The ultrasound system is indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans- vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac. In addition, the ultrasound diagnostic system and associated transducers are intended for: - Measurements of abdominal anatomical structures, - Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen, - Measurements of brightness ratio between liver and kidney, - Visualization of abdominal vascularization, microvascularization and perfusion, The shear wave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion maybe used as an aid to clinical management of adult and pediatric patients with liver disease. This device is intended for use by, or by the order of, and under the supervision of a licensed physician qualified to use or direct the use of the device. This system should only be used by trained sonographers who are knowledgeable about the risk of excessive acoustic energy in the body, particularly in the case where a great amount of fluid is present in the scanning area. This device is intended to be used in a hospital or medical clinic.
Device Story
Cart-based diagnostic ultrasound system; utilizes scan converter and various transducers (linear, convex, micro-convex, phased array) to acquire acoustic signals; processes data for B-mode, M-mode, Doppler (Color, Pulsed Wave, Continuous Wave, Power, Directional Power), Harmonic, Contrast-Enhanced, and Elasticity imaging; displays images on LCD monitor. Operated by trained sonographers/physicians in hospitals or clinics. Provides anatomical imaging, tissue stiffness/viscosity measurements, and vascular perfusion visualization to aid clinical management of liver disease. Features include spatial compounding, panoramic imaging, biopsy guidance, and automated TGC adjustment. Output used by clinicians for diagnostic assessment and clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including acoustic output, biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and software validation (IEC 62304).
Technological Characteristics
Cart-based ultrasound system; supports linear, convex, micro-convex, and phased array transducers. Materials meet ISO 10993-1. Imaging modes: B, M, Color/Power/CW/PW Doppler, Harmonic, CEUS, Elasticity. Connectivity: DICOM, wired network, USB, DisplayPort. Software lifecycle per IEC 62304. Acoustic output complies with IEC 62359 (Track 3, MI ≤ 1.9, Ispta.3 ≤ 720 mW/cm²).
Indications for Use
Indicated for imaging and measurement of anatomical structures in abdominal, small organs, musculoskeletal, superficial musculoskeletal, vascular, peripheral vascular, OB-GYN, pelvic, pediatric, urology, trans-rectal, trans-vaginal, neonatal/adult cephalic, and non-invasive cardiac applications. Includes liver/spleen stiffness and shear wave speed measurements, liver/kidney brightness ratio, and abdominal vascularization/perfusion visualization as an aid to clinical management of adult and pediatric patients with liver disease.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
Esonic Technology (Wuhan) Co., Ltd.
July 23, 2026
Guo Dongyun
Regulatory Affairs Manager
Floor 7, Bldg. B1, Block B, Phase Ii, High-Tech Medical Devices Park, #818 Gaoxin Ave., E. Lake High-Tech Development Zone
Wuhan, Hubei 430000
China
Re: K261272
Trade/Device Name: ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic ultrasound system
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX,
Dated: June 23, 2026
Received: June 23, 2026
Dear Guo Dongyun:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261272 - Guo Dongyun
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261272 - Guo Dongyun
Page 3
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.07.23 14:23:57 -04'00' For
Yanna Kang, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K261272
Please provide the device trade name(s).
ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic ultrasound system
Please provide your Indications for Use below.
The ultrasound system is indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans- vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac.
In addition, the ultrasound diagnostic system and associated transducers are intended for:
- Measurements of abdominal anatomical structures,
- Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen,
- Measurements of brightness ratio between liver and kidney,
- Visualization of abdominal vascularization, microvascularization and perfusion,
The shear wave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion maybe used as an aid to clinical management of adult and pediatric patients with liver disease.
This device is intended for use by, or by the order of, and under the supervision of a licensed physician qualified to use or direct the use of the device.
This system should only be used by trained sonographers who are knowledgeable about the risk of excessive acoustic energy in the body, particularly in the case where a great amount of fluid is present in the scanning area. This device is intended to be used in a hospital or medical clinic.
