K261609 · Maxx Orthopedics, Inc. · HSX · Aug 4, 2026 · Orthopedic
Device Facts
Record ID
K261609
Device Name
Freedom Partial Knee System
Applicant
Maxx Orthopedics, Inc.
Product Code
HSX · Orthopedic
Decision Date
Aug 4, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Maxx Orthopedics Freedom® Partial Knee comprising the femoral component, tibial component and tibial insert is designed for a single compartment replacement of the natural knee joint. The Freedom Partial Knee is indicated for cemented use in partial/unicompartmental knee arthroplasty procedures. Partial replacement of the articulating surfaces of the knee is indicated only when only one compartment of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
Device Story
Modular, cemented, unconstrained partial knee prosthesis; replaces single compartment of tibiofemoral joint. Components: Cobalt-Chromium-Molybdenum femoral component; Titanium alloy tibial baseplate; UHMWPE tibial insert. Used in operating theater by qualified surgeons for partial/unicompartmental knee arthroplasty. Device resurfaces joint to restore mobility and alleviate pain. Subject device adds 8mm tibial insert thickness to existing 9-11mm range; provides intraoperative flexibility for joint line restoration and soft-tissue balance. Implanted via standard surgical technique; provides low-friction articulation. Benefits patient through bone-conserving design and restoration of joint function.
Clinical Evidence
Bench testing only. Wear simulation performed per ISO 14243-1 & 2 for 5 million cycles on a four-station knee simulator. Results showed wear rates for the 8mm insert were lower than two contemporary unicondylar knee replacement designs tested under identical conditions.
Technological Characteristics
Materials: Cobalt-Chromium-Molybdenum (femoral), Titanium alloy (tibial baseplate), UHMWPE (tibial insert). Fixation: Cemented. Sterilization: Ethylene oxide. Design: Modular, unconstrained, single-compartment prosthesis. Modification: Added 8mm tibial insert thickness to existing 9-11mm range. No changes to articulating surface geometry or contact mechanics.
Indications for Use
Indicated for patients requiring single-compartment knee replacement due to primary or post-traumatic degenerative disease, previous tibial condyle/plateau fractures, deformity, or revision of previous arthroplasty, where only one compartment of the joint is affected.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 4, 2026
Maxx Orthopedics, Inc.
Asawari Hare
Regulatory Associate
2460 General Armistead Ave. #100
Norristown, Pennsylvania 19403
Re: K261609
Trade/Device Name: Freedom Partial Knee System
Regulation Number: 21 CFR 888.3520
Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: HSX
Dated: June 3, 2026
Received: June 4, 2026
Dear Asawari Hare:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261609 - Asawari Hare
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K261609 - Asawari Hare
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LIXIN LIU -S
Lixin Liu, Ph.D
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261609
Device Name
Freedom Partial Knee System
Indications for Use (Describe)
The Maxx Orthopedics Freedom® Partial Knee comprising the femoral component, tibial component and tibial insert is designed for a single compartment replacement of the natural knee joint. The Freedom Partial Knee is indicated for cemented use in partial/unicompartmental knee arthroplasty procedures. Partial replacement of the articulating surfaces of the knee is indicated only when only one compartment of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
510(k) #: K261609
510(k) Summary
Prepared on: 2026-08-03
# Contact Details
21 CFR 807.92(a)(1)
Applicant Name
Maxx Orthopedics, Inc.
Applicant Address
2460 General Armistead Ave #100 Norristown PA 19403 United States
Applicant Contact Telephone
6177087388
Applicant Contact
Ms. Asawari Hare
Applicant Contact Email
asawari.hare@maxxortho.com
# Device Name
21 CFR 807.92(a)(2)
Device Trade Name
Freedom Partial Knee System
Common Name
Unicompartmental Knee Prosthesis
Classification Name
Knee joint femorotibial metal/polymer non-constrained cemented prosthesis
Regulation Number
888.3520
Product Code(s)
HSX
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K212870
Freedom Partial Knee
HSX
# Device Description Summary
21 CFR 807.92(a)(4)
# General Description
- The Freedom® Partial Knee System is a modular, cemented, unconstrained partial knee prosthesis intended to replace a single compartment of the deficient and damaged tibiofemoral joint surfaces using metal femoral and tibial components, and provide a low-friction articulation with a polyethylene bearing. The system consists of a Cobalt-Chromium-Molybdenum femoral component, an Ultra High Molecular Weight Polyethylene (UHMWPE) tibial insert, and a Titanium alloy tibial baseplate.
