LOGIQ Focus is a general-purpose diagnostic ultrasound system intended for qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. The system is intended for use for the following Part of Body or Type of Tissue: Fetal/Obstetrical, Abdominal (including Renal, Gynecology/Pelvic, Thoracic/Pleural), Pediatric, Small Organ (Breast, Testes, Thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional and Superficial, Urology (including prostate), Transrectal, Transvaginal, Transesophageal and Intraoperative (Vascular), Interventional Guidance (including tissue biopsy, fluid drainage, vascular and nonvascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, Shear Wave Elastography, Attenuation Imaging and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD. LOGIQ Focus is intended to be used in a hospital or medical clinic.
Device Story
Mobile diagnostic ultrasound system; console with display, control panel, and various transducers (convex, linear, sector, mechanical 4D, wireless). Inputs: B-mode, Doppler, and other ultrasound signals. Operation: Physician/trained professional in hospital/clinic. Features AI-based Auto Preset Assistant and Auto Abdominal Color Assistant 2.0 for image optimization and anatomical detection. Outputs: Real-time ultrasound images, measurements, and analysis. Clinical decision-making: Assists in diagnosis, interventional guidance (biopsy/drainage), and tissue characterization. Benefits: Improved workflow efficiency, standardized imaging presets, and automated anatomical identification.
Clinical Evidence
No clinical studies required. Substantial equivalence supported by bench testing, including acoustic output, biocompatibility, electrical/electromagnetic safety, and software verification/validation. AI performance metrics provided: Auto Preset Assistant achieved >80% success rate; Auto Abdominal Color Assistant 2.0 achieved 93.8% detection accuracy, 0.94 sensitivity, 0.99 specificity, and 0.85 DICE score.
Technological Characteristics
Mobile ultrasound console; HDU or LCD display; various RS-connector transducers. Energy: Electrical. Connectivity: Wi-Fi, Bluetooth, DICOM. Software: Rule-based and AI-based (Auto Preset Assistant, Auto Abdominal Color Assistant 2.0). Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA PS 3.1-3.20, IEC 62359.
Indications for Use
Indicated for neonates, infants, children, adolescents, and adults for diagnostic ultrasound imaging, measurement, and analysis of human body and fluid across fetal, abdominal, pediatric, small organ, cephalic, cardiac, vascular, musculoskeletal, urological, and interventional applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}

GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC
Lee Bush
Regulatory Affairs Director
3200 N Grandview Blvd.
Waukesha, Wisconsin 53188
July 30, 2026
Re: K261835
Trade/Device Name: LOGIQ Focus
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, QIH
Dated: June 2, 2026
Received: June 2, 2026
Dear Lee Bush:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261835 - Lee Bush
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261835 - Lee Bush
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D. O'hara -S
Date: 2026.07.30 14:23:31 -04'00'
For
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261835 ? |
| Please provide the device trade name(s). | | ? |
| LOGIQ Focus | | |
| Please provide your Indications for Use below. | | ? |
| LOGIQ Focus is a general-purpose diagnostic ultrasound system intended for qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. The system is intended for use for the following Part of Body or Type of Tissue: Fetal/Obstetrical, Abdominal (including Renal, Gynecology/Pelvic, Thoracic/Pleural), Pediatric, Small Organ (Breast, Testes, Thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional and Superficial, Urology (including prostate), Transrectal, Transvaginal, Transesophageal and Intraoperative (Vascular), Interventional Guidance (including tissue biopsy, fluid drainage, vascular and nonvascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, Shear Wave Elastography, Attenuation Imaging and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD. LOGIQ Focus is intended to be used in a hospital or medical clinic. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
1/1
{4}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
510(k) Summary - K261835
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: May 29, 2026
Submitter: GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC
3200 N Grandview Blvd
Waukesha, WI 53188, USA
Manufacturer: GE Medical Systems (China) Co., Ltd.
