K261636 · Canyon Medical, Inc. · NEY · Oct 28, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261636
Device Name
Microwave Ablation Antennas
Applicant
Canyon Medical, Inc.
Product Code
NEY · General, Plastic Surgery
Decision Date
Oct 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Microwave Ablation Antennas, in conjunction with the compatible Microwave Ablation Generator, is intended for coagulation (ablation) of soft tissue. This device is not intended for cardiac use.
Device Story
Microwave ablation antenna system; utilizes electromagnetic waves (300 MHz to 300 GHz) to induce frictional heating in polar tissue molecules; causes irreversible tissue necrosis. Device consists of antenna and cable; requires connection to compatible microwave ablation generator. Used in clinical settings by physicians; antenna placed into target tissue under imaging guidance (CT, US). Coagulated zones monitored via imaging. Benefits include targeted soft tissue destruction. Single-use; EO sterilized.
Clinical Evidence
No clinical data. Bench testing only. Performance and functionality evaluated against specifications using recognized consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-6, ISO 10993, ISO 11607, ASTM F 1980).
Technological Characteristics
Microwave ablation antenna; 13G-20G diameter; 70-1200mm shaft length; 3-31mm active tip. Operates at microwave frequencies (300 MHz-300 GHz); max power 150W. Pumped coolant system for antenna cooling. Biocompatible per ISO 10993. EO sterilized. Single-use. Complies with IEC 60601-1, 60601-1-2, 60601-2-6.
Indications for Use
Indicated for coagulation (ablation) of soft tissue in patients requiring thermal ablation therapy. Not indicated for cardiac use.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
October 28, 2026
Canyon Medical, Inc.
Huiqi Jiang
RA Manager
Bldg. 3, Phase 2 Accelerator, #11 Yaogu Ave., Jiangbei New Area, 210032
Nanjing, Jiangsu Province, People'S Republic Of China
Nanjing,
China
Re: K261636
Trade/Device Name: Microwave Ablation Antennas
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: NEY
Dated: September 29, 2026
Received: September 29, 2026
Dear Huiqi Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261636 - Huiqi Jiang
Page 2
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K261636 - Huiqi Jiang
Page 3
Office of Industry and Consumer Education (OICE) to ask a question about a specific regulatory topic. See the OICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-office-industry-and-consumer-education-oice) for more information or contact OICE by email (AskOICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K.
Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2026.10.28
17:56:13 -04'00'
Colin Kejing Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261636 | ? |
| Please provide the device trade name(s). | | ? |
| Microwave Ablation Antennas | | |
| Please provide your Indications for Use below. | | ? |
| Microwave Ablation Antennas, in conjunction with the compatible Microwave Ablation Generator, is intended for coagulation (ablation) of soft tissue. This device is not intended for cardiac use. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? |
{5}
CANYON
510(k) Summary - K261636
### 1. 510(k) Owner
Canyon Medical Inc.
### 2. Submission Correspondent
Submitter Name: Canyon Medical Inc.
Address: Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue, Jiangbei New Area, 210032, Nanjing, Jiangsu Province, People's Republic of China
Official Correspondent: Ms. Huiqi Jiang
Position: RA Manager
E-mail: jhq@canyonmedical.com.cn
Telephone: +86-13241196637
### 3. Date of Preparation: April 24th, 2026
### 4. Subject Device(s) Information
| Common Name: | Microwave Ablation Antennas |
| --- | --- |
| Trade Name: | Microwave Ablation Antennas |
| Regulation Number: | 21 CFR 878.4400 |
| Classification Name: | System, Ablation, Microwave and Accessories |
| Device Class: | II |
| Product Code: | NEY |
| Review Panel: | General & Plastic Surgery |
### 5. Predicate Device(s) Information
| Manufacturer: | Canyon Medical Inc. |
| --- | --- |
| 510(K) Number: | K241825 |
| Trade Name: | Microwave Ablation Antennas |
| Regulation Number: | 21 CFR 878.4400 |
| Classification Name: | System, Ablation, Microwave and Accessories |
| Device Class: | II |
| Product Code: | NEY |
| Review Panel: | General & Plastic Surgery |
### 6. Device Description
Microwave ablation is a kind of thermal ablation technology used in the treatment
2
{6}
CANON
510(k) Summary - K261636
of soft tissue, utilizing electromagnetic waves in the microwave energy spectrum (300 MHz to 300 GHz) to produce tissue-heating effects. The oscillation of polar molecules produces frictional heating, ultimately generating tissue necrosis.
Microwave Ablation Antennas generally consist of the Microwave Ablation Antennas, Microwave Ablation Cable. This device is single-use and sterilized by EO.
During ablation, Microwave Ablation Antennas will be punctured into the target tissue of the patients. In order to achieve the intended use, Microwave Ablation Antennas shall be connected with Microwave Ablation Generator which deliver microwave energy to Microwave Ablation Antennas through Microwave Ablation Cable (sterile or non-sterile cable). When reaching the target temperature, the soft tissue will be damaged leading to irreversible necrosis. Coagulated zones are monitored during treatment using appropriate imaging techniques.
### 7. Indications for Use
Microwave Ablation Antennas, in conjunction with the compatible Microwave Ablation Generator, is intended for coagulation (ablation) of soft tissue.
This device is not intended for cardiac use.
### 8. Technical Characteristics
The subject devices are substantially equivalent to the predicate devices in terms of intended use and technological characteristics. The differences between the subject devices and predicate devices do not affect the basic design principle, usage, effectiveness and safety of the subject device.
