K261919 · Zhenjiang Assure Medical Equipment Co., Ltd. · IOR · Sep 16, 2026 · Physical Medicine
Device Facts
Record ID
K261919
Device Name
Multifunctional wheelchair (YJ-059)
Applicant
Zhenjiang Assure Medical Equipment Co., Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Sep 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The device is to provide mobility to persons limited to a sitting position.
Device Story
Manual wheelchair (YJ-059) for passive users; provides mobility to individuals limited to sitting position. Features seat tilting mechanism, reclining backrest, adjustable-angle foldable footrests, and calf support. Operated manually by occupant or caregiver; includes both occupant-operated and caregiver-operated brakes. Used in indoor/outdoor environments with level, accessible ground. Enables postural adjustments (tilting/reclining) without transfers to enhance user comfort. Quick-release wheels facilitate transport. Output is physical mobility and postural support for the patient.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified via ISO 7176 series (stability, brakes, dimensions, strength, friction, labeling, set-up) and ISO 16840-10 (ignition resistance). Biocompatibility verified via ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-23 (irritation).
Technological Characteristics
Manual, four-wheeled, foldable cross-brace wheelchair. Materials evaluated for biocompatibility per ISO 10993-1. Performance standards: ISO 7176-1, -3, -5, -7, -8, -11, -13, -15, -22; ISO 16840-10. Mechanical operation; no energy source. Features adjustable seat/backrest angles, calf support, and dual-brake system (occupant/caregiver).
Indications for Use
Indicated for adults (22 years and older) limited to a sitting position who require mobility assistance. Intended for use by the occupant or a caregiver. Maximum weight capacity 300 lbs.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 16, 2026
Zhenjiang Assure Medical Equipment Co., Ltd.
% Eva Li
Consultant
Shanghai Sungo Management Consulting Co., Ltd.
Rm. 1401, Dongfang Bldg., 1500# Century Ave.
Shanghai, 200122
China
Re: K261919
Trade/Device Name: Multifunctional wheelchair (YJ-059)
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: August 18, 2026
Received: August 19, 2026
Dear Eva Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}
K261919 - Eva Li
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{3}
K261919 - Eva Li
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261919 | ? |
| Please provide the device trade name(s). | | ? |
| Multifunctional wheelchair (YJ-059) | | |
| Please provide your Indications for Use below. | | ? |
| The device is to provide mobility to persons limited to a sitting position. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
510(K) Summary - K261919
# 1. Submitter
Zhenjiang Assure Medical Equipment Co.,Ltd.
Address: No.799, Chuqiao Road, Zhenjiang City, Jiangsu Province, China
Contact Person: Christine Dong
Tel: +86 13451917629
Email: christine@assuremedical.com.cn
Submission Correspondent
Contact: Eva Li
Phone: + 86 (215) 881-7802
Email: jia.li@sungoglobal.com
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave, Shanghai, 200122 CHN
Prepared Date: Feb 24,2026
# 2. Device
Name of Device: Multifunctional wheelchair
Common Name: Manual Wheelchair
Model: YJ-059
Regulatory Information
Classification Name: Mechanical Wheelchair
Regulatory Class: I
Product code: IOR
Regulation Number: 890.3850
Review Panel: Physical Medicine
# 3. Predicate device:
K201461
Ningbo Shenyu Medical Equipment Co.,Ltd.
Manual Wheelchair (A011)
This predicate has not been subject to a design-related recall.
# 4. Device description
YJ-059 manual wheelchair is a wheelchair for passive users, featuring a seat tilting mechanism, a reclining backrest, and foldable footrests with adjustable angles. It is intended for users who require postural adjustments, such as seat and backrest reclining, to enhance comfort.
It is suitable for both indoor and outdoor use, provided that the ground is level and accessible. This wheelchair enables controlled tilting, which facilitates changes in the user's position without the need for transfers. The wheels of the device can be quick-released.
1 / 5
{6}
510(K) Summary - K261919
# 5. Indication for use
The device is to provide mobility to persons limited to a sitting position.
6. Application: An adult user weight of no more than 300LBS, and the passenger capacity is only 1. The intended users are the wheelchair occupant and/or a caregiver.
