K261503 · 3-D Matrix, Inc. · QAU · Aug 3, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261503
Device Name
PuraStat GI
Applicant
3-D Matrix, Inc.
Product Code
QAU · General, Plastic Surgery
Decision Date
Aug 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4456
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.
Device Story
PuraStat GI is a sterile, synthetic peptide hydrogel (2.5% peptide in water) used for hemostasis in gastrointestinal procedures. Delivered via prefilled syringe through an endoscopic catheter to the bleeding site; the peptide solution self-assembles into a transparent hydrogel barrier upon contact with physiological fluids (blood/ions). This mechanical barrier covers the wound, supporting natural coagulation. Used by healthcare professionals (surgeons, endoscopists) in clinical settings. The device is non-animal/non-plant derived, resorbable within 30 days, and provides clear visualization of the treatment site. It benefits patients by providing effective, temporary hemostasis for mild-to-moderate GI bleeding.
Clinical Evidence
Bench testing only. Evidence includes Gamma and EtO sterilization validation (ISO 11137, ISO 11135), packaging validation (ISO 11607), product stability, biocompatibility (ISO 10993-1, ISO 10993-7), and mechanical property verification of syringe components. No clinical data presented.
Technological Characteristics
Synthetic 16-amino acid (RADA-16) peptide hydrogel (2.5% concentration). Supplied in 1ml, 3ml, 5ml cyclo-olefin polymer syringes with HDPE plunger. Self-assembles into a scaffold barrier upon contact with physiological fluids. Resorbable within 30 days. Sterilized via Gamma radiation or terminal EtO. Standalone device; requires separate endoscopic catheter/adapter.
Indications for Use
Indicated for hemostasis of mild/moderate bleeding post ESD or EMR, as adjunct/bridge/prophylactic/rescue therapy for intraprocedural venous bleeding, or prophylactic therapy to prevent post-procedure bleeding, and for primary non-variceal GI bleeding. Contraindicated for arterial Forrest 1a bleeding.
Regulatory Classification
Identification
A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance data must support the sterility and pyrogenicity of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
(i) The ability to deliver the hemostatic material to the bleeding site;
(ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and
(iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids.
(ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
(6) Labeling must include:
(i) Information identifying and explaining how to use the device and its components; and
(ii) A shelf life.
{0}

August 03, 2026
3-D Matrix, Inc.
Alan MISKEWICZ
RAQA Director
1234 Chestnut St., Suite 205
Newton, Massachusetts 02464
Re: K261503
Trade/Device Name: PuraStat GI
Regulation Number: 21 CFR 878.4456
Regulation Name: Hemostatic Device For Intraluminal Gastrointestinal Use
Regulatory Class: Class II
Product Code: QAU
Dated: May 6, 2026
Received: May 6, 2026
Dear Alan MISKEWICZ:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261503 - Alan MISKEWICZ
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261503 - Alan MISKEWICZ
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE
-S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261503
Device Name
PuraStat GI
Indications for Use (Describe)
PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) SUMMARY
K261503
## 1. SUBMITTER
3-D Matrix Inc.
1234 Chestnut Street – Suite 205
Newton -- MA 02464
USA
Contact Person : Alan MISKEWICZ
RAQA Director - 3-D Matrix, Inc.
Phone: 617-291-0586
Email: amiskewicz@3dmatrix.com
Date Prepared: June 11, 2026
## 2. DEVICE
Name of Device: PuraStat GI
Common Name: Hemostatic device for intraluminal gastrointestinal use
Classification Regulation: 21 CFR 878.4456
Regulatory Class: II
Product Code: QAU
Panel: General & Plastic Surgery
## 3. PREDICATE DEVICES
Primary Predicate Device: 3-D Matrix Europe SAS PuraStat (K242250, K253924)
## 4. DEVICE DESCRIPTION
PuraStat GI is a sterile gel composed of synthetic peptide and sterile water for injection. It is provided as a prefilled syringe (2.5% peptide content) ready for use either with the separately packaged adapter and a commercially available luer lock endoscopic catheter, or with a direct connect endoscopic catheter, to deliver the gel. Adapter and endoscopic catheters are sold separately.
PuraStat GI is completely non-animal and non-plant derived and contains no preservatives that might present a risk of allergic reaction or skin irritation.