The technological characteristics of the subject devices are identical to those of predicate devices, with the exception of the following changes:
1) Add new models.
2) Change the shaft length of one existing model.
3) Change the material of bushing of part of models; and the bushing actually is used in other unchanged models cleared in K241825
Equivalence has been identified as follows:
3
{7}
CAN'ON
510(k) Summary - K261636
| Item | Subject Device-K261636 | Predicate Device-K241825 |
| --- | --- | --- |
| Device | Microwave Ablation Antennas | Microwave Ablation Antennas |
| Classification | Class II | Class II |
| Product Code | NEY | NEY |
| Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Indications for Use | Microwave Ablation Antennas, in conjunction with the compatible Microwave Ablation Generator, is intended for coagulation (ablation) of soft tissue. This device is not intended for cardiac use. | Microwave Ablation Antennas, in conjunction with the compatible Microwave Ablation Generator, is intended for coagulation (ablation) of soft tissue. This device is not intended for cardiac use. |
| Operating Principle | During the surgery, the Microwave Ablation Antennas is accurately placed in the tumor target area by imaging techniques (such as CT, US, etc.). The microwave energy generated by the microwave generator transmits to the Microwave Ablation Antennas through Microwave Ablation Cable. When reaching the target temperature, the soft tissue will be damaged leading to irreversible necrosis. Coagulated zones are monitored during treatment using appropriate imaging techniques. | During the surgery, the Microwave Ablation Antennas is accurately placed in the tumor target area by imaging techniques (such as CT, US, etc.). The microwave energy generated by the microwave generator transmits to the Microwave Ablation Antennas through Microwave Ablation Cable. When reaching the target temperature, the soft tissue will be damaged leading to irreversible necrosis. Coagulated zones are monitored during treatment using appropriate imaging techniques. |
| Shaft length(mm) | 70, 80, 100, 130, 150, 160, 180, 200, 250, 300, 1200mm | 70, 80, 100, 130, 150, 180, 200, 250, 300, 1200mm |
| Antenna outer diameter(G) | 13G, 14G, 15G, 16G, 17G, 18G, 19G, 20G | 13G, 14G, 15G, 16G, 17G, 18G, 19G, 20G |
| Antenna active tip(mm) | 3, 5, 11, 17, 25, 28, 31mm | 3, 5, 11, 25, 28mm |
| Max tolerable power | Up to 150W | Up to 150W |
| Sterilization method | EO gas | EO gas |
| Single Use | Yes | Yes |
| Device Temperature Monitoring | Temperature monitoring features used to ensure | Temperature monitoring features used to ensure |
4
{8}
CAN'ON
510(k) Summary - K261636
| Item | Subject Device | Predicate Device-K241825 |
| --- | --- | --- |
| Device | Microwave Ablation Antennas | Microwave Ablation Antennas |
| | safety | safety |
| Device Cooling | Pumped, coolant is used to cool the antenna. | Pumped, coolant is used to cool the antenna. |
| Patient-contacting information | Externally communicating device, in contact with the patient for a limited duration (no more than 24 hours) | Externally communicating device, in contact with the patient for a limited duration (no more than 24 hours) |
| Biocompatibility | Biocompatible according to ISO 10993 applicable parts | Biocompatible according to ISO 10993 applicable parts |
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- Similar indications for use
- Same material types that meet ISO 10993 biocompatibility requirements
- Same sterilization methods
- Same fundamental technology
There is no significant risk raised by the difference.
### 9. Non-Clinical Test
Non-clinical performance data is performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-6, ISO 10993 series standards, ISO 11607-1, ISO 11607-2, ASTM F 1980, and specific FDA guidance etc. Results from testing performed confirm the design requirements.
1) Electrical Safety as per IEC 60601-1
2) Electromagnetic Compatibility as per IEC 60601-1-2
3) Specific Performance as per IEC 60601-2-6
5
{9}
CAN'ON
510(k) Summary - K261636
4) Thermal Effects Test and Temperature Monitoring Test:
FDA Guidance: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery
5) Biocompatibility
The subject devices are externally communicating medical devices with tissue less than 24h which means limited duration. Relevant tests are conducted as per ISO 10993 series standards.
6) Package Validation
Primary package of the proposed device is intended to maintain the sterility of the product during its claimed shelf life. The integrity performance as per ISO 11607-1 and ISO 11607-2 is carried out.
7) Transport
Subject devices are processed by compression, first vibration, shock and second vibration, and then check. The result indicates that packaging is not damaged.
8) Shelf Life
Accelerated aging was used to simulate the storage of 3 years and the test results demonstrate that the aged samples complied with the pre-determined acceptance criteria.
9) EO/ECH Residual
Examination of the EO and ECH residuals have been conducted in accordance with ISO 10993-7 to evaluate whether the sterilized proposed device comply with the above selected allowable limits, and the results meet the requirements.
### 10. Clinical Test
No. Use of microwave ablation therapy is proven technology and is well accepted by the medical community. Performance test is sufficient to demonstrate safety and effectiveness of the subject devices with the predicate devices.
### 11. Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed subject devices are as safe, as effective, and performs as well as
6
{10}
CANYON
510(k) Summary - K261636
the legally marketed predicate devices and raises no new questions of safety or effectiveness.
The differences between both devices are insignificant in terms of safety and effectiveness.
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.