K261571 · Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. · IOR · Aug 6, 2026 · Physical Medicine
Device Facts
Record ID
K261571
Device Name
Manual Wheelchair (H008, PCK7)
Applicant
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Aug 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The Manual Wheelchair is to provide mobility to people limited to sitting position.
Device Story
Manual, mechanical wheelchair; provides mobility for users limited to sitting position. Propelled by occupant via hand rims or pushed by caregiver via rear handgrips. Features foldable steel frame, removable armrests, elevating legrests, and manual wheel locks. Used in indoor/outdoor environments. Provides mobility assistance; enhances patient independence. No electronic or software components.
Clinical Evidence
No clinical data. Bench testing only, including performance testing per ISO 7176-1, -3, -5, -7, -8, -11, -13, -15, -22, and ISO 16840-10. Biocompatibility testing performed per ISO 10993-5, -10, and -23.
Technological Characteristics
Mechanical, manually operated wheelchair. Materials: Steel frame, armrest, wheel lock; PVC leather upholstery. Foldable X-brace design. Features: removable armrests, elevating legrests, manual push-to-lock brakes. Complies with ISO 7176 series for stability, strength, and dimensions. Biocompatible per ISO 10993.
Indications for Use
Indicated for adults (22 years and older) with mobility limitations requiring a sitting position.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 6, 2026
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Fang (Fawn) Zhang
Director of Regulatory and Compliance
Yunyang Industrial Park Danyang City
#1 Baisheng Rd. Development Zone
Danyang Jiangsu, / 212300
China
Re: K261571
Trade/Device Name: Manual Wheelchair (H008, PCK7)
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: May 11, 2026
Received: May 12, 2026
Dear Fang (Fawn) Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261571 - Fang (Fawn) Zhang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K261571 - Fang (Fawn) Zhang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation
and Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261571 | ? |
| Please provide the device trade name(s). | | ? |
| Manual Wheelchair (H008, PCK7) | | |
| Please provide your Indications for Use below. | | ? |
| The Manual Wheelchair is to provide mobility to people limited to sitting position. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
510(k) Summary - K261571
# 510(k) Summary - K261571
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Submitter's Name: Jiangsu Yuyue Medical Equipment & Supply Co., LTD.
Address: Yunyang Industrial Park Danyang City, Jiangsu 212300 China
Contact Person: Fang (Fawn) Zhang
Contact Title: JD. PharmD, RAC Director of Regulatory and Compliance
Contact E-mail Address: Fawn.zhang@yuyue.com.cn
Zip code: 212300
Telephone: (206) 639-1311
Date of Preparation: Jan.7th, 2025
## 2.0 Subject Device Information
Common Name: Manual Wheelchair
Proprietary Name: Manual Wheelchair
Model:H008, PCK7
Regulation Name: Wheelchair, Mechanical
Product Code: IOR
Regulation Number:890.3850
Regulation Class:1
## 3.0 Predicate Device Information
Manufacturer: Zhenjiang Assure Medical Equipment Co., Ltd.
Trade/Device: Reclining wheelchair (YJ-011S 16"D, YJ-011S 16"DF, YJ-011S 18"D, YJ-011S 18"DF, YJ-011S 20"D, YJ-011S 20"DF)
510(k) number: K232198
## 4.0 Indication for Use Statement
The Manual Wheelchair is to provide mobility to people limited to sitting position.
## 5.0 Subject Device Description
The subject device, mechanical wheelchair, is a manually operated device with wheels that is intended for medical purposes to provide mobility to people limited to
1 of page 6
{6}
510(k) Summary - K261571
sitting position.
The mechanical wheelchair is a chair with wheels, designed to be a replacement for walking, where it is propelled by the seated occupant turning the rear wheels by hand. There are also handgrips behind the seat for someone else to do the pushing.
The components include a frame, a backrest, a seat cushion, two handgrips, two removeable armrests, two armrest pads, two side panels, two rear wheels, two hand rims, two wheel locks, two front wheels with forks, and footrests with footplates.
The subject device has two models: Model H008 and Model PCK7
- H008: two 8 inch front wheels and two 24 inch rear wheels, a steel frame and a PVC upholstery that is flame resistant. Armrest is removeable. It has elevating legrest.
Main materials: Steel (frame, armrest, wheel lock), PVC leather (armrest pad, backrest, seat cushion).
Its maximum loading can not be over 150kgs (around 330lbs).
- PCK7: two 8 inch front wheels and two 24 inch rear wheels, a steel frame and a PVC upholstery that is flame resistant. Armrest is removeable. It has elevating legrest.
Main materials: Steel (frame, armrest, wheel lock), PVC leather (armrest pad, backrest, seat cushion).
Its maximum loading can not be over 225kgs(around 496lbs).
