K261807 · Bio Compression Systems, Inc. · JOW · Sep 11, 2026 · Cardiovascular
Device Facts
Record ID
K261807
Device Name
Sequential Circulator (SC-4004-DL, SC-4008-DL)
Applicant
Bio Compression Systems, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Sep 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sequential Circulators are sequential pneumatic compression devices intended for either primary or adjunctive treatment of lymphedema, phlebolymphedema, post-mastectomy edema, peripheral edema, lipedema, venous insufficiency, and venous stasis ulcers. Sequential Circulators are also intended for the prophylaxis of deep vein thrombosis (DVT). Intended for use in a home or healthcare setting.
Device Story
Sequential pneumatic compression system consisting of segmented garment and pneumatic pump; pump cyclically inflates garment chambers from distal to proximal to stimulate interstitial fluid and blood flow. Used in home or healthcare settings by patients or clinicians. Pump components include motor, air compressor, disc valves, and micro switch; rotating disc valve directs airflow to chambers. Provides adjustable pressure (20-120 mmHg), cycle times, and treatment modes (e.g., Focus Therapy, Pre-Therapy). Output is mechanical compression of limb; assists in edema reduction and DVT prevention.
Clinical Evidence
No clinical or animal studies were submitted. Substantial equivalence is supported by bench testing, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1), EMC (IEC 60601-1-2), and functional verification (pressure testing, cycle/treatment time validation, and mode operation).
Technological Characteristics
Class II pneumatic compression device. Materials: 200 Denier Nylon Oxford coated with 3 mil Polyurethane. Power: 100-240V, 50/60 Hz; 12VDC input. Applied part: Type BF. Ingress protection: IP21. Operation: Motor-driven rotating disc valve for sequential inflation. Connectivity: None. Software safety class: A (IEC 62304).
Indications for Use
Indicated for patients requiring treatment for lymphedema, phlebolymphedema, post-mastectomy edema, peripheral edema, lipedema, venous insufficiency, venous stasis ulcers, or DVT prophylaxis. Contraindicated for limb infections (e.g., cellulitis) without antibiotics, lymphangiosarcoma, suspected/confirmed DVT, inflammatory phlebitis, pulmonary embolism, or uncontrolled congestive heart failure.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 11, 2026
Bio Compression Systems, Inc.
Marc Somelofski
Vice President of RA/QA
120 W. Commercial Ave.
Moonachie, New Jersey 07074
Re: K261807
Trade/Device Name: Sequential Circulator (SC-4104-DL, SC-4108-DL)
Regulation Number: 21 CFR 870.5800
Regulation Name: Compressible Limb Sleeve
Regulatory Class: Class II
Product Code: JOW
Dated: August 7, 2026
Received: August 7, 2026
Dear Marc Somelofski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261807 - Marc Somelofski
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K261807 - Marc Somelofski
Page 3
Sincerely,
**NICOLE M. GILLETTE -S**
Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261807
Device Name
Sequential Circulators (SC-4104-DL, SC-4108-DL)
Indications for Use (Describe)
Sequential Circulators are sequential pneumatic compression devices intended for either primary or adjunctive treatment of lymphedema, phlebolymphedema, post-mastectomy edema, peripheral edema, lipedema, venous insufficiency, and venous stasis ulcers. Sequential Circulators are also intended for the prophylaxis of deep vein thrombosis (DVT). Intended for use in a home or healthcare setting.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
BIO
COMPRESSION
SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
# Summary
Date Prepared: August 8, 2026
# I. SUBMITTER
Bio Compression Systems, Inc.
120 West Commercial Avenue
Moonachie, NJ 07074, USA
Phone: +1-201-939-0716
Contact Person: Marc L. Somelofski
# II. DEVICE
| Device Name | Sequential Circulators (SC-4104-DL, SC-4108-DL) |
| --- | --- |
| Trade/Proprietary Name | Sequential Circulators (SC-4104-DL, SC-4108-DL) |
| Common/Usual Name | Sequential Circulators (SC-4104-DL, SC-4108-DL) |
| Classification Name | Compressible Limb Sleeve (21 CFR 870.5800) |
| Product Code | JOW |
| Class | Class II |
# III. PREDICATE DEVICE
Sequential Circulator models SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL (K210417).
Reference device K203178 is mentioned.
