Calcium Hydroxide Temporary Filling Material (KP-Cal)
K261514 · Guilin Kevin Peter Technology Co., Ltd. · KIF · Aug 25, 2026 · Dental
Device Facts
Record ID
K261514
Device Name
Calcium Hydroxide Temporary Filling Material (KP-Cal)
Applicant
Guilin Kevin Peter Technology Co., Ltd.
Product Code
KIF · Dental
Decision Date
Aug 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Calcium Hydroxide Temporary Filling Material is suitable for several indications including : -Temporary root canal filling; -Vital pulpotomy -Temporary pulp capping.
Device Story
KP-Cal is a premixed, hydrophilic calcium hydroxide paste for temporary root canal dressings; supplied in pre-loaded syringes for direct application by dental professionals. Material does not solidify in the root canal, facilitating thorough flushing/removal post-treatment. Radiopaque for clinical confirmation of placement. Used in dental clinics for temporary root canal filling, vital pulpotomy, and temporary pulp capping. Benefits include ease of use, biocompatibility, and effective temporary disinfection/treatment of root canals.
Clinical Evidence
No clinical data. Evidence consists of bench testing (appearance, pH, calcium hydroxide content, loading volume, flowability, radiopacity, film thickness per ISO 6876) and biocompatibility testing (ISO 10993-1, -3, -5, -6, -10, -11, -23 and ISO 7405).
Technological Characteristics
Composition: Calcium hydroxide, zirconium dioxide, propylene glycol, polyethylene glycol. Form: Paste in pre-loaded syringe. Non-sterile. Shelf life: 2 years. Physical properties: pH ~12.32, radiopacity 4.5, flowability 19.9. Standards: ISO 6876 (performance), ISO 10993/7405 (biocompatibility).
Indications for Use
Indicated for temporary root canal filling, vital pulpotomy, and temporary pulp capping in patients requiring endodontic treatment.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 25, 2026
Guilin Kevin Peter Technology Co., Ltd.
Libin Zhang
Regulatory Affairs Manager
North Side Of Linsu Rd., Lingui Town, Lingui Dist.
Guilin, Guangxi 541100
CHINA
Re: K261514
Trade/Device Name: Calcium Hydroxide Temporary Filling Material (KP-Cal)
Regulation Number: 21 CFR 872.3820
Regulation Name: Root canal filling resin
Regulatory Class: Class II
Product Code: KIF, EJK
Dated: July 21, 2026
Received: July 21, 2026
Dear Libin Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K261514 - Libin Zhang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K261514 - Libin Zhang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MICHAEL E. ADJODHA -S**
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261415 | ? |
| Please provide the device trade name(s). | | ? |
| Calcium Hydroxide Temporary Filling Material (KP-Cal) | | . |
| Please provide your Indications for Use below. | | ? |
| Calcium Hydroxide Temporary Filling Material is suitable for several indications including : -Temporary root canal filling; -Vital pulpotomy -Temporary pulp capping. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
[LOGO]
桂林凯文彼德科技有限公司
Guilin Kevin Peter Technology Co., Ltd.
# 510(k) Summary-K261514
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Guilin Kevin Peter Technology Co., Ltd.
Address: North Side of Linsu Rd, Lingui Town, Lingui Dist., Guilin, Guangxi, 541100,P. R. China
Telephone: +86-773-2352685
Contact person: Zhang Libin
Contact title: Regulatory Affairs Manager
E-mail: zhanglb@kvpeter.com
Date of Preparation: August 19, 2026
## 2.0 Device Information
Trade/Device Name: Calcium Hydroxide Temporary Filling Material
Common Name: Resin, root canal filling
Model: KP-Cal
Panel: Dental
Regulation Number: 21 CFR 872.3820
Device Class: Class II
Product Code: KIF
## 3.0 Identification of Predicate Device
Primary Predicate Device
510(k) Number: K210904
Device Name: Metapaste Plus
Common Name: Root Canal Filling Resin
Panel: Dental
Regulation Number: 21 CFR 872.3820
Device Class: Class II
Product Code: KIF
Manufacturer: Meta Biomed Co.,Ltd
1 / 6
{6}
fp
桂林凯文彼德科技有限公司
Guilin Kevin Peter Technology Co., Ltd.
Reference Predicate Device:
510(k) Number: K060365
Device Name: Calcium hydroxide paste
Common Name: Calcium hydroxide cavity liner
Model: ApexCal
Panel: Dental
Regulation Number: 21 CFR 872.3250
Device Class: Class II
Product Code: EJK
Manufacturer: Ivoclar Vivadent, Incorporated
### 4.0 Indication for Use Statement
Calcium Hydroxide Temporary Filling Material is suitable for several indications including:
-Temporary root canal filling;
-Vital pulpotomy
-Temporary pulp capping.
### 5.0 Device Description
Calcium Hydroxide Temporary Filling Material is a premixed, immediately applicable, hydrophilic root canal temporary dressings. KP Cal does not solidify in the root canal and is easy to thoroughly flush out after treatment.
### 6.0 Summary of Non-Clinical Testing
The following testing was conducted on subject device:
- Performance testing of Appearance, pH range, Calcium hydroxide content, Loading volume according to manufacturer standard.
Performance testing of Flowability, Radio-opacity, Film thickness according to ISO 6876.
- Biocompatibility tests were performed according to ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, ISO 10993-23.
- Shelf life test: Appearance, pH range, Calcium hydroxide content, Loading volume (manufacturer standard), Flowability, Radio-opacity, Film thickness (ISO 6876).
### 7.0 Summary of Clinical Testing
Clinical testing was not required for this submission.
