Eye Recovery Pro (PH-e03)

K261602 · Shenzhen Goodwind Technology Development Co., Ltd. · OHS · Aug 10, 2026 · General, Plastic Surgery

Device Facts

Record IDK261602
Device NameEye Recovery Pro (PH-e03)
ApplicantShenzhen Goodwind Technology Development Co., Ltd.
Product CodeOHS · General, Plastic Surgery
Decision DateAug 10, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Eye Recovery Pro (PH-e03) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region.

Device Story

Eye Recovery Pro (PH-e03) is a wearable, home-use LED phototherapy mask; treats periorbital wrinkles. Device utilizes 80 dual-core LEDs emitting amber (605nm), red (630nm, 660nm), and infrared (830nm) light. User wears mask for 3-minute treatment sessions. Powered by internal 3.7V lithium battery; rechargeable via USB-C. Includes detachable protective eyewear and adjustable head straps. Device operates via simple user-controlled interface; no complex software analysis or AI/ML algorithms involved. Intended for patient self-use in home environment. Benefits include non-invasive wrinkle reduction.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), battery safety (IEC 62133-2), and photobiological safety (IEC 62471).

Technological Characteristics

Wearable LED mask; 80 dual-core LEDs; wavelengths 605nm, 630nm, 660nm, 830nm; irradiance 65 mW/cm². Powered by 3.7V 500mAh lithium battery. Biocompatibility per ISO 10993-5, 10993-10, 10993-23. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, and IEC 62471.

