Customised crowns, inlays, onlays, veneers and implant-retained crowns, 3-unit bridges as well as veneers of frameworks of indirect restorations in the anterior and posterior regions.
Device Story
Grandio disc multicolor is a nano-hybrid composite CAD/CAM disc; 86% inorganic filler content. Used by dental professionals to manufacture permanent single-tooth restorations, 3-unit bridges, and framework-supported prosthetics. Material is milled via dental CAD/CAM systems into final restoration geometry; then polished for clinical placement. Device addresses tooth substance loss and edentulism. Benefits include customized, durable, aesthetic dental restorations. Operates as a restorative material; no active electronic or software components.
Clinical Evidence
Bench testing only. Evaluated flexural strength, flexural modulus, compressive strength, water absorption, and water solubility per ISO 4049. Biocompatibility evaluated per ISO 7405 and ISO 10993-1 (cytotoxicity, irritation, sensitization, genotoxicity). Results confirm subject device meets standards and is substantially equivalent to the predicate.
Technological Characteristics
Nano-hybrid composite CAD/CAM disc. Composition: 86% inorganic silica-based glass/silica fillers; resin matrix includes EtoBG, UDMA, BisGMA, TEGDMA. Form factor: Discs (12, 15, 18 mm). Non-sterile. Standards: ISO 4049 (restorative materials), ISO 10993-1 (biocompatibility), ISO 7405 (dental biocompatibility).
Indications for Use
Indicated for patients with lack of tooth substance, partial edentulism, or complete edentulism requiring customized crowns, inlays, onlays, veneers, implant-retained crowns, or 3-unit bridges (excluding inlay, cantilever, and Maryland bridges) in anterior and posterior regions.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 27, 2026
Voco GmbH
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120a
Madison, New Jersey 07940
Re: K261357
Trade/Device Name: Grandio disc multicolor
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth shade resin material
Regulatory Class: Class II
Product Code: EBF
Dated: April 24, 2026
Received: April 24, 2026
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261357 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261357 - Dave Yungvirt
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (303) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K261357 | |
| Device Name Grandio disc multicolor | |
| Indications for Use (Describe) Intended purpose: Customised crowns, inlays, onlays, veneers and implant-retained crowns, 3-unit bridges as well as veneers of frameworks of indirect restorations in the anterior and posterior regions. Indication: Lack of tooth substance, partial edentulism and complete edentulism. Fields of application: Without framework - Inlays, onlays, veneers - Crowns, implant-retained crowns - 3-unit bridges (not Inlay bridges, Cantilever bridges and Maryland bridges) With supporting framework - Veneers of supporting frameworks irrespective of the selected manufacturing process. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
510(k) Number: K261357
Submitter: VOCO GmbH, Anton-Flettner-Str. 1-3, 27472
Cuxhaven, Germany
Contact: Dr. T. Gerkensmeier, Regulatory Affairs, Tel: +49 (0) 4721 719 200, t.gerkensmeier@voco.de
Prepared By: K. Fröhlich, Regulatory Affairs, k.froehllich@voco.de
Date Prepared: 24 April, 2026
## Device Identification
Trade Name: Grandio disc multicolor
Common Name: Restorative resin material
Classification Name: Tooth shade resin material
Regulation Number: 872.3690
Product Code: EBF
Class: Class II
Classification Panel: Dental
## Legally Marketed Predicate Device:
AMBARINO® High-Class, (K222723); Manufacturer: Creamed GmbH & Co. Produktions- und Handels KG
Device Description: Grandio disc multicolor is a nano-hybrid composite CAD/CAM disc with 86% by weight of inorganic fillers for the manufacture of customised, permanent single-tooth restorations, 3-unit bridges and/or framework-supported prosthetic restorations.
Statement of Intended Use: Customised crowns, inlays, onlays, veneers and implant-retained crowns, 3-unit bridges as well as veneers of frameworks of indirect restorations in the anterior and posterior regions.
Substantial Equivalence: Information provided in this application shows that the product is substantially equivalent to the predicate device. Comparisons of the physical properties of the Grandio disc multicolor to the predicate devices are included in this application.
{5}
The table below compares Grandio disc multicolor to the predicate device with respect to its physical state, structure, materials, mechanical properties, indications for use, biocompatibility and performance testing, and provides detailed information regarding the basis for the determination of substantial equivalence.