### 8.0 Technological Characteristics and Substantial Equivalence
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
2 / 6
{7}
桂林凯文彼德科技有限公司
Guilin Kevin Peter Technology Co., Ltd.
Table 1 General Device Characteristics Comparison Table
| Item | Subject device | Primary predicate device | Secondary predicate device K060365 | Discuss/Justify the difference |
| --- | --- | --- | --- | --- |
| Product Name | KP-Cal Calcium Hydroxide Temporary Filling Material | Metapaste Plus Calcium Hydroxide Temporary Filling Material | ApexCal Calcium hydroxide paste | -- |
| Product Code | KIF | KIF | EJK | -- |
| Regulation No. | 21 CFR 872.3820 | 21 CFR 872.3820 | 21 CFR 872.3250 | -- |
| Class | II | II | II | -- |
| Device description | Calcium Hydroxide Temporary Filling Material is a premixed, immediately applicable, hydrophilic root canal temporary dressings. KP Cal does not solidify in the root canal and is easy to thoroughly flush out after treatment. | Metapaste Plus is a water-based calcium hydroxide paste material for root canal temporary filling. It is pre-mixed in a syringe for convenient delivery in the root canal. When applied to root canal, it is alkaline by calcium hydroxide. The state of the treatment can be confirmed by the radiopacity of barium sulfate contained. Also, it is water-soluble and easy to remove. | ApexCal is a creamy radiopaque,ready-to-use calcium hydroxide paste with a pH above 12.5. | Equivalent* (Analysis 1) |
| Indication for use | Calcium Hydroxide Temporary Filling Material is suitable for several indications including: -Temporary root canal filling; -Vital pulpotomy -Temporary pulp capping. | Metapaste plus is a biocompatible root canal filling used for the temporary filling of root canals after endodontic surgery. Metapaste Plus can be used on its own and for vital pulpectomy in deciduous teeth.Metapaste Plus is intended for use by qualified healthcare personnel trained in its use. | ApexCAL is a Calcium Hydroxide paste that has a creamy consistency and is suitable for several indications including: -Temporary disinfectant dressings in the obturation of root canals; | Equivalent* (Analysis 2) |
3 / 6
{8}
桂林凯文彼德科技有限公司
Guilin Kevin Peter Technology Co., Ltd.
| | | | -Indirect pulp capping or management of deep caries lesions; or -Direct pulp capping. | |
| --- | --- | --- | --- | --- |
| Prescription Use | Yes | Yes | Yes | Equivalent |
| Composition of Materials | This product consists of calcium hydroxide, zirconium dioxide, propylene glycol, polyethylene glycol. | Calcium hydroxide, Barium sulfate, Aluminum oxide, Titanium oxide, Polyethylene glycol, Dihydrogen oxide | ApexCal contains calcium hydroxide and bismuth carbonate in a mixture of water, glycerine, polyethylene glycol and other supplementary agents. | Similar (Analysis 3) |
| Liquid Formula | Paste | Paste | Paste | Equivalent |
| Packaging | Pre-loaded syringe | Pre-loaded syringe | Pre-loaded syringe | Equivalent |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Equivalent |
| Shelf Life | 2 years | 2 years | 2 years | Equivalent |
| Physical Properties | The subject device and the predicate device have substantially equivalent physical properties including flowability,Radio-opacity,Film thickness,pH range,Calcium hydroxide content according to ISO 6876 and manufacturer standard. | | | Similar (Analysis 4) |
| pH range | 12.32 | 12.40 | 12.46 | |
| Calcium hydroxide concentration | 26% | 25% | 24% | |
| Flowability | 19.9 | 20.8 | 20.4 | |
| Radiopacity | 4.5 | 4.5 | 4.6 | |
| Film thickness | 19 | 18 | 22 | |
| Biocompatibility | Conformed to ISO 10993 and ISO 7405. | Conformed to ISO 10993 and ISO 7405. | Conformed to ISO 10993 and ISO 7405. | Equivalent |
4 / 6
{9}
桂林凯文彼德科技有限公司
Guilin Kevin Peter Technology Co., Ltd.
# Analysis:
1) It's simply a difference in terminology. The subject device and predicate device are all pre-mixed, ready-to-use, strongly alkaline calcium hydroxide paste.
2) KP-Cal and the Primary predicate device(Metapaste plus) can be used for temporary root canal filling and pulpotomy. The indication for temporary pulp capping is only found in the KP-Cal's indication for use and not in the Metapaste Plus's. This technique is commonly used for temporary root canal fillings, and, as such, does not affect the equivalence of the indications of the subject and predicate devices.
3) The main component is same, but some other components are different. The main difference is in the composition of the retardant and solvent, but it does not affect the clinical performance and safety of the product. The biocompatibility and the performance test results supported the equivalence, the subject device is substantially equivalent to the predicate device.
4) Performance comparison testing was conducted to the predicate device and current subject device to compare their performance according to ISO 6876 and manufacturer standard. The performance technological parameters are little different with those of the predicate, and all meet the requirement. So the difference will not cause the difference in safety or efficacy.
Therefore, it is concluded that the subject device is substantially equivalent to the predicate device.
5 / 6
{10}
[LOGO]
桂林凯文彼德科技有限公司
Guilin Kevin Peter Technology Co., Ltd.
## 9.0 Conclusion
KP-Cal Calcium Hydroxide Temporary Filling Material has the similar intended use as the predicate devices. It has similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices.
The chemical compositions might slightly different from the predicate devices, both are used calcium hydroxide main composition as medicament, and additional component is used to make its appropriate workability.
The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that KP-Cal is substantially equivalent to the predicate devices. Hence, its equivalent is acceptable because it shows no clinically significant difference in the performance and safety to the device.
6 / 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.