Plenity-V (MC0979 / GS110)

K261783 · Gelesis SRL · QFQ · Oct 2, 2026 · Gastroenterology, Urology

Device Facts

Record IDK261783
Device NamePlenity-V (MC0979 / GS110)
ApplicantGelesis SRL
Product CodeQFQ · Gastroenterology, Urology
Decision DateOct 2, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5982
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Plenity-V is indicated to aid in weight management in adults with excess weight or obesity, with a Body Mass Index (BMI) of 25-40 kg/m2, when used in conjunction with diet and exercise

Device Story

Plenity-V is an oral, non-systemic, space-occupying weight management device. It consists of HPMC capsules containing thousands of cellulose and citric acid hydrogel particles. Patients ingest three capsules with water 20-30 minutes before lunch and dinner. In the stomach, capsules disintegrate, releasing particles that hydrate up to 100 times their weight, occupying approximately 25% of stomach volume. The resulting gel mixes with food to increase volume, elasticity, and viscosity, creating a sensation of fullness. The device is non-absorbable, passes through the GI tract, and is eliminated via normal bowel movements. It is intended for OTC use by adults. The device provides a mechanical satiety aid to support weight loss efforts.

Clinical Evidence

No new clinical studies were conducted for the HPMC capsule modification. Clinical evidence relies on the original De Novo and GLOW/GLOW-EX studies for the predicate. The GLOW pivotal study (n=not specified) showed 58.6% of patients lost ≥5% total body weight at 6 months (95% CI 51.8-65.4%). Safety profile showed no significant difference in TEAEs between device (71.3%) and placebo (70.6%) groups, with no unanticipated adverse device effects or SAEs associated with the device.

Technological Characteristics

Hydrogel particles composed of cellulose and citric acid; HPMC capsule shell. Dimensions: particles approx. size of salt grain. Energy source: none (mechanical). Connectivity: none. Sterilization: non-sterile. Biocompatibility: ISO 10993-1:2025 compliant. Performance: disintegration time NMT 15 min; media uptake NLT 70 g/g; elastic modulus NLT 500 Pa. Packaging: ASTM D7709-12 (water vapor transmission), ASTM D4169:2016 and ASTM D999-08:2015 (shipping/vibration/compression).

Indications for Use

Indicated for weight management in adults with excess weight or obesity (BMI 25-40 kg/m2) as an adjunct to diet and exercise.

Regulatory Classification

Identification

This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.

Special Controls

In combination with the general controls of the FD&C Act, the ingested, transient, space occupying device for weight management and/or weight loss is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows: (i) Performance bench testing in a simulated use model must evaluate device disintegration and device hydration state throughout the gastrointestinal tract; (ii) Bioburden and moisture content assessments must evaluate device infection risk throughout the labeled shelf life; and (iii) Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life. (3) Clinical performance testing must demonstrate the device performs as intended and evaluate the following: (i) Weight change; (ii) All adverse events, including obstruction, dilation, diarrhea, constipation, and dehydration; and (iii) Interaction with representative medications. (4) Physician and patient device labeling must state: (i) The clinical benefit of the device as assessed by using percent total body weight loss; (ii) Treatment must be offered in combination with diet and exercise; (iii) Instructions on how to use the device as intended including how to avoid interaction with medication; and (iv) The shelf life of the device.

Predicate Devices

Submission Summary (Full Text)