### 6.0 Technological Characteristic Comparison Table
Table 1 - General Comparison
| Device | Subject Device | Predicate Device | Results |
| --- | --- | --- | --- |
| 510(k) Number | K261571 | K232198 | -- |
| Manufacturer | Jiangsu Yuyue Medical Equipment & Supply CO., LTD. | Zhenjiang Assure Medical Equipment Co., Ltd. | -- |
| Proprietary Name | Manual wheelchair | Reclining wheelchair | Same |
| Model | H008, PCK7 | YJ-011S 16"D, YJ-011S 16"DF, YJ-011S 18"D, YJ-011S 18"DF, YJ-011S 20"D, YJ-011S 20"DF | |
| Classification | I | I | Same |
| Indications for use | Manual Wheelchair is to provide mobility to people limited to a sitting position. | The YJ-011S reclining wheelchair is to provide mobility to persons limited to a sitting position. | Same |
2 of page 6
{7}
510(k) Summary - K261571
| Design Characteristic | Manual Operation, Four-Wheels, Foldable, X-Brace, Push-to-Lock, Armrest, Backrest, Footrest, skirt guard | Manual Operation, Four-Wheels, Foldable, Cross-Brace, Pull-to-Lock, Armrest, Backrest, Foot rest, skirt guard | Similar |
| --- | --- | --- | --- |
| Brake control | Occupant-operated brake only | Occupant-operated brake only | Same |
| Operation Environment | For indoor/outdoor use | For indoor/outdoor use | Same |
| Control Mode | Mechanical | Mechanical | Same |
| Size(unfold) (mm) | H008: 1270×685×1280mm PCK7: 1320×770×1320mm (L*W*H) | 1340 (L)* 1258 (H) *16" 625mm(W) *18" 675mm(W) *20" 725mm(W) | NOTE1 |
| Stowage length/width/height (mm) | H008:1270×290×1280(L*W*H) PCK7:1320×400×1320(L*W*H) | 1340 (L)* 1258 (H)*323(W) | NOTE2 |
| Weight (Total) | H008:24kg PCK7:30kg | 25.5kg | NOTE3 |
| Weight Capacity | H008:150kg (around 330lbs) PCK7:225kg (around 496lbs) | 350lbs | NOTE4 |
| Seat Width | H008:500mm PCK7:610mm | 410-510mm | NOTE5 |
| Seat height | H008:500mm PCK7:480mm | 525mm | NOTE6 |
| Seat depth | H008:385mm PCK7:510mm | 460mm | NOTE7 |
| Back type | Adjustable | Adjustable | Same |
| Tires | Front: 193mm Rear:610mm | Front: 195mm Rear:613mm | NOTE8 |
| Armrest | Removeable | fixed | NOTE9 |
| Footrest | elevating legrest | Optional/ swing away Optional/ swing away | NOTE10 |
| Rear Axle Position | Single | Single | Same |
| Frame Construction | Foldable frame Push inward from left and right | Foldable frame Push inward from left and | Same |
3 of page 6
{8}
510(k) Summary - K261571
| | sides to fold | right sides to fold | |
| --- | --- | --- | --- |
| Safety Feature | Manual Wheel Lock | Manual Wheel Lock | Same |
| Performance | Comply with: ISO7176-1 ISO7176-3 ISO7176-5 ISO7176-7 ISO7176-8 ISO7176-11 ISO7176-13 ISO7176-15 ISO7176-22 ISO16840-10 | Comply with: ISO7176-1 ISO7176-3 ISO7176-5 ISO7176-7 ISO7176-8 ISO7176-11 ISO7176-13 ISO7176-15 ISO7176-16 | Same |
| Biocompatibility | Comply with: ISO10993-5 ISO10993-10 ISO10993-23 | Comply with: ISO10993-5 ISO10993-10 ISO10993-23 | Same |
#### Summary of Substantial Equivalence Discussion:
NOTE1、2、3、5、6、7、8: Compared to the predicate device, the subject device has different value on the unfold size, stowage size, device weight, seat height, seat depth, and tire size. However, the subject devices have passed the ISO 7176-7-1998 and ISO 7176-5-2008, so the above different will not raise any new risk of safety or effectiveness.
NOTE 4: The Maximum loading is different with the predicate device. The subject manual wheelchair has already been tested for performance according to ISO 7176-1/-3/-8/-11/-13/-15 and complies with the standards. Therefore, the subject manual wheelchair is as safe, effective and performs as well as the legally marketed predicate device.
NOTE9: Compare the predicate device, the subject's armrest is Removeable. However, the subject device passed the test of ISO 7176-1/-8, so this difference will not raise any new risk of safety or effectiveness.
NOTE10: Compare the predicate device, the subject's foot rest is elevating legrest. However, the subject device passed the test of ISO 7176-1/-8, so this difference will not raise any new risk of safety or effectiveness.
Similar: Compare the predicate device, the subject's Wheel Lock is a push lock. However, the subject device passed the test of ISO 7176-3, so this difference will not raise any new risk of safety or effectiveness.
Despite of the above slight differences, the two devices all completed the
4 of page 6
{9}
510(k) Summary - K261571
performance tests in accordance with ISO 7176 series standards. There are no safety and effectiveness aspects concerned.
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The slight differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
### 7.0 Summary of Non-Clinical Testing
Non clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
ISO 7176-5:2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
ISO 7176-11: 2012 Wheelchairs - Part 11: Test dummies
ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method
ISO 7176-22:2014 Wheelchairs - Part 22: Set-up procedures
### Biocompatibility of User-contacting Parts
User-contacting material are carried out biocompatibility assessment in accordance with ISO 10993-1: 2018, including:
Cytotoxicity per ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
Skin Sensitization per ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization.
Irritation per ISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for Irritation.
### 8.0 Summary of Clinical Testing
No clinical study implemented for the wheelchair.
5 of page 6
{10}
510(k) Summary - K261571
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is as safe and effective as the legally marketed predicated device in K232198 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
6 of page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.