MONODEK™ Sutures are indicated for use in all types of soft tissue approximation. MONODEK™ suture is not indicated in adult cardiovascular tissue, microsurgery and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable.
Device Story
Monodek™ is a dyed violet monofilament synthetic absorbable surgical suture; used for soft tissue approximation; provides extended wound support up to six weeks. Device is provided sterile (EO) for single use; available in USP 0 size with or without 300/400 series stainless steel needles (taper point or taper-cutting). Operated by surgeons in clinical settings. Output is physical wound closure; healthcare providers use the suture to approximate tissues during surgery. Benefits include absorbable material properties and extended support duration. Modifications include updated ethylene oxide sterilization cycle (MC-1) and new Tyvek/Foil single-barrier packaging.
Clinical Evidence
No clinical data. Bench testing only. Verification activities included sterilization validation (SAL 10^-6), EO residual testing (ISO 10993-7), packaging integrity validation (ISO 11607-1/2), and 3-year shelf-life stability testing (USP <861> Diameter, <871> Needle Attachment, <881> Tensile Strength, and Resorption).
Technological Characteristics
Monofilament synthetic absorbable polydioxanone suture. USP 0 size. Needles: 300/400 series stainless steel. Sterilization: Ethylene Oxide (EO) per BS EN ISO 11135:2014. Packaging: Tyvek/Foil single-barrier system. Shelf-life: 3 years. Conforms to USP requirements for synthetic absorbable sutures (except oversized diameter) and E.P. diameter requirements.
Indications for Use
Indicated for soft tissue approximation in patients requiring absorbable sutures with extended wound support (up to six weeks). Contraindicated for use in adult cardiovascular tissue, microsurgery, and neural tissue.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Force Fiber White and White/Blue, White/Black, Blue, and White/Green sutures; Force Fiber OrthoTape suture; Bondek suture and Bondek Plus suture; Monodek suture; Polyglytone*6211 suture (K181774)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 06, 2026
Teleflex Medical, LLC
Rebecca Walker
Sr. Regulatory Affairs Specialist
3015 Carrington Mill Blvd.
Morrisville, North Carolina 27560
Re: K261988
Trade/Device Name: Monodek™ (POLYDIOXANONE) Dyed Violet Monofilament Synthetic Absorbable Surgical Suture
Regulation Number: 21 CFR 878.4840
Regulation Name: Absorbable polydioxanone surgical suture
Regulatory Class: Class II
Product Code: NEW
Dated: June 12, 2026
Received: June 12, 2026
Dear Rebecca Walker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261988 - Rebecca Walker
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261988 - Rebecca Walker
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# TEK N.
LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261988
Device Name
MONODEK™ (POLYDIOXANONE) Dyed Violet Monofilament Synthetic Absorbable Surgical Suture
Indications for Use (Describe)
MONODEK™ Sutures are indicated for use in all types of soft tissue approximation. MONODEK™ suture is not indicated in adult cardiovascular tissue, microsurgery and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
510(k) Summary
K261988 Page 1 of 3
### 510(k) Summary
### Teleflex
### MONODEK™ (POLYDIOXANONE) Dyed Violet Monofilament Synthetic Absorbable Surgical Suture
### K261988
Prepared on: July 09, 2026
Applicant Name: Teleflex Medical, LLC
3015 Carrington Mill Blvd.
Morrisville, NC 27560
Applicant Contact: Laura Medlin
Laura.Medlin@teleflex.com
919-544-8000
Correspondent: Rebecca Walker
Rebecca.Walker@teleflex.com
919-544-8000
### 1. Device Name
Device Trade Name: Teleflex MONODEK™ (POLYDIOXANONE) Dyed Violet Monofilament Synthetic Absorbable Surgical Suture
Common Name: Absorbable polydioxanone surgical suture
Classification Name: Suture, surgical, absorbable, polydioxanone
Regulation Number: 21 CFR 878.4840
Product Code(s): NEW
### 2. Legally Marketed Predicate Devices
| Predicate 510(k) # | Predicate Trade Name (Primary Predicate is listed first) | Predicate Product Code(s) |
| --- | --- | --- |
| K181774 | Force Fiber White and White/Blue, White/Black, Blue, and White/Green sutures; Force Fiber OrthoTape suture; Bondek suture and Bondek Plus suture; Monodek suture; Polyglytone*6211 suture | GAT, GAM, NEW |
### 3. Device Description Summary
The Teleflex Monodek™ (POLYDIOXANONE) Dyed Violet Monofilament Synthetic Absorbable Surgical Sutures are provided sterile (EO), for single use only. Monodek is available exclusively with
{5}
K261988 Page 2 of 3
510(k) Summary
USP 0 suture in a variety of suture lengths and configurations, with or without attached needles. The needles are manufactured from 300 or 400 series stainless steel. Attached needles may include a ½-circle taper point single flat body style curved needle that conforms to the applicable USP requirements for needle attachment strength, or a bullet needle with a taper-cutting point and a straight round-body style. The sutures have a shelf-life of 3 years. The finished suture may be packaged in cartons as single packs, multi-packs, or procedure packs.