{6}
| Elements of Comparison | Subject Device | Predicate Device | Similarities / Differences |
| --- | --- | --- | --- |
| Device Name | Grandio disc multicolor | AMBARINO® High-Class | - |
| Manufacturer | VOCO GmbH | Creamed | - |
| 510(k) # | pending | K222723 | - |
| Product Code | EBF | EBF | Same |
| Regulation | §872.3690 | §872.3690 | Same |
| Class | II | II | Same |
| Review Panel | Dental | Dental | Same |
| Device Image | | | - |
| Indications for Use | Fields of application: Without framework
– Inlays, onlays, veneers
– Crowns, implant-retained crowns
– 3-unit bridges (not Inlaybridges, Cantileverbridges and Marylandbridges)
With supporting framework
– Veneers of supporting frameworks irrespective of the selected manufacturing process. | For fabrication of inlays/ onlays, laminate veneers, anterior and posterior full crown restorations, and implant crowns and bridges by dental professionals and manufacturers using a dental CAD/CAM system. | Similar |
3
{7}
| Physical State | Cured discs in a variety of shapes and shades | Cured blocks and discs in a variety of shapes and shades | Same |
| --- | --- | --- | --- |
| Structure | Polymer resin / ceramic hybrid composite | Polymer resin / ceramic hybrid composite | Same |
| Methacrylate-based Resin Matrix | 14.44% EtoBG, UDMA, BisGMA, TEGDMA | 29% UDMA, BDDMA | Similar. The subject and predicate devices contain UDMA in their resin matrix. |
| Filler Content | 86% inorganic silica-based glass and silica | 70% inorganic silica-based glass and barium glass | Similar |
| Sizes | 12, 15, 18 | 14, 40, 98, 100 | The size ranges are comparable. |
| Flexural Strength | 315.9 MPa | 192.61 MPa | Similar |
| Modulus of Elasticity | 12.95 GPa | 6.10 GPa | Similar |
| Compressive Strength | 524.06 MPa | 562.91 MPa | Similar |
| Water absorption (EN ISO 10477) | 14.67 [μg/mm³] | 12.64 [μg/mm³] | Similar |
| Water solubility (EN ISO 10477) | -0.18 [μg/mm³] | -2.11 [μg/mm³] | Same |
| Packaging | One disc per box. | One disc per box, or for mandrel mounted blocks, six small or two large to a box | Same for disc. |
| Usage | Single Patient, multiple use | Single Patient, multiple use | Same |
| Sterility | Non-Sterile | Non-Sterile | Same |
| Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | Same |
{8}
5
| Performance | Conforms with ISO 4049 | Conforms with ISO 4049 | Same |
| --- | --- | --- | --- |
Non-Clinical Performance Data: As part of demonstrating substantial equivalence of Grandio disc multicolor to the predicate device, VOCO GmbH submitted final finished devices for extensive testing in accordance with the applicable parts of the following voluntary standards, as well as to the company's own internal test protocols:
- ISO 4049, Dentistry – Polymer-based RestorativeMaterials
- ISO 7405, Dentistry – Evaluation of biocompatibility of medical devices used in dentistry
- ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process
Testing evaluated flexural strength, flexural modulus, compressive strength, water absorption, water solubility and biocompatibility of the subject device, as well as other related physical properties. The subject device exceeded all minimum test standards, and all documented test results were substantially equivalent to the predicate device.
Design Characteristics: The geometry of the Grandio disc multicolor is a disc form as defined by the CAD/CAM manufacturer. The material will be inserted into the CAM machine and milled into its final form, then polished and ready for placement thereafter. This design is substantially equivalent to the predicate device.
Biocompatibility: Biocompatibility tests were conducted on Grandio disc multicolor with regard to cytotoxicity, irritation, sensitization, and genotoxicity. All tests were inconclusive.
Conclusion: Grandio disc multicolor has the same or similar intended use, indications for use, physical attributes, and are fabricated into permanent tooth restorations using the same CAD/CAM manufacturing methods as AMBARINO® High-Class. Different sizes are available for the submitted device. All performance data of the two devices are similar, and both exceed the minimum requirements. Any minor differences in the materials used to make the subject device when compared to the predicate device have been successfully evaluated by the manufacturer through extensive performance and biocompatibility testing on their device, such that the information submitted to the FDA demonstrates that the subject device is as safe and effective as the predicate device and does not raise any new questions of safety and effectiveness.
Grandio disc multicolor, as designed and manufactured by VOCO GmbH, have been determined to be substantially equivalent to AMBARINO® High-Class.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.