The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm to 56cm.
Device Story
Arm-type oscillometric blood pressure monitor; uses piezoelectric pump for cuff inflation and electric valve for rapid deflation; extracts pulse waveform data during inflation. Microprocessor software analyzes waveforms to calculate systolic/diastolic pressure and pulse rate; detects irregular heartbeats if pulse interval variance exceeds 25%. Used in home or clinical environments by patients or clinicians. Models with Bluetooth (2.4GHz BLE) transmit data to external terminals/apps for display/storage. Output includes LCD results, irregular heartbeat warnings, WHO classification, and trend charts. Benefits include non-invasive, automated monitoring of cardiovascular metrics to support clinical decision-making.
Clinical Evidence
No new clinical testing performed. Accuracy leveraged from previously cleared devices (K230566, K250548). Leveraged studies used ISO 81060-2:2018+AMD2020 protocol with manual mercury sphygmomanometer reference. Study 1 (n=95) and Study 2 (n=85) confirmed accuracy within acceptable limits. Bench testing verified equivalence of core algorithms and performance characteristics.
Technological Characteristics
Oscillometric measurement; piezoelectric pump; solenoid valve; 4 AAA batteries or AC adapter. Upper arm cuff (22-42cm or 40-56cm). Connectivity: 2.4GHz BLE module (select models). Standards: IEC 60601-1, IEC 80601-2-30, IEC 60601-1-11, IEC 60601-1-2, ISO 10993-1/5/10/23. Software: Embedded microprocessor algorithm for waveform analysis.
Indications for Use
Indicated for adults and adolescents over 12 years of age with upper arm circumference 22-42cm or 40-56cm for non-invasive measurement of systolic/diastolic blood pressure and pulse rate in home or clinical settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 26, 2026
JOYTECH Healthcare Co., Ltd.
Yingting Lan
#502, Shunda Rd., Hangzhou, Zhejiang 311100, China
#365, Wuzhou Rd., Hangzhou, Zhejiang 311100, China
Hangzhou, Zhejiang 311100
China
Re: K261394
Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: July 31, 2026
Received: July 31, 2026
Dear Yingting Lan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261394 - Yingting Lan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261394 - Yingting Lan
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# STEPHEN C. BROWNING -S
CDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics and Monitoring
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261394 | ? |
| Please provide the device trade name(s). | | ? |
| Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus) | | |
| Please provide your Indications for Use below. | | ? |
| The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm to 56cm. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
Page 1 of 1
{4}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
# 510(k) Summary
The assigned 510(k) number is: K261394
## 1. Sponsor's Identification
Name: JOYTECH Healthcare Co., Ltd.
Address 1: No.502, Shunda Road, Hangzhou, Zhejiang 311100, China.
Address 2: No.365, Wuzhou Road, Hangzhou, Zhejiang 311100, China.
Contact Person: Yingting Lan
Email:lanyt@sejoy.com
## 2. Name of the Device:
Trade Name: Arm-type Fully Automatic Digital Blood Pressure Monitor
| Arm-type | DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus |
| --- | --- |
Common Name: Blood Pressure Monitor
Classification name: Non-invasive blood pressure measurement System
21 CFR 870-1130, Class II, 74-DXN.
## 3. Classification Information:
Regulation Number: 870.1130
Product Code: DXN
Device Class: II
Panel: 74 Cardiovascular
## 4. Predicate/Reference Device Information:
The Arm-Type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following devices:
Page 1 of 9
{5}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
| Device | 510(k) number | Predicate device model | Product code | Manufacturer |
| --- | --- | --- | --- | --- |
| Predicate | K250548 | DBP-6286B, DBP-6186 | DXN | JOYTECH Healthcare Co., Ltd. |
## 5. Device Description:
The Arm-type Fully Automatic Digital Blood Pressure Monitors(DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus) are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm to 56cm.
They are automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use at home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During inflation measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min. The pulse rate measurement is compare the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.
DBP-6286B Plus, DBP-62E3B Plus, DBP-62E2B Plus with bluetooth function can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD display of the device appears results. And the device will start to transmit data to the pair-up terminal automatically.
With the use of software (including APP) and wireless communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without
Page 2 of 9
{6}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled ‘Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices’ (issued on September 28, 2022.), this software function is belong to Non-device-MDDS, and the hardware function is belong to Device-MDDS, they are not subject to FDA laws and regulations applicable to devices.