Indications for Use

Indicated for adults (22 years and older) for the treatment of wrinkles in the periorbital region.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA U.S. FOOD & DRUG** ADMINISTRATION August 10, 2026 Shenzhen Goodwind Technology Development Co., Ltd. Ziming Yang Head of firm Rm. 801, Bldg. 1, Baolong Intelligent Manufacturing Park, # 34, Baotong Rd., Baolong Community, Longgang District Shenzhen, Guangdong 518116 China Re: K261602 Trade/Device Name: Eye Recovery Pro (PH-e03) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: May 14, 2026 Received: May 14, 2026 Dear Ziming Yang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K261602 - Ziming Yang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K261602 - Ziming Yang Page 3 See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA Digitally signed by TANISHA L. HITHE -S L. HITHE -S Date: 2026.08.10 16:21:52 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261602 | ? | | Please provide the device trade name(s). | | ? | | Eye Recovery Pro (PH-e03) | | | | Please provide your Indications for Use below. | | ? | | The Eye Recovery Pro (PH-e03) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region. | | | | Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? | {4} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary--- # 510(k) Summary - K261602 **'510(k) Summary" as required by 21 CFR Part 807.92.** ## I. Submitter Shenzhen Goodwind Technology Development CO., LTD Address: Room 801, Building 1, Baolong Intelligent Manufacturing Park, No. 34, Baotong Road, Baolong community, Longgang District, Shenzhen, Guangdong, China Contact person: Ziming Yang Title: General Manager Title: Head of firm Tel: +86 13802283843 E-mail: admin@goodwind.com.cn Shenzhen Goodwind Technology Development CO., LTD Date: 2026-05-14 ## II. Subject Device Name of Device: Eye Recovery Pro Model(s): PH-e03 Common or Usual Name: Light based over the counter wrinkle reduction Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHS Regulation Number: 21 CFR 878.4810 ## III. Predicate Device 510(k) number: K233556 Trade name: LED Eye Mask, model: EY-20R, A20, EY-20N Product Codes: OHS Classification: Class II Regulation Number: 21 CFR 878.4810 Regulation Name: Light Based Over The Counter Wrinkle Reduction (OHS) Manufacturer: Shenzhen Kaiyan Medical Equipment Co., Ltd ## IV. Device Description Eye Recovery Pro (PH-e03) is a home use wearable LED phototherapy device which can help reduce wrinkles within the periorbital region. Eye Recovery Pro is consisting of main unit (mask), Protective Eyewear (Detachable), Adjustable Head Straps, USB-C to USB-A Charging Cable, Storage Bag. There are 4 kinds of light, which include Amber light(wavelength 605nm), Red light (wavelength 630nm, 660nm) and Infrared light (wavelength 830nm). ## V. Indications for Use Page 1 of 7 {5} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary--- The Eye Recovery Pro (PH-e03) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region. ## VI. Comparison of Technological Characteristics With the Predicate Device The Eye Recovery Pro has the same intended use as the predicate device. The technological characteristics, features, specifications, materials are similar to the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the Eye Recovery Pro (PH-e03) may be found substantially equivalent to its predicate device. Page 2 of 7 {6} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary--- | Item | Subject device | Predicate device | Remark | | --- | --- | --- | --- | | 510 (k) number | K261602 | K233556 | / | | Trade name | Eye Recovery Pro (Model: PH-e03) | LED Eye Mask, model: EY-20R, A20, EY-20N | / | | Manufacturer | Shenzhen Goodwind Technology Development CO., LTD | Shenzhen Kaiyan Medical Equipment Co., Ltd | / | | Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same | | Regulation name | Light Based Over The Counter Wrinkle Reduction (OHS); | Light Based Over The Counter Wrinkle Reduction (OHS); | Same | | Product code | OHS | OHS | Same | | Class | Class II | Class II | Same | | Indications for use/ Intended use | The Eye Recovery Pro (PH-e03) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region. | The LED Eye mask (Model: EY-20R, 20A, EY-20N) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region. | Same | | Location for use | Periorbital region | Periorbital region | Same | | OTC or prescription | OTC | OTC | Same | | Power supply | DC 5V, 1A Lithium battery 3.7V, 500 mAh | Main unit: 3.7V, 300mAh lithium battery, 1.11Wh Adapter Input: 100 - 240Va.c., 50/60Hz Adapter Output: 5Vd.c, 2A | **Note 1** | Page 3 of 7 {7} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary | Item | Subject device | Predicate device | Remark | | --- | --- | --- | --- | | Light source | Light emitting diodes | LED | Same | | LED array | 80 (40 dual-core) | 605nm: 20 633nm: 20 660nm: 20 830nm: 20 Total: 40 (Double lamp beads) | Note 2 | | Wavelength | 605nm, 630nm, 660nm, 830nm | 605nm, 630nm, 660nm, 830nm | Same | | Irradiance(mW/cm²) | 65 | 605nm: 26.2±3 633nm: 13.8±3 660nm: 20.3±3 830nm: 8.5±3 Total: 65 | Same | | Treatment time | 3 minutes | 3 minutes | Same | | Dimensions (mm) | 300 x 76 x 9.5mm | Not publicly available | Note 3 | | Weight | Mask: 110g (not including straps) | Not publicly available | Note 3 | | Compliance with voluntary standards | IEC 60601-1; IEC 60601-1-2; IEC 60601-1-11; IEC 60601-2-83; IEC 62471 | IEC 60601-1; IEC 60601-1-2; IEC 60601-1-11; IEC 60601-2-57; IEC 62471 | Same | Page 4 of 7 {8} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary | Item | Subject device | Predicate device | Remark | | --- | --- | --- | --- | | Biocompatibility feature | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 | Same | # **Note 1:** The power supply for the subject device is a little different from the predicate device, however the lithium battery of the subject device has passed IEC 62133-2 test, so this difference should not raise any safety/effectiveness problems. # **Note 2:** Though the LED array is different from the predicate device, the subject device adopts the dual core beads, actually the number of each type of light used in the subject product and the predicate device is the same, also the subject device has passed the IEC 62471 test, so this difference should not raise any safety and effectiveness problems. # **Note 3:** Though the dimension and weight are different from the predicate device, this difference is insignificant and do not raise any safety or effectiveness problems. Page 5 of 7 {9} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary ### VII. Performance Data The following performance data were provided in support of the substantial equivalence determination. #### 1) Biocompatibility Testing The biocompatibility evaluation for the body-contacting components of the Eye Recovery Pro was conducted in accordance with the “Use of International Standard ISO 10993-1, ‘Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020”, as recommended by FDA. The following testing was performed to, and passed, including: - ISO 10993-5: 2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity - ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization - ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation #### 2) Electrical Safety - IEC 60601-1:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - IEC TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems - IEC 60601-1-11:2020 Medical Electrical Equipment –Part 1-11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment - IEC 62133-2:2021, Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems #### 3) Eye Safety - IEC 62471:2006 Photobiological safety of lamps and lamp systems #### 4) Software Verification and Validation Software verification and validation testing was conducted and basic documentation provided as recommended by FDA’s Guidance for Industry and FDA Staff “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” ### VIII. Conclusions Page 6 of 7 {10} Shenzhen Goodwind Technology Development CO., LTD 510(K) Summary Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device. Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...