No. 19, Changjiang Road,
Wuxi National High-Tech Development Zone,
214028 Jiangsu P. R. China
Primary Contact Person: Lee Bush
Regulatory Affairs Director GE
HealthCare
T:(262) 309-9429
Alternate Contact Person: Jian Sophie Xie
Regulatory Affairs Manager
GE HealthCare
T: +86-13338113925
Device Trade Name: LOGIQ Focus
Common/Usual Name: Diagnostic Ultrasound System
Classification Names: Class II
Product Code: IYN (primary), IYO, ITX, QIH (secondary)
Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN;
Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;
Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Medical Image Management and Processing System, 21 CFR 892.2050, 90-QIH
Primary Predicate Device: K253366 LOGIQ Fortis
Reference Device(s): K242005 Versana Premier, Versana Premier Lotus, LOGIQ F
K214039 LOGIQ P10, LOGIQ P9, LOGIQ P8
K253370 LOGIQ Totus
-Public-
Page 1 of 7
{5}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
# Device Description:
The proposed LOGIQ Focus system is a general-purpose, track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
The system consists of a mobile console, operator control panel, display monitor and some transducers. The operator control panel includes an ultrasound keyboard, an alfa-numeric key and a Touch panel as input sources of the device.
The variety of compatible transducers include convex, linear, sector, mechanical 4D, and wireless probes.
LOGIQ Focus is available in two configurations, LOGIQ Focus X and LOGIQ Focus S, which share an identical user interface design and layout. The primary differences between the two configurations are the display type and trackball design: LOGIQ Focus X features an HDU display with an illuminated trackball, while LOGIQ Focus S is equipped with an LCD display and a non-illuminated trackball.
# Intended Use/Indications for Use:
LOGIQ Focus is a general-purpose diagnostic ultrasound system intended for qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid.
The system is intended for use for the following Part of Body or Type of Tissue: Fetal/Obstetrical, Abdominal (including Renal, Gynecology/Pelvic, Thoracic/Pleural), Pediatric, Small Organ (Breast, Testes, Thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional and Superficial, Urology (including prostate), Transrectal, Transvaginal, Transesophageal, Intraoperative (Vascular), Interventional Guidance (including tissue biopsy, fluid drainage, vascular and non-vascular access).
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, Shear Wave Elastography, Attenuation Imaging and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD.
LOGIQ Focus is intended to be used in a hospital or medical clinic.
# Technology:
The LOGIQ Focus employs the same fundamental scientific technology as its predicate device(s).
-Public-
Page 2 of 7
{6}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
# Determination of Substantial Equivalence:
The proposed LOGIQ Focus is substantially equivalent to the predicate device, The following is an overview of the differences between the LOGIQ Focus and the predicate device LOGIQ Fortis (K253366).
# Indications for Use
The proposed LOGIQ Focus and predicate device LOGIQ Fortis (K253366) have similar clinical indications for use. Addition of Thoracic/Pleural under Abdominal and Interventional Guidance (including tissue biopsy, fluid drainage, vascular and non-vascular access) in the indications for use. Although Thoracic/Pleural and interventional guidance applications were not explicitly called out in the indications for use of LOGIQ Fortis, the capability to perform pleural imaging has already been clinically established and cleared on predicate device LOGIQ Fortis (K253366), and Interventional Guidance (including tissue biopsy, fluid drainage, vascular and non-vascular access) is enabled by existing probe capabilities that support biopsy and needle guidance.
# Transducers:
The proposed LOGIQ Focus and predicate device LOGIQ Fortis (K253366) transducers are similar, except for:
- Addition of 9L-RS, L3-12-RS, L4-20t-RS, L8-18i-RS and M5Sc-RS with identical indications for use and modes of operation as the predicate transducers 9L-D, L3-12-D, L4-20t-D, L8-18i-D and M5Sc-D cleared with LOGIQ Fortis (K253366), These transducers are identical to the corresponding predicate transducers except for the use of an RS connector instead of a D connector and were previously cleared on the reference device Versana Premier (K242005).
- Addition of ML6-15-RS with identical indications for use and modes of operation as the predicate transducer ML6-15-D cleared with LOGIQ Fortis (K253366), The ML6-15-RS transducer is the same as the ML6-15-D transducer except it uses RS connector instead of D connector. ML6-15-RS was previously cleared on reference device Cleared on LOGIQ P10, LOGIQ P9, LOGIQ P8 (K214039).
