ERBECRYO 2 Cryosurgical Unit and Accessories

K261806 · Erbe Elektromedizin GmbH · GEH · Jun 9, 2026 · General, Plastic Surgery

Device Facts

Record IDK261806
Device NameERBECRYO 2 Cryosurgical Unit and Accessories
ApplicantErbe Elektromedizin GmbH
Product CodeGEH · General, Plastic Surgery
Decision DateJun 9, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization (destruction) of tissue by the application of extreme cold.

Device Story

Cryosurgical unit; utilizes CO2 gas source; Joule-Thomson effect for cooling. Components: steel/aluminum casing, LED display, membrane keypad, gas flow/control module, flow sensor, footswitch, instrument connection sockets. Operated by clinician via footswitch. Delivers regulated CO2 flow to connected cryoprobes for tissue destruction or adhesion. Software version 1.0.4 includes safety feature to detect cryoprobe defects via unusual pressure rise monitoring, triggering operation stop. Device is non-sterile and reusable.

Clinical Evidence

No clinical data. Bench testing only. Software verification performed per IEC 62304; functional testing conducted per ISO 14971 and ISO 13485 to validate the software update for cryoprobe defect detection.

Technological Characteristics

Materials: steel, aluminum, copper, plastic, electronic components. Energy: electrical (100-240V, 50-60Hz) and CO2 gas. Principle: Joule-Thomson effect. Dimensions: 410x130x370mm. Weight: 6.7kg. Connectivity: footswitch, instrument sockets, ECB (Erbe communication bus). Software: version 1.0.4, includes automated pressure-based defect detection. Sterilization: non-sterile, reusable.