The function for DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus are listed below:
| Function | DBP-6186 Plus | DBP-6286B Plus | DBP-61E3 Plus | DBP-61E2 Plus | DBP-62E3B Plus | DBP-62E2B Plus |
| --- | --- | --- | --- | --- | --- | --- |
| Blood Pressure measurement | Y | Y | Y | Y | Y | Y |
| Pulse rate measurement | Y | Y | Y | Y | Y | Y |
| Irregular Heartbeat Indicator | Y | Y | Y | Y | Y | Y |
| Memory | Y | Y | Y | Y | Y | Y |
| WHO Classification Indicator | Y | Y | Y | Y | Y | Y |
| Last 3 Test Average | Y | Y | Y | Y | Y | Y |
| MVM function | Y | Y | Y | Y | Y | Y |
| Low Battery Detection | Y | Y | Y | Y | Y | Y |
| Automatic Power-Off | Y | Y | Y | Y | Y | Y |
| Voice | Y | Y | Y | Y | Y | Y |
| Backlight | Y | Y | Y | Y | Y | Y |
| Arm shake indicator | Y | Y | Y | Y | Y | Y |
| Cuff loose indicator | Y | Y | Y | Y | Y | Y |
| Bluetooth function | N | Y | N | N | Y | Y |
| Trend Chart | Y | Y | Y | Y | Y | Y |
Note: Y: Yes; N: No
Page 3 of 9
{7}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
# 6. Indication for use/Intended Use:
The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm
# 7. Comparison of Technological Characteristics with Predicate/Reference Device:
The arm-type blood pressure monitor for model DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus manufactured by JOYTECH have the same indication for use and similar technological characteristics with the DBP-6286B, DBP-6186(K250548). And the main change is the subject device include extra function. Therefore, we select DBP-6286B, DBP-6186 in K250548 as Predicate device. The detailed comparison of technical characteristic is as below:
| Comparison item | Subject Device (DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus) | Primary Predicate device(DBP-6286B, DBP-6186, K250548) | Comparison result / Explanation |
| --- | --- | --- | --- |
| The trade name | Arm-type Fully Automatic Digital Blood Pressure Monitor | Blood pressure monitor | / |
| Manufacturer | JOYTECH Healthcare Co., Ltd. | JOYTECH Healthcare Co., Ltd. | / |
| Recommended classification regulation | 21CFR 870.1130, Noninvasive Blood Pressure Measurement System | 21CFR 870.1130, Noninvasive Blood Pressure Measurement System | Identical |
| Regulatory class | II | II | Identical |
| Panel | 74 Cardiovascular | 74 Cardiovascular | Identical |
| Product code | DXN | DXN | Identical |
| Indication for use/Intended use | The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm | The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm | Identical |
| Intended use environment | Home and Professional healthcare facility environment. | Home and Professional healthcare facility environment. | Identical |
| Measuring principle | Oscillometric method | Oscillometric method | Identical |
| Measurement type | Determined during inflation | Determined during inflation | Identical |
| Cuff location | Upper arm | Upper arm | Identical |
| Specification | | | |
Page 4 of 9
{8}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
| Comparison item | Subject Device (DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus) | Primary Predicate device(DBP-6286B, DBP-6186, K250548) | Comparison result / Explanation |
| --- | --- | --- | --- |
| Measuring range | Systolic Pressure: 60mmHg~260 mmHg; Diastolic Pressure: 40mmHg~200 mmHg; Cuff pressure: 0mmHg~299 mmHg; Pulse rate: 30~180 Beats/Minute; | Systolic Pressure: 60mmHg~260 mmHg; Diastolic Pressure: 40mmHg~200 mmHg; Cuff pressure: 0mmHg~299 mmHg; Pulse rate: 30~180 Beats/Minute; | Identical |
| Accuracy | Pressure deviation:±3 mmHg; Pulse deviation: 40-180Beats/Minute, ±5%; 30-39Beats/Minute, ±5bpm | Pressure deviation: ±3 mmHg Pulse deviation: 40-180Beats/Minute, ±5%; 30-39Beats/Minute, ±5bpm | Identical |
| Pressure release | By solenoid valve | By solenoid valve | Identical |
| Operating Temp. & humidity | Temp.: 10°C~40°C Humidity: 15~93%RH Atmospheric pressure:80kPa~106kPa | Temp.: 10°C~40°C Humidity: 15~93%RH Atmospheric pressure:80kPa~106kPa | Identical |
| Storage Temp. & humidity | Temp.: -25°C~55°C Humidity: ≤93% RH | Temp.: -25°C~55°C Humidity: ≤93% RH | Identical |
| Cuff circumference | 22-42 cm, 40-56cm | 22-42cm, 40-56cm | Identical |
| Sterilization | Not applicable | Not applicable | Identical |
| Supply power source | 4 AAA batteries or Medical AC Adapter | 4 AAA batteries or Medical AC Adapter | Identical |
| **Function** | | | |
| Wireless transmission function | DBP-6186 Plus, DBP-61E3 Plus, DBP-61E2 Plus: No DBP-62E3B Plus, DBP-62E2B Plus, DBP-6286B Plus: Bluetooth | DBP-6186: No DBP-6286B: Bluetooth | Similar. See Note 1 |
| Memory | 2*150 Memories in Two Groups | 2*150 Memories in Two Groups | Identical |
| Other function | Blood Pressure measurement Pulse rate measurement Irregular Heartbeat Indicator WHO Classification Indicator Last 3 Test Average MVM function Low Battery Detection Automatic Power-Off Voice Backlight Arm shake indicator Cuff loose indicator Trend chart | Blood Pressure measurement Pulse rate measurement Irregular Heartbeat Indicator WHO Classification Indicator Last 3 Test Average MVM function Low Battery Detection Automatic Power-Off Voice Backlight Arm shake indicator Cuff loose indicator | Similar. See note 2 |
Note 1: The wireless function is different between the subject devices and the predicate devices. However, it serves the same purpose, that is to transfer measurement results. As for the wireless technology of the subject device, it is same, which has also been validated according to IEC 60601-1-2, Part 15 Subpart C, ANSI C63.27. As demonstrated in relevant test reports as well, the difference here does not raise different questions of safety and effectiveness, and subject devices are as safe and effective as the predicate device.