# IV. DEVICE DESCRIPTION
Summary
Page 1
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{6}
BIO COMPRESSION SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
Bio Compression Systems' Sequential Circulators are sequential pneumatic compression device which consists of a segmented pneumatic compression sleeve ("garment") connected to a pneumatic compression pump ("pump"). The pump cyclically inflates the garment's segments ("chambers") in sequence from the distal end toward the trunk of the body. The inflation of the garment compresses the area of the body on which it is worn, stimulating the movement of interstitial fluid and blood flow.
The core components of the pump are the motor, air compressor, disc valves, and micro switch. The air compressor generates air flow into a stationary disc valve. The motor moves a rotating disc valve. The geometry of the disc valve directs air flow and cyclically triggers the micro switch.
The device uses the Predicate Device's garments.
# V. INDICATIONS FOR USE
| | Subject Device | Predicate Device (K210417) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | Sequential Circulators are sequential pneumatic compression devices intended for either primary or adjunctive treatment of lymphedema, phlebolymphedema, post-mastectomy edema, peripheral edema, lipedema, venous insufficiency, and venous stasis ulcers. Sequential Circulators are also intended for the prophylaxis of deep vein thrombosis (DVT). Intended for use in a home or healthcare setting. | The Bio Compression Systems' Sequential Circulators are sequential, pneumatic compression devices intended for either primary or adjunctive treatment of lymphedema, peripheral edema, lipedema, venous insufficiency, and venous stasis ulcers. Sequential Circulators are also intended for the prophylaxis of deep vein thrombosis (DVT). Intended for use in a home or healthcare setting. | Equivalent. The addition of the phlebolymphedema and post-mastectomy edema indications are not a new use. Devices in the category with similar characteristics have been cleared for this indication (K203178). |
Summary
Page 2
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{7}
BIO COMPRESSION SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
| | Subject Device | Predicate Device (K210417) | Comparison |
| --- | --- | --- | --- |
| Contra - indications | Compression IS NOT recommended in the following conditions: • Infections in the limb, including cellulitis, without appropriate antibiotic coverage • The presence of lymphangiosarcoma • Suspicion or confirmation of the presence of Deep Vein Thrombosis (DVT) • Inflammatory phlebitis or episodes of pulmonary embolism • Congestive Heart Failure (CHF) unless controlled by medication • Other indications as identified by the treating physician | Compression IS NOT recommended in the following conditions: • Infections in the limb, including cellulitis, without appropriate antibiotic coverage • The presence of lymphangiosarcoma • Suspicion or confirmation of the presence of Deep Vein Thrombosis (DVT) • Inflammatory phlebitis or episodes of pulmonary embolism • Congestive Heart Failure (CHF) unless controlled by medication • Other indications as identified by the treating physician | Identical |
| Prescription/ over-the-counter use | Prescription | Prescription | Identical |
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | Subject Device | Predicate Device (K210417) | Comparison |
| --- | --- | --- | --- |
| Operating Principle | See Device Description | See Device Description | Identical |
| Power Supply Rating | 100-240V, 50/60 Hz | 100-240V, 50/60 Hz | Identical |
| Rated Input | 12VDC, 3A | 12VDC, 3A | Identical |
| Electrical Classification | Class II | Class II | Identical |
| Applied Part | Type BF | Type BF | Identical |
| Ingress Protection | IP21 | IP21 | Identical |
Summary
Page 3
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{8}
BIO COMPRESSION SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
| | Subject Device | Predicate Device (K210417) | Comparison |
| --- | --- | --- | --- |
| Cycle Time | 60-120 seconds in 15-second increments | • 1000 series: 60-120 seconds in 30-second increments • 2000/4000 series: 60-120 seconds in 15-second increments | Identical |
| Treatment Time | Continuous or adjustable from 10-120 minutes in 5-minute increments | Continuous or adjustable as follows: • 1000 series: 30-120 minutes adjustable in 30-minute increments • 2000/4000 series: 10-120 minutes in 5-minute increments | Identical |
