The Altis Full Face Mask is intended to provide an interface for the application of non-invasive positive airway pressure (PAP) therapy. The mask is intended for single-patient reuse in the home environment and healthcare facilities and/or hospital environments. The mask is to be used by adult patients aged 22 years and older weighing more than 66lb (30kg) for whom non-invasive PAP therapy has been prescribed. The device is not intended for critical care patients.
Device Story
Altis Full Face Mask; over-the-nose/mouth interface for non-invasive PAP therapy. Components: cushion (polycarbonate faceplate, silicone seal), frame, exhalation elbow, headgear, magnetic clips. Connects to standard ISO 5356-1 22mm hoses. Used in home and hospital settings; operated by patients or healthcare providers. Delivers pressurized air to airways. Bench-tested for flow, resistance, CO2 rebreathing, and biocompatibility. No clinical data required.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via non-clinical bench testing, including ISO 17510:2015 performance testing (exhaust flow, resistance, CO2 rebreathing, anti-asphyxia valve, noise), biocompatibility (ISO 10993, ISO 18562), and cleaning validation.
Technological Characteristics
Materials: polycarbonate, silicone, polypropylene, woven/laminate fabrics, magnets. Interface: full face mask with exhalation elbow. Connectivity: ISO 5356-1 22mm conical hose connection. Pressure range: 4-40 cm H2O. Non-sterile. Biocompatibility per ISO 10993 and ISO 18562. Performance per ISO 17510:2015.
Indications for Use
Indicated for adult patients aged 22+ years weighing >66lb (30kg) requiring non-invasive PAP therapy. Not for critical care patients.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 25, 2026
Respironics, Inc.
Viralyne Lipps
Principal Regulatory Affairs Specialist
1001 Murry Ridge Ln.
Murrysville, Pennsylvania 15668
Re: K261262
Trade/Device Name: Altis Full Face Mask
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD
Dated: August 28, 2026
Received: August 28, 2026
Dear Viralyne Lipps:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261262 - Viralyne Lipps
Page 2
When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K261262 - Viralyne Lipps
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J.
Mathews -S
Digitally signed by
Binoy J. Mathews -S
Date: 2026.09.25
12:02:32 -04'00'
For
James Lee, PhD
Division Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261262 | ? |
| Please provide the device trade name(s). | | ? |
| Altis Full Face Mask | | |
| Please provide your Indications for Use below. | | ? |
| The Altis Full Face Mask is intended to provide an interface for the application of non-invasive positive airway pressure (PAP) therapy. The mask is intended for single-patient reuse in the home environment and healthcare facilities and/or hospital environments. The mask is to be used by adult patients aged 22 years and older weighing more than 66lb (30kg) for whom non-invasive PAP therapy has been prescribed. The device is not intended for critical care patients. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
PHILIPS
Altis Full Face Mask
Traditional 510(k)
K261262
### 510(k) Summary
[As required by 21 CFR 807.92(c)]
Date Prepared: September 24, 2026
### I. Submitter
Company Name/Owner: Respironics, Inc.
1001 Murry Ridge Lane
Murrysville, PA 15668, USA
Official Contact: Viralyne Lipps
Principal Regulatory Affairs Specialist
### II. Device
Device Trade Name: Altis Full Face Mask
Device Common Name: Full Face Mask
Regulation Number: 21 CFR 868.5905
Classification Name: Noncontinuous ventilator (IPPB)
Product Code: BZD
Device Classification: Class II
### III. Predicate Device
Predicate Device: Resmed AirFit F20 Mask System; AirFit F20 NM Mask System (K242547)
Regulation Number: 21 CFR 868.5905
Classification Name: Noncontinuous ventilator (IPPB)
Product Code: BZD
This predicate has not been the subject of a design-related recall.
### IV. Device Description
The Altis Full Face Mask is an over-the-nose, full-face mask with the hose connection in the front. The mask covers both the nose and mouth, ensuring pressurized air is delivered directly into the patient's airways with the use of a positive airway pressure (PAP) therapy device; it is intended for single-patient reuse in the home and healthcare facilities and/or hospital environments.
