FlexTract Deflectable-Rotational Dilator Sheath Set
K261310 · Merit Medical Systems, Inc. · DRE · Jul 27, 2026 · Cardiovascular
Device Facts
Record ID
K261310
Device Name
FlexTract Deflectable-Rotational Dilator Sheath Set
Applicant
Merit Medical Systems, Inc.
Product Code
DRE · Cardiovascular
Decision Date
Jul 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FlexTract Sheath Set is indicated for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads.
Device Story
Handheld, manually operated, mechanically rotating and deflectable lead extraction tool; used during transvenous lead extraction (TLE) procedures. Input: manual user control of handle/trigger. Transformation: mechanical rotation of distal stainless-steel dilating tip; manual deflection of distal sheath section to navigate tortuous anatomy. Output: dilated tissue/fibrous attachments around cardiac leads. Used in clinical settings by physicians. Benefits: facilitates removal of anchored cardiac leads. Includes radiopaque PTFE outer sheath for manual dilation or stiffening. Available in 9, 11, 13, 16 Fr sizes; 46 cm length.
Clinical Evidence
No clinical data. Evidence consists of bench-top performance testing, including dimensional verification, cyclical testing, critical function testing (deflection, rotation, trigger force), and design validation via clinician-simulated use with phantom models. Biocompatibility testing performed per ISO 10993-1 for limited (<24h) blood contact.
Technological Characteristics
Materials: polymer/stainless steel braided sheath, stainless-steel tip, PTFE outer sheath. Energy: manual mechanical. Dimensions: 9-16 Fr, 46 cm length. Connectivity: none. Sterilization: ethylene oxide. Design: manual handle/trigger for bi/uni-directional tip rotation and distal deflection.
Indications for Use
Indicated for patients requiring percutaneous dilation of tissue surrounding cardiac leads.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
Predicate Devices
Evolution RL Controlled-Rotation Dilator Sheath Set (K141148)
Submission Summary (Full Text)
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
July 27, 2026
Merit Medical Systems, Inc.
Kirk Mcintosh
Senior Regulatory Affairs Specialist
1600 W. Merit Pkwy.
South Jordan, Utah 84095
Re: K261310
Trade/Device Name: FlexTract Deflectable-Rotational Dilator Sheath Set
Regulation Number: 21 CFR 870.1310
Regulation Name: Vessel dilator for percutaneous catheterization
Regulatory Class: Class II
Product Code: DRE
Dated: June 26, 2026
Received: June 26, 2026
Dear Kirk Mcintosh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261310 - Kirk Mcintosh
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261310 - Kirk Mcintosh
Page 3
Sincerely,
JESSICA L. BATTISTA -S
Digitally signed by JESSICA L. BATISTA -S
Date: 2026.07.27
13:01:39 -04'00'
Jessica Batista Bertolini
Acting Assistant Director
DHT2A: Division of Cardiac
Electrophysiology, Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261310
Device Name
FlexTract Deflectable-Rotational Dilator Sheath Set
Indications for Use (Describe)
The FlexTract Sheath Set is indicated for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# FlexTract Deflectable-Rotational Dilator Sheath Set 510(k) Summary
| General Provisions | Submitter Name: Merit Medical Systems, Inc. Address: 1600 West Merit Parkway South Jordan, UT 84095 Telephone Number: (801) 316-3695 Contact Person: Kirk McIntosh Date Prepared: April 15, 2026 Registration Number: 1721504 |
| --- | --- |
| Subject Device | Trade Name: FlexTract Deflectable-Rotational Dilator Sheath Set Common/Usual Name: Dilator, Vessel, for Percutaneous Catheterization Classification Name: Dilator, Vessel, for Percutaneous Catheterization Regulatory Class: II Product Code: DRE 21 CFR §: 870.1310 Review Panel: Cardiovascular |
| Predicate Device | Trade Name: Evolution RL Controlled-Rotation Dilator Sheath Set Classification Name: Dilator, Vessel, for Percutaneous Catheterization Premarket Notification: K141148 (submitted by Cook Vascular Inc.) Manufacturer: Merit Medical Systems, Inc.* * In 2024, Merit Medical acquired K141148 portfolio of devices from Cook Medical This predicate has not been subject to a design-related recall. |
| Reference Device | No reference devices were used in this submission. |
| Device Description | The FlexTract Deflectable-Rotational Dilator Sheath Set is a handheld, manually operated, mechanically rotating and deflectable lead extraction tool for use during transvenous lead extraction (TLE) procedures. Its primary function is to dilate tissue and fibrous attachments which may be anchoring the targeted lead to the inner vascular wall, thereby facilitating removal of the lead. The FlexTract Sheath Set is comprised of a polymer/stainless steel braided sheath connected to a polymeric ergonomic handle/trigger capable of mechanically rotating a stainless-steel dilating tip at the distal terminus of the device. It is designed for the user to be able to rotate the distal tip in either a bi-directional or uni-directional manner. To facilitate passage through tortuous anatomy, the FlexTract is designed with manual controls to allow the user to deflect the distal section of the sheath and choose the direction of deflection. A radiopaque polytetrafluoroethylene (PTFE) Outer Sheath accessory is provided with the FlexTract device and may be used to provide manual dilation or as a stiffening aid of the FlexTract shaft during use. The FlexTract Sheath Set is available in 9 Fr, 11 Fr, 13 Fr and 16 Fr configurations with a working length of 46 cm. It is labeled for single-use and provided sterile (ethylene oxide). |
K261310
Page 1 of 4
{5}