7. The specification table is as below:
| Model | YJ-059 | | |
| --- | --- | --- | --- |
| Overall dimension(L*W*H)(mm) | (1190-1265)*670*(900-1170) | | |
| Folded dimension | width 1000mm,height 760mm | | |
| Seat width | 425mm | Seat plane angle | 56° |
| Seat depth | 440mm | Seat height from edge | 530-660mm |
| Backrest angle | 67° | Backrest height | 520mm |
| Backrest width | 370-420mm | Footrest-to-seat distance | 520-430mm |
| Footrest clearance | 10mm | Footrest length | 160mm |
| Total mass | 44.65kg | Mass of heaviest part | 31kg |
| Armrest-to-seat distance | 190-230mm | Footrest-leg-angle | 90° |
| Leg-to-seat-surface angle | 108° -180° | Front-armrest-to-backrest distance | 470mm |
| Armrest height | 190-230mm | Armrest length | 450mm |
| Armrest width | 80mm | Armrest angle | 3° |
| Distance between armrests | 410mm | Front location of armrest structure | 740-790mm |
| Hand rim diameter | 530mm | Castor wheel diameter | 20mm |
| Propelling wheel diameter | 600mm | Static stability downhill | 10.05° |
| Static stability uphill | 10.15° | Static stability sideways | 10.2° |
| Parking brake | | | |
| Max slope uphill | | 7.10° | |
| Max slope downhill | | 7.05° | |
# 7. Comparison of technological characteristics with the predicate device
| Device | Predicate Device | Subject Device | Results |
| --- | --- | --- | --- |
| 510K Number | K201461 | | — — |
| Manufacturer | Ningbo Shenyu Medical Equipment Co., Ltd. | Jiangsu Assure Medical Equipment Co., Ltd. | — — |
| Proprietary Name | Manual Wheelchair | Multifunctional wheelchair | — — |
| Model | A011 | YJ-059 | — — |
| Classification | I | I | same |
| Indications for use | The A011 Manual wheelchair is to | The device is to provide mobility | same |
2 / 5
{7}
# 510(K) Summary - K261919
| | provide mobility to persons limited to a sitting position. | to persons limited to a sitting position. | |
| --- | --- | --- | --- |
| Design Characteristic | Manual Operation, Four-Wheels, Foldable, Cross-Brace, Pull-to-Lock, Armrest, Backrest, Foot rest | Manual Operation, Four-Wheels, foldable,Cross-Brace, Armrest, Backrest, Foot rest, calf support | Similar*1 |
| Brake control | occupant-operated brake only | occupant-operated brake nursing people operated brake | Different*1 |
| Operation Environment | For indoor/outdoor use | For indoor/outdoor use | same |
| Control Mode | Mechanical | Mechanical | same |
| Size(unfold) | 1100 (L) *660 (W) * 910mm (H) | (1190-1265)(L)*670(W)*(900-1170)(H)mm | Different*2 |
| Stowage length/width/height | 810 (L) X 320 (W) X 930mm (H) | 1000 (L)* 670 (W)*760mm(H) | different*2 |
| Weight(Total) | 16kg(35.2lbs) | 44.65kg | different*2 |
| Weight Capacity | 136Kg(300lbs) | 300LBS | same |
| Seat Width | 480mm | 425mm | different*2 |
| Seat height | 540mm | 530-660mm | different*2 |
| Seat depth | 420mm | 440mm | different*2 |
| Back type | Fixed | Adjustable | Different*3 |
| Tires | Front: 200mm Rear:610mm | Front: 20mm Rear:600mm | different*2 |
| Armrest | Flip back armrest | adjustable | different*3 |
| Foot rest | Optional/ swing away Optional/ swing away | adjustable | different*3 |
| Rear Axle Position | Single | Single | Same |
| Frame Construction | Foldable frame Push inward from left and right sides to fold | Back rest flip | Different*3 |
| Safety Feature | Manual Wheel Lock | Manual Wheel Lock | Same |
| Performance | Comply with: ISO7176-1 ISO7176-3 ISO7176-5 ISO7176-7 ISO7176-8 ISO7176-11 ISO7176-13 ISO7176-15 ISO7176-16 | Comply with: ISO7176-1 ISO7176-3 ISO7176-5 ISO7176-7 ISO7176-8 ISO7176-11 ISO7176-13 ISO7176-15 ISO7176-16 | Same |
| Biocompatibility | Comply with: ISO10993-1 ISO10993-5 ISO10993-10 | Comply with: ISO10993-1 ISO10993-5 ISO10993-10 ISO10993-23 | Similar*2 |
3 / 5
{8}
510(K) Summary - K261919
# **Disscussion:**
| Similar*1: | Compare the predicate device, the subject device don't have the pull-to-lock feature, this will not raise any new risk of safety or effectiveness. |
| --- | --- |
| Different*1: | Compare the predicate device, the subject device adds a brake method, this feature will not raise any new risk of safety or effectiveness. |
| Different*2: | Compare the predicate device, the subject device have different value on the unfold size, device weight, Seat Width, Seat height, Seat depth, Tires size and so on. However the subject has pass the ISO 7176 related standard, so the above different will not raise any new risk of safety or effectiveness. |
| Different*3: | Compare the predicate device, the subject device' armrest, footrest is adjustable, and the backrest can be flip. These difference will not raise any new risk of safety or effectiveness. |
| Similar*2 | The stimulation test in ISO 10993-23 replaces the stimulation test in ISO 10993-10:2010, it will not raise any new risk of safety or effectiveness. |
# **8. Summary of Non-Clinic Performance Testing**
# **Performance Testing**
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ISO 7176-1:2014, Wheelchairs - Part 1: Determination of static stability
ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
ISO 7176-5: 2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and maneuvering space
ISO 7176-7: 1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
ISO 7176-8: 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
ISO 7176-11: 2012 Wheelchairs - Part 11: Test dummies
ISO 7176-13: 1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
ISO 7176-22:2014 Wheelchairs — Part 22: Set-up procedures
ISO 16840-10:2024 Wheelchair seating-Part 10: Resistance to ignition of postural support devices-Requirements and test method, heat source: heating cartridge
# **Biocompatibility**
Biocompatibility testing was conducted in accordance with the 2020 FDA guidance *Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process."* Testing included:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010)
- Irritation (ISO 10993-23:2010)
4 / 5
{9}
510(K) Summary - - K261919---
The testing supports the biocompatibility of the patient-contacting device materials that were shown to be non-cytotoxic, non-irritating, and non-sensitizing. The biocompatibility of the subject device is based on the use of low-biocompatibility risk materials and supporting information in accordance with Attachment G of FDA’s 2023 Biocompatibility Guidance.
# **9. Clinical Test Conclusion**
No clinical study is included in this submission.
# **10. Conclusion**
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the subject device is as safe, as effective, as well as the legally marketed predicate device cleared under K201461.
5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.