Mr. TAHA CIBIR |
| Applicant Contact Email | ortlerfda@gmail.com |
| Correspondent Name | Mr. Victor Weaver |
| Correspondent Address | null United States |
| Correspondent Contact Telephone | +905323973043 |
| Correspondent Contact | Mr. Victor Weaver |
| Correspondent Contact Email | 9342mdd@gmail.com |
| Device Name | 21 CFR 807.92(a)(2) |
| --- | --- |
| Device Trade Name | FRAXION Spinal Fixation System |
| Common Name | Thoracolumbosacral pedicle screw system |
| Classification Name | Thoracolumbosacral Pedicle Screw System |
| Regulation Number | 888.3070 |
| Product Code(s) | NKB, KWP, KWQ |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| --- | --- | --- |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K212220 | ArtFX Spinal Fixation System | NKB |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| --- | --- |
The FRAXION Spinal Fixation System is a top-loading, multiple-component posterior spinal fixation system intended to provide immobilization and stabilization of non-cervical spinal segments in skeletally mature patients as an adjunct to fusion using autograft or allograft. The system is used for anterior/anterolateral and posterior non-cervical pedicle and non-pedicle fixation in the treatment of acute and chronic spinal instabilities or deformities, including degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion.
The device functions by creating a rigid spinal construct using bone-anchored screws, rods, setscrews, connectors, and hooks. Pedicle screws or hooks are placed into or onto the vertebral anatomy, longitudinal rods are seated into the screw or hook heads, and setscrews are tightened to lock the rod-screw or rod-hook interface. Once assembled, the construct limits motion between affected spinal segments, maintains alignment, supports load sharing, and provides mechanical stability during the spinal fusion process. The top-
{6}
K262091, Page 2 of 3
loading design allows the rod and setscrew to be inserted from above the screw head, facilitating intraoperative assembly and reducing procedural complexity.
The scientific concept underlying the device is mechanical stabilization of the spine to support biological fusion. By immobilizing unstable or deformed spinal segments and maintaining the desired anatomical relationship between vertebrae, the system provides an environment that supports osseous fusion using autograft or allograft. The device does not achieve its primary intended effect by chemical action, metabolism, or pharmacological means; its function is based on mechanical fixation, load support, and stabilization.
The system includes polyaxial pedicle screws, monoaxial pedicle screws, spondylolisthesis/reduction screws, cannulated screws, fenestrated screws, straight rods, pre-bent rods, transverse connectors, multiaxial transverse connectors, lateral connectors, axial extension connectors, domino connectors, laminar hooks, pedicular hooks, and setscrews. Polyaxial screws include a screw body, tulip head, and fastening component, allowing angular variability for construct alignment. Monoaxial screws have a fixed head. Cannulated screws allow placement over a guidewire. Reduction screws include an extended tulip head to assist rod reduction in spondylolisthesis cases. Rods are provided in straight and pre-bent configurations and may be contoured intraoperatively to match patient anatomy. Connectors and hooks provide additional construct options to improve fixation, accommodate anatomy, or increase construct stability.
All implantable components are manufactured from titanium alloy Ti6Al4V-ELI conforming to ASTM F136. Components may be anodized for color coding and identification; anodizing forms a titanium dioxide surface layer. Components are also laser-marked for product identification. The system is supplied non-sterile, for single use only, and is available in multiple sizes to accommodate patient anatomy and surgical requirements.
Significant physical characteristics include screw diameters generally ranging from 4.5 mm to 8.0 mm depending on screw type, screw lengths varying by configuration, compatibility with 5.5 mm and/or 6.0 mm rods, smooth cylindrical rods in various lengths, and threaded setscrews that lock the rod within the screw or hook head. The screw heads include U-shaped slots and threaded internal walls to receive the rod and setscrew. The screw bodies include self-tapping, tapered threaded designs to support insertion into vertebral bone. The polyaxial screw design allows angular adjustment before final locking, while the setscrew design secures the rod and minimizes the potential for misalignment or cross-threading.
