The Arterial Pressure Monitoring Set/Tray is intended for arterial blood pressure monitoring and blood sampling in adult and pediatric patients. • 2.5 French catheters are intended for patients from birth and older. • 3.0 and 4.0 French catheters are intended for patients aged 1 year and older. • 5.0 French catheters are intended for patients aged 12 years and older.
Device Story
Arterial Pressure Monitoring Set/Tray facilitates arterial blood pressure monitoring and blood sampling. Device components include pressure monitoring catheter, wire guide, entry access needle, and syringe. Clinician inserts catheter into radial artery, femoral artery, or distal aorta using Seldinger technique. Device is sterile, single-use, and non-electrical. Output is direct arterial pressure signal transmitted to external monitoring equipment. Clinical benefit includes accurate, real-time hemodynamic monitoring and access for blood sampling.
Clinical Evidence
No clinical data. Bench testing only, specifically Ethylene Oxide Sterilization Performance Qualification to validate changes in sterilization cycle parameters.
Technological Characteristics
Catheter materials: Polyethylene, Ethylene-Vinyl Acetate, Polyurethane, Nylon. Wire guide: Stainless steel. Single lumen, straight tip. Sizes: 2.5F to 5.0F. Sterilization: Ethylene Oxide (EtO) to 10^-6 SAL. No electrical components or software.
Indications for Use
Indicated for arterial blood pressure monitoring and blood sampling in adult and pediatric patients. 2.5F catheters for birth and older; 3.0F and 4.0F for 1 year and older; 5.0F for 12 years and older.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Xcela Hybrid PICC with PASV Valve Technology (K252373)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 26, 2026
Spectrum Vascular
Sharon Klugewicz
Sr. VP Quality & Regulatory Affairs
50 Main St.
Suite 1000
White Plains, New York 10606
Re: K261455
Trade/Device Name: Arterial Pressure Monitoring Set/Tray
Regulation Number: 21 CFR 870.1200
Regulation Name: Diagnostic Intravascular Catheter
Regulatory Class: Class II
Product Code: DQO
Dated: May 1, 2026
Received: May 1, 2026
Dear Sharon Klugewicz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261455 - Sharon Klugewicz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261455 - Sharon Klugewicz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**HETAL B. ODOBASIC -S**
for
Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology
Diagnostics and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261455
Device Name
Arterial Pressure Monitoring Set/Tray
Indications for Use (Describe)
The Arterial Pressure Monitoring Set/Tray is intended for arterial blood pressure monitoring and blood sampling in adult and pediatric patients.
• 2.5 French catheters are intended for patients from birth and older.
• 3.0 and 4.0 French catheters are intended for patients aged 1 year and older.
• 5.0 French catheters are intended for patients aged 12 years and older.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
510(k) Summary
Spectrum Vascular
K261455
# **1. SUBMITTER INFORMATION**
Applicant: Spectrum Vascular
Contact: Ms. Sharon Klugewicz
Phone: 516-425-4446
Email: sklugewicz@spectrumvascular.com
Address: 50 Main Street, Suite 1000 White
Plains, NY 10606
# **2. CORRESPONDENT INFORMATION**
Contact: Ms. Sharon Klugewicz
Title: Sr. VP, Quality & Regulatory Affairs
Firm: Spectrum Vascular
50 Main Street, Suite 1000
White Plains, NY 10606
Phone: 516-425-4446
Email: sklugewicz@spectrumvascular.com
# **3. DATE PREPARED: JUNE 25, 2026**
# **4. DEVICE INFORMATION**
Device Name: Arterial Pressure Monitoring Set/Tray
Common Name: Diagnostic intravascular catheter
Regulation Number: 21 CFR 870.1200
Regulation Name: Diagnostic intravascular catheter
Product Code: DQO
Regulatory Class: Class II
# **5. PREDICATE DEVICE INFORMATION**
Primary Predicate Device Name: Arterial Pressure Monitoring Set/Tray 510(k) Number: K254278
Manufacturer: Spectrum Vascular
Page 1 of 4
{5}
510(k) Summary
Spectrum Vascular
K261455
| Reference Device Name: | BioFlo Hybrid PICC with ENDEXO Technology and PASV Valve Technology; BioFlo PICC with ENDEXO Technology and PASV Valve Technology; BioFlo PICC with ENDEXO Technology; BioFlo Midline Catheter; Xcela Power Injectable PICC; Xcela PICC with PASV Valve Technology; Xcela Hybrid PICC with PASV Valve Technology |
| --- | --- |
| 510(k) Number: | K252373 |
| Manufacturer: | Spectrum Vascular |
## 6. DEVICE DESCRIPTION
The Arterial Pressure Monitoring Set/Tray is intended to be used for arterial pressure monitoring and blood sampling. The pressure monitoring catheter is inserted into the vasculature using the Seldinger technique. Based on the device length, the subject device may gain access via the radial artery, femoral artery, or distal aorta.