# Change Introduced
- This submission introduces one additional Partial Tibial Insert thickness (8mm), expanding the existing Partial Tibial Insert range from 9-11mm to 8-11mm. The subject 8mm Partial Tibial Insert has been developed as a line extension of the cleared Freedom® Partial Knee System (K212870) with the intention to provide a more bone conserving implant, expanding upon the already cleared Freedom® Partial Knee.
- The material, articulating surface geometry, and fixation method are equivalent to the cleared predicate device, Freedom® Partial Tibial Insert (K212870). There are no changes to articulation surfaces or condylar geometry.
# Principle of Operation
- The device functions in the same manner as the predicate, resurfacing the joint surfaces to restore mobility and alleviate pain associated with degenerative joint disease, and is intended for the same patient population and surgical technique.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Maxx Orthopedics Freedom® Partial Knee comprising the femoral component, tibial component and tibial insert is designed for a
{5}
single compartment replacement of the natural knee joint. The Freedom Partial Knee is indicated for cemented use in partial/unicompartmental knee arthroplasty procedures. Partial replacement of the articulating surfaces of the knee is indicated only when only one compartment of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Indications for use are the same.
## Technological Comparison
21 CFR 807.92(a)(6)
### Predicate Device
- Predicate: Freedom Partial Knee (K212870)
- Rationale: The subject device is substantially equivalent to the predicate in intended use, material, articulating surface geometry, and method of fixation. The only modification is the addition of one tibial insert thickness, 8mm.
### Discussion
- The subject 8mm Partial Tibial Insert is intended for use with the currently available, FDA Cleared, Freedom® Partial Knee System as part of partial knee replacement surgery. Partial knee replacement surgery is a procedure where worn, diseased, or damaged surfaces of one compartment of the knee joint are removed and replaced with artificial surfaces. The subject implant will be a prescription product consisting of single use only, implantable device for implanting into patients in an operating theatre by a qualified surgeon.
- The device is to be packaged and terminally sterilized by ethylene oxide, identical to the predicate device. The packaging is to keep devices sterilized for 10 years. The packaging will be able to be opened easily in an operating theatre without instruments and in a manner that allows the implant to remain sterile.
- To expand the existing Partial Tibial Insert portfolio, an 8mm thickness is being added for each size offering. The Anterior-Posterior (AP) and Medial-Lateral (ML) profiles are identical, and the overall size ranges are also identical.
- The expansion of available Partial Tibial Insert thicknesses does not alter clinical functionality, mechanical performance, or patient exposure to new materials. Therefore, the differences do not raise different questions of safety or effectiveness.
### Design and Specifications
- The subject device shares equivalent articulating surface geometry, material, tolerances, and manufacturing processes with the predicate device. The only modification is the introduction of one additional thickness (8mm), expanding the range from 9-11mm to 8-11mm with the intention to provide surgeons with additional intraoperative flexibility in restoring joint line and soft-tissue balance.
- The articulating surface, contact geometry, and fixation method are unchanged. The subject 8mm Partial Tibial Insert are fabricated from equivalent Ultra High Molecular Weight Polyethylene (UHMWPE) using equivalent machining, cleaning, packaging, and ethylene oxide sterilization processes as the predicate device.
### Sterilization and Biocompatibility
- Sterilization, packaging materials, and biocompatibility characteristics are equivalent to those previously cleared for the predicate device.
Accordingly, the added Partial Tibial Insert thickness is fully compatible with existing system components and meet the applicable performance expectations established for the cleared predicate.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
A wear simulation test was performed at the laboratories of the University of Nebraska Medical Center on the Freedom Partial Knee Replacement design utilizing the new 8mm tibial insert. The standard force-control wear testing method (ISO 14243-1 & 2) was used on a four-station Instron-Stanmore Knee Simulator set up to run as right knees. The test was run for 5 million cycles which loosely approximated to a patient walking for 5 years.
A force-controlled simulation and wear test was conducted to 5 million cycles on the Freedom Partial Knee Replacement components. Medial and lateral systems were paired together to facilitate testing similar to a full Total Knee Replacement. Compared to two other contemporary unicondylar knee replacement designs tested identically at the same laboratory, the wear rates of the Freedom Partial Knee Replacement design utilizing the 8mm tibial insert were lower than the wear rates of the other two systems.
Therefore, similarities in technological characteristics and results from non-clinical performance testing indicate that the Freedom Partial Knee System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.