The principal performance characteristics of the FRAXION Spinal Fixation System are its ability to provide spinal immobilization, stabilization, construct rigidity, rod-screw locking strength, and resistance to expected mechanical loads during the fusion period. The system is designed to maintain spinal alignment, support fusion, and withstand loads associated with non-cervical spinal fixation when used according to the surgical technique and instructions for use.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
FRAXION SPINAL FIXATION SYSTEM is intended for use in the non-cervical spine. When used as an anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the FRAXION SPINAL FIXATION SYSTEM is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:
- Degenerative disc disease (DDD), defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies;
- Spondylolisthesis;
- Trauma, including fracture or dislocation;
- Spinal stenosis;
- Curvatures, including scoliosis, kyphosis, and/or lordosis;
- Tumor;
- Pseudoarthrosis; and
- Failed previous fusion.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use of the FRAXION Spinal Fixation System are the same as the predicate device, Artfx Spinal Fixation System (K212220). The subject device and predicate device are both intended for use in the non-cervical spine as anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation systems to provide additional support during fusion using autograft or allograft in skeletally mature patients for the treatment of the same acute and chronic instabilities or deformities, including degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion.
There are no differences in indications for use between the subject device and the predicate device. In addition, the subject device is identical to the predicate device with respect to intended use, technological characteristics, design, materials, manufacturing process, performance specifications, and method of use. The predicate device manufacturer, Artfx, will manufacture the FRAXION Spinal Fixation System using the same design, materials, manufacturing processes, equipment, quality system controls, technological characteristics, and performance specifications as the device cleared under K212220.
{7}
K262091, Page 3 of 3
Therefore, the indications for use are the same as the predicate device and do not constitute a new intended use.
## Technological Comparison
21 CFR 807.92(a)(6)
The FRAXION Spinal Fixation System has the same technological characteristics as the predicate device, Artfx Spinal Fixation System (K212220). The subject device and predicate device have the same design, materials, chemical composition, principle of operation, sterility status, method of use, and performance specifications.
Both devices are top-loading, multiple-component spinal fixation systems consisting of pedicle screws, rods, connectors, hooks, and setscrews. The subject device uses the same rod-screw-setscrew interconnection mechanism as the predicate device. After the screws are placed in the spine, the rod is seated into the U-shaped slot of the screw head and locked using a setscrew to create a spinal fixation construct.
The subject device and predicate device are manufactured from the same implant material, Ti6Al4V-ELI titanium alloy conforming to ASTM F136. The material composition, biocompatibility profile, corrosion resistance, mechanical strength, and fatigue strength are the same as the predicate device.
The device is non-powered and does not use an energy source. Its principle of operation is mechanical stabilization and immobilization of non-cervical spinal segments as an adjunct to fusion. The function of the subject device is the same as the predicate device: to act as a spinal implant construct to stabilize and promote spinal fusion.
The subject device is identical to the predicate device cleared under K212220 with respect to design, materials, technological characteristics, interconnection mechanism, performance specifications, and intended use. Therefore, the FRAXION Spinal Fixation System has the same technological characteristics as the predicate device. No technological differences exist that raise different questions of safety or effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The FRAXION Spinal Fixation System is identical to the predicate device, Artfx Spinal Fixation System (K212220), with respect to intended use, indications for use, design, materials, technological characteristics, principle of operation, interconnection mechanism, manufacturing approach, sterility status, and performance specifications. Therefore, the nonclinical performance information relied upon for substantial equivalence is based on the previously established performance data for the predicate device and the mechanical evaluation provided in K212220, including ASTM F1717 (static and dynamic compression bending, static torsion) and ASTM F1798 (axial grip, torsional grip, static flexion-extension).
The information presented demonstrates the substantial equivalence of the subject device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.