Modes of operation include: B-Mode, M-mode, Color Doppler imaging, Pulsed Wave Doppler, Continuous Wave, Harmonic Imaging, Power Doppler imaging, Directional Power Doppler imaging, Contrast Enhanced Ultrasound Imaging, Elasticity Imaging and combined modes: B+ Color Flow, B+ Shear Wave Elastography, B+ Pulsed Wave, B+ Pulsed Wave + Color Flow, B+ M mode, B+ Continuous Wave.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
# Classification
Add a primary product code and any associated product codes below. You may type in the primary product code directly (only the product code field is required) or you may filter down by choosing first a medical specialty, regulation, then product code. Please verify whether a device specific guidance or special control is available for the product code at the Product Classification website linked to below. The regulation should be consulted to determine if any special controls need to be considered (e.g. PAE and 21 CFR 890.3450). Use this website to verify information for associated product codes as well.
| Medical Specialty | Radiology |
| --- | --- |
| Regulation | 892.1550 - Ultrasonic pulsed doppler imaging system |
| Product Code | IYN (Class 2) - System, Imaging, Pulsed Doppler, Ultrasonic |
Information about your primary product code, such as whether any applicable guidances or standards should be consulted, is available at the website below.
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPCD/classification.cfm?ID=IYN
Associated Product Code(s) IYO, ITX
/
Page 11 of 58
{5}
eSonic
K261272
### 510k Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
### 1. Submitter's name, address, telephone number, contact person
Submitted by:
Manufacturer Name: eSonic Technology (Wuhan) Co., LTD
Address: Floor 7, Building B1, Block B, Phase II, High-tech Medical Devices Park, No.818 Gaoxin Avenue, East Lake High-tech Development Zone, Wuhan, Hubei, China
Primary Correspondent: Dongyun Guo/ Regulatory Affairs Manager
E-mail: registration@esonicimage.com
Telephone: (+86)13476081662
Date: 2026.4.16
### 2. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Trade/Device Name: ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic ultrasound system
Common Name: Diagnostic Ultrasound System
Regulatory Class: II
Review Panel: Radiology
| Classification Name | 21 CFR Section | Product Code |
| --- | --- | --- |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
### 3. Substantially Equivalent/Predicate Devices
AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007).
### 4. Device Description
The proposed ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound System are cart based ultrasound imaging system used to perform diagnostic
- 1 -
{6}
eSonic
general purpose ultrasound imaging studies.
The system contains a scan converter and can be coupled to a variety of linear, convex, micro-convex and phased array transducers to produce images, which are displayed on a LCD monitor.
An adjustable control panel with integrated touch screen allows the user to perform an ultrasound exam quickly and efficiently in accordance with ALARA principles.
The system also allows the user to perform measurements and associated calculations, capture images to digital memory or to an external device (such as a printer), and review diagnostic studies in the form of a report.
### 5. Intended Use/Indications for Use
The ultrasound system is indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans- vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac.
In addition, the ultrasound diagnostic system and associated transducers are intended for:
- Measurements of abdominal anatomical structures,
- Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen,
- Measurements of brightness ratio between liver and kidney,
- Visualization of abdominal vascularization, microvascularization and perfusion
The shear wave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion may be used as an aid to clinical management of adult and pediatric patients with liver disease.
This device is intended for use by, or by the order of, and under the supervision of a licensed physician qualified to use or direct the use of the device.
This system should only be used by trained sonographers who are knowledgeable about the risk of excessive acoustic energy in the body, particularly in the case where a great amount of fluid is present in the scanning area. This device is intended to be used in a hospital or medical clinic.
Modes of operation include: B-Mode, M-mode, Color Doppler imaging, Pulsed
- 2 -
{7}
eSonic
Wave Doppler, Continuous Wave, Harmonic Imaging, Power Doppler imaging, Directional Power Doppler imaging, Contrast Enhanced Ultrasound Imaging, Elasticity Imaging and combined modes: B+ Color Flow, B+ Shear Wave Elastography, B+ Pulsed Wave, B+ Pulsed Wave + Color Flow, B+ M mode, B+ Continuous Wave.