- Addition of RAB2-6-RS with identical indications for use and modes of operation as the predicate transducer RAB6-D cleared with LOGIQ Fortis (K253366). RAB2-6-RS was previously cleared on reference device Versana Premier (K242005).
- Addition of C1-5-RS with identical indications for use and similar modes of operation as previously cleared on reference device Versana Premier (K242005). Additional modes Shear Wave Elastography and Attenuation Imaging are being added, these modes were previously cleared on the similar transducer C1-6-D on LOGIQ Fortis (K253366).
- Addition of IC9-RS with identical indications for use and modes of operation as previously cleared on reference device LOGIQ P10, LOGIQ P9, LOGIQ P8 (K214039).
- Addition of E8Cs-RS, 12L-RS and 3Sc-RS with identical indications for use and modes of operation as previously cleared on reference device Versana Premier (K242005).
# Software Features/Functionality:
The software features supported in proposed LOGIQ Focus and the predicate LOGIQ Fortis (K253366) are similar except:
- Auto Abdominal Color Assistant 2.0: migrated from predicate device LOGIQ Fortis (K253366) and expanded to support probe C1-5-RS, which is similar to the previously cleared C1-6-D
-Public-
Page 3 of 7
{7}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
transducer on the predicate device LOGIQ Fortis (K253366).
- Auto Preset Assistant: migrated from predicate device LOGIQ Fortis (K253366) and expanded to support the 9L-RS, L3-12-RS, and ML6-15-RS transducers. The supported transducers L3-12-RS, and ML6-15-RS differ only in connector type compared to those used with predicate device LOGIQ Fortis (K253366) and the supported transducer 9L-RS differs only in connector type compared to 9L-D on reference device LOGIQ Totus (K253370).
- Ultrasound-Guided Fat Fraction (UGFF)/Ultrasound-Guided Attenuation Parameter (UGAP): migrated from reference device LOGIQ Totus (K253370) and expanded to support C1-5-RS, which is similar to the previously cleared C1-6-D transducer on the predicate device LOGIQ Totus (K253370).
- IOTA (International Ovarian Tumor Analysis): a worksheet-based ovary measurement and analysis tool, same as cleared on LOGIQ P10, LOGIQ P9, LOGIQ P8 (K214039)
# Accessories:
- Added the SONY UP-D899DC printer
- Added compatible OEM biopsy guide accessory compatibility for the C1-5-RS, 3Sc-RS, 12L-RS, IC9-RS Transducers
- Added new hardware options: Gel Warmer, Charge Box with Battery Kit, Battery Pack, Charge Box, 4D Module, ECG Module, Lemo Module
- Added new connectivity option Wi-Fi and Bluetooth 2 in 1 USB Stick Kit
- Added new peripherals: UP-D899DC Printer Shelf, UP-D899DC Printer Kit, DVD Shelf, DVD Kit
- Added new accessories: 64G USB Stick, USB Barcode Reader Honeywell, Operator Panel Rear Storage, Trolley Base Storage, Trolley Rear Storage, Trolley Top Storage, P6D Probe Holder, Vertical Endo Probe Holder Adapter for Cardiac Probe
# AI Testing Summary:
Auto Preset Assistant
| Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance | - The overall model success rate of the Abdomen, Air, Breast, Carotid, Leg, MSK, Scrotal, Thyroid and Carotid/Thyroid (Mixed) view suggestion is expected to be 80% or higher. - The number of individual patients' images were collected from: 67 patients - The number of samples, if different from above, and the relationship between the two: 1253 images |
| --- | --- |
| Information about clinical subgroups and confounders present in the dataset | - Gender: Male & Female: 32 (48%), 35 (52%) - Country: USA (100%) - Age: 58±17.1 - BMI: 27.1±8.9 - Ethnicity: Hispanic (5), Not Hispanic (61) - Race: Asian (8), Black (14), Hispanic/Latino (1), White (44) |
-Public-
Page 4 of 7
{8}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
| Information about equipment and protocols used to collect images | Mix of data from across 3 different probe models (L3-12-RS, ML6-15-RS, 9L-RS) and 1 Console (LOGIQ Focus) variants. The data collection protocol was standardized across all data collection sites. |
| --- | --- |
| Information about how the reference standard was derived from the dataset (i.e. the “truthing” process) | - Before the process of data annotation, all information displayed on the device is removed and performed on information extracted purely from Ultrasound B-mode images. - Readers [7] to ground truth the “anatomy” visible in static B-Mode image. (Before running AI) - Ran AI and created confusion matrix of ground truth vs AI predictions. - Calculated the accuracies of the algorithm against each class. |