| Pressure Range | 20-120 mmHg, adjustable in 1 mmHg increments | • 1000 series: 20-100 mmHg, adjustable in 5 mmHg increments • 2000/4000 series: 10-120 mmHg, adjustable in 1 mmHg increments | Identical |
| Pressure Accuracy | ± 20% | ± 20% | Identical |
| Pressure Gradient (decrease/proximal chamber) | –1 mmHg (SC-4108-DL) or –2 mmHg (SC-4104-DL) | –1 mmHg (8-chamber models) or –2 mmHg (4-chamber models) | Identical |
| Software Safety Class (IEC 62304) | A | A | Identical |
| Weight | • SC-4104-DL: 3.5 lbs. (1.59 kg) • SC-4108-DL: 3.85 lbs. (1.75 kg) | • SC-1004-DL, SC-2004-DL: 3.3 lbs. (1.5 kg) • SC-1008-DL, SC-2008-DL: 3.65 lbs. (1.66 kg) • SC-4004-DL: 3.5 lbs. (1.59 kg) • SC-4008-DL: 3.85 lbs. (1.75 kg) | Identical |
| Dimensions | 4.5" x 12" x 7.34" (114 mm x 304 mm x 186 mm) | • 1000 series: 4.5" x 11.75" x 7.75" (114 mm x 298 mm x 197 mm) • 2000/4000 series: 4.5" x 12" x 7.34" (114 mm x 304 mm x 186 mm) | Identical |
| Compression Sleeves ("Garments") | Uses Predicate's garments | 200 Denier Nylon Oxford, coated with 3 mil of Polyurethane | Identical |
Summary
Page 4
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{9}
BIO COMPRESSION SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
| | Subject Device | Predicate Device (K210417) | Comparison |
| --- | --- | --- | --- |
| User Interface | Soft key buttons | - 2000 series: soft key buttons - 4000 series: touch screen | Identical |
| Features | - Adjustable Cycle Time - Adjustable Pressure - Compliance/Usage Meter - Focus Therapy Mode - Pause - Pre-Therapy Mode (SC-4108-DL) - Timed Treatment - Individual Chamber Adjustment | - Adjustable Cycle Time - Adjustable Pressure - Compliance/Usage Meter - Focus Therapy Mode (2000/4000 series) - Pause - Pre-Therapy Mode (SC-4008-DL) - Timed Treatment - Individual Chamber Adjustment (4000 series) | Identical |
The Subject Device is based upon the Predicate Device and identical technological characteristics with respect to design, materials used, and construction.
- Subject Device and Predicate Device have the same operating principle and use the same disc valve assembly for operation
- Components used in Subject and Predicate Devices are identical, the same material, or have the same specifications
- The Predicate Device uses the Subject Device's compression garments
- Subject Device and Predicate Device have the same performance specifications
The differences between the Subject Device and Predicate Device exist in similar cleared devices and do not raise any different questions of safety and/or effectiveness.
- The addition of the phlebolymphedema and post-mastectomy edema indications are not a new use. Devices in the category with similar characteristics have been cleared for this indication (K203178)
- The Subject Device introduces modifications to the user interface, using soft keys over the previous version's touch screen
## VII. PERFORMANCE DATA
The following testing and performance data were provided in support of the substantial equivalence determination.
### Electrical safety and electromagnetic compatibility (EMC)
Summary
Page 5
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{10}
BIO COMPRESSION SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012 and A2:2021
- IEC 60601-1:2005/AMD1:2012 and A2:2020
- IEC 60601-1-2:2014 and A1:2020
- IEC 60601-1-6:2010/AMD1:2013 and A2:2020
- IEC 60601-1-11:2015 and A1:2020
**Predicate Device routine acceptance tests conducted on Subject Device**
- Observation of continuous and timed operation
- Pressure testing
- HiPot (dielectric withstand test) testing
**Functional verification and validation testing**
- Cycle time verification and validation
- Treatment time verification and validation
- Pressure setting endpoint testing
- Operation to confirm all modes, settings, and mode/setting changes function as intended
**Software Verification and Validation Testing**
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". The software documentation level for this device is basic and there are no cybersecurity risks.
**Clinical/Animal Studies**
Clinical/animal studies were not submitted, referenced, or relied on in this premarket notification submission for a determination of substantial equivalence.
**VIII. CONCLUSION**
The data included in this submission demonstrates that the Subject Device is substantially equivalent to the legally marketed Predicate Device and performs comparably to the Predicate Device that is currently marketed for the same intended use.
Summary
Page 6
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.