The mask consists of five main parts: a cushion (polycarbonate faceplate and bond ring with silicone seal), frame, exhalation elbow, headgear, and headgear clips. The elbow connects to standard ISO 5356-1 22 mm conical hoses.
### V. Indications For Use
The Altis Full Face Mask is intended to provide an interface for the application of non-invasive positive airway pressure (PAP) therapy. The mask is intended for single-patient reuse in the home environment and healthcare facilities and/or hospital environments. The mask is to be used by adult patients aged 22 years and older weighing more than 66 lb (30 kg) for whom non-invasive PAP therapy has been prescribed. The device is not intended for critical care patients.
Page 1 of 6
{6}
PHILIPS
Altis Full Face Mask
Traditional 510(k)
K261262
# VI. Comparison with Predicate Device
| Design/Technological Characteristic | Subject Device Altis Full Face Mask | Predicate Device AirFit F20 and Air Fit NM Mask Systems (K242547) | Comments |
| --- | --- | --- | --- |
| Product Code | BZD | BZD | Identical |
| Regulation Number | 21 CFR 868.5905 | 12 CFR 868.5905 | Identical |
| Classification Name | Non-continuous ventilator | Non-continuous ventilator | Identical |
| Regulatory Class | Class II | Class II | Identical |
| Classification Panel | Anesthesiology | Anesthesiology | Identical |
| Indications for Use / Intended Use | The Altis Full Face Mask is intended to provide an interface for the application of non-invasive positive airway pressure (PAP) therapy. The mask is intended for single-patient reuse in the home environment and healthcare facilities and/or hospital environments. The mask is to be used by adult patients aged 22 years and older weighing more than 66 lb (30 kg) for whom non-invasive PAP therapy has been prescribed . The device is not intended for critical care patients. | The AirFit F20 masks have two product variants: - The AirFit F20 variants are intended for single-patient reuse in the home environment - The AirFit F20 SLM (Sleep Lab Masks) variants are intended for multi-patient reuse in the hospital/institutional environment.These masks are intended for patients weighing more than 66lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Masks are the only variants that are validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients | Equivalent |
| Availability | Prescription use only | Prescription use only | Identical |
| Patient Population | Adult patients weighing more than 66 lb (30 kg) | Patients weighing more than 66 lb (30 kg) | Equivalent |
| Anatomical Sites | Nose and mouth | Nose and mouth | Identical |
| Operating Environment | Home or healthcare facilities and/or hospitals | Home or health institution | Equivalent |
| Operating and Storage Temperature Range | Operating: 5 °C to 40 °C Storage: -20 °C to +60 °C | Operating: 5 °C to 40 °C Storage: -20 °C to +60 °C | Identical |
| Patient Circuit Connection | ISO 5356-1 (22 mm) | ISO 5356-1 (22 mm) | Identical |
| Sizes | Cushion available in four sizes Headgear available in two sizes Frame available in one size | Cushion available in three sizes Headgear available in three sizes Frame available in one size | Equivalent |
| Exhalation Vent | Multi-hole vent array | Multi-hole vent | Equivalent |
| Sterile State as Provided | Non-sterile | Non-sterile | Identical |
Page 2 of 6
{7}
PHILIPS
Altis Full Face Mask
Traditional 510(k)
K261262
| Design/Technological Characteristic | Subject Device Altis Full Face Mask | | Predicate Device AirFit F20 and Air Fit NM Mask Systems (K242547) | | Comments |
| --- | --- | --- | --- | --- | --- |
| Pressure Range | 4 to 40 cm H_{2}O | | 3 to 40 cm H_{2}O | | Equivalent |
| Mask Exhaust Flow (nominal) | Pressure (cmH_{2}O) | SLPM | Pressure (cmH_{2}O) | SLPM | Equivalent Disclosed per ISO 17510:2015 Annex B |
| | 4 | 16 | 3 | 19 | |
| | 13 | 32 | 20 | 54 | |
| | 22 | 44 | 40 | 82 | |
| | 31 | 55 | | | |
| | 40 | 64 | | | |