| **Indications for Use** | The FlexTract Sheath Set is indicated for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads. The indications of the subject device are only a subset of those of the predicate. The change does not represent an increase in the level of specificity of the indication for use. The difference in indications is not critical to the intended surgical use of the device. |
| --- | --- |
| **Comparison to Predicate Device** | The subject FlexTract Sheath Set is similar in design and technological characteristics to the predicate Evolution RL Sheath Set. The comparison between the subject device and predicate devices is based on the following: - The Indications for Use are a subset of those of the predicate - Same fundamental technology - Same sterilization methods - Same labeling The following technological differences exist between the subject device and predicate device: - Minor dimensional differences - Addition of a larger (16 Fr) configuration - Addition of design feature to allow deflection and positioning of the distal section of the device - Different patient contacting materials that meet ISO 10993-1 requirements - Packaging configuration includes a protective tray within the sealed pouch The design, material, dimensional, and packaging differences are considered technological differences but do not raise new or different questions of safety and effectiveness and can be evaluated with performance testing. |
| **Performance Data** | No specific FDA guidance documents or recognized performance standards have been established for the subject device classification. A battery of tests was performed based upon risk analysis and the requirements of the following internationally recognized standards and guidance documents pertaining to the device performance, as well as biocompatibility, sterilization, and labeling standards and guidance. Conformity to these standards demonstrates that the proposed FlexTract Sheath Set meets the acceptance criteria established by the standards. - FDA guidance document: Use of International Standard ISO 10993-1 (2023), 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process' - ISO 11607-1 (2023), Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems - ASTM D4169-22, Standard Practice for Performance Testing of Shipping Containers and Systems - ISO 2233 (2001), Packaging – Complete, Filled Transport Packages and Unit Loads – Conditioning for Testing - ASTM F1980-21, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices |
K261310
Page 2 of 4
{6}
# **Performance Data**
**(Continued)**
- ASTM F1929-23, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM F2096-19 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- ASTM F88/F88M-23, Standard Test Method for Seal Strength of Flexible Barrier Materials
- ISO 10993-1 (2025), Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process
- ISO 10993-4 (2017), Biological evaluation of medical devices – Part 4: Selection of tests for interaction with blood
- ISO 10993-5 (2009), Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 (2021), Biological evaluation of medical devices – Part 10: Tests for irritation and delayed type hypersensitivity
- ISO 10993-11 (2017), Biological evaluation of medical devices – Part 11: Tests for systemic toxicity
- ISO 10993-12 (2025), Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- ISO 10993-23 (2025), Biological evaluation of medical devices - Part 23: Tests for irritation
- ASTM F756-17, Standard Practice for Assessment of Hemolytic Properties of Materials
- USP 43-NF38:2020, <151> Pyrogen Test (USP Rabbit Test)
- ANSI/AAMI HE75 (2018), Human factors engineering - Design of medical devices
- IEC 62366-1 (2020), Medical devices - Part 1: Application of usability engineering to medical devices
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Testing
The biocompatibility evaluation for the FlexTract Sheath Set was conducted in accordance with the FDA guidance document: Use of International Standard ISO-10993-1, “Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process” and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA.
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Pyrogenicity
- Hemolysis
- Thrombogenicity
- Complement Activation
The FlexTract Sheath Set meets the biocompatibility requirements for externally communicating devices with circulating blood contact for a limited (< 24 hours) duration.
K261310
Page 3 of 4
{7}
| Performance Data (Continued) | The results of the following performance tests demonstrated that the subject FlexTract Sheath Set met the acceptance criteria applicable to the device. Performance / Design Verification Testing: (including shelf-life stability and distribution testing) - Dimension verification - Visual inspection - Cyclical testing - Critical functions (Deflection, Tab rotation, Tip rotation, Trigger pull) - Bend without deformation - Tip rotation testing - 3-Point bend stiffness - Dial rotation - Maximum deflection - Tip to Tab Alignment - Coaxiality when relaxed - Trigger force testing - Sheath retraction while deflected - Dial torque - Tip to Handle force comparison - Device Torsional testing - Radiopacity - Visual Inspection – Packaging - Dye Penetration – Packaging - Underwater Bubble Emission Testing – Packaging - Tensile Strength – Packaging Design Validation Testing: - Clinician simulated use, bench-top testing with phantom model - Summative and HFE activities |
| --- | --- |
| Summary of Substantial Equivalence | Based on the indications for use, design, safety, and performance testing, the subject FlexTract Deflectable-Rotational Dilator Sheath Set meets the requirements that are considered essential for its intended use, does not raise new questions of safety or effectiveness, and is therefore substantially equivalent to the predicate device, the Evolution RL Controlled-Rotation Dilator Sheath Set (K141148). |
K261310
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.