{1} [LOGO] FDA U.S. FOOD & DRUG ADMINISTRATION October 2, 2026 Gelesis SRL Giorgio Sorcini Sr. Dir. Design and RA Via Europa 187 Calimera, LE 73021 Italy Re: K261783 Trade/Device Name: Plenity-V (MC0979/GS110) Regulation Number: 21 CFR 876.5982 Regulation Name: Ingested, Transient, Space Occupying Device For Weight Management and/or Weight Loss Regulatory Class: Class II Product Code: QFQ Dated: July 3, 2026 Received: July 6, 2026 Dear Giorgio Sorcini: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K261783 - Giorgio Sorcini Page 2 may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) {3} K261783 - Giorgio Sorcini Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, STEPHANIE COLE -S *for* Anthony Lee, Ph.D., M.B.A Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K261783 Device Name Plenity-V (MC0979/GS110) Indications for Use (Describe) Plenity-V is indicated to aid in weight management in adults with excess weight or obesity, with a Body Mass Index (BMI) of 25-40 kg/m2, when used in conjunction with diet and exercise Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) # **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. # **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} GELESIS K261783 page 1 of 5 # 510(k) Summary # Regulatory Information Classification Name: Ingested, transient, space occupying device for weight management and/or weight loss. Common Name: Weight management device Regulation Number: 21 CFR 876.5982 Classification: Class II Product Code: QFQ Device Trade Name: Plenity-V (MC0979/GS110) Indications for Use: Plenity-V is indicated to aid in weight management in adults with excess weight or obesity, with a Body Mass Index (BMI) of 25 to 40 kg/m², when used in conjunction with diet and exercise. # Sponsor Information: Gelesis Srl Via Europa, 187 73021 Calimera (LE) Italy # Contact Person: Giorgio Sorcini Sr. Design and Regulatory Affairs Director Phone: +39 347 5047105 email: gsorcini@gelesis.com # Date Prepared: September 10th, 2026 # Device description Plenity-V is an absorbent hydrogel, delivered by a Hydroxypropyl methylcellulose (HPMC) capsule. Each capsule contains thousands of hydrogel particles (0.75 grams [g] per capsule); each particle is approximately the size of a grain of salt. Plenity-V is non-systemic and works directly {6} GELESIS K261783 page 2 of 5 in the gastrointestinal (GI) tract. Plenity-V hydrogel is made from two naturally derived ingredients, cellulose and citric acid, forming a three-dimensional matrix designed to occupy volume in the stomach and small intestine, to create a sensation of fullness. Plenity-V is provided non-sterile. ![img-0.jpeg](img-0.jpeg) ### Plenity-V capsules with hydrogel exposed The capsules disintegrate in the stomach and release the Plenity-V particles, which can hydrate up to 100 times their original weight. When fully hydrated, the individual non-clustering Plenity-V particles from the 2.25 g/dose occupy about a quarter of the average stomach volume. The gel particles mix with ingested foods, creating a larger volume with higher elasticity and viscosity in the stomach and small intestine. Plenity-V passes through the digestive system, maintaining its three-dimensional structure in the stomach and small intestine before breaking down in the colon. The water is then released and reabsorbed by the body. Plenity-V particles are eliminated through normal bowel movements; they are not absorbed by the body. Patients consume three (3) capsules (2.25 g/dose) with water before both lunch and dinner. Plenity-V is supplied in blisters holding 8 doses of 3 capsules each, in separable peelable single-dose containers. As the patients take two doses daily, 7 blisters are provided in each Plenity-V sell unit, to provide the therapy needs for 28 day (4 weeks). ### Predicate Device The Plenity-V device is identical to the currently cleared device Plenity® (K230133) in terms of its technological characteristics and mechanism of action, apart from the capsule material, as the white porcine gelatin capsule has been replaced by a transparent HPMC capsule, same size. Plenity-V has the same intended use as its predicate, Plenity® (K230133). ### Technological Characteristics As noted above, Plenity-V remains technologically identical to its predicate device. See the table below for the device specifications. {7} GELESIS K261783 page 3 of 5 Table: Plenity-V Specifications | | Plenity-V / Plenity® K230133 | | --- | --- | | **General Specifications** | | | **Capsule disintegration time** | NMT 15 minutes | | **Media uptake after disintegration** | NLT 70 g/g | | **Capsule content uniformity** | 760 mg ± 10% | | **Media uptake** | NLT 70 g/g | | **Elastic modulus** | NLT 500 Pa | | **Loss on drying** | NMT 10% | | **Dosing** | Plenity-V should be taken with water twice a day, 20 to 30 minutes before lunch and 20 to 30 minutes before dinner. Each dose includes 3 capsules of Plenity-V (each dose provided in peelable separable blisters) | ### Summary of Nonclinical/Bench Studies No non-clinical/bench studies were specifically conducted in support of the replacement of capsule's material as the device is the same as the predicate. The device labeling interface cleared by K230133, supporting self-selection ad use without the intervention of a prescriber, has not been modified apart from the description of package and the list of ingredients. #### Biocompatibility The Plenity® devices biocompatibility assessment has been updated to comply with ISO 10993-1:2025 requirements and includes Plenity-V capsule new material. Devices have contact with intact mucosal membranes for long-term exposure (> 30 days). All