The Arterial Pressure Monitoring Set/Tray is comprised of a pressure monitoring catheter, wire guide, entry access needle, and syringe. The Arterial Pressure Monitoring Set/Tray is supplied sterile and intended for one-time use.
## 7. INDICATIONS FOR USE
The Arterial Pressure Monitoring Set/Tray is intended for arterial blood pressure monitoring and blood sampling in adult and pediatric patients.
- 2.5 French catheters are intended for patients from birth and older.
- 3.0 and 4.0 French catheters are intended for patients aged 1 year and older.
- 5.0 French catheters are intended for patients aged 12 years and older.
## 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Table 1 contains the technological comparison of the subject device to the primary predicate device.
Table 1: Technological Comparison to Predicate Devices
| | Predicate Device (K254278) | Subject Device |
| --- | --- | --- |
| Regulation | 21 CFR 870.1200 | 21 CFR 870.1200 |
| Product Code | DQO | DQO |
| Classification | Class II | Class II |
| Catheter Insertion Method | Percutaneously via Seldinger technique | Percutaneously via Seldinger technique |
| Catheter Shaft Material | Polyethylene | Polyethylene |
Page 2 of 4
{6}
510(k) Summary
Spectrum Vascular
K261455
| | Predicate Device (K254278) | Subject Device |
| --- | --- | --- |
| | Ethylene-Vinyl Acetate Polyurethane Nylon | Ethylene-Vinyl Acetate Polyurethane Nylon |
| Catheter Hub Material | Polyethylene Polyurethane Nylon | Polyethylene Polyurethane Nylon |
| Catheter Outer Diameter | 2.5, 3.0, 4.0, 5.0 French | 2.5, 3.0, 4.0, 5.0 French |
| Catheter Length | 2.5, 4.0, 5.0, 6.0, 8.0, 12.0, 15.0 cm | 2.5, 4.0, 5.0, 6.0, 8.0, 12.0, 15.0 cm |
| Catheter Lumen Design | Single lumen | Single lumen |
| Catheter Distal End | Straight tip | Straight tip |
| Wire Guide Material | Stainless Steel | Stainless Steel |
| Wire Guide Diameter | 0.015, 0.018, 0.021, 0.035 inch | 0.015, 0.018, 0.021, 0.035 inch |
| Packaging | Tray w/ Tyvek Lidstock | Tray w/ Tyvek Lidstock |
| Single Use? | Yes | Yes |
| Sterilization Method | EtO | EtO |
| Sterility Assurance Level | 10^{-6} | 10^{-6} |
## 8.1 DISCUSSION OF SIMILARITIES AND DIFFERENCES
Other than the change in the sterilization location and sterilization cycle parameters, the subject device is identical to the predicate device. There are no changes to catheter design or materials from the predicate device. The differences in sterilization location and parameters have been validated to achieve the same sterility assurance level of 10$^{-6}$ with acceptable EO and ECH residuals and thus does not impact either the safety or the efficacy of the subject devices.
## 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
### Biocompatibility Testing
The medical devices in their final finished form are identical to the predicate device (K254278) in formulation, processing, sterilization method and geometry. No other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
The subject device is sterilized using the same ethylene oxide (EtO) method as the predicate device, with a reduced EtO gas concentration. This represents a less aggressive exposure condition than the predicate's validated cycle. Because the predicate's biocompatibility testing was conducted under the higher EtO concentration, that testing represents the worst-case residual scenario and remains applicable to the subject device. No additional biocompatibility testing was conducted.
### Electrical Safety
Page 3 of 4
{7}
510(k) Summary
Spectrum Vascular
K261455
Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type.
#### **Electromagnetic Compatibility (EMC)**
Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type.
#### **Software**
Not applicable. The subject device contains no software.
#### **Performance Testing**
The following performance testing was conducted to validate changes to the sterilization cycle parameters:
- Ethylene Oxide Sterilization Performance Qualification
### **10. CONCLUSION**
The results of the performance testing described above demonstrate that the Arterial Pressure Monitoring Set/Tray performs equivalently to the predicate device. These results support a determination of substantial equivalence and demonstrate that the subject device is as safe and effective as the predicate device.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.