K261626 · Guangzhou Heygears IMC., Inc. · EBI · Aug 13, 2026 · Dental
Device Facts
Record ID
K261626
Device Name
UltraPrint Flexible Denture UV
Applicant
Guangzhou Heygears IMC., Inc.
Product Code
EBI · Dental
Decision Date
Aug 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
UltraPrint Flexible Denture UV is indicated for the fabrication of partial removable dentures and intended for continuous use not to exceed 30 days and until the permanent denture is fabricated and placed. It is intended exclusively for professional dental work.
Device Story
Light-cured methacrylate-based resin for CAD/CAM additive manufacturing of patient-specific temporary partial dentures; used by dental professionals (dentists/technicians). Workflow: dental CAD software designs framework; STL dataset generated; resin printed via validated 3D printer and UV post-curing (385 nm). Output: temporary partial denture. Benefits: provides flexible, biocompatible temporary prosthetic solution for patients awaiting permanent dentures.
Clinical Evidence
No clinical data. Evidence consists of bench performance testing and biocompatibility testing. Bench testing included material property testing per ISO 20795-1, manufacturing validation, recycle/re-use, service life, and shelf life. Biocompatibility testing included ISO 10993-1, -5 (cytotoxicity), -10 (sensitization), -11 (systemic toxicity), and -23 (irritation). All testing met acceptance criteria.
Technological Characteristics
Methacrylate-based photo-curable resin; 385 nm UV light polymerization; 3D printing additive manufacturing; four shades (Pink, Dark Pink, Opaque Pink, Transparent); non-sterile; ISO 20795-1 compliant; biocompatibility per ISO 10993 series.
Indications for Use
Indicated for fabrication of partial removable dentures for professional dental use; intended for continuous use not exceeding 30 days until permanent denture placement.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 13, 2026
Guangzhou Heygears IMC., Inc.
Mingyang Qu
Co-Founder
Block B2, 501,601, Enterprise Accelerator, Kaifa District
Guangzhou, Guangdong 510700
CHINA
Re: K261626
Trade/Device Name: UltraPrint Flexible Denture UV
Regulation Number: 21 CFR 872.3760
Regulation Name: Denture Relining, Repairing, Or Rebasing Resin
Regulatory Class: Class II
Product Code: EBI
Dated: May 15, 2026
Received: May 15, 2026
Dear Mingyang Qu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261626 - Mingyang Qu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261626 - Mingyang Qu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261626
Device Name
UltraPrint Flexible Denture UV
Indications for Use (Describe)
UltraPrint Flexible Denture UV is indicated for the fabrication of partial removable dentures and intended for continuous use not to exceed 30 days and until the permanent denture is fabricated and placed. It is intended exclusively for professional dental work.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# **K261626 - 510(k) Summary
GUANGZHOU HEYGEARS IMC.INC
UltraPrint Flexible Denture UV
8/10/2026**
# **ADMINISTRATIVE INFORMATION**
Manufacturer Name GUANGZHOU HEYGEARS IMC.INC
Block B2, 501, 601
Enterprise Accelerator, Kaifa District
Guangzhou, Guangdong, 510700, China
Telephone: +86 020-28186347
Official Contact Mingyang Qu, Co-Founder
# **DEVICE NAME AND CLASSIFICATION**
Proprietary/Device Name: UltraPrint Flexible Denture UV
Common Name: Resin, Denture, Relining, Repairing, Rebasing
Classification Name: Denture relining, repairing, or rebasing resin
Classification Regulation: 21 CFR 872.3760
Device Class: Class II
Product Code: EBI
Review Panel: Dental
Reviewing Branch: Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)
Dental Devices (DHT1B)
# **PREDICATE DEVICE INFORMATION**
The Subject device in this submission is substantially equivalent in indications, use and design principles to the following Predicate device.
| 510(k) | Predicate Device Name | Company Name |
| --- | --- | --- |
| K250617 | Apex Flex | SprintRay Inc. |
# **INDICATIONS FOR USE**
UltraPrint Flexible Denture UV is indicated for the fabrication of partial removable dentures and intended for continuous use not to exceed 30 days and until the permanent denture is fabricated and placed. It is intended exclusively for professional dental work.
# **DEVICE DESCRIPTION**
The Subject device is a light-cured methacrylate-based resin used by a dental professional (dentist or dental technician) for the CAD/CAM manufacturing of patient-specific temporary partial dentures. There are four models or “shades” of the Subject device which are referred to as Pink, Dark Pink, Opaque Pink and Transparent. The “Pink” models include additional pigments to provide a degree of pink color to the final printed denture. Methacrylates are known materials, commonly used in the dental industry for fixed and removable prosthetic devices due to their physical-chemical, mechanical and biocompatible properties.
Page 1 of 4
{5}
Commonly used dental CAD software is used by dental professionals to virtually design a flexible denture framework and generate an industry-standard “STL” 3D dataset which reflects the intended shape and contour. The Subject device resin is used within a validated additive manufacturing workflow to create the intended framework.