| Description of how independence of test data from training data was ensured | The exams used for test/training validation purposes are separated from the ones used during training process and there is no overlap between the two. |
Auto Abdominal Color Assistant 2.0:
| Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance | - The overall model detection accuracy (sensitivity and specificity) and DICE score for the Aorta, Kidney, Liver/Spleen/inferior vena cava (IVC), Gallbladder (GB)/Urinary Bladder, Pancreas, and Air view is expected to be as follows: - Detection accuracy ≥ 80% (0.80) - Sensitivity (True Positive Rate): ≥ 80% (0.80) - Specificity (True Negative Rate): ≥ 80% (0.80) - DICE Similarity Coefficient (Segmentation Accuracy): ≥ 0.80 - The number of individual subjects: 67 - The number of annotation images: 728 - The model achieved a detection accuracy of 93.8%, with sensitivity of 0.94, specificity of 0.99, and a DICE score of 0.85, all of which meet the predefined acceptance criteria. |
| --- | --- |
| Information about clinical subgroups and confounders present in the dataset | - Gender: Male & Female: 32 (48%), 35 (52%) - Country: USA (100%) - Age: 58±17.1 - BMI: 27.1±8.9 - Ethnicity: Hispanic (5), Not Hispanic (61) - Race: Asian (8), Black (14), Hispanic/Latino (1), White (44) |
| Information about equipment and protocols used to collect images | Data from 1 probe model (C1-5-RS) and 1 Console (LOGIQ Focus) variants. The data collection protocol was standardized. |
-Public-
Page 5 of 7
{9}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
| Information about how the reference standard was derived from the dataset (i.e. the “truthing” process) | Reference Standard and Ground Truthing Process for Testing Dataset • Prior to annotation, all device-displayed information was removed to ensure unbiased evaluation. Only raw B-mode ultrasound images were used for analysis. • Independent clinical readers reviewed the static B-mode images to identify and annotate anatomical structures, establishing the reference standard (“ground truth”) before any AI inference was performed. • The AI model was then executed on the same dataset, and its predictions were compared against the ground truth annotations. • A confusion matrix was generated to evaluate the model’s performance across each anatomical class. • Detection accuracy metrics—including sensitivity, specificity, and DICE similarity coefficient—were calculated based on this comparison. |
| --- | --- |
| Description of how independence of test data from training data was ensured | The exams used for test/training validation purposes are separated from the ones used during training process and there is no overlap between the two. |
# Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIQ Focus complies with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety, 2005/A2:2021
- IEC 60601-1-2 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbance - Requirements and Tests, Edition 4.1, 2020
- IEC 60601-2-37, Medical Electrical Equipment – Part 2-37: Particular Requirements for the Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment, Edition 3.0, 2024
- ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within Risk Management Process, Fifth edition, 2018
- ISO 14971, Application of risk management to medical devices, 2019
- NEMA PS 3.1 - 3.20, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology), 2024e
- IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, Edition 2.1, 2017
- AAMI TIR69, Technical Information Report Risk management of radio-frequency wireless
-Public-
Page 6 of 7
{10}
[LOGO]
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
coexistence for medical devices and systems, 2017(R2020)
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification)
- Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
# Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ Focus, did not require clinical studies to support substantial equivalence.
# Conclusion:
Based on the equipment design similarities, conformance to recognized performance standards, and performance testing, GE HealthCare considers the proposed LOGIQ Focus to be as safe, effective, and performs in a substantially equivalent manner as the predicate LOGIQ Fortis (K253366).
-Public-
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.