Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique, it will be filled with air and/or saline during use. The LOR Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. The intended patient population includes children, adolescents, and adults.
Device Story
Manual syringe used by clinicians during epidural procedures; verifies needle tip placement in epidural space via Loss of Resistance (LOR) technique. Device filled with air or saline; not for injection or aspiration. Available in 5mL, 7mL, 10mL volumes. Models feature either NRFit (ISO 80369-6) or Luer (ISO 80369-7) connectors. Single-use; supplied sterile or non-sterile for kit integration. Clinician observes tactile resistance feedback during plunger advancement to confirm needle location. Benefits include standardized, reliable verification of epidural space entry.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench performance testing and biocompatibility evaluation.
Technological Characteristics
Manual syringe; materials include polypropylene, silicone, and polyisoprene rubber. Connectors comply with ISO 80369-6 (NRFit) and ISO 80369-7 (Luer). Sterilized via ethylene oxide (SAL 10^-6). Volume sizes: 5mL, 7mL, 10mL. One-piece or three-piece plunger assemblies.
Indications for Use
Indicated for use in children, adolescents, and adults to verify epidural needle tip placement via Loss of Resistance technique using air or saline.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
{1}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 25, 2026
CMT Health PTE., Ltd.
Monica Ma
Official Correspondent
150 Beach Rd., #28-05, Gateway W.,
Singapore, 189720
Singapore
Re: K261372
Trade/Device Name: Profoject™ LOR Syringe
Regulation Number: 21 CFR 868.5140
Regulation Name: Anesthesia conduction kit
Regulatory Class: Class II
Product Code: CAZ
Dated: August 21, 2026
Received: August 21, 2026
Dear Monica Ma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261372 - Monica Ma
Page 2
may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K261372 - Monica Ma
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JAMES J. LEE -S
for Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261372
Device Name
Profoject™ LOR Syringe
Indications for Use (Describe)
Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique, it will be filled with air and/or saline during use. The LOR Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. The intended patient population includes children, adolescents, and adults.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
# 510(k) Summary
[As required by 21 CFR 807.92]
1. Submission Information
| 510(k) Number: | K261372 |
| --- | --- |
| Date: | April 27, 2026 |
| Type of 510(k) Submission: | Traditional |
| Submitter: | CMT HEALTH PTE. LTD. 6001 BEACH ROAD, #19-13A, GOLDEN MILE TOWER, SINGAPORE 199589 Contact Person: Monica Ma E-mail: ra@cmthealth.com Tel: +65 6846 1379 |
2. Device Description
| Trade Name: | Profoject™ LOR Syringe |
| --- | --- |
| Common Name: | Loss of Resistance Syringe |
| Regulation Name: | Anesthesia Conduction Kit |
| Product Code: | CAZ |
| Device Class: | II |
| Regulation Number: | 21 CFR 868.5140 |
| Review Panel: | Anesthesiology |
| Indications for use: | Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique, it will be filled with air and/or saline during use. The LOR Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. The intended patient population includes children, adolescents, and adults. |
| Device Description: | The Profoject™ LOR Syringe is available in two models: Model A and Model B. Model A is a syringe equipped with an NRFit slip or NRFit lock connector that complies with ISO 80369-6, while Model B is a syringe equipped with a Luer slip or Luer lock connector that complies with ISO 80369-7. Depending on the type of plunger assembly, Model B can be further subdivided into Model B1 and Model B2. Model A has only one type of plunger assembly. The product is offered in three volume sizes: 5 mL, 7 mL, and 10 mL. ● Model A: Available with NRFit lock or NRFit slip male connectors (compliant with ISO 80369-6). The type of plunger is one-piece plunger. ● Model B: Model B1: Available with Luer lock or Luer slip male connectors (compliant with ISO 80369-7); and the type of plunger is one-piece plunger. Model B2: Fitted with Luer slip male connector only (compliant with |
1
{6}
| | ISO 80369-7). The type of plunger is three-piece plunger.The devices are single-use and are supplied either sterile and individually packaged, or non-sterile and bulk-packed. The sterile devices are sterilized with ethylene oxide gas to achieve an SAL of \( 10^{-6} \) and are provided with a sterility maintenance package that maintains sterility throughout the shelf life. The non-sterile, bulk-packed syringes are supplied to anesthetic conduction kit manufacturers for incorporation into the kit and terminal sterilization. |
| --- | --- |
## 3. Predicate Device Identification
K061737 - LOSS OF RESISTANCE SYRINGE
## 4. Non-Clinical Test Conclusion
Non-clinical verification of the Profoject™ LOR Syringe has been conducted to evaluate their safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods, and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the effectiveness of the proposed device in accordance with relevant test methods cited below:
### 4.1. Performance Testing
1) ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
2) ISO 80369-1 Second edition 2018-11 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
3) ISO 80369-6 First Edition 2016-03-15 Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications
4) ISO 80369-7 Second edition 2021-05 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
5) ISO 80369-20 First edition 2015-05-15 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
6) ISO 11737-1 Third edition 2018-01 [Including AMD1:2021] Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
7) ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
8) ASTM F1886/F1886M-16 (2024) Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
9) ASTM F1140/F1140M-13 (2020) Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
10) ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
11) ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
12) USP - NF <71> Sterility Tests
13) USP - NF <788> Particulate Matter in Injections
2
{7}
14) USP - NF <85> Bacterial Endotoxin Testing
15) ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)]
16) ISTA 3A: 2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
17) USP - NF <788> Particulate Matter in Injections
18) Testing for Scale Marking Stability
### 4.2. Biocompatibility Testing
The biocompatibility evaluation for the Profoject™ LOR Syringe was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as an externally communicating device.
1) ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2) ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
3) ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
4) USP <151> Rabbit Pyrogen Study
5) ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
6) ISO 10993-4 Third edition 2017-04 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood
7) ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
## 5. Usability Testing
Simulated Clinical Study Report
## 6. Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data was not included in this submission.
## 7. Substantially Equivalent Comparison Conclusion
### 7.1. Comparison of Technology Characteristics
Table 1 Comparison of Technology Characteristics
| Parameters | | Proposed Device | Predicate Device | Remark |
| --- | --- | --- | --- | --- |
| 1 | 510(k) Number | K261372 | K061737 | / |
| 2 | 510(k) Holder | CMT HEALTH PTE. LTD. | Robert Busse & Co., Inc. | / |
| 3 | Trade Name | ProfojectTM LOR Syringe | LOSS OF RESISTANCE SYRINGE | / |
| 4 | Product Code | CAZ | CAZ | Same |
| 5 | Regulation Number | 21CFR 868.5140 | 21 CFR 868.5140 | |
3
{8}
| 6 | Review Panel | Anesthesiology | Anesthesiology | |
| --- | --- | --- | --- | --- |
| 7 | Device Class | II | II | Same |
| 8 | Indications for use | The Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the loss of Resistance technique, it will be filled with air and/or saline during use. The loss of resistance syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. | The Busse Loss of Resistance Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the loss of Resistance technique, it will be filled with air and/or saline during use. The loss of Resistance Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. | Same |
| 9 | Overall design | Barrel; Plunger | Barrel; Plunger | Same |
| 10 | Single Use | Single Use | Single Use | Same |
| 11 | Priming Volume | 5mL, 7mL, 10mL | 10mL | Different See Note 1 |
| 12 | Connectivity | ISO 80369-6 ISO 80369-7 | ISO 80369-7 | Different See Note 2 |
| 13 | Materials | Model A Barrel: Polypropylene Plunger: Polypropylene, yellow color additive Plunger stopper: Silicone compound and Light blue color additive | Plastic materials | Different See Note 3 |
| | | Model B1 Barrel: Polypropylene Plunger: Polypropylene, blue color additive Plunger stopper: Silicone compound and Light blue color additive | | |
| | | Model B2 Barrel: Polypropylene Plunger: Polypropylene, white color additive and blue color additive Plunger stopper: Polyisoprene rubber | | |
| 14 | Biocompatibility | Complies with ISO 10993-1 | Complies with ISO | Same |
4
{9}
| | | | 10993-1 | |
| --- | --- | --- | --- | --- |
| 15 | Sterile | Sterile or non-sterile | Sterile or non-sterile | Same |
| 16 | Sterile Method | EO Sterilized | EO Sterilized | Same |
| 17 | Sterilization | \( SAL=10^{-6} \) | \( SAL=10^{-6} \) | Same |
### 7.2. Substantial Equivalence Discussion
Table 2 Substantial Equivalence Discussion
| Discussion | Note 1 - Syringe Volume:The syringe volume of the proposed device differs from that of the predicate device. However, this difference is just in capacity. In clinical practice, physicians select the appropriate syringe specification based on individual patient conditions, and this difference does not affect the intended use. Furthermore, bench performance testing was conducted on the proposed device in accordance with ISO 7886-1, ISO 80369-7, and ISO 80369-6, and all results met the requirements of these standards. Therefore, this difference does not raise new questions of safety or effectiveness.Note 2 - Connectivity:The proposed device includes both ISO 80369-7 and ISO 80369-6 connectors, whereas Predicate Device includes only ISO 80369-7 connector. All connectors on the proposed device have been tested and conform to ISO 80369-6 and ISO 80369-7 standards, and labeling addresses misconnection risks. Therefore, this difference does not raise new questions of safety or effectiveness.Note 3 - Materials:The specific materials of the predicate devices are not publicly available. However, the proposed devices were tested, and the results demonstrated that they complied with the ISO 10993 series standards. Predicate devices are legally marketed and have been determined to meet equivalent biocompatibility requirements. Therefore, the lack of publicly available predicate material information does not raise different questions of safety and effectiveness. In addition, bench performance testing per ISO 7886-1, ISO 80369-7, and ISO 80369-6 was conducted on the proposed device, and all results met the standard requirements. Accordingly, the proposed device is substantially equivalent to the predicate devices. |
| --- | --- |
| Conclusion | The ProfojectTM LOR Syringe is substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device. |
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.