The Subject device is provided in HPDE bottles. Dental devices fabricated using the Subject device are one-time use, prescription-only devices. The Subject device is intended for continuous use not to exceed 30 days and until the permanent denture is fabricated and placed.
## PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included:
Biocompatibility testing according to:
- ISO 10993-1 Evaluation and testing within a risk management process
- ISO 10993-5 Cytotoxicity
- ISO 10993-10 Sensitization
- ISO 10993-11 Material Mediated Pyrogenicity/Systemic Toxicity
- ISO 10993-23 Irritation/Intracutaneous Reactivity
Bench Performance testing:
- Material Property Testing of each shade to ISO 20795-1
- Comparative Testing with the Predicate device to specific ISO 20795-1 endpoints
- Manufacturing Validation/Placement Orientation/Printer Compatibility
- Recycle/Re-use
- Service Life
- Shelf Life
No animal testing data is included in this submission.
Non-clinical performance testing of the Subject device met the acceptance criteria for each validation and test described above. The non-clinical performance testing performed demonstrates that the Subject device is suitable for intended use.
## EQUIVALENCE TO MARKETED DEVICE
The Subject device is substantially equivalent in Indications for Use, Technological Characteristics, and design principles to Predicate device listed above. The Summary tables at the end of this section compare the Subject and Predicate devices.
### Indications for Use
Predicate Device K250617, Apex Flex
The Subject device Indications for Use Statement (IFUS) is similar to that of the K250617 Predicate device. The device name differs, but that does not impact the indications or intended use of the devices. They both indicate the devices are intended for use in fabricating partial dentures. The wording is slightly different but expresses the same intent. The Subject device IFUS includes additional clarifying language regarding the duration of use, but that does not change the intended use of partial denture fabrication.
Page 2 of 4
{6}
# **Technological Characteristics**
*Predicate Device K250617, Apex Flex*
The Subject and Predicate devices are highly similar or the same in all Technological Characteristics. The intended use of the Subject and Predicate devices are the same. The Predicate device is leveraged for comparative ISO 20795-1 material performance testing endpoints to demonstrate substantial equivalence. The Predicate device is leveraged for the use of 3D printing technology in conjunction with denture base resin. The Subject device offers additional denture base shades relative to the Predicate device, but those differences are mitigated through performance testing on the additional Subject device shades. The Subject device was tested to biocompatibility endpoints related to the intended device use and duration some of which may differ than the Predicate device tested endpoints, but that does not change the intended use of the device or impact substantial equivalence.
# **Conclusion**
Overall, the Indications for Use statements of the Subject device and the K250617 Predicate are highly similar, differing in the device name and descriptive wording. The intended use is the same.
Overall, the Technological Characteristics of the Subject device are highly similar or the same as the Predicate device, including material performance to ISO 20795-1 testing endpoints.
Overall, the Subject device is substantially equivalent to the Predicate device.
Page 3 of 4
{7}
# Comparison - Indications for Use Statement
| Subject Device UltraPrint Flexible Denture UV HeyGears | Predicate Device Apex Flex SprintRay Inc. K250617 |
| --- | --- |
| UltraPrint Flexible Denture UV is indicated for the fabrication of partial removable dentures and intended for continuous use not to exceed 30 days and until the permanent denture is fabricated and placed. It is intended exclusively for professional dental work. | SprintRay Apex Flex is an alternative to traditional thermoplastic material used for the fabrication and repair of partial dentures. It is intended exclusively for professional dental work. |
# Comparison of Technological Characteristics
| Comparison | Subject Device UltraPrint Flexible Denture UV HeyGears | Predicate Device Apex Flex SprintRay Inc. K250617 |
| --- | --- | --- |
| Product Code | EBI | EBI |
| Regulation | 21 CFR 872.3760 | 21 CFR 872.3760 |
| Intended Use | Used in the fabrication of flexible partial removable dentures. | Used in the fabrication of flexible partial removable dentures. |
| Reason for Predicate/Reference | Not Applicable | Intended use, manufacturing process, material properties |
| Material Technology | 3D liquid (light-cured) print resin for dental CAD/CAM | 3D liquid (light-cured) print resin for dental CAD/CAM |
| Material | Methacrylate resin, photo initiator | Methacrylate resin, photo initiator |
| Shades | Pink, Dark Pink, Opaque Pink and Transparent | Light Pink, Standard Pink |
| Polymerization (Curing) Method | UV light, 385 nm w/post curing | UV light, 385 nm w/post curing |
| Prescription device | Yes | Yes |
| Manufacturing Equip | Validated 3D-Printer and post curing devices | Validated 3D-Printer and post curing devices |
| Performance Testing | ISO 20795-1 | ISO 20795-1 |
| Biocompatibility Testing | ISO 10993-1 ISO 10993-5 ISO 10993-10 ISO 10993-11 ISO 10993-23 | ISO 7405 ISO 10993-3 ISO 10993-5 ISO 10993-10 ISO 10993-11 ISO 10993-23 |
| Biocompatibility | Yes | Yes |
| Sterile | No | No |
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.