### 6. Summary of the technological characteristics- new device compared to the predicate device
The ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems and its predicate device, AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007), have the equivalent intended use, and similar physical characteristics.
| Description | Subject Device | Predicate Device | Comments |
| --- | --- | --- | --- |
| | ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007) | |
| Indications for use | The ultrasound system is indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans- vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac. In addition, the ultrasound diagnostic system and associated | The SuperSonic Imagine AIXPLORER MACH 30 ultrasound system is indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, Intraoperative, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac. In addition, the SuperSonic Imagine AIXPLORER MACH 30 ultrasound diagnostic system and associated | Equivalent See Note 1 Note 3 |
- 3 -
{8}
eSonic
| Description | Subject Device ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | Predicate Device AIXPLORER MACH 30 Ultrasound Diagnostic System **(K191007)** | Comments |
| --- | --- | --- | --- |
| | transducers are intended for: • Measurements of abdominal anatomical structures, • Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen, • Measurements of brightness ratio between liver and kidney, • Visualization of abdominal vascularization, microvascularization and perfusion The shear wave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion may be used as an aid to clinical management of adult and pediatric patients with liver disease. | transducers are intended for: • Measurements of abdominal anatomical structures, • Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen, • Measurements of brightness ratio between liver and kidney, • Visualization of abdominal vascularization, microvascularization and perfusion, • **Quantification of abdominal vascularization and perfusion** The shear wave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion, **the quantification of vascularization and perfusion** may be used as an aid to clinical management of adult and pediatric patients with liver disease. | |
- 4 -
{9}
[LOGO]
| Description | Subject Device | Predicate Device | Comments |
| --- | --- | --- | --- |
| | ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | AIXPLORER MACH 30 Ultrasound Diagnostic System **(K191007)** | |
| | This device is intended for use by, or by the order of, and under the supervision of a licensed physician qualified to use or direct the use of the device. This system should only be used by trained sonographers who are knowledgeable about the risk of excessive acoustic energy in the body, particularly in the case where a great amount of fluid is present in the scanning area. **This device is intended to be used in a hospital or medical clinic.** **Modes of operation include: B-Mode, M-mode, Color Doppler imaging, Pulsed Wave Doppler, Continuous Wave, Harmonic Imaging, Power Doppler imaging, Directional Power Doppler imaging, Contrast Enhanced Ultrasound Imaging, Elasticity Imaging and combined modes: B+ Color Flow, B+ Shear Wave** | This device is intended for use by, or by the order of, and under the supervision of a licensed physician qualified to use or direct the use of the device. This system should only be used by trained sonographers who are knowledgeable about the risk of excessive acoustic energy in the body, particularly in the case where a great amount of fluid is present in the scanning area | |
- 5 -
{10}
[LOGO]
| Description | Subject Device | Predicate Device | Comments |
| --- | --- | --- | --- |
| | ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | AIXPLORER MACH 30 Ultrasound Diagnostic System **(K191007)** | |
| | **Elastography, B+ Pulsed Wave, B+ Pulsed Wave + Color Flow, B+ M mode, B+ Continuous Wave.** | | |
| User | Qualified and trained healthcare professionals | Qualified and trained healthcare professionals | Same |
