The BOBBY Balloon Guide Catheter is intended for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
Device Story
Co-axial, braid-reinforced, variable stiffness catheter; external hydrophilic coating; compliant balloon at distal end; bifurcated luer hub at proximal end. Operates by insertion into vasculature; balloon inflated/deflated via syringe with contrast media through co-axial outer lumen; inner lumen allows guidewire/catheter passage. Used in clinical settings (e.g., neurointerventional suites) by physicians. Radiopaque markers facilitate fluoroscopic visualization for positioning. Balloon provides temporary vascular occlusion; catheter acts as conduit for retrieval devices. Benefits include facilitated access to target vessels and controlled occlusion during procedures.
Clinical Evidence
No clinical data or animal testing was required. Substantial equivalence supported by bench testing, including simulated use, balloon burst/fatigue/compliance, kink resistance, torque strength, and leakage testing. All results met acceptance criteria.
Indicated for patients requiring intravascular catheter insertion and guidance in peripheral and neurovascular systems, temporary vascular occlusion during angiographic procedures, and as a conduit for retrieval devices.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 9, 2026
MicroVention, Inc. d/b/a Terumo Neuro
Miranda Beach
Senior Regulatory Affairs Specialist
35 Enterprise
Aliso Viejo, California 92656
Re: K261549
Trade/Device Name: BOBBY Balloon Guide Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP, DQY
Dated: May 8, 2026
Received: May 11, 2026
Dear Miranda Beach:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261549 - Miranda Beach
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261549 - Miranda Beach
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261549
Device Name
BOBBY Balloon Guide Catheter
Indications for Use (Describe)
The BOBBY Balloon Guide Catheter is intended for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
**TERUMO**
NEURO
## 510(k) Summary
**K261549**
### 510(k) Owner Information:
510(k) Owner: MicroVention, Inc. d/b/a Terumo Neuro
Address: 35 Enterprise, Aliso Viejo, California 92656, USA
Phone Number: (714)-247-8000
Contact Person: Miranda Beach
Phone number: (949)-989-0580
**Date Prepared:** June 8, 2026
**Device Trade Name:** BOBBY™ Balloon Guide Catheter
**Common Name:** Balloon Guide Catheter
**Device Class:** Class II
**Classification Name:** Percutaneous Catheter
**Product Codes:** QJP, DQY
**CFR Classification:** 21 CFR 870.1250
**Predicate Device:** BOBBY™ Balloon Guide Catheter (K193607)
### Indications for Use:
The BOBBY Balloon Guide Catheter is intended for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
### Device Description:
The BOBBY™ Balloon Guide Catheter is a co-axial, braid-reinforced, variable stiffness catheter with an external hydrophilic coating. The BOBBY Balloon Guide Catheter incorporates a compliant balloon, radiopaque markers, and a bifurcated luer hub on the proximal end. Both inflation lumen and guidewire lumen are constructed with a PTFE inner liner.
The BOBBY Balloon Guide Catheter has an inner lumen through which a guidewire and catheter can be inserted, and a co-axial outer lumen that is used to inflate and deflate the balloon with a syringe filled with contrast media. A bifurcated luer hub is attached to the proximal end of the balloon guide catheter to provide access to both the inner and outer lumens. In addition, a
35 Enterprise, Aliso Viejo, CA 92656, USA
terumoneuro.com
Page 1 of 4
{5}
hydrophilic coating is applied to the distal end of the balloon guide catheter to provide a lubricious outer surface for catheter advancement in the vasculature. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. The balloon incorporates a distal air-purging system to purge air from the inflation lumen prior to use. The balloon catheter also incorporates radiopaque markers to facilitate fluoroscopic visualization and indication of the balloon position.
