The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Ophthalmic, Trans-esophageal (Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MV-Flow mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. (Note: V5/V4 models include Dermatology, exclude Ophthalmic).
Device Story
Mobile, software-controlled diagnostic ultrasound systems; acquire ultrasound data via transducers; display images in multiple modes (2D, Doppler, M-mode, 3D/4D, ElastoScan, MV-Flow). Operated by trained healthcare professionals in clinical settings (hospitals, private practices). Provides anatomical measurements and analysis packages to assist clinical diagnosis. Features include voice command, automated risk assessment (IOTA-SRrisk), structure/flow analysis (EzStructure, EzFlow), and 3D printing support. Real-time acoustic output (mechanical/thermal indices) displayed for safety. Output used by clinicians for diagnostic decision-making; benefits include non-invasive visualization and quantification of anatomical structures and fluid dynamics.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including acoustic output measurement, biocompatibility, software verification/validation, and electrical/electromagnetic safety testing per IEC/ISO standards.
Technological Characteristics
Mobile ultrasound system; Windows 11 OS; transducers (e.g., BCC2-13, PA2-9S, CA2-13M, PA1-5AED); acoustic output per NEMA UD 2-2004; electrical safety per IEC 60601-1; electromagnetic compatibility per IEC 60601-1-2. Supports 2D, Doppler, M-mode, 3D/4D, ElastoScan, and MV-Flow modes.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients of all ages (neonates to adults) across various clinical applications including fetal/obstetrics, abdominal, cardiac, vascular, musculoskeletal, and others as specified per model.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999)
EVO Q30/EVO Q20/EVO Q10/ EVO XQ30/EVO XQ20/EVO XQ10/EVO QH30/ EVO QH20/EVO QH10 Diagnostic Ultrasound System (K254099)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
Samsung Medison Co., Ltd.
Dayoung Choi
Regulatory Affairs Specialist
3366, Hanseo-Ro, Nam-Myeon
Hongcheon-Gun, Gangwon-do 25108
Republic Of Korea
July 28, 2026
Re: K261412
Trade/Device Name: V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System; V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System; V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System; V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System; V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, LLZ, QIH
Dated: July 6, 2026
Received: July 6, 2026
Dear Dayoung Choi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261412 - Dayoung Choi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261412 - Dayoung Choi
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D. O'hara - S
Date: 2026.07.28 10:35:07 -04'00' For
Yanna Kang
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K261412
Please provide the device trade name(s).
V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System; V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System; V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System; V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System; V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System;
Please provide your Indications for Use below.
For V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System:
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Ophthalmic, Trans-esophageal (Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MV-Flow mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
For V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System:
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac), Peripheral vessel and Dermatology.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MV-Flow Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
Please select the age group(s) for which the device(s) is to be used.
☑ Neonates/Newborns (Birth to < 29 days old)
☑ Infants (29 days old to < 2 years old)
☑ Children (2 years old to < 12 years old)
☑ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater)
?
{5}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
## 510(K) Summary: K261412
In accordance with 21 CFR 807.92, the following summary of information is provided:
1. Date Prepared – April 30th, 2026
2. Manufacturer
SAMSUNG MEDISON CO., LTD.
3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, Republic of Korea
3. Contact Person
Da-young Choi
Regulatory Affairs Specialist
Phone: +82.2.2194.1407
Fax: +82.2.2194.0273
Email: dy1217.choi@samsung.com
4. Secondary Contact Person
Ninad Gujar
Vice President
Phone: +1.978.564.8632
Fax: +1.978.564.8677 Email: ngujar@samsunghme.com
5. Proposed Device
- Common/Usual Name : Diagnostic Ultrasound System and Accessories
- Proprietary Name :
V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System;
V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System;
V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System;
V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System;
V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System;
- Regulation Name : Ultrasonic pulsed doppler imaging system
- Regulatory Class : Class II
- Product Code : IYN, IYO, ITX, LLZ, QIH
- Regulation Number : 21 CFR 892.1550, 892.1560, 892.1570, 892.2050
6. Predicate / Reference Device
- Predicate device
• V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999)
- Reference Device
• HERA Z20, HERA Z20e, HERA Z20s, R20, HERA Z30, R30 Diagnostic Ultrasound System (K252018)
1
{6}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
• RS85 Diagnostic Ultrasound System (K240516)
• EVO Q30/EVO Q20/EVO Q10/ EVO XQ30/EVO XQ20/EVO XQ10/EVO QH30/ EVO QH20/EVO QH10 Diagnostic Ultrasound System (K254099)
# 7. Device Description
■ V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System:
The V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System are general purpose, mobile, software controlled, diagnostic ultrasound systems. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MV-Flow Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
■ V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System:
The V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System are general purpose, mobile, software controlled, diagnostic ultrasound systems. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MVFlow Mode. Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V5/XV5/XH5/H5/cV5
2
{7}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
# 8. Intended Use
■ V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System:
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Ophthalmic, Trans-esophageal (Cardiac), and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MV-Flow Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
■ V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System:
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac), Dermatology, and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, MV-Flow Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
# 9. Technological Comparison to Predicate Device
The V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6
3
{8}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System employ the same fundamental scientific technology as its predicate device V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999).
