K262007 · PRIMED Medical Products, Inc. · QPC · Sep 11, 2026 · General, Plastic Surgery

Device Facts

Record IDK262007
Device NamePRIMED AAMI Level 3 Open Back Gowns
ApplicantPRIMED Medical Products, Inc.
Product CodeQPC · General, Plastic Surgery
Decision DateSep 11, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

PRIMED AAMI Level 3 Open Back Gowns are non-sterile, single use gowns intended to be worn by healthcare personnel to protect both the patient and the wearer from transfer of microorganisms, body fluids, and particulate material in non-surgical, non-isolation settings.

Device Story

Single-use, non-sterile protective gowns; overhead open-back design with thumbloops and waist ties. Available in two material configurations: poly-coated polypropylene (Model PM4-68363) or film (CPE) (Model PM4-88372). Used by healthcare personnel in non-surgical, non-isolation clinical settings to provide AAMI Level 3 barrier protection. Protects both wearer and patient from fluid and particulate transfer. Device is donned by the user; provides physical barrier; disposed of after use.

Clinical Evidence

No clinical data. Bench testing only, including barrier protection (AATCC 42, AATCC 127), physical properties (ASTM D5034, ASTM D1683, ASTM D5587), biocompatibility (ISO 10993-5, -10, -23), and flammability (16 CFR 1610/1611).

Technological Characteristics

Materials: Poly-coated polypropylene or film (CPE). Design: Open-back, overhead, thumbloops, waist tie. Barrier performance: AAMI Level 3 (AATCC 42, AATCC 127). Physical standards: ASTM D5034 (tensile), ASTM D1683 (seam), ASTM D5587 (tear). Biocompatibility: ISO 10993-5, 10993-10, 10993-23. Flammability: 16 CFR 1610/1611 Class 1. Non-sterile, single-use.

