K261674 · Beijing Nubway S&T Co., Ltd. · GEX · Jul 16, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261674
Device Name
Skin Air Cooling System (QALF-01)
Applicant
Beijing Nubway S&T Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Skin Air Cooling System is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.
Device Story
Skin Air Cooling System (QALF-01) provides localized skin cooling via forced convection; utilizes compressor, evaporator, and fan to generate cold air (down to -30°C); delivered to skin surface via interchangeable air outlets (4 sizes) without direct contact. Operated by clinicians in dermatological/surgical settings; used in conjunction with laser or injection procedures to mitigate thermal damage and pain. System includes main unit with LCD touch interface for parameter control (temperature, air volume, defrosting). Cooling cycle involves refrigerant compression, condensation, and evaporation. Benefits include patient comfort and protection of skin during heat-generating procedures.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (AAMI/ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance validation of temperature and air speed specifications.
Technological Characteristics
Closed-loop cooling system using R404a refrigerant; compressor-based. Materials: internal metal chassis, ABS plastic shell, ABS+PC air outlets. Dimensions: 955mm x 400mm x 840mm. Power: 110-120V AC. Interface: RS232, TTL. Software: basic level of concern, embedded control for system initialization, UI, and cooling parameters.
Indications for Use
Indicated for adults (22 years and older) to minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
July 16, 2026
Beijing Nubway S&T Co., Ltd.
% Lena Zhang
RA Manager
Tianjin Xinnuocheng Medical Technology Co., LTD.
Rm. 1505, Wanhai Bldg., Dazhigu Sub-District
Hedong District
Tianjin, 300170
China
Re: K261674
Trade/Device Name: Skin Air Cooling System (QALF-01)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: May 20, 2026
Received: May 21, 2026
Dear Lena Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261674 - Lena Zhang
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K261674 - Lena Zhang
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.07.16 16:00:18 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261674 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Skin Air Cooling System (QALF-01) | | |
| Please provide your Indications for Use below. | | ? |
| Skin Air Cooling System is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
510(k) Summary
K261674
### 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
1. Date of Preparation: 2026/06/28
2. Sponsor Identification
**Beijing Nubway S&T Co., Ltd.**
202,No.5 workshop, No.1 caida 3rd Road Nancai Shunyi District, Beijing China. 101300.
Contact Person: Xiting Fan
Position: Registration Manager
Tel: +86-13716933250
Email: 1015932471@qq.com
3. Designated Submission Correspondent
Ms. Lena Zhang
Tianjin Xinnuocheng Medical Technology Co., LTD.
Room 1505, Wanhai Building, Dazhigu Sub-district, Hedong District Tianjin, Hedong 300170 CHINA
Tel: +86-13821206320
Email: ra@tianjinxinnuochengyi.com.cn
4. Identification of Proposed Device
Trade Name: Skin Air Cooling System (QALF-01)
Common Name: Powered Laser Surgical Instrument
Regulatory Information
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Classification: II
Product Code: GEX
Regulation Number: 878.4810
Review Panel: General & Plastic Surgery
1 / 7
{5}
510(k) Summary
K261674
# 5. Identification of Predicate Device(s)
Predicate device 1:
510(k) Number: K233416
Product Name: Cryojet Plus (SHE-ACP001-1)
Manufacturer: Beijing Sano Laser S&T Development Co., Ltd.
Predicate Device 2:
510(k) Number: K220020
Product Name: Cryo 7
Manufacturer: Zimmer MedizinSysteme GmbH
# 6. Device Description
The Skin Air Cooling System is composed of main unit (including power supply system, control system, cooling system), 4 different sizes of air outlet, air duct, air duct support rod, power cord, storage platform, casters and other accessories. The power supply system is composed of power cable, air switch, key switch and emergency stop switch power; the control system consists of LCD screen display and control panel, etc; the cooling system includes compressor, evaporator, radiator, fan, water tank, etc.
How it works: Gaseous refrigerant is compressed into high-temperature and high-pressure gas by high-power compressor and sent to condenser for cooling and liquefaction. Refrigerant can produce cold air by condenser and evaporation system for skin cooling through conduction, evaporation and forced convection. Evaporator absorbs heat in air and vaporizes the heat, then heat is gaseous, then the gas is back to compressor and continue to be compressed. Thus cooling is done through such circulation. Defrosting system melts the frost inside evaporation system into liquid water and discharge the water through drain valve.