| Resistance to Flow (pressure drop) | @50l/min | @100l/min | @50l/min | @100l/min | Equivalent Disclosed per ISO 17510:2015 Annex C |
| | 0.2 cmH_{2}O | 0.9 cmH_{2}O | 0.2 cmH_{2}O | 0.6 cmH_{2}O | |
| Anti-Asphyxia Valve Operation | Open to atmosphere | Closed to atmosphere | Open to atmosphere | Closed to atmosphere | Equivalent (<min. rated pressure) ISO 17510:2015 Annex D |
| | 1.5 cmH_{2}O | 1.8 cmH_{2}O | 1.5 cmH_{2}O | 2.0 cmH_{2}O | |
| Inspiratory and Expiratory Resistance | Resistance | @50SLPM | Resistance | @50SLPM | Equivalent (<10 cmH_{2}O) ISO 17510:2015 Annex E |
| | Inspiratory | 1.6 cmH_{2}O | Inspiratory | 0.60 cmH_{2}O | |
| | Expiratory | 1.3 cmH_{2}O | Expiratory | 0.80 cmH_{2}O | |
| CO_{2} Rebreathing (normal) | Pressure (cmH_{2}O) | Relative CO_{2} increase | Pressure (cmH_{2}O) | Relative CO_{2} increase | Equivalent (<20%) ISO 17510:2015 Annex F |
| | 4 (min.) | 12.4% | 3 (min.) | 11.7% | |
| | 5 | 10.9% | 5 | 7.3% | |
| | 10 | 7.0% | 10 | 2.9% | |
| CO_{2} Rebreathing (single fault) | Fault | Relative CO_{2} increase | Fault | Relative CO_{2} increase | Equivalent (<60%) ISO 17510:2015 Annex F |
| | Case 1 | 27.8% | Case 1 | 40.5% | |
| | Case 2 | 26.6% | Case 2 | 52.7% | |
| Sound | Power level: 25 dBA Pressure level: 18 dBA | | Power level: 30 dBA Pressure level: 23 dBA | | Equivalent Disclosed per ISO 17510:2015 Annex G |
| Dead Space | ≤ 248 mL | | ≤ 240 mL | | Equivalent Disclosed characterization |
| Device Design | • Elbow • Frame • Cushion • Headgear | | • Elbow • Frame • Cushion • Headgear | | Identical |
| Construction Materials | Polycarbonate, silicone, polypropylene, woven/ laminate fabrics, magnets | | Polymeric materials, Textile and magnets | | Equivalent |
Page 3 of 6
{8}
PHILIPS
Altis Full Face Mask
Traditional 510(k)
K261262
| Design/Technological Characteristic | | Subject Device Altis Full Face Mask | Predicate Device AirFit F20 and Air Fit NM Mask Systems (K242547) | Comments |
| --- | --- | --- | --- | --- |
| **Faceplate / Cushion Type** | | Covers nose and mouth | Covers nose and mouth | Identical |
| **Exhalation Device Design** | | Exhalation Elbow | Exhalation Elbow | Equivalent |
| **Headgear Type** | | Adjustable headgear straps | Adjustable headgear straps | Equivalent |
| **Validated Home Cleaning** | | Mild liquid detergent in warm water | Mild liquid detergent in warm water | Equivalent |
| **Magnetic Clips** | Number of Magnets | 4 | 4 | Equivalent Disclosed characterization |
| | Strength on the surface | <400 mT | <400 mT | |
| | Distance where strength <0.5mT | 76mm | 50 mm | |
### VII. Performance Data
#### Non-Clinical Tests
The Altis Full Face Mask subject device was designed in accordance with ISO 17510:2015, Medical devices – Sleep apnoea breathing therapy – Masks and application accessories and performance testing has demonstrated that the device meets the industry standard performance requirements. The following non-clinical bench performance tests were conducted:
- Exhaust flow (ISO 17510:2015 Annex B)
- Resistance to flow (ISO 17510:2015 Annex C)
- Anti-Asphyxia valve operating pressures (ISO 17510:2015 Annex D)
- Breathing resistance (ISO 17510:2015 Annex E)
- CO₂ rebreathing (ISO 17510:2015 Annex F)
- Vibration and noise (ISO 17510:2015 Annex G)
- Mask connectors (ISO 17510: 2015 Clause 5.1)
Other bench tests to verify product specifications were also completed:
- Minimum/maximum pressure performance testing
- Dead space volume determination
- Magnetic flux density determination for magnetic components
- Simulated transport and storage of packaging (ISTA 3A)
- Intentional mask leak and patient pressure drop
- Home cleaning validation
The biocompatibility of the Altis Full Face Mask was evaluated in accordance with ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, ISO 18562-1:2024 – Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process, and the FDA guidance document Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, Issued September 8, 2023.