Note 2: The subject devices have additional trend chart function. Also, the subject devices with such design has been verified according to IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO
Page 5 of 9
{9}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
80601-2-30, and internal bench testing. Therefore, the difference here does not raise different questions of safety and effectiveness, and subject devices are as safe and effective as the predicate device
# 8. Performance Data:
Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted.
The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.
The following National and International Standards were utilized for testing the subject device.
# Electrical Safety and performance requirements:
- IEC 60601-1:2005+AMD1:2012+AMD2:2020, Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.
AAMI ES60601-1:2005+AMD1:2012+AMD2:2021, Medical Electrical Equipment.
- IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.
# Home-used medical equipment requirements and environmental test:
- IEC 60601-1-11:2015+AMD1:2020, General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
# Electromagnetic Compatibility Requirements:
- IEC 60601-1-2:2014+AMD1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
# Bio-compatibility Evaluation for patient contacting components:
- ISO 10993-1:2018, Biological evaluation of medical devices--Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- ISO 10993-23:2021, Biological evaluation of medical devices--Part 23: Tests for irritation
Page 6 of 9
{10}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
# FCC Test
FCC Part15 Subpart C
RF Exposure Evaluation
# Wireless Coexistence
ANSI C63.27-2021
# Guidance Document:
- The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions
- Biological evaluation was made in accordance with "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
- Electromagnetic Compatibility (EMC) of Medical Devices
The test results all meet the requirement of these standards.
# 9. Discussion of Clinical Tests Performed:
# Clinical Validation for 22-42cm Cuff (Leveraged from Previously Cleared Device):
The accuracy and performance of the subject devices intended for the use with 22-42cm cuff are leveraged and supported by clinical data from the model DBP-6279B, previously cleared device in K230566, which is unchanged from subject device models intended for the general population. No new clinical testing was conducted for these models on the general populations.
The leveraged clinical study was performed according to ISO 81060-2:2018+AMD2020, "Non-invasive sphygmomanometers — Part 2: Clinical investigation of intermittent automated measurement type." This study involved 95 subjects aged over 12 years using the same-arm sequential method with a manual mercury sphygmomanometer as reference. The results
Page 7 of 9
{11}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
demonstrated that the accuracy of the model DBP-6279B was within the acceptable limits specified by the standard.
# Clinical Validation for 40-56cm Cuff(Leveraged from Previously Cleared Device)
The accuracy and performance of the subject devices intended for use with 40-56cm cuff are leveraged and supported by clinical data from the model DBP-6286B, previously cleared device in K250548. No new clinical testing was conducted for these models with 40-56cm cuff.
Model DBP-6286B was model for clinical study. It was conducted according to ISO 81060-2: 2018+AMD2020 standard. Same arm sequential method was adopted during clinical testing.
The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The subject demographics include 85 valid subjects data were used to analysis. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.
# Equivalence Justification via Bench Testing:
Comprehensive bench testing and internal simulation analyses were conducted to demonstrate equivalence. The testing confirmed:
Equivalence between the subject device models and the previously cleared model DBP-6286B from K250548, DBP-6279B from K230566, as they share the same core algorithm and essential performance characteristics.
# Clinical Coverage Discussion:
The subject devices share the same upper-arm application site as previous portable models DBP-6279B, DBP-6286B cleared by JOYTECH. Combined with the bench testing equivalence proof between the subject models, the previously cleared model DBP-6279B, DBP-6286B, the safety and effectiveness of all subject devices are adequately demonstrated. The above clinical studies can cover the subject devices. No additional study is needed.
# 10. Conclusions:
The subject device and predicate devices have same Measuring principle, measuring method, are designed for the measurement of blood pressure, pulse rate and detection of irregular pulses
Page 8 of 9
{12}
JOYTECH 健拓
JOYTECH. Healthcare CO., LTD.
K261394
for home use. The minor difference between the subject devices and the predicate devices have been evaluated and determined to not raise any new issues of safety or effectiveness.
Therefore, the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
Page 9 of 9
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.