### **Comparison of Technological Characteristics:**
The subject and predicate devices have the same intended use and fundamental technological characteristics. The device changes do not alter device dimensions, materials composition, or operating principle.
| Feature | Predicate Device | Subject device |
| --- | --- | --- |
| | BOBBY Balloon Guide Catheter K193607 | BOBBY Balloon Guide Catheter K261549 |
| FDA Classification | Class II | Same |
| Product Code(s) | DQY, QJP | Same |
| Regulation Number | 21 CFR 870.1250 | Same |
| Regulation Name | Percutaneous Catheter | Same |
| Indications for Use | The BOBBY Balloon Guide Catheter is intended for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices. | Same |
| Anatomical Locations | Peripheral and neuro vasculature | Same |
| Materials | Commonly used medical grade plastics (e.g. nylon, PTFE, polyethylene, polyolefin) and stainless steel | Same |
| Reinforced Catheter Shaft | Stainless steel braid and coil reinforced shaft | Same |
| Injection Port | Yes | Same |
| Radiopacity | Yes, shaft is visible due to distal tip Pt-Ir marker bands | Same |
| Radiopaque Marker Bands | 2 | Same |
| Compliant Balloon | Yes, polyurethane | Same |
| Labeled Shaft Outer Diameter | 0.110 in (2.8 mm) 8Fr | Same |
| Labeled Shaft Inner Diameter | 0.086 in (2.18 mm) 6.5Fr | Same |
35 Enterprise, Aliso Viejo, CA 92656, USA
terumoneuro.com
Page 2 of 4
{6}
TERUMO
NEURO
| Feature | Predicate Device | Subject device |
| --- | --- | --- |
| | BOBBY Balloon Guide Catheter K193607 | BOBBY Balloon Guide Catheter K261549 |
| Effective Length | 95 cm (37.4 in) | Same |
| Tip Shape | Straight | Same |
| Coating | Hydrophilic Coating – distal portion of shaft | Same |
| Internal Construction | Coaxial lumen | Same |
| Accessories Supplied | Peel Away Sheath | Same |
| How Supplied | Sterile, single use | Same |
| Sterilization Method | EtO | Same |
| Sterility Assurance Level | 10^{-6} | Same |
### **Summary of Bench Testing:**
Bench testing was conducted to demonstrate the subject device performs as intended and support substantial equivalence to the predicate device. All testing met previously established acceptance criteria or results were comparable to the predicate device. Testing included the following:
- Simulated Use
- Compatibility with Devices and Agents
- Balloon Rated Burst
- Balloon Inflation / Deflation Time
- Balloon Fatigue
- Balloon Decay
- Atraumatic Tip and Flexible Distal Tip
- Lumen Collapse During Aspiration
- Kink Resistance
- Force at Break
- Freedom from Leakage – Liquid
- Freedom from Leakage – Air
- Dynamic Burst Pressure
- Burst Pressure of Catheter
- Flexural Fatigue
- Balloon Compliance
- Catheter Particle Testing
- Torque Strength Testing
35 Enterprise, Aliso Viejo, CA 92656, USA
terumoneuro.com
Page 3 of 4
{7}
**TERUMO**
NEURO
### Animal Testing:
An animal study was not deemed necessary to support the substantial equivalence of the subject device to the predicate device.
### Clinical Study:
A clinical study was not deemed necessary to support the substantial equivalence of the subject device.
### Sterilization and Shelf Life:
The subject device, as packaged with included accessories, is sterilized using a validated 100% Ethylene Oxide sterilization process to ensure sterility assurance level (SAL) of $10^{-6}$ in accordance with ISO 11135. The modifications described in this submission do not impact sterilization as compared to the predicate device.
The subject device’s shelf-life and packaging configuration remain identical to that of the predicate device. Therefore, no additional shelf life or packaging validation testing was required to demonstrate substantial equivalence between the subject and predicate device.
### Biocompatibility:
The biological safety of the BOBBY Balloon Guide Catheter was established in accordance with the ISO 10993-1, “Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a Risk Management Process.” The balloon guide catheter is classified as an externally communicating device, circulating blood, limited contact ($\leq 24$ hrs.).
The subject device does not introduce new materials or alter any patient-contacting surfaces. Materials of construction, sterilization method, sterility assurance level, and packaging configuration remain unchanged from the predicate device (K193607). Accordingly, no additional biocompatibility testing was conducted, and the previously established biocompatibility data remains applicable to the subject device.
### Conclusion:
The subject BOBBY Balloon Guide Catheter has the same intended use and similar technological characteristics as the predicate device BOBBY Balloon Guide Catheter (K193607). The differences between the subject device and predicate device do not raise new questions of safety and effectiveness. The nonclinical performance testing demonstrates that the subject device performs as intended and is substantially equivalent to the predicate device.
35 Enterprise, Aliso Viejo, CA 92656, USA
terumoneuro.com
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.