Page 4 of 6
{9}
PHILIPS
Altis Full Face Mask
Traditional 510(k)
K261262
Validation of single-patient re-use reprocessing claims was evaluated in accordance with ISO 17664-1:2021 – Processing of health care products – Information to be provided by the medical device manufacturer for the processing of medical devices – Part 1: Critical and semi-critical medical devices, ANSI AAMI ST98:2022 – Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices, AAMI TIR12:2020 – Designing, testing, and labeling medical devices intended for processing by health care facilities: A guide for device manufacturers, and the FDA guidance document, Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, updated June 9, 2017, included a combination of cleaning efficacy, residual toxicity, and performance testing.
The Altis Full Face Mask has been tested in accordance with the applicable requirements of the following standards:
- ISO 5356-1:2015 – Anaesthetic respiratory equipment – Conical connectors – Part 1: Cones and sockets
- ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 – Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 – Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-12:2021 – Biological evaluation of medical devices – Part 12: Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- ISO 10993-17:2023 – Biological evaluation of medical devices – Part 17: Toxicological risk assessment of medical device constituents
- ISO 10993-18:2020 + A1:2022 – Biological evaluation of medical devices – Part 18: Chemical characterization of medical device materials within a risk management process – Amendment 1: Determination of the uncertainty factor
- ISO 10993-23:2021 – Biological evaluation of medical devices – Part 23: Tests for irritation
- ISO 15223-1:2021 – Medical devices – Symbols to be used with information to be supplied by the manufacturer – Part 1: General requirements
- ISO 17510:2015 – Medical devices – Sleep apnoea breathing therapy – Masks and application accessories
- ISO 17664-1:2021 – Processing of health care products – Information to be provided by the medical device manufacturer for the processing of medical devices – Part 1: Critical and semi-critical medical devices
- ISO 18562-1:2024 – Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2024 – Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2024 – Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds
- ISO 18562-4:2024 – Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 4: Tests for leachables in condensate
- ISO 20417:2021 – Medical devices – Information to be supplied by the manufacturer
- IEC 62366-1:2015+AMD1:2020 – Medical devices – Part 1: Application of usability engineering to medical devices
- ASTM F2503-23 – Standard Practice for Marking Medical Devices and other Items for Safety in the Magnetic Resonance Environment
Page 5 of 6
{10}
PHILIPS
Altis Full Face Mask
Traditional 510(k)
K261262
- ISTA 3A:2018 – Packaged Products for Parcel Delivery System Shipment 70kg (150lb. or less)
- ANSI AAMI ST98:2022 – Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices
- AAMI TIR12:2020 – Designing, testing, and labeling medical devices intended for processing by health care facilities: A guide for device manufacturers
### VIII. Clinical Tests
Clinical tests were not required to demonstrate the safety and effectiveness of the Altis Full Face Mask. All risks have been sufficiently mitigated and product functionality has been adequately assessed by non-clinical tests.
### IX. Conclusion
The Altis Full Face Mask is substantially equivalent to the predicate device, the Resmed AirFit F20 and Air Fit NM Mask Systems (K242547) based on its intended use and technological characteristics. The results of the verification testing also confirm compliance with FDA recognized consensus standards that were also evaluated by the predicate device. These evaluations support the conclusion that the subject device does not raise new questions of safety or efficacy.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.