Exposure to physiological fluids such as blood causes the peptide solution to quickly form a transparent gel without expansion in volume. PuraStat GI achieves hemostatic effects by forming a hydrogel matrix barrier that blocks the flow of blood at the site of application.
Page 1 of 7
{5}
## 5. INDICATIONS FOR USE
PuraStat GI is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat GI is not indicated for arterial Forrest 1a bleeding.
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject PuraStat GI is substantially equivalent to the primary predicate PuraStat (K242250, K253924) representing alternative configurations of the same technology.
The subject device includes cumulative non-significant changes that are consistent with those previously cleared for the predicate devices. These variations do not affect the technological principles, mode of action, or performance of the device.
Performance data, including verification and validation activities, demonstrate that the subject device performs as intended and is as safe and effective as the predicate device.
Page 2 of 7
{6}
| | Subject Device | Primary Predicate | Difference |
| --- | --- | --- | --- |
| Device and Manufacturer | PuraStat GI 3-D MATRIX Inc. | PuraStat 3-D MATRIX EUROPE SAS | / |
| 510k N° | K261503 | K242250 & K253924 | / |
| Intended use | PuraStat GI is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary gastrointestinal (GI) bleeding. | 3-D MATRIX Europe SAS PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary gastrointestinal (GI) bleeding. | Identical |
| Indications for use | PuraStat GI is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding. | 3-D MATRIX Europe SAS PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding. | Identical |
| Regulation | 878.4456 | 878.4456 | Identical |
| Product code | QAU | QAU | Identical |
| Class | II | II | Identical |
Page 3 of 7
{7}
| | Technological Characteristics | | |
| --- | --- | --- | --- |
| | Subject Device | Primary Predicate | Difference |
| Device Design Description | 3-D MATRIX Inc. PuraStat is a synthetic haemostatic product supplied as a prefilled syringe containing a clear aqueous peptide solution. PuraStat GI is made of synthetic, absorbable peptides that form a hydrogel when applied. These peptides are designed to self-assemble into a structured network under physiological conditions. This process is driven by interactions between different parts of the peptide molecules, allowing them to organize into a gel that helps control bleeding. The self-assembly is promoted by conditions typically found in the body, such as neutral pH and the presence of common ions. | 3-D MATRIX Europe SAS PuraStat is a synthetic haemostatic product supplied as a prefilled syringe containing a clear aqueous peptide solution. PuraStat is made of synthetic, absorbable peptides that form a hydrogel when applied. These peptides are designed to self-assemble into a structured network under physiological conditions. This process is driven by interactions between different parts of the peptide molecules, allowing them to organize into a gel that helps control bleeding. The self-assembly is promoted by conditions typically found in the body, such as neutral pH and the presence of common ions. | Identical |
| Chemical description | 3-D Matrix Inc. PuraStat GI consists of a 2.5% solution of a synthetic peptide in sterile water for injection. The peptide is manufactured using standard chemical synthesis methods and does not contain any materials of animal or biological origin. | 3-D Matrix Europe SAS PuraStat consists of a 2.5% solution of a synthetic peptide in sterile water for injection. The peptide is manufactured using standard chemical synthesis methods and does not contain any materials of animal or biological origin. | Identical |
| Materials | Synthetic peptide hydrogel | Synthetic peptide hydrogel | Identical |
| Absorbable | Resorbed within 30 days | Resorbed within 30 days | Identical |
| Appearance | Clear, viscous gel | Clear, viscous gel | Identical |
| Product range | Three volumes are available: 1ml, 3ml, 5ml | Three volumes are available: 1ml, 3ml, 5ml | Identical |
Page 4 of 7
{8}
| | Technological Characteristics | | |
| --- | --- | --- | --- |
| | Subject Device | Primary Predicate | Difference |