Indications for Use

Indicated for cryoadhesion and tissue devitalization via extreme cold application. Prescription use only.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION Erbe Elektromedizin GmbH Matthias Kollek Regulatory Affairs Specialist Waldhoernlstrasse 17 Tuebingen, 72072 Germany June 9, 2026 Re: K261806 Trade/Device Name: ERBECRYO 2 Cryosurgical Unit and Accessories Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: June 1, 2026 Received: June 1, 2026 Dear Matthias Kollek: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download). {1} K261806 - Matthias Kollek Page 2 Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, {2} K261806 - Matthias Kollek Page 3 Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2026.06.09 08:54:46 -04'00' Colin K. Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. # Indications for Use 510(k) Number (if known) K261806 Device Name ERBECRYO 2 Cryosurgical Unit and Accessories Indications for Use (Describe) The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization (destruction) of tissue by the application of extreme cold. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} Erbe Elektromedizin GmbH Special 510(k) for ERBECRYO 2--- K261806 # **510(k) SUMMARY** | **Applicant** | Erbe Elektromedizin GmbH Waldhoernlestrasse 17 72072 Tuebingen Germany Tel: 0049-7071-755-0 Fax: 0049-7071-755-179 | | --- | --- | | **Contact Person** | Dr. Matthias Kollek Regulatory Affairs Specialist E-Mail: Matthias.Kollek@erbegroup.com | | **Date Prepared** | June 01, 2026 | | **Device Information** | | | Trade/Proprietary Name: | ERBECRYO 2 Cryosurgical Unit and Accessories | | Common Name: | Cryosurgical Unit | | Classification Name | Cryosurgical Unit & Accessories | | Regulation Number: | 21 CFR 878.4350 | | Class: | II | | Product Code: | GEH | | **Legally Marketed Predicate Devices** | ERBECRYO® 2 Cryosurgical Unit and Accessories - K190651 & K253230 | Page 1 of 4 {5} Erbe Elektromedizin GmbH Special 510(k) for ERBECRYO 2 K261806 ### Device Description The ERBECRYO 2 unit consists of a steel casing with an aluminum casing button. The plastic face frame comprises the power switch, a plastic receptacle, and a LED display with a membrane keypad. Inside the housing the hardware components are installed. These are the electrical mains supply, the main board (central control unit), the user interface, the gas flow and control module, the flow sensor module and connection sockets for compatible instruments, the footswitch and ECB (Erbe communication bus). The Unit is connected to an electrical power source, a Carbon Dioxide (CO₂) source, the footswitch, and an instrument. Upon activation via the pedal of the footswitch, the Unit delivers the regulated CO₂ flow to the end of the connected instrument. ### Indications for Use The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization (destruction) of tissue by the application of extreme cold. ### Comparison of Technological Characteristics #### ERBECRYO 2 | Characteristics | Subject Device ERBECRYO 2 Cryosurgical Unit and Accessories | Predicate Device ERBECRYO 2 Cryosurgical Unit and Accessories | Discussion | | --- | --- | --- | --- | | **510(k) Number** | N/A | K190651 & K253230 | Different | | **Model Name** | ERBECRYO® 2 Cryosurgical Unit and Accessories | ERBECRYO® 2 Cryosurgical Unit and Accessories | Same | | **Manufacturer** | Erbe Elektromedizin GmbH | Erbe Elektromedizin GmbH | Same | | **Common Name** | Cryosurgical Unit and Accessories | Cryosurgical Unit and Accessories | Same | | **Classification Regulation** | 21 CFR 878.4350 | 21 CFR 878.4350 | Same | | **Product Code** | GEH | GEH | Same | | **Device Class** | II | II | Same | | **Indications for Use** | The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization (destruction) of tissue by the application of extreme cold. | The ERBECRYO® 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization (destruction) of tissue by the application of extreme cold. | Same | | **Intended Use** | Supply of CO₂ to a connected instrument for | Supply of CO₂ to a connected instrument for | Same | Page 2 of 4 {6} Erbe Elektromedizin GmbH Special 510(k) for ERBECRYO 2 K261806 | Characteristics | Subject Device ERBECRYO 2 Cryosurgical Unit and Accessories | Predicate Device ERBECRYO 2 Cryosurgical Unit and Accessories | Discussion | | --- | --- | --- | --- | | | cooling via the Joule-Thompson effect. | cooling via the Joule-Thompson effect. | | | **Prescription or OTC** | Prescription | Prescription | Same | | **Materials** | Metal sheet, aluminum, copper pipes, plastics and wiring, electronic components | Metal sheet, aluminum, copper pipes, plastics and wiring, electronic components | Same | | | | | | | **Dimensions of Unit** | 16.1'x5.1'x14.6' 410x130x370mm) | 16.1'x5.1'x14.6' 410x130x370mm) | Same | | **Weight of Unit** | 14.8lbs (6.7kg) | 14.8lbs (6.7kg) | Same | | **Technical Data of Unit** | Rated supply voltage: 100 - 240V (±10%) | Rated supply voltage: 100 - 240V (±10%) | Same | | | Rated supply frequency: 50 - 60Hz | Rated supply frequency: 50 - 60Hz | Same | | | Line current: 0.4 - 0.8A | Line current: 0.4 - 0.8A | Same | | **Software Version** | 1.0.4 | 1.0.3 | Different | | **Performance Data of Unit** | No. of effect levels: 1 - 5 (depending on instrument) | No. of effect levels: 1 - 5 (depending on instrument) | Same | | | Activation: via footswitch | Activation: via footswitch | Same | | | Cooling gas used: CO_{2} | Cooling gas used: CO_{2} | Same | | | Input Pressure: 653 - 943psi (45 - 65bar) | Input Pressure: 653 - 943psi (45 - 65bar) | Same | | **Provided Condition** | Non-sterile, reusable | Non-sterile, reusable | Same | The intended use of the subject and predicate device is the same. Compared to previously cleared device, the software of the subject device was updated to detect defects in connected Cryoprobes to stop operation in cases of an unusual pressure rise. Besides the updated software, the subject device has the same design, operating principle, and (performance) specifications compared to the cleared device. ### Summary of non-clinical bench performance testing To show safety and effectiveness of the subject device with updated software version, software verification in compliance with FDA recognized consensus standard IEC 62304 was performed to demonstrate that the updated software works as intended. Functional testing to validate the software update was conducted based on the risk analysis according to ISO 14971 in compliance with ISO 13485, clause 7.3 to ensure that the subject device performs as intended and meets design specifications. Page 3 of 4 {7} Erbe Elektromedizin GmbH Special 510(k) for ERBECRYO 2--- K261806 ## Conclusion The subject device has the same intended use, the same fundamental design, substantially equivalent performance characteristics, and the same energy source as the predicate device. The software update was tested as described above and assessed with regards to safety and effectiveness. Taken together, the software update of the subject device does not raise new or different questions of safety and effectiveness, and the subject device is substantially equivalent to the predicate device. Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...