SmartGuard Technology; Predictive Low Glucose Technology

K261962 · Medtronic MiniMed · QJI · Jul 9, 2026 · Clinical Chemistry

Device Facts

Record IDK261962
Device NameSmartGuard Technology; Predictive Low Glucose Technology
ApplicantMedtronic MiniMed
Product CodeQJI · Clinical Chemistry
Decision DateJul 9, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1356
Device ClassClass 2
AttributesSoftware as a Medical Device, Therapeutic, PCCP, Pediatric

Indications for Use

SmartGuard technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGM), alternate controller enabled (ACE) pumps, and digitally connected devices to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose (SG) values. SmartGuard technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin. SmartGuard technology is intended for single patient use and requires a prescription. Predictive Low Glucose technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGM), alternate controller enabled (ACE) pumps, and digitally connected devices to automatically suspend delivery of insulin when the sensor glucose (SG) value falls below or is predicted to fall below predefined threshold values. Predictive Low Glucose technology suspends and resumes insulin delivery in Manual mode. Manual mode contains a bolus calculator that calculates an insulin dose based on user-entered data. Predictive Low Glucose technology is intended for the management of type 1 diabetes mellitus in persons 7 years of age and older, and type 2 diabetes mellitus in persons 18 years of age and older requiring insulin. Predictive Low Glucose technology is intended for single patient use and requires a prescription.

Device Story

Software-only interoperable automated glycemic controller (iAGC) embedded in ACE pump firmware; receives sensor glucose (SG) data via Bluetooth Low Energy (BLE) and user inputs via pump or app. SmartGuard (AHCL) algorithm uses adaptive control to adjust basal insulin every 5 minutes and deliver automatic correction boluses to maintain user-selectable glucose targets (100, 110, 120 mg/dL). Predictive Low Glucose (PLGM) algorithm monitors SG values to automatically suspend/resume insulin delivery in Manual mode when glucose is predicted to fall below thresholds. Integrated bolus calculator assists with meal dosing. Used in home and professional settings by patients/caregivers. Output commands sent to ACE pump to modulate insulin delivery. Benefits include improved glycemic control and reduced hypoglycemia risk.

Clinical Evidence

No clinical data; substantial equivalence based on design control activities, risk analysis, and verification/validation of software modifications.

Technological Characteristics

Software-only iAGC; embedded firmware in ACE pumps. Connectivity via Bluetooth Low Energy (BLE). Adaptive control algorithm (SmartGuard) and threshold-based suspension algorithm (PLGM). Supports U-100 insulin. Interoperable with iCGM/CGM devices. Complies with 21 CFR 862.1356 special controls.

Indications for Use

Indicated for management of Type 1 diabetes (age 7+) and Type 2 diabetes (age 18+) requiring insulin. Requires prescription. For use with compatible iCGM/CGM and ACE pumps.

Regulatory Classification

Identification

An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include: (i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use. (A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use. (B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect ( *e.g.,* more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events. (ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use. (iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices. (iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to ( *e.g.,* specification of glucose sensor performance) or accept commands from (*e.g.,* specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices. (vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place ( *e.g.,* validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller. (2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following: (i) Risk control measures to address device system hazards; (ii) Design decisions related to how the risk control measures impact essential performance; and (iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design. (3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following: (i) Secure authentication (pairing) to connected devices; (ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices; (iii) Sharing of necessary state information between the controller and any connected devices ( *e.g.,* battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified; (v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and (vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance. (4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include: (i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices; (ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices ( *e.g.,* infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and (iv) Connectivity events ( *e.g.,* establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section. (6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section. (7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable. (8) The labeling required under § 809.10(b) of this chapter must include: (i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part. (ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe. (iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate. (iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section. (v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate. (vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section. (vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device. (viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.

Submission Summary (Full Text)

{1} **FDA** U.S. FOOD & DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY I Background Information: A 510(k) Number K261962 B Applicant Medtronic Minimed Inc. C Proprietary and Established Names SmartGuard Technology Predictive Low Glucose Technology D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QJI | II | 21 CFR 862.1356 | Clinical Chemistry | | QJS | II | 21 CFR 862.1356 | Clinical Chemistry | | NDC | II | 21 CFR 868.1890 | Clinical Chemistry | II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own Class II device requiring 510(k). The following items are present and acceptable. 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} 3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device **has not changed. This change was for modifications to the PCCP to include updates to the Patient App and AHCL software to support therapy transfer between two pump reusables.** 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. K261962 - Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...