| **Principle of Operations** | The product is a transparent aqueous peptide solution that, when applied to a bleeding site, rapidly forms a hydrogel barrier. This barrier helps control bleeding by covering the affected area and supporting the body's natural coagulation process. Its transparency allows clear visualization of the treatment site during use. Upon contact with physiological fluids, such as blood, the solution undergoes a change that triggers the peptides to self-assemble into a gel. This gel adheres to the tissue and forms a protective layer over the bleeding surface, contributing to effective hemostasis. Any remaining product is gradually absorbed by the body over time, although small amounts may persist for an extended period. | The product is a transparent aqueous peptide solution that, when applied to a bleeding site, rapidly forms a hydrogel barrier. This barrier helps control bleeding by covering the affected area and supporting the body's natural coagulation process. Its transparency allows clear visualization of the treatment site during use. Upon contact with physiological fluids, such as blood, the solution undergoes a change that triggers the peptides to self-assemble into a gel. This gel adheres to the tissue and forms a protective layer over the bleeding surface, contributing to effective hemostasis. Any remaining product is gradually absorbed by the body over time, although small amounts may persist for an extended period. | **Identical** |
| **Single use?** | Yes | Yes | **Identical** |
| **Provided sterile?** | Yes | Yes | **Identical** |
| **Sterilization method** | **Contract Sterilizers #1:** Gamma radiation between 17,5 kGy and 25 kGy of the device followed by a terminal EtO sterilization for final packaging sterilization **Contract Sterilizer #2:** Terminal Gamma radiation between 17,5 kGy and 25 kGy of the finished packaged device. | **Contract Sterilizers #1 (K253924):** Gamma radiation between 17,5 kGy and 25 kGy of the device followed by a terminal EtO sterilization for final packaging sterilization **Contract Sterilizer #2 (K242250):** Terminal Gamma radiation between 17,5 kGy and 25 kGy of the finished packaged device. | **Identical** |
| **Shelf life** | 3 years | 3 years | **Identical** |
| **User population** | PuraStat GI must be used by healthcare professionals (e.g surgeons, endoscopists, nurses, etc.) | 3-D MATRIX Europe SAS PuraStat must be used by healthcare professionals (e.g surgeons, endoscopists, nurses, etc.) | **Identical** |
| **How the gel is delivered** | Gel in a syringe, connected to catheter and delivered endoscopically. The gel solution is applied as close as possible to the bleeding point. | Gel in a syringe, connected to catheter and delivered endoscopically. The gel solution is applied as close as possible to the bleeding point. | **Identical** |
Page 5 of 7
{9}
The final product specifications are equivalent.
PuraStat GI is comprised of a synthetic repeating 16-amino acid (RADA-16) oligopeptide in sterile water for injection. In the presence of a bleeding wound, the peptides assemble into a scaffold structure that forms a mechanical barrier over the bleeding site. The solution is filtered and filled into 5-ml syringes made of cyclo-olefin polymers with a high-density polyethylene plunger and a butyl rubber head cap and gasket. Each syringe is filled with either 1, 3, or 5 ml of gel. The device is terminally sterilized, and the resorbable gel is delivered to the intended application site(s) via a commercially available adapter connected to a commercially available catheter, or with a direct connect endoscopic catheter, to deliver the gel. Adapter and endoscopic catheters are sold separately. The device is delivered endoscopically via a catheter to the patient.
## 7. PERFORMANCE DATA
The subject PuraStat GI is substantially equivalent to the primary predicate PuraStat (K242250, K253924) representing alternative configurations of the same technology.
The subject device has the same intended use and indications for use as the predicate device.
The determination of substantial equivalence is based on an assessment of non-clinical performance data.
PuraStat underwent the following performance testing.
- Gamma Sterilization Validation / ISO 11137
- Product performance
- Packaging Validation / ISO 11607
- Product stability over shelf life
- Biocompatibility / ISO 10993-1
The tests were performed on the subject device using the same methods, and acceptance criteria used for the primary predicate device.
Additional performance testing has been conducted to assess the changes introduced:
- EtO Sterilization Validation / ISO 11135
- EO/ECH residual analysis / ISO 10993-7
- Equivalence in mechanical properties of syringe component by comparing predicate and subject device components.
The results demonstrate that the subject device performs as intended and that any differences in technological characteristics do not raise new questions of safety and effectiveness.
Page 6 of 7
{10}
## **8. CONCLUSIONS**
The subject PuraStat GI is substantially equivalent to the primary predicate device 3-D Matrix Europe SAS PuraStat (K242250, K253924) in intended use, raw materials, specifications, chemical composition and performance. All those products aim to achieve hemostasis for the same indications through the same mechanism and contact the same human tissues and fluids.
PuraStat complies with the special controls for a hemostatic device for intraluminal gastrointestinal internal use.
In conclusion, PuraStat GI is substantially equivalent to the primary predicate.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.