| Environment | Hospital or medical clinic | Hospital or medical clinic | Same |
| Design | Autocorrelation for color processing and FFT for pulse Doppler processing. Supporting Linear, Convex, Phase array probes. | Autocorrelation for color processing and FFT for pulse Doppler processing. Supporting Linear, Convex, Phase array probes. | Same |
| Patient Contact Materials | Material meet ISO 10993-1 and FDA guidance | Material meet ISO 10993-1 and FDA guidance | Equivalent **See Note2** |
| Mode of operation | Modes of operation include: B-Mode, M-mode, Color Doppler imaging, Pulsed Wave Doppler, Continuous Wave, Harmonic Imaging, *Power Doppler imaging*, *Directional Power Doppler imaging*, Contrast Enhanced Ultrasound Imaging, Elasticity Imaging. Combined modes: B+ Color Flow, B+ Shear Wave | Modes of operation include: B-Mode, M-mode, Color Doppler imaging, Pulsed Wave Doppler, Continuous Wave, Harmonic Imaging, *Amplitude Power Doppler imaging*, *Directional Amplitude Power imaging*, Contrast Enhanced Ultrasound Imaging, Elasticity Imaging and **3D imaging**. Combined modes: B+ Color Flow, B+ Shear Wave | Equivalent **See Note 3** |
- 6 -
{11}
[LOGO]
eSonic
| Description | Subject Device | Predicate Device | Comments |
| --- | --- | --- | --- |
| | ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | AIXPLORER MACH 30 Ultrasound Diagnostic System **(K191007)** | |
| | Elastography, B+ Pulsed Wave, B+ Pulsed Wave + Color Flow, B+ M mode, B+ Continuous Wave. | Elastography, B+ Pulsed Wave, B+ Pulsed Wave + Color Flow, B+ M mode, B+ Continuous Wave, **B+Color flow +Shear wave Elastography, B+ M mode + Color flow, B + Strain + Shear Wave Elastography** | |
| Functions | ➢ Tissue Harmonic Imaging ➢ *eCompound* (Spatial compounding) ➢ Panoramic imaging ➢ Biopsy guideline ➢ *Needle TS* ➢ ShearWave Elastography ➢ *Viscosity (dispersion nature of shear wave)* ➢ *ATI (attenuation imaging)* ➢ *QSOS(quantitative imaging of sound speed)* ➢ *CFI, PDI, dPDI* ➢ *iPlane PW* ➢ *iPlane Vascular (Color Doppler improvement)* ➢ *TDI* ➢ *CEUS* ➢ *ECG* ➢ *eTuner (Automatically adjust TGC)* ➢ *Optimize* | ➢ Tissue Harmonic Imaging ➢ *SuperCompound (Spatial compounding)* ➢ Panoramic imaging ➢ Biopsy guideline ➢ *Needle PL.U. S* ➢ ShearWave Elastography ➢ *Vi PLUS (dispersion nature of shear wave)* ➢ *Att PLUS (attenuation imaging)* ➢ *SSp PLUS(quantitative imaging of sound speed)* ➢ *CFI, CPI, dCPI* ➢ *UltraFast™* ➢ *Angio PL.U.S (Color Doppler improvement)* ➢ *TDI* ➢ *CEUS* ➢ *ECG* ➢ *Auto (Automatically adjust TGC)* ➢ *Optimization* | Equivalent **See Note 4** |
- 7 -
{12}
[LOGO]
eSonic
| Description | Subject Device ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | Predicate Device AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007) | Comments |
| --- | --- | --- | --- |
| | ➢ Width ➢ eTissue ➢ eClear ➢ Expand (trapezoid) ➢ USB ports ➢ Displayport ➢ CD/DVD Drive ➢ Printing ➢ Network(wired) ➢ DICOM ➢ Dual/ Quad ➢ Cine ➢ Bodymarker ➢ Annotation ➢ Report | ➢ Sector Size ➢ TissueTuner (Speed of Sound control) ➢ SuperRes (Adaptive Filtering) ➢ Wide Image (trapezoid) ➢ USB ports ➢ Displayport ➢ CD/DVD Drive ➢ Printing ➢ Network(wired and wireless) ➢ DICOM ➢ Dual-split/ Quad-split ➢ Cine ➢ Bodymarker ➢ Annotation ➢ Report ➢ Strain Elastography ➢ 3D ➢ Barcode Scanner | |
| Measurements | **Basic measurement:** ➢ Measurement in B Mode: distance, ellipse, trace, depth, volume, hip angle, d:D, IMT, B-mode ratio, % diameter reduction, % area reduction, distance ratio ➢ Measurement in D Mode: doppler time, PW autotrace, velocity, PSV/EDV, vessel diameter, doppler slope, doppler trace, Heart rate, TAMV, volume flow | **Basic measurement:** ➢ Measurement in B Mode: distance, ellipse, trace, depth, volume, hip angle, d:D, IMT, B-mode ratio, % diameter reduction, % area reduction, distance ratio ➢ Measurement in D Mode: doppler time, PW autotrace, velocity, PSV/EDV, Doppler 2 Velocity Ratio, vessel diameter, doppler slope, doppler time-slope, | Equivalent See Note 5 |