Monodek™ (POLYDIOXANONE) Dyed Violet Monofilament Synthetic Absorbable Surgical Sutures conform to all requirement established by the USP for synthetic absorbable surgical sutures, except for oversized diameter, and the European Pharmacopeia (E.P.) for diameter for sterile synthetic absorbable sutures. The product labeling lists the specifications for the oversized diameter sutures.
### 4. Intended Use/Indications for Use
MONODEK™ Sutures are indicated for use in all types of soft tissue approximation. MONODEK™ suture is not indicated in adult cardiovascular tissue, microsurgery and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable.
### 5. Substantial Equivalence
#### Indications for Use Comparison
The indications of the subject Monodek Suture have been narrowed, as compared to the previous clearance for the same suture to align with the sizes marketed. This change does not alter the device's intended use or technological characteristics in a way that raises new safety or effectiveness questions.
#### Technological Comparison
An alternate sterilization cycle (MC-1) has been qualified for the Monodek Sutures to comply with the requirements of the updated EPA regulations regarding air toxics standards for ethylene oxide commercial sterilization facilities (March 14, 2024). Additionally, this Special 510(k) includes a change in the primary packaging of the Monodek suture. This change implements a new Tyvek/Foil pouch single-barrier system, replacing the double barrier, Foil Paper/Clear film outer pouch.
The modified Monodek Suture has the same technological characteristics as the previously cleared suture. Monodek is available exclusively with USP 0 suture in a variety of suture lengths and configurations, with or without attached needles. Differences in primary packaging (sterile barrier) and sterilization cycle parameters do not introduce new risks and packaging. Sterilization and stability studies have been conducted in accordance with recognized standards. Additional minor modifications in labeling are not the result of a design change and do not have an impact on the intended use.
### 6. Non-Clinical Testing
Verification and validation activities conducted based on the risk assessment for the proposed changes are summarized below. The results demonstrate that the proposed changes do not adversely affect the safety or effectiveness of the subject device. The Teleflex Monodek Sutures continue to meet all applicable design and performance requirements and remain substantially equivalent to the predicate device.
{6}
K261988 Page 3 of 3
# **510(k) Summary**
- Sterilization validation in accordance with BS EN ISO 11135:2014 Sterilization of healthcare products – Ethylene oxide: Requirements for development, validation and routine control of a sterilization process for medical devices was performed and EO residual testing was performed to demonstrate the cycle achieves a sterility assurance level (SAL) of 10-6 and the device meets the EO residual requirements as defined in BS EN ISO 10993-7:2008+A1:2022.
- The modified packaging was validated according to the same test methods as previously performed for the Monodek Sutures cleared under K181774, in accordance with BS EN ISO 11607-1 Packaging for terminally sterilized medical devices – Part 1: requirements for materials, sterile barrier systems and packaging systems and ANSI/AAMI/ISO 11607-2 Packaging for terminally sterilized medical devices — Part 2: Validation requirements for forming, sealing and assembly processes.
- To validate the 3-year shelf life in the modified packaging, product stability testing in accordance with USP (United States Pharmacopeia) Absorbable Sutures, 43 <861> Sutures – Diameter, <871> Sutures – Needle Attachment, <881> Tensile Strength, and Resorption was performed. Testing was conducted following the same test methods as previously performed on the Monodek Suture submitted under K181774.
# **7. Conclusions**
The changes presented in this 510(k), which do not alter the intended use or fundamental scientific technology, are supported by the risk analysis and performance data following FDA-recognized consensus standards and the same testing as previously performed for the Monodek Sutures, most recently cleared under K181774, to demonstrate the modified device is substantially equivalent to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.