K261437 · The Surgical Company International B.V., doing business as TSC Life · DWJ · Aug 7, 2026 · Cardiovascular
Device Facts
Record ID
K261437
Device Name
Mistral-Air® Warming System
Applicant
The Surgical Company International B.V., doing business as TSC Life
Product Code
DWJ · Cardiovascular
Decision Date
Aug 7, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Mistral-Air® Warming System is a forced air warming system that comprises of a warming unit and a variety of blankets. It is intended to raise and maintain patient temperature by means of surface warming.
Device Story
Forced air warming system; comprises reusable warming unit (MA1400) and disposable blankets. Unit draws air through filter; heats via internal element; propels warm air through flexible hose into blanket draped on, around, or under patient. Heat transfer via convection; blanket provides insulation. Operated manually via control panel by clinicians in healthcare settings. Maintains patient temperature; prevents inadvertent hypothermia. System requires use of specific Mistral-Air blankets for safety and performance.
Clinical Evidence
No clinical data. Evidence consists of nonclinical safety data including electrical safety, electromagnetic compatibility, software verification and validation, mechanical and durability assessments, lifecycle and packaging evaluations, human factors evaluation, and risk management activities.
Technological Characteristics
Electrically powered (100-125 V AC, 50/60Hz). Components: reusable warming unit, flexible hose, disposable blankets. Materials: metal, electronics, ABS/PC. Convection-based warming. Manual control panel operation. Software-controlled heating and airflow regulation.
Indications for Use
Indicated for patient warming in situations where prevention of inadvertent hypothermia is required. Applicable to all age groups: neonates, infants, children, adolescents, and adults.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 7, 2026
The Surgical Company International B.V., doing business as TSC Life
Gianpaolo Soares
RA Project Manager
Paalbergweg 3
Noord-Holland
Amsterdam, 1105 AG
Netherlands
Re: K261437
Trade/Device Name: Mistral-Air® Warming System
Regulation Number: 21 CFR 870.5900
Regulation Name: Thermal Regulating System
Regulatory Class: Class II
Product Code: DWJ
Dated: July 10, 2026
Received: July 10, 2026
Dear Gianpaolo Soares:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261437 - Gianpaolo Soares
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261437 - Gianpaolo Soares
Page 3
Sincerely,
**KATHLEEN M.
GRUNDER -S**
for Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support,
Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261437 | ? |
| Please provide the device trade name(s). | | ? |
| Mistral-Air® Warming System | | |
| Please provide your Indications for Use below. | | ? |
| The Mistral-Air® Warming System is indicated for patient warming in situations where prevention of inadvertent hypothermia is required. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
TSC LIFE
K261437
Innovation with purpose
Submitter Name and Address
The Surgical Company International B.V., doing business as TSC Life
Paalbergweg 3
Amsterdam, 1105 AG
The Netherlands
Contact Person
Name: Gianpaolo Soares
Title: Regulatory Affairs Project Manager, TSC Life
Company Telephone: 765-205-6030
Email: fda.tsclife-us@tsc-life.com
Device Information
Name of Device: Mistral-Air® Warming System
Model: MA1400 (Warming Unit)
Common or Usual Name: Thermal regulating system
Classification Name: System, Thermal Regulating, 21 CFR 870.5900
Regulatory Class: II
Product Code: DWJ
Predicate Information
Same device as required for Special 510(k) pathway.
Name of Predicate: Mistral-Air® Warming System
Model: MA1200 (Warming Unit)
510(k) Number: K171234
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
Device Description
The Mistral-Air® Warming System is comprised of a warming unit and blankets and is designed to warm patients using temperature-controlled filtered warm forced airflow. The device's blanket, which can be placed on top, around, or underneath a patient, guides the airflow towards the skin of the patient, where heat is transferred by convection. Additionally, this combination keeps the patient insulated. The warming unit generates and directs the warm air into the blankets. This controlled airflow ensures consistent and effective warming. For optimal safety and performance, warming unit and hose shall only be used with the Mistral-Air® blankets.
Intended Use
The Mistral-Air® Warming System is a forced air warming system that comprises of a warming unit and a variety of blankets. It is intended to raise and maintain patient temperature by means of surface warming.
Indications for Use
The Mistral-Air® Warming System is indicated for patient warming in situations where prevention of inadvertent hypothermia is required.
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
Page | 2
{5}
TSC LIFE
Innovation with purpose
Table 1: Comparison of Technological Characteristics
| Subject | Predicate Device Mistral-Air® Warming System (with MA1200) | Subject Device Mistral-Air® Warming System (with MA1400) | Substantial Equivalence |
| --- | --- | --- | --- |
| Mode of Operation | Manual operation via control panel | Manual operation via control panel | Same |
| Material of Mistral-Air® Warming Unit | Metal, electronics and ABS/PC | Metal, electronics and ABS/PC | Same |
| Operating Principle | Electrically powered device (consisting of a control circuit, air filter, fan, heating element and temperature sensors) that propels warmed air via a flexible hose to a blanket (Sterile or Non-sterile, depending on the model) draped on or under the patient. | Electrically powered device (consisting of a control circuit, air filter, fan, heating element and temperature sensors) that propels warmed air via a flexible hose to a blanket (Sterile or Non-sterile, depending on the model) draped on or under the patient. | Same |
| System Components | • Reuseable Warming unit • Disposable sterile and non-sterile blankets/suits | • Reuseable Warming unit • Disposable sterile and non-sterile blankets/suits | Same |
| Warming Technology | Convection | Convection | Same |
| Energy Source | 100-125 V AC, 50Hz/60Hz | 100-125 V AC, 50Hz/60Hz | Same |
| Blanket Compatibility | Yes - Compatible with Mistral-Air® Warming Blankets | Yes - Compatible with Mistral-Air® Warming Blankets | Same |
### Performance Data
Nonclinical safety data included electrical safety and electromagnetic compatibility evaluation, software verification and validation, mechanical and durability assessments, lifecycle and packaging evaluations, human factors evaluation, and risk management activities. Cybersecurity documentation was prepared in accordance with FDA's guidance 'Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions' (issued February 3, 2026).
### Conclusion
The purpose of this Special 510(k) premarket notification is to introduce a new model of the Mistral-Air® Warming Unit (part of the Mistral-Air® Warming System). The results of the nonclinical evaluations conducted for this premarket notification demonstrate that the modified device is as safe and as effective as the legally marketed predicate device and performs as well as or better than the predicate. The proposed model change does not alter the device's intended use or fundamental technological characteristics, nor does it raise new questions of safety or effectiveness. The information provided supports a determination that the Mistral-Air® Warming System, including the MA1400 Warming Unit, is substantially equivalent to the predicate Mistral-Air® Warming System with the MA1200 Warming Unit cleared under K171234.
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
Page | 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.