the biological endpoints have been addressed by biocompatibility studies and/or rationales based on scientific literature, EU and US food regulations, as the device components have long and documented story of use with the same contact. Endpoints considered are: - Cytotoxicity, tested in compliance with ISO 10993-5 (Elution and Agar mediated assays). - Sensitization, tested in compliance with ISO 10993-10 (Guinea Pig Maximization Test). - Irritation: tested in compliance with ISO 10993-10 (Oral Mucosal Irritation Study in Hamsters - Collar Method - 14 Days). {8} GELESIS K261783 page 4 of 5 - Acute Systemic Toxicity: tested in compliance with ISO 10993-11 (Oral Toxicity Study in Mice, 7 Days). - Genotoxicity: tested in compliance with ISO 10993-3 (Bacterial Reverse Mutation Study, Genotoxicity - Mouse Lymphoma Assay and Mouse Peripheral Blood Micronucleus Study) - Sub-chronic Toxicity: tested in compliance with ISO 10993-11 (Safety and Tolerability 90 Days Study in Rats with Recovery) - Chronic Systemic Toxicity: assessed based on chemical characterization according to ISO 10993-18 and literature data. - Reproductive/Developmental Toxicity: assessed by literature on known experience with the constituent materials. - Biodegradation and toxicokinetic: assessed based on degradation products studies according to ISO 10993-13 compared with literature data. - Implantation: assessed by literature data on constituent substances, but not required for devices administered orally. # Bench models Studies were performed on a bench model to evaluate swelling properties of the hydrogel at each relevant point in a simulated gastro-intestinal transit. A specific Plenity-V assay was performed on bench to assess, by an internally developed method (RD-0518.01), correct hydration with no lumps of the hydrogel in simulated clinical conditions. # Stability and shipping studies Stability studies in compliance with ICH guidances have been performed with Plenity pods (Plenity® primary packaging level). Comparison by ASTM D7709-12 Water Vapor Transmission Rate test showed that results are fully applicable to Plenity-V blisters. An ongoing stability study with multiple lots of Plenity-V has been started to confirm on bench a shelf-life of 30 months for Plenity-V. Similarly, ship testing was performed to verify that the packaging configurations for Plenity® met the expected shipping and storage conditions during commercial use. The shipping/environmental testing was performed per ASTM D4169:2016 and ASTM D999-08:2015. The test environment included exposure to vibration, stacking, compression and impact simulations. As HPMC capsules are less brittle than gelatin components, and the blister provides a closer containment to the capsules, the tests performed should represent a worst condition, thus results are applicable to validate Plenity-V shipping features. # Usability studies Usability studies in compliance with IEC 62366-1:2015/AMD1:2020 have been performed to support transition of Plenity® to OTC distribution in K230133 submission, which demonstrated an effective communication of requirements for a safe device self-selection and use. No changes have {9} GELESIS K261783 page 5 of 5 been made to the labelling contents in the Plenity-V box and IFU with respect to the cleared interface. ## Summary of Clinical Information No clinical studies were specifically conducted in support of the change of capsule's material in Plenity-V, as the new component fulfills the disintegration time requirement. The Plenity was originally granted by De Novo submission based on results of GLOW and GLOW-EX studies. It demonstrated robust efficacy in GLOW pivotal study, with 58.6% of patients losing ≥5% total body weight in 6 months (95% CI 51.8 to 65.4%), and 27% of patient loosing 10% or more. The incidence of treatment emergent adverse events (TEAEs) in the GLOW study (71.3% in Plenity arm versus 70.6% Placebo) and the incidence of AEs related to Plenity (39.5% in Plenity arm versus 30.3% in Placebo) were no different in the device group compared to Placebo. There were no unanticipated adverse device effects or deaths in the study. There were no SAEs associated with the use of Plenity. There was one SAE in the Placebo treatment group. This exemplary safety profile was maintained for the 21 subjects exposed to Plenity for one year in the GLOE-EX study. ## Conclusion Gelesis has demonstrated that the proposed Plenity-V device is substantially equivalent to its predicate device. The intended use and mechanism of action are identical. The device characteristics, specifications, design and performance are unchanged apart from replacement of the white porcine gelatin capsule with transparent HPMC components. This design replaces a component which is derived from animal tissues with a safer plant-based material and eliminates the withe opacifier (Titanium Dioxide), thus reducing the biological risk of the device. Plenity-V meets the same special controls regarding biocompatibility and non-clinical performance as the predicate device. Plenity OTC was tested and shown to perform according to its specifications that are the same as those of the previously cleared Plenity® devices (K230133 and DEN180060). The new capsule material fulfills the same design requirements. After review of the Plenity-V risk analysis, it was concluded that there were no new unacceptable safety and effectiveness risks identified, as a result of replacing the porcine gelatine capsule with a HPMC component, that would increase risk to patients and/or present unacceptable residual risks.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...