- 8 -
{13}
[LOGO]
eSonic
| Description | Subject Device ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | Predicate Device AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007) | Comments |
| --- | --- | --- | --- |
| | ➤ **Measurement in M mode** Heart rate, M distance, M time slope ➤ **Measurement in E mode** *Q-MEAS, Q- ratio, Q- MEAS trace, MULTI Q- MEAS* **Measurement package:** ➤ Abdomen measurement package ➤ Gynecology measurement package ➤ Obstetric measurement package ➤ Cardiac measurement package ➤ Vascular measurement package ➤ Urology measurement package ➤ Small organ measurement package | doppler trace, TAMV, **congestion index**, Heart rate, volume flow ➤ **Measurement in M mode** Heart rate, M-distance, M time-slope ➤ **Measurement in E mode** *Q-BOX™, Q-BOX™ ratio, Q-BOX™ trace, MULTI Q-BOX™* **Measurement package:** ➤ Abdomen measurement package ➤ Gynecology measurement package ➤ Obstetric measurement package ➤ Cardiac measurement package ➤ Vascular measurement package ➤ Urology measurement package ➤ Small organ measurement package | |
| Transducer Types | Convex array Phased array Linear array Micro convex array | Convex array Phased array Linear array Micro convex array | Same |
| Acoustic Output within FDA | Track 3; Ispta.3 ≤ 720 mW/cm² MI ≤ 1.9 | Track 3; Ispta.3 ≤ 720 mW/cm² MI ≤ 1.9 | Same |
- 9 -
{14}
eSonic
| Description | Subject Device ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound Systems | Predicate Device AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007) | Comments |
| --- | --- | --- | --- |
| guidelines | TI ≤ 6.0 | TI ≤ 6.0 | |
| General safety and effectiveness information | ANSI/AAMI ES60601-1 IEC60601-2-37 IEC60601-1-2 ISO 10993-1 | ANSI/AAMI ES60601-1 IEC60601-2-37 IEC60601-1-2 ISO 10993-1 | Same |
| Labeling | Conforms to 21 CFR Part 801 | Conforms to 21 CFR Part 801 | Same |
| Accessories | ➤ Foot switch ➤ ECG cables ➤ Printer | ➤ Foot switch ➤ ECG cables ➤ Printer ➤ Displayport to HDMI, VGA and DVI adapter ➤ Barcode scanner ➤ Ethernet cable ➤ Wifi dongle ➤ Gels ➤ Sheaths ➤ Biopsy Guide ➤ Cleaners | Equivalent See Note 6 |
### Technology:
The ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Ultrasound Diagnostic System employs the same fundamental scientific technology as its predicate device.
### Determination of substantial equivalence:
The Proposed ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Ultrasound Diagnostic System are substantially equivalent to the predicate AIXPLORER MACH 30 Ultrasound Diagnostic System (K191007) with regards to intended use, indication
- 10 -
{15}
eSonic
for use, image capabilities, technological characteristics, image mode.
The following is an overview of the differences between the proposed Ultrasound Diagnostic System and its predicate device.
# **Comparison Analysis:**
# **Note 1: Indications for use**
The Indication for use of subject device does not support application of intraoperative, quantification of abdominal vascularization and perfusion, and not support operation modes of 3D imaging, B+Color flow +Shear wave Elastography, B+M mode+Color flow, B+Strain+Shear Wave Elastography. As the subject device does not support specific application and mode of operation or combined mode, it is less than predicate device. They can demonstrate substantial equivalence with predicate device.
# **Note 2: Patient Contact Materials**
The Patient Contact Materials of probes is equivalent to the predicate device, both the subject and predicate device probe material meet ISO 10993-1 and FDA guidance requirement. They can demonstrate substantial equivalence with predicate device.
# **Note 3: Mode of operation**
The subject device does not support operation modes of 3D imaging, B+Color flow +Shear wave Elastography, B+M mode+Color flow, B+Strain+Shear Wave Elastography. As the subject device does not support specific application and mode of operation or combined mode, it is less than predicate device. They can demonstrate substantial equivalence with predicate device.