Indications for Use

Indicated for healthcare personnel in non-surgical, non-isolation settings to provide moderate barrier protection against microorganisms, body fluids, and particulate material.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{1} **FDA** U.S. FOOD & DRUG ADMINISTRATION September 11, 2026 PRIMED Medical Products, Inc. Mitra Fard Senior Manager, Regulatory Affairs 200, 2003-91 St. SW Edmonton, AB T6X 0W8 Canada Re: K262007 Trade/Device Name: PRIMED AAMI Level 3 Open Back Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: QPC Dated: June 15, 2026 Received: June 15, 2026 Dear Mitra Fard: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K262007 - Mitra Fard Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K262007 - Mitra Fard Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # BIFENG QIAN -S BiFeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 *See PRA Statement below.* 510(k) Number (*if known*) K262007 Device Name PRIMED AAMI Level 3 Open Back Gowns Indications for Use (*Describe*) PRIMED AAMI Level 3 Open Back Gowns are non-sterile, single use gowns intended to be worn by healthcare personnel to protect both the patient and the wearer from transfer of microorganisms, body fluids, and particulate material in non-surgical, non-isolation settings. Type of Use (*Select one or both, as applicable*) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} PRIMED AAMI Level 3 Open Back Gowns (K262007) 510 (k) Summary | Summary date | September 8, 2026 | | --- | --- | | Submitter name | PRIMED Medical Products Inc. | | Submitter Address: | 200, 2003-91 St. SW, Edmonton, AB, Canada, T6X 0W8 | | Phone Number | +1.877.877.4633 | | Fax Number | +1.780.497.7670 | | Email | mitra.fard@primed.com Mitra Fard | | Contact person: | | | Common name of the device | Gown, Non-Sterile, Non-Isolation, intended to provide moderate or high barrier protection | | Trade or Proprietary Name: | PRIMED AAMI Level 3 Open Back Gowns | | Regulation Description | Surgical Apparel | | Review Panel | General Hospital | | Regulation number | 878.4040 | | Device Class: | 2 | | Product Code: | QPC | # Device Description: PRIMED AAMI Level 3 Open Back Gowns are single-use, non-sterile gowns intended to provide moderate barrier protection in non-surgical and non-isolation settings. Model PM4-68363 is a poly-coated (PP-PE) gown and Model PM4-88372 is a film (CPE) gown. The gowns feature an overhead open back design and are available in a universal size. # Indication for Use: PRIMED AAMI Level 3 Open Back Gowns are non-sterile, single use gowns intended to be worn by healthcare personnel to protect both the patient and the wearer from transfer of microorganisms, body fluids, and particulate material in non-surgical, non-isolation settings. Page 1 of 3 {6} PRIMED AAMI Level 3 Open Back Gowns (K262007) 510 (k) Summary Here below is the summary of the technological characteristics and the performance of the proposed device compared to the predicate device. | Elements of Comparison | Proposed device | Predicate | comments | | --- | --- | --- | --- | | Trade name | PRIMED AAMI Level 3 Open Back Gowns | Medline Open-Back Level 3 Protective Gown | - | | 510k number | K262007 | K233526 | - | | Manufacturer | PRIMED Medical Products Inc. | Medline Industries LP | - | | Regulation number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | classification | 2 | 2 | same | | Product code | QPC | QPC | same | | Indication for use | PRIMED AAMI Level 3 Open Back Gowns are non-sterile, single use gowns intended to be worn by healthcare personnel to protect both the patient and the wearer from transfer of microorganisms, body fluids, and particulate material in non-surgical, non-isolation settings. | Gown is a single use, disposable, non-sterile, non-isolation gown intended to be worn by healthcare personnel to provide moderate barrier protection (AAMI Level 3) in non-sterile and non-patient isolation situation. | similar | | Single use vs. reusable | Single Use | Single Use | Same | | Sterility | Non-sterile | Non-sterile | Same | | Material composition | Film or Poly-coated polypropylene | Polypropylene | Different^{1} | | Design features | Open Back, Over-Head, Thumbloops, Back Perforation and waist tie | Design includes Open Back, thumbloops, back perforation for easy removal and waist tie | Same | | Color | Blue | Blue | Same | | Use | Over the Counter | Over the counter | Same | | Contact duration | Surface, <24 hours | Surface, <24 hours | Same | | Performance per AAMI PB70 | AAMI Level 3 | AAMI level 3 | Same | | Biocompatibility | Under the conditions of the testing, not cytotoxic, not a sensitizer, not an irritant per ISO 10993-5, ISO 10993-10, ISO 10993-23 | Under the test conditions, the subject device was shown to be noncytotoxic, non-irritating, and non-sensitizing per ISO 10993-5, ISO 10993-10, ISO 10993-23 | Same | | Tensile Strength | Average tensile for MD and CD ≥ 30 N | 15.98±0.68N (Machine Direction) 9.41±0.46 (Transverse Direction) | Similar | | Seam Strength | Average seam strength ≥ 30 N | 35.01±3.91 N | Similar | | Tear Resistance | Average tear strength for both MD and CD ≥10 N | 41.74±2.74N (Machine Direction) 30.29±3.52N (Transverse Direction) | Similar | | Shelf life | 3 Years | N/A | - | $^{1}$ Different – Proposed device has undergone biocompatibility and barrier performance testing and has been demonstrated to be non-cytotoxic, non-irritant and not a sensitizer under the condictio of the study it. The difference is raw material type does not raise new questions of safety or effectiveness. Page 2 of 3 {7} PRIMED AAMI Level 3 Open Back Gowns (K262007) 510 (k) Summary # Non-Clinical Testing: The PRIMED AAMI Level 3 Open Back Gowns have been tested for barrier protection according to ANSI/AAMI PB70:2022 requirements. The testing results demonstrate that the PRIMED AAMI Level 3 Open Back Gowns meet the claimed barrier protection requirements as defined by ANSI/AAMI PB70:2022. Summary of Non- Clinical Testing: | Test Description | Standard | Results | Remark | | --- | --- | --- | --- | | Impact Penetration | AATCC42 | ≤1.0 g | Meets requirements | | Hydrostatic Pressure | AATCC 127 | ≥50 cmH2O | Meets requirements | | Tensile Strength | ASTM D5034 | Average tensile for MD and CD ≥ 30 N | Meets requirements | | Seam Strength | ASTM D1683 | Average seam strength ≥ 30 N | Meets requirements | | Tear Resistance | ASTM D5587 | Average tear strength for both MD and CD ≥10 N | Meets requirements | | Cytotoxicity | ISO 10993-5 | No cytotoxic under the test condition | Meets requirements | | Sensitization | ISO 10993-10 | Non sensitizing under the test condition | Meets requirements | | Irritation | ISO 10993-23 | Non irritating under the test condition | Meets requirements | | Flammability | 16CFR1610 | Poly coated gowns meet the requirements for Class 1 in accordance with 16CFR1610 | Meets requirements | | | 16CFR1611 | Film gowns meet the requirements for Class 1 in accordance with 16CFR1611 | | # Summary of Clinical Testing: No clinical testing was performed in support of this 510(k). # Conclusion: The conclusion drawn from the nonclinical tests demonstrate that the PRIMED AAMI Level 3 Open Back Gowns are as safe, as effective and perform as well as or better than the legally marketed predicate device, K233526, Medline Open Back Level 3 Gown. Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...