Software is used for system initialization, displaying user interface and cooling system control. Parameter setting (temperature and air volume level) can be done by touching LCD screen, and the Stand by/ready and defrosting of device can be controlled. The software can be used in conjunction with hardware system during treatment to minimize thermal damage to tissues and reduce pain.
# 7. Indication For Use Statement:
Skin Air Cooling System is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.
# 8. Substantially Equivalent (SE) Comparison
Table 1 General Comparison
2 / 7
{6}
510(k) Summary
K261674
| Item | Proposed Device | Predicate Device 1 K233416 | Predicate Device 2 K220020 | Remark |
| --- | --- | --- | --- | --- |
| Device name | Skin Air Cooling System | Cryojet Plus (SHE-ACP001-1) | Cryo 7 | / |
| Classification Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Classification | II | II | II | SAME |
| Product Code | GEX | GEX | GEX | SAME |
| Regulation Name | Laser surgical instrument for use in general and plastic surgery and in dermatology | Laser surgical instrument for use in general and plastic surgery and in dermatology | Laser surgical instrument for use in general and plastic surgery and in dermatology | SAME |
| Indications for use | Skin Air Cooling System is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections. | Cryojet Plus is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections. | The Cryo 7 is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections. | SAME |
| Device Description | The Skin Air Cooling System consists of a compressor, evaporator and fan to cool the patient's skin locally without contact to skin. | The Cryojet Plus Air Device consists of a compressor, evaporator and fan to cool the patient's skin locally without contact to skin. | The Cryo 7 Cold Air Device consists of a compressor, evaporator and fan to cool the patient's skin locally without contact to skin. | SAME |
| Air outlets | 1. Small Sized Round Air outlet: φ 10mmFor small area treatment and skin cooling meanwhile (e.g. eyebrow, armpits, forehead.)2. Medium Sized Square Air outlet: 12 * 4mm | 1. Small Sized Round Adapter:For small area treatment and skin cooling meanwhile (e.g. eyebrow, armpits, forehead.)2. Medium Sized Square Adapter: | there is no adapter for choice, wind is blown out from end of wind tube directly. | Analysis(1) |
3/7
{7}
510(k) Summary
K261674
| | For medium area treatment and skin cooling meanwhile (especially used for hair removal on arms, armpits and legs, in conjunction with hair removal machine.)3. Large Sized Square Air outlet: 27 * 5mm, 43* 5mmFor large area treatment and skin cooling meanwhile (especially used for hair removal on thighs, abdomen, in conjunction with hair removal machine.) | For medium area treatment and skin cooling meanwhile (especially used for hair removal on arms, armpits and legs, in conjunction with hair removal machine.)3.Large Sized Square Adapter:For large area treatment and skin cooling meanwhile (especially used for hair removal on thighs, abdomen, in conjunction with hair removal machine.) | | |
| --- | --- | --- | --- | --- |
| Performance Characteristics | -30°C ± 3°C, At an air speed of 200±50 to 620 ± | -25°C ± 3°C, At an air speed of 200 ± 50 to 620 ± | -30°C ± 3°C, At an air speed of 210 ± 40 to | The temperature is |
4/7
{8}
510(k) Summary
K261674
| Measurement at device outlet | 60 liters/minute. | 60 liters/minute. | 600 ± 100 liters/minute. | SAME with Predicate Device 2, the air speed is SAME with Predicate Device 1 |
| --- | --- | --- | --- | --- |
| Cooling Method, Operation Principle | Cooling agent-based condenser and evaporation system. Cold air is produced which is then delivered to skin for skin cooling by conduction, evaporation, and forced convection. | Cooling agent-based condenser and evaporation system. Cold air is produced which is then delivered to skin for skin cooling by conduction, evaporation, and forced convection. | Cooling agent-based condenser and evaporation system. Cold air is produced which is then delivered to skin for skin cooling by conduction, evaporation, and forced convection. | SAME |
| Cooling Material | Gas (Air) | Gas (Air) | Gas (Air) | SAME |
| Mains Voltage | AC110V ~ 120V, 50/60Hz | 100V - 120V / 50-60Hz | 100V - 120V / 50-60Hz | SAME |
| Main Fuse | Without circuit breaker in main switch | Without circuit breaker in main switch | 16A circuit breaker in main switch | SAME with Predicate Device 1 |
| Power Consumption | Max. 12A | Max. 8A | Max. 10A | Analysis(2) |