# **Note 4: Functions**
The subject device does not support functions of wireless, strain elastography, 3D and barcode scanner. As the subject device does not support specific functions, it is less than predicate device. And also functions eCompound (Spatial compounding), Needle TS, Viscosity (dispersion nature of shear wave), ATI (attenuation imaging), QSOS(quantitative imaging of sound speed), PDI, dPDI, iPlane PW, iPlane Vascular (Color Doppler improvement), eTuner, Optimize, Width, eTissue, eClear, Expand (trapezoid), Dual/ Quad of subject device are equivalent as SuperCompound (Spatial compounding), Needle PL.U. S, Vi PLUS, Att PLUS, SSp PLUS, CPI, dCPI, UltraFast™, Angio PL.U.S (Color Doppler improvement), Auto (Automatically adjust TGC), Optimization, Sector Size, TissueTuner (Speed of Sound control), SuperRes (Adaptive Filtering), Wide Image (trapezoid), Dual-
- 11 -
{16}
eSonic
split/ Quad-split of predicate device respectively, just different name.
They can demonstrate substantial equivalence with predicate device.
# Note 5: Measurements
The subject device does not support Doppler 2 Velocity Ratio, and congestion index. As the subject device function is less than Predicate device. And also Q-MEAS, Q- ratio, Q- MEAS trace, MULTI Q- MEAS of subject device are equivalent as Q-BOX™, Q-BOX™ ratio, Q-BOX™ trace, MULTI Q-BOX™ of predicate device respectively, just different name.
They can demonstrate substantial equivalence with predicate device.
# Note 6: Accessories
The subject device does not support Accessories of Display port to HDMI, VGA and DVI adapter, Barcode scanner, Ethernet cable, Wifi dongle, Gels, Sheaths, Biopsy Guide, Cleaners. As the subject device accessories are less than predicate device, they can demonstrate substantial equivalence with predicate device.
The ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Ultrasound Diagnostic Systems employ the same fundamental scientific technology as its predicate device. The Proposed Ultrasound Diagnostic Systems are substantially equivalent to the predicate AIXPLORER MACH 30 (K191007) with regards to intended use, indication for use, image capabilities, technological characteristics, image mode. The patient contact materials, functions are demonstrated substantial equivalent with predicate device.
eSonic Technology (Wuhan) Co., LTD. believes that the ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Ultrasound Diagnostic Systems are substantially equivalent with the predicate device.
# 7. A brief discussion of the nonclinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence
# Summary of Non-clinical test:
The ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Ultrasound Diagnostic System were evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical
- 12 -
{17}
eSonic
safety, and have been found to comply with applicable medical device safety standard. The Ultrasound Diagnostic System complies with voluntary standards:
1. (1) ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
2. (2) IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
3. (3) IEC TS 60601-4-2: 2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
4. (4) IEC 60601-2-37 Edition 3.0 2024-07 Medical electrical equipment – Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
5. (5) IEC62359 Edition 2.1 2017-09 Ultrasonics-Field characterization- Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017.
6. (6) ISO10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
7. (7) ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
8. (8) ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
9. (9) ISO10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
10. (10) ISO14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
11. (11) NEMA PS 3.1 - 3.20 2024e Digital Imaging and Communications in Medicine (DICOM) Set.
12. (12) IEC60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
13. (13) IEC62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION Medical devices - Part 1: Application of usability engineering to medical devices
14. (14) IEC62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes
- 13 -
{18}
[LOGO]
(15) ISTA 3E 2017 Similar Packaged-Products in Unitized Loads of Truckload Shipment
The above testing confirmed that ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound System perform according to the stated intended use. All data fell within pre-determined product specifications and external standard requirements. Results of non-clinical testing confirmed the substantial equivalence of the Diagnostic Ultrasound System to the predicate device.
The performance tests of the Diagnostic Ultrasound System demonstrate substantial equivalence (SE) to the predicate device.
8. A brief discussion of the clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence
Not Applicable. The subject of this premarket submission, did not require clinical studies to support substantial equivalence.
9. Conclusion
For testing, all pre-determined acceptance criteria were met. Results of these tests show that the proposed ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound System meet their intended use.
The proposed ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound System is similar to the predicate AIXPLORER MACH 30 (K191007) in terms of indications for use, design, technological characteristics, modes of operations. eSonic considers the ePascal S/ ePascal SOR/ ePascal S5/ ePascal SA/ ePascal SE/ ePascal 80Elite/ ePascal 80/ MOZI-X1/ MOZI-X12/ DF-41E Diagnostic Ultrasound System to be substantially equivalent, with respect to safety and effectiveness, to the predicate device.
- 14 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.