# 10. Determination of Substantial Equivalence
Comparison to Predicate: The V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System, V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, and V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System are substantially equivalent to the predicate device with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
■ V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System:
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System and the predicate V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999) have the same intended use, and modes of operation.
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have introduced a clinical application for ophthalmic use.
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have been updated to Windows 11 operating system.
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7
4
{9}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have migrated the following software features from the reference device HERA Z20, HERA Z20e, HERA Z20s, R20, HERA Z30, R30 Diagnostic Ultrasound System (K252018): Voice Command, IOTA-SRrisk, EzStructure, EzFlow, EzCheck, PortraitVue, IDEA, IETA, EzPictogram, Luminant, Slice A, Live Q-Scan, 3D Printing.
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have included a new transducer, BCC2-13.
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have migrated transducers PA2-9S, CA2-13M from the reference device HERA Z20, HERA Z20e, HERA Z20s, R20, HERA Z30, R30 Diagnostic Ultrasound System (K252018).
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System have migrated Prostate Fusion, a sub-function of the previously cleared S-Fusion in the reference device RS85 Diagnostic Ultrasound System (K240516).
- For marketing purposes, we've included the following product names for the XV, XH, H, cV, V ONE, XV ONE, XH ONE, H ONE, cV ONE series: (XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE, XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE, XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE), which have the same specifications as the V series (V8, V7 and V6).
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System and the primary predicate device V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999) have been designed in compliance with approved electrical and physical safety standards.
5
{10}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- The proposed V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, and V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System and the primary predicate device V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999) have the same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The system is manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The system has acoustic power levels which are below the applicable FDA level.
■ V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System:
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The proposed V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System and the predicate V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999) have the same intended use, and modes of operation.
- The proposed V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System have been updated to Windows 11 operating system.
- The proposed V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System have migrated the following software features from the reference device HERA Z20, HERA Z20e, HERA Z20s, R20, HERA Z30, R30 Diagnostic Ultrasound System (K252018): A-Focus, Voice Command, IOTA-SRrisk, EzStructure, EzFlow, EzCheck, PortraitVue, IDEA, IETA, EzPictogram, Luminant, Slice A, 3D Printing.
- The proposed V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System have migrated transducer PA2-9S from the reference device HERA Z20, HERA Z20e, HERA Z20s, R20, HERA Z30, R30 Diagnostic Ultrasound System (K252018), and the PA1-5AED from the reference device EVO Q30/EVO Q20/EVO Q10/EVO XQ30/EVO XQ20/EVO XQ10/EVO QH30/ EVO QH20/EVO QH10 Diagnostic Ultrasound System (K254099).
6
{11}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- For marketing purposes, we’ve included the following product names for the XV, XH, H, cV series: (XV5/XH5/H5/cV5, XV4/XH4/H4/cV4), which have the same specifications as the V series (V5, V4).
- The proposed V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound and the primary predicate device V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999) have been designed in compliance with approved electrical and physical safety standards.
- The proposed V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound and the primary predicate device V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 Diagnostic Ultrasound System (K250999) have the same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The system is manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The system has acoustic power levels which are below the applicable FDA levels.
# 11. Summary of Non-Clinical Testing
The device has been evaluated for acoustic output, biocompatibility, software function, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable FDA guidances and medical device safety standards. The V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System, V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, and V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System and its applications comply with the following FDA-recognized standards.
| Reference No. | Title |
| --- | --- |
| IEC 60601-1 | IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 |
| IEC 60601-1-2 | IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020 |
| IEC 60601-2-37 | IEC 60601-2-37:2024 |
| IEC 60601-4-2 | IEC TS 60601-4-2:2024 |
| ISO10993-1 | ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
| ISO14971 | ISO 14971:2019, Medical devices - Application of risk management to medical devices |
7
{12}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
| NEMA UD 2-2004 | NEMA UD 2-2004 (R2009) Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3 |
| --- | --- |
All software tests met the acceptance criteria for verification and validation testing.
# 12. Summary of Clinical Tests
The proposed device V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System, V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, and V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System did not require clinical studies to demonstrate substantial equivalence.
# 13. Conclusion
Since the predicate device and the proposed device have a similar intended use and key technological features, the non-clinical data support the safety of the device and demonstrate that the V8/XV8/XH8/H8/cV8/V8 ONE/XV8 ONE/XH8 ONE/H8 ONE/cV8 ONE Diagnostic Ultrasound System, V7/XV7/XH7/H7/cV7/V7 ONE/XV7 ONE/H7 ONE/XH7 ONE/cV7 ONE Diagnostic Ultrasound System, V6/XV6/XH6/H6/cV6/V6 ONE/XV6 ONE/H6 ONE/XH6 ONE/cV6 ONE Diagnostic Ultrasound System, V5/XV5/XH5/H5/cV5 Diagnostic Ultrasound System, and V4/XV4/XH4/H4/cV4 Diagnostic Ultrasound System should perform as intended in the specified use conditions. Therefore, SAMSUNG MEDISON CO., LTD. considers the proposed device to be as safe, as effective, and substantially equivalent in performance to the predicate device that is currently marketed for the same intended use.
- END of 510(k) Summary -
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.