| IP Classification | IPX0 | IPX0 | IP20 | SAME with Predicate Device 1 |
| Design | Internal metal chassis and framework by industrial designed ABS plastic shell, adapter material is ABS+PC. | Internal metal chassis and framework by industrial designed ABS plastic shell, adapter material is ABS+PC. | Internal metal chassis covered by industrial designed plastic shell. | SAME with Predicate Device 1 |
| Materials | Environmentally approved R404a, closed loop cooling system. | Environmentally approved R404a, closed loop cooling system. | Environmentally approved R452a, closed loop cooling system. | SAME with Predicate Device 1 |
| Cooling Spot Size | Varies in cooling spot size (coverage area) with a distance of 5cm from wind outlet part to skin - 10cm². | Varies in cooling spot size (coverage area) with a distance of 5cm from wind outlet part to skin – 10cm² | Varies in cooling spot size (coverage area) with a distance of 5cm from wind outlet part to skin – 10cm² | SAME |
| Fan Speed Settings | 1-6 | 1-10 | 1-9 | Analysis(3) |
5 / 7
{9}
510(k) Summary
K261674
| Interface | RS232、TTL | TTL, PWM | USB 2.0, RS232 | Analysis(4) |
| --- | --- | --- | --- | --- |
| Therapy time, Treatment time | adjustable for users | 1:00 – 100:00 min – adjustable for users | 1:00 – 100:00 min – adjustable for users | Analysis(5) |
| Time Delay to Operate | Immediate from pressing START button when system turns to standby mode | Immediate from pressing START button when system turns to standby mode | Immediate from pressing START button when system turns to standby mode | SAME |
| Compatibility with the environment | R404 Ozone destruction level 0 | R404 Ozone destruction level 0 | R452a Ozone destruction level 0 | SAME with Predicate Device 1 |
| Display | 8.0 inch touch screen display | 5.0 inch touch screen display | 10.1 inch touch screen display | Analysis(6) |
| Net Weight without accessories | 87kg | 65kg | 60kg | Analysis(6) |
| Dimensions | H955 mm x W400 mm x D840mm | H810 mm x W400 mm x D845mm | H1060 mm x W500 mm x D560 mm | Analysis(6) |
#### Analysis:
Analysis(1)
The air outlets is similar with predicate device 1, the proposed device has 4 sizes of air outlet, while predicate device 1 has 3 sizes of air outlet, but as it has similar specifications and the same intended use, the difference has no impact on device safety and effectiveness.
Analysis(2)
The proposed device has similar power consumption with the predicate devices, with no significant difference, the difference will not affect the safety and effectiveness of the proposed device.
Analysis(3)
There is difference on the fan speed setting, but the proposed device has the same air speed range (200±50 to 620±60 liters/minute) with the predicate device 1, the speed setting difference will not affect the safety and effectiveness of the proposed device.
Analysis(4)
There is difference on the interface, the proposed device complies with electrical safety standard IEC 60601-1, the interface difference will not affect the device safety and effectiveness.
Analysis(5)
The proposed device and the predicate devices all can adjust the treatment time as needed, the proposed device has no time limit, and the difference in range limit of the time adjustment doesn't affect device safety and effectiveness.
6/7
{10}
510(k) Summary
K261674
# Analysis(6)
There are differences on the screen sizes, net weight and dimensions, and the different screen sizes, net weight and dimensions have no impact on the device safety and effectiveness.
# 9. Non-Clinical Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications and to support the Substantial Equivalence determination.
# Electrical safety and electromagnetic compatibility (EMC)
Testing was conducted to demonstrate that the proposed device complies with the following standards:
AAMI/ANSI ES60601-1:2005(R) 2012&A1:2012 and A2:2021, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014+A1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” The software documentation level for this device was determined to be “basic.”
# Specification Testing
Testing was conducted to validate that the proposed device meets design specifications for temperature range and air speed using each air outlet at each speed setting.
# 10. Clinical Testing
No clinical study is included in this submission.
# 11. Conclusion
The data obtained from the nonclinical tests support that the device is as safe, as effective, and performs as well as